Quantitative analysis of near-infrared tomography: sensitivity to the tissue-simulating precalibration phantom

[+] Author Affiliations
Shudong Jiang, Brian W. Pogue, Troy O. McBride, Keith D. Paulsen

Dartmouth College, Thayer School of Engineering, Hanover, New Hampshire 03755 Shudong.Jiang@Dartmouth.edu

J. Biomed. Opt. 8(2), 308-315 (Apr 01, 2003). doi:10.1117/1.1559692
History: Received Nov. 8, 2001; Revised May 29, 2002; Accepted Sep. 23, 2002; Online April 01, 2003
Text Size: A A A

A near-infrared (NIR) imaging system is evaluated as a diagnostic clinical tool to image total hemoglobin concentration and oxygen saturation within tissue. Calibration of this type of system requires measurement of the response at each detector and source location from a homogeneous tissue-simulating phantom. The effect of using calibration phantoms of varying composition, size, and optical properties is examined to determine how it affects the overall image accuracy. All of the calibration phantoms investigated result in accurate reconstruction of absorbing heterogeneities due to increased blood concentration with less than 4% standard deviation. Images from a patient with a biopsy-confirmed ductal carcinoma are also evaluated and found to be insensitive to the choice of calibration object, with only 1% variation between images generated with different calibration objects. The tumor total hemoglobin contrast is approximately 240% higher than the average total hemoglobin concentration in contralateral breast. Soft calibration phantoms, which mimic the elastic properties of human breast tissue, are also considered and found to diminish positioning errors in the fibers relative to the actual breast exam, thereby reducing the artifacts in the periphery of the reconstructed image. © 2003 Society of Photo-Optical Instrumentation Engineers.

© 2003 Society of Photo-Optical Instrumentation Engineers

Citation

Shudong Jiang ; Brian W. Pogue ; Troy O. McBride and Keith D. Paulsen
"Quantitative analysis of near-infrared tomography: sensitivity to the tissue-simulating precalibration phantom", J. Biomed. Opt. 8(2), 308-315 (Apr 01, 2003). ; http://dx.doi.org/10.1117/1.1559692


Tables

Access This Article
Sign in or Create a personal account to Buy this article ($20 for members, $25 for non-members).

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Related Book Chapters

Topic Collections

PubMed Articles
Advertisement
  • Don't have an account?
  • Subscribe to the SPIE Digital Library
  • Create a FREE account to sign up for Digital Library content alerts and gain access to institutional subscriptions remotely.
Access This Article
Sign in or Create a personal account to Buy this article ($20 for members, $25 for non-members).
Access This Proceeding
Sign in or Create a personal account to Buy this article ($15 for members, $18 for non-members).
Access This Chapter

Access to SPIE eBooks is limited to subscribing institutions and is not available as part of a personal subscription. Print or electronic versions of individual SPIE books may be purchased via SPIE.org.