0
Research Papers

Delivery of midinfrared (6 to 7-μm) laser radiation in a liquid environment using infrared-transmitting optical fibers

[+] Author Affiliations
Mark A. Mackanos, E. Duco Jansen

Department of Biomedical Engineering, Vanderbilt University, Nashville, TennesseeE-mail: duco.jansen@vanderbilt.edu

Brandon L. Shaw, Jas S. Sanghera, Ishwar Aggarwal

Naval Research Laboratory, Washington, D.C.

Abraham Katzir

School of Physics and Astronomy, Tel Aviv University, Tel Aviv, Israel

J. Biomed. Opt. 8(4), 583-593 (Oct 01, 2003). doi:10.1117/1.1606460
History: Received Aug. 14, 2002; Revised Mar. 3, 2003; Accepted Apr. 4, 2003; Online October 15, 2003
Text Size: A A A

Ablation at wavelengths near λ=6.45μm results in tissue ablation with minimal collateral damage (<40μm) yet yields a high ablation rate that is useful for human surgery. However, delivery of this wavelength has been limited to that in air and thus to applications in which the target tissue can be readily exposed. The goal of this study is to investigate the potential of a pulsed infrared laser at λ=6.45μm for noncontact ablation in a liquid environment. To this end we investigated fiber delivery in combination with the use of infrared transparent liquids. Transmission characteristics and damage thresholds for two types of fiber materials (silver halide and arsenic sulfide), for high-power pulsed laser radiation were determined using the Mark III free electron laser. Both fibers had comparable bulk losses (0.54 dB/m and 0.62 dB/m, respectively) while the arsenic sulfide fibers showed more coupling losses (37 versus 27%). Damage thresholds were higher in arsenic sulfide fibers than in silver halide fibers (1.12GW/cm2 versus 0.54GW/cm2), but both fibers were sufficient to deliver radiant exposures well above the ablation threshold in tissue. Seven different perfluorocarbon liquids (PFCLs), known for their transparency at λ=2.94μm, were investigated and their optical transmission was determined using Fourier transform infrared and direct Beer’s law measurements. All of the PFCLs tested had similar values for an absorption coefficient μa at a given wavelength (μa=0.05mm1 at λ=2.94μm and μa3mm1 at λ=6.45 μm). Pump-probe imaging showed the ablation sequence (λ=6.45μm) at the fiber tip in a water environment, which revealed a fast expanding and collapsing bubble. In contrast, the volatile PF-5060 showed no fast bubble expansion and collapse, but rather formation of nontransient gas bubbles. Perfluorodecalin did not show any bubble formation at the radiant exposures used. It was shown that using the λ=6.45 μm wavelength delivered via fiber optics in combination with perfluorodecalin allows a noncontact laser surgical procedure. Deeper structures, however, are effectively shielded because the radiant exposure of the beam will fall below the ablation threshold owing to the absorption by perfluorodecalin. This may optimize the efficacy and safety of laser-based vitreoretinal surgery. © 2003 Society of Photo-Optical Instrumentation Engineers.

Figures in this Article
© 2003 Society of Photo-Optical Instrumentation Engineers

Citation

Mark A. Mackanos ; Brandon L. Shaw ; Abraham Katzir ; Jas S. Sanghera ; E. Duco Jansen, et al.
"Delivery of midinfrared (6 to 7-μm) laser radiation in a liquid environment using infrared-transmitting optical fibers", J. Biomed. Opt. 8(4), 583-593 (Oct 01, 2003). ; http://dx.doi.org/10.1117/1.1606460


Access This Article
Sign In to Access Full Content
Please Wait... Processing your request... Please Wait.
Sign in or Create a personal account to Buy this article ($20 for members, $25 for non-members).
 
Your Session has timed out. Please sign back in to continue.
Sign In to Access Full Content

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Related Book Chapters

Topic Collections

PubMed Articles
Advertisement

Buy this article ($18 for members, $25 for non-members).
Sign In