1 January 2006 Tissue-like phantoms for near-infrared fluorescence imaging system assessment and the training of surgeons
Alec M. Degrand, Stephen J. Lomnes, Deborah S. Lee, Matthew Pietrzykowski, Shunsuke Ohnishi, Timothy Morgan, Andrew Gogbashian, Rita G. Laurence, John V. Frangioni M.D.
Author Affiliations +
Abstract
We demonstrate how to construct calibrated, stable, and inexpensive tissue-like phantoms for near-IR (NIR) fluorescence imaging applications. The bulk phantom material is composed of gelatin, intralipid, hemoglobin, and indocyanine green (ICG). Absorbance, scatter, background fluorescence, and texture can be tuned as desired. NIR fluorescent inclusions are comprised of ICG-labeled polystyrene divinylbenzene beads and Pam78-labeled hydroxyapatite crystals. The former mimic tumor masses of controllable size and contrast agent concentration, and the latter mimic microcalcifications in breast cancer. NIR-fluorescent inclusions can be positioned precisely in phantoms, with one or more regions having different optical properties, and their position can be verified independently using microcomputed tomography. We demonstrate how these phantoms can be used to calibrate and compare imaging systems, and to train surgeons to operate under NIR fluorescence image guidance.

1.

Introduction

Living tissue has relatively low photon absorption in the near-IR (NIR) range of the electromagnetic spectrum (700 to 1000nm ). Endogenous fluorophores are strong contributors to autofluorescence in the UV and visible regions of the spectrum, but are significantly weaker in the NIR. Hence, from an absorbance and autofluorescence standpoint, NIR light has distinct advantages. Scatter is significant at all wavelengths, and is a complex function of tissue composition (discussed in Ref. 1). However, for many tissues, the absolute value of scatter tends to be lower in the NIR.

Recently, there has been a dramatic rise in the publication of imaging systems that exploit NIR light, and in some cases, exogenous NIR fluorophores. The three basic types of imaging systems are steady state (also known as continuous wave or cw), time domain, and frequency domain. Within each, source/detector geometry can be based on point or area illumination, and point or area detection, leading to over 12 different imaging system configurations (reviewed in Ref. 2).

Paralleling the explosion of imaging systems is a similar one for exogenous NIR fluorophores. Several comprehensive reviews of NIR fluorophores used for in vivo imaging were recently published.3, 4, 5 Briefly, NIR fluorescent contrast agents are now available for quantitation of vascular mapping, tissue perfusion, tissue calcification, protease activity, cell injury, tissue response to injury, and for specific applications such as sentinel lymph node mapping and tumor imaging.

As imaging systems and contrast agents converge, there is an increasing need for standardized NIR fluorescent phantoms that can be compared among laboratories, and that assist with the training of surgeons and technologists. Cubeddu 6 described agar-based phantoms utilizing intralipid for scattering and India ink for absorption. Wagnieres 7 described agarose-based phantoms containing silica powder and intralipid for scattering and India ink and erythrocyte concentrate for absorption. The former optical phantom utilizes nonphysiological absorbers, the latter is difficult to construct, and both utilize a galactose polymer as base material. Quan 8 and Kelly 9 utilized gelatin derived from the hydrolysis of vertebrate connective tissue as a phantom base material for photoacoustic and radiation dosimetry studies, respectively. Two recent studies10, 11 utilized gelatin-based phantoms for NIR fluorescence studies; however, neither describes the construction of precise NIR fluorescent inclusions or a general-purpose phantom platform. Other approaches to NIR imaging phantoms include hard resin or soft room temperature vulcanizing (RTV) silicon materials12 into which holes are drilled; however, this system is inflexible with respect to inclusion size, placement, and geometry, and multiple interfaces are formed at each hole site. In this paper, we demonstrate how to construct simple, yet stable, gelatin-based phantoms that can be tailored to the optical properties of various tissues using readily available materials. We also describe two different types of NIR fluorescent inclusions that simulate tumor spheroids and/or breast cancer microcalcifications.

2.

Materials and Methods

2.1.

Reagents

National Formulary (NF) grade gelatin (catalog #EM-GX0048-1) was from VWR (West Chester, Pennsylvania). Sterile IntralipidTM ( 20% w/v) was purchased from Baxter (Deerfield, Illinois). Bovine hemoglobin (catalog #H-2625) and indocyanine green (ICG; catalog #I-2633) were from Sigma (St. Louis, Missouri). AG1-X8 anion exchange resin (catalog #140-1422) was from Bio-Rad (Hercules, California). Pam78 was prepared as described (see Zaheer 13). The 100μm hydroxyapatite (HA) crystals (catalog #391947) were from Calbiochem (La Jolla, California).

2.2.

Phantom Mixture Construction

The buffer for all phantoms was Tris-buffered saline (TBS) composed of 50mM Tris, pH 7.4, and 150mM sodium chloride. Sodium azide was added to a final concentration of 0.1% ( 15mM ; TBS/azide) to bind methemoglobin, block reoxygenation, and to act as a preservative. Gelatin was suspended at the desired concentration (% w/v) in TBS (GTS), and warmed to 37 to 50°C with constant stirring until completely dissolved. GTS was then cooled to 25°C and the desired amounts of hemoglobin and Intralipid added, with constant stirring. Solidification of the GTS mixture should be performed as rapidly as possible to avoid separation of the Intralipid. We found that pouring the room temperature solution into the desired prechilled mold at 4°C will minimize this effect. Depending on the size and shape of the mold, solidification should take 5 to 20min .

2.3.

NIR Fluorescent Inclusions

NIR fluorescent inclusions were made using ICG-labeled polystyrene divinylbenzene beads (AG1-X8) or Pam78-labeled hydroxyapatite crystals. These latter inclusions were used to simulate microcalcifications found in breast cancer.14 The concentration of fluorophore associated with each inclusion was measured precisely, and expressed as “ICG equivalence” (defined in the following). Beads were sieved twice to a diameter of 1000±100μm using a 20-mesh ( 850μm square holes) brass sieve (VWR catalog #57334-112). For bead loading, 10mM ICG in DMSO was diluted in TBS to twice the desired loading concentration and mixed 1:1 with a 50% bead slurry, vortexed for 3s , incubated 20min , and washed 4× with TBS. Pam78 was loaded onto HA in a similar fashion. For dual-labeling with iodine contrast and ICG, 60% iothalamate meglumine (stock concentration 2.22M iodine; ConrayTM , Mallinckrodt, St. Louis, Missouri) was diluted in TBS to a final iodine concentration of 88.8mM prior to the addition of ICG and bead loading. Prior to embedding into phantoms, NIR fluorescent inclusions were calibrated against ICG standards in DMSO having the same path length to determine their “ICG equivalence” (see later).

Fluorescent inclusions, having ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) from 0 to 5μM , could be varied in size from 75 to 1000μm . The 3-D positioning of fluorescent inclusions was verified independently using an eXplore Locus microcomputed tomography system (GE Healthcare Biosciences, Waukesha, Wisconsin) to within ±90μm .

2.4.

Phantom Construction

For phantoms with uniform bulk optical properties, a mold is filled and gelled to the desired level of the inclusion, then the inclusions are mixed with a 30°C phantom solution (5 to 100μL depending on inclusion) and pipetted onto the surface of the solidified mold. The warmth of the solution softens the top of the mold and allows the inclusion solution to adhere to the surface. Once the inclusion solution has set, more of the same warm phantom mix can be poured on top, thus removing any interfaces. Different levels of the same phantom can have different inclusions inserted into it without having to create separate molds for each level (Fig. 1 ). Moreover, various regions of the same phantom can have different properties, without artificial refractive interfaces.

Fig. 1

Tissue-like phantoms for NIR fluorescence imaging. By choosing the scatter, absorbance, and background fluorescence of the phantom medium, simple (left) or complex (right) phantoms can be created. In complex phantoms, optical properties can be varied in three dimensions, including the use of gradients (e.g., Gel3 ), without refractive interfaces. For optical tomography, probe holes are predrilled and sealed before filling, or the container is coated with vegetable oil for easy removal of the entire phantom. Fluorescent inclusions can be placed at any position in 3-D within the phantom.

014007_1_035601jbo1.jpg

Phantoms with different optical properties in different compartments can be produced by placing smaller molds inside the larger main mold. These smaller submolds can then be filled with different phantom mixtures, with the inclusions inserted into these molds in a similar manner as already described. Alternatively, different compartments can be constructed separately, then added to the main mold using warm gelatin as already described for inclusions. The remaining gaps are filled in using the desired warm phantom mixture. To minimize glare in color images due to uneven surfaces, complex phantoms are often covered with an additional 1- to 3-mm layer of GTS, which has negligible effect on scatter and absorption. Once a phantom has been constructed, it can be stored for long periods in a sealed container, preferably a humidified one, at 4°C .

2.5.

Optical Measurements

Absorbance spectrometry was performed using a 1-cm path length quartz cuvette (Starna, Atascadero, California), HR2000 fiber optic spectrometer and a DH2000-BAL light source (Ocean Optics, Dunedin, Florida). Spectral reflectance measurements were made using the same spectrometer and light source equipped with a R400-7-VIS/NIR probe (Ocean Optics), and normalization was performed using a homogeneously reflecting surface.

Absolute absorption and reduced scattering were measured on a Lambda 19 (Perkin Elmer, Boston, Massachusetts) UV/Vis/NIR spectrophotometer equipped with a 150-mm integrating sphere. For long-term stability experiments, phantoms were measured in this apparatus over a 4-week period, being stored at 4°C in a humidified chamber on intervening days. To assess diffusion during long-term stability testing, samples were prepared between two 2- ×3in. glass slides, and were composed of a central 0.5- ×1.385in. rectangle (area of measurement) of one material with the remaining periphery composed of a second material. The edges were sealed with epoxy. The scattering and absorbing properties of samples were computed using the inverse adding-doubling method15 from 400 to 1000nm (the algorithm is available for download at http://omlc.ogi.edu/software/iad/).

We compared the results provided by the integrating sphere and inverse adding-doubling method to Mie theory using samples of known properties. We purchased 903nm polystyrene (PS) beads (Catalog #3900A) from Duke Scientific (Palo Alto, California). The Mie theory calculator was accessed at http://omlc.ogi.edu/calc/mie_calc.html. Inputs to the calculator included an index of refraction of 1.33; real indices of refraction of spheres of 1.6, 1.582, 1.575, and 1.57 for 400 , 600 , 800 , and 1000nm , respectively; imaginary indices of refraction of 0.00045,0.0008,0.001 , and 0.002 for 400, 600, 800, and 1000nm , respectively; 200 angles; and, a density of 0.0073spheresμm3 (Ref. 16).

Fluorescence spectrometry was performed with a 225μL , three-sided quartz cuvette (Starna), HR2000 fiber optic spectrometer and CUV-ALL-UV four-way cuvette holder (Ocean Optics), and 770nm NIR laser diode light source (Electro Optical Components, Santa Rosa, California) set to 8mW with output through a 300μm core diameter, numerical aperture (NA) 0.22 fiber (Fiberguide Industries, Stirling, New Jersey).

Intraoperative NIR fluorescence imaging was performed as described in detail previously,17 using filtered (Chroma #HHQ750/50x) 725- to 775nm NIR excitation light, and custom light collection optics equipped with a long-pass (795nm) emission filter (Chroma #HQ795LP).

2.6.

Quantification of Autofluorescence in Living Tissues

Animal protocols were in accordance with Institutional Animal Care and Use Committee Guidelines. Thirty-five-kilogram adult male Yorkshire pigs (n=3) were purchased from E.M. Parsons and Sons (Hadley, Massachusetts). All animals acclimated to the animal facility for at least 48h prior to experimentation. Pig anesthesia was induced with 4.4mgkg intramuscular Telazol (Fort Dodge Labs, Fort Dodge, Iowa) and anesthesia maintained through a 7-mm endotracheal tube with 1.5% isoflurane/98.5% O2 at 5Lmin .

Tissues and organs were surgically exposed, then imaged using NIR excitation light of 725 to 775nm , with an excitation fluence rate of 5mWcm2 (as measured with a calibrated Orion-TH digital light power meter equipped with an Orion model 2A-SH thermopile detector head). In each field of view, and on nonabsorbing and nonreflecting black matte paper (Canson Mi-Teintes, Stygian Black #425), were placed 1mm path length ICG standards in DMSO of 0 , 31 , 63 , 125 , 250 , 500 , 1000 , and 2000nM , and calibration curves were created for camera exposure times of 10 , 34 , 67 , 134 , 500 and 1000ms . From these six curves, the average autofluorescence of the tissue or organ under study was calculated, and expressed as ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) .

After each study, anesthetized animals are euthanized by rapid intravenous injection of 10mL of Fatal-Plus (Vortech Pharmaceuticals, Dearborn, Michigan). This method of euthanasia is consistent with the recommendations of the Panel on Euthanasia of the American Veterinary Medical Association.

3.

Results

3.1.

Preparation of Phantoms

Phantoms were prepared in covered plastic food storage containers of the desired size and geometry. The basic building block is a gelatin-containing medium having the desired absorbance, scatter, and background fluorescence. Absorbance of the medium is controlled by the concentration of oxy- and deoxyhemoglobin (the ratio of which can be adjusted using azide), and/or other biomolecules. Scatter is controlled by the concentration of Intralipid, with virtually no contribution from the gelatin, even at high concentrations.

3.2.

Characterization of Phantom Optical Properties

Optical properties of the phantoms were measured using an integrating sphere as described in Sec. 2. Validation of the inverse adding-doubling algorithm employed was performed by comparing results of Mie scattering theory with calibrated polystyrene beads of defined optical properties. As shown in Fig. 2A , theoretical and empirical values for μA,μS , and anisotropy were closely correlated over the wavelength range 400 to 1000nm . Optical properties of the various phantom components over this wavelength range are shown in Fig. 2B (various concentrations of intralipid in 10% GTS) and Fig. 2C (complete phantom). Note that our values for μS of 1% Intralipid in GTS are approximately twofold lower than previously published values for 1% Intralipid in water, which likely reflects lot-to-lot variation in particle size. We also found unexpectedly high scatter at 400nm in the presence of concentrated Hb [Fig. 2C], possibly due to Rayleigh scatter from the Hb molecules themselves. Absorbance of the phantom is controlled by the hemoglobin concentration, with a negligible contribution by gelatin at blue wavelengths [Fig. 2D].

Fig. 2

Optical characterization of phantom components. (A) Validation of the employed inverse adding-doubling method by comparing measurements using polystyrene (PS) beads of defined optical properties (PS beads) to Mie scattering theory (Mie theory). Shown are μA (left), μS (middle), and anisotropy (right). (B) Optical properties of various Intralipid concentrations as measured using an integrating sphere and the inverse adding-doubling method. Shown are μA (left), μS (middle), and anisotropy (right). (C) Optical properties of a typical phantom mixture composed of 1% Intralipid and 170μM Hb in 10% GTS. This composition most closely matches scattering human tissue with an 8% blood volume. Shown are μA (left), μS (middle), and anisotropy (right). (D) Absorbance of various gelatin concentrations in TBS/azide (left), and Hb in 10% GTS (right; diluted 1:100 in TBS/azide prior to measurement). (E) Absorbance (left) and “background” fluorescence (ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) , described in text; right) of ICG in 10% GTS. (F) Homogeneity of a 1% Intralipid170μM Hb phantom in 10% GTS measured in 5mm increments in the X (left) and Y (directions), over a total distance of 10cm . Shown are the mean ±SEM (standard error of the mean) of the 20 reflectance spectra at each wavelength from 400 to 1000nm .

014007_1_035601jbo2.jpg

Varying the concentration of ICG in the medium controls background fluorescence, and also adds a NIR component to absorption [Fig. 2E]. Background fluorescence could be varied from 0 to 1.4μM ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) by varying ICG concentration from 0 to 1μM . The enhancement of ICG quantum yield after adsorption to proteins, such as gelatin, is well known.18

The absorbance, scatter, and background fluorescence of the phantom can be varied in three dimensions; hence, complex phantoms having any desired optical properties and geometry can be constructed inexpensively and easily using this system (Fig. 1; see also Fig. 6 in Sec. 3.8). Using the same base material for all compartments minimizes refractive interfaces. Interestingly, phantoms can be remelted, NIR fluorescent inclusions retrieved, and all materials recycled into new phantoms as needed.

Fig. 6

Complex phantom for the calibration and assessment of a NIR fluorescence imaging system, and independent confirmation by microcomputed tomography. (A) Schematic design of a complex phantom containing four compartments with optical properties shown in the table. Each compartment contained a 1mm diameter bead of 880nM ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) and 88.8mM ConrayTM placed at the desired depth. (B) Simultaneous color video/ NIR fluorescence imaging (top view) of the phantom using a NIR excitation fluence rate of 5mWcm2 and 100ms exposure time. Shown are color video (left), NIR fluorescence (middle), and a pseudocolored (lime green) merge of the two (right). All pixel values were within the linear range of the NIR camera. Below each NIR fluorescent bead (arrows) is shown its SBR. (C) Confirmation of precise position of NIR fluorescent inclusions using microcomputed tomography. Shown are the maximum intensity projections (MIPs) for top view and side view, along with a 3-D rendering of the phantom with beads in yellow.

014007_1_035601jbo6.jpg

3.3.

Characterization of Phantom Homogeneity

Phantom homogeneity was tested by taking 20 full-spectrum (400 to 1000-nm) reflectance measurements each in the X and Y directions, at 5-mm intervals. As shown in Fig. 2F, variation over this 10-cm distance, in either direction, was less than 2%.

3.4.

Characterization of Phantom Stability

The stability of the phantom over time was measured at 400, 600, 800, and 1000nm over a 4-week period. Shown in Fig. 3A is the variation in transmission, and in Fig. 3B, the variation in reflectance. At wavelengths from 600 to 1000nm , there was virtually no change in optical properties of the phantom components. However, at 400nm and starting at approximately 2 weeks, there was evidence of Hb transition from the oxy-, deoxy-, and possibly met-Hb states, as well as evidence of diffusion of Hb within the phantom. Although these results should have no impact on NIR fluorescence experiments, use of multicompartment phantoms for measuring Hb concentration over periods greater than 2 weeks must be performed with these results in mind.

Fig. 3

Long-term stability of tissue-like phantoms and their NIR fluorescent inclusions. (A) Transmission over a 4-week interval for 1% Intralipid in 10% GTS (top left), 1% Intralipid in a central location with Hb around the periphery, both in 10% GTS (top right), 170μM Hb in 10% GTS (bottom left), and 170μM Hb in a central location with 1% Intralipid around the periphery, both in 10% GTS (bottom right). (B) Reflectance over a 4-week interval for the samples described in (A). (C) Stability of NIR fluorescent inclusions. NIR fluorescent beads and HA inclusions, each with a 1μM ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) , were placed randomly throughout a phantom (1% Intralipid and 170μM Hb in 10% GTS) at a depth of 5mm , and measurements were taken weekly. Between measurements, the phantom was sealed tightly in a storage container at 4°C . On the ordinate is signal-to-background ratio (SBR) expressed as mean ± SEM.

014007_1_035601jbo3.jpg

3.5.

Concept of “ICG Equivalence”

To solve the problem of extrapolating results from one embedded fluorescent inclusion to another, and from one phantom to another, we report all fluorescence emissions in terms of “ICG equivalence,” that is, the measured fluorescence emission over a defined wavelength range, excited with NIR light of a defined wavelength and fluence rate, which is equivalent to an ICG standard in DMSO of a defined path length and concentration. For simplicity, the following notation is employed: ICG equivalenceExcitation Wavelength(s), Fluence Rate, Emission Wavelength(s), Path Length.. DMSO is one of the few solvents in which ICG is stable,19 and for which an absolute quantum yield (13%) has been reported.20 Hence, by expressing measured fluorescence as ICG equivalence, the performance of virtually any other NIR fluorophore, with similar excitation/emission wavelengths, can be extrapolated. For example, IRDye78 (LI-COR, Lincoln, Nebraska), an NIR fluorophore commonly used by our laboratory, has a quantum yield of 14% in aqueous buffer,21 hence fluorescent inclusion results expressed as ICG Equivalence would be roughly equivalent to the expected performance of IRDye78.

3.6.

Autofluorescence of Living Tissue

Living tissue has a low, but measurable, autofluorescence, which must be known before construction of a phantom with the proper amount of “background” fluorescence. Figure 4 shows the autofluorescence of virtually every tissue and organ in the body, expressed as ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) . The path length of the ICG standards in DMSO used to make these measurements must be specified since the reflectance imaging system employed in these studies is sampling an unknown depth of tissue. Since fluorescence emission often scales nonlinearly with inclusion size, it is best to match the inclusion size to the typical tumor size of the model system employed. The values in Fig. 4 can be used to add the appropriate level of phantom background fluorescence for the tissue type under study. This is especially important since the relative autofluorescence of different tissues varies by over ninefold.

Fig. 4

Relative and absolute NIR autofluorescence of living tissues. ICG equivalence of major tissue types of the pig. Results shown are the average of three separate animals (mean ± SEM), and are expressed as ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) (nM).

014007_1_035601jbo4.jpg

3.7.

NIR Fluorescent Inclusions

Similar to the submillimeter radioscintigraphic calibration sources previously reported by our group,22 the beads used in this study are available in wet bead diameters from 45 to 1180μm , which can be sieved to any desired size, and are also available with cross-linking ranging from 2 to 8%. The latter value is important, since it determines the exclusion limit of the bead interior. The 8% bead used in this study does not exclude 775-Da ICG, resulting in relatively homogeneous fluorescence over the entire volume of the bead. Bead loading with ICG is distinctly nonlinear (Fig. 5 ), with stacking and quenching limiting the ICG Equivalence(725775nm,5mWcm2,> 795nm,1mm) per bead to approximately 4.2μM . Due to high viscosity, ConrayTM iodine contrast is diluted to 88.8mM prior to bead loading. As expected, iodine contrast shifts the ICG loading curve shown in Fig. 5 to the right (data not shown), emphasizing again the importance of confirming the ICG equivalence of each fluorescent inclusion prior to placement in the phantom.

Fig. 5

Characterization of NIR fluorescent inclusions. (A) 0.79μL ICG standards in DMSO. (B) ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) [ μM ; mean ± SD (standard deviation)] of individual 1-mm (0.79μL) ICG-loaded anion exchange beads expressed as a function of loading concentration. (C) ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) ( μM ; mean ± SD) of individual 100μm HA crystals loaded with Pam78 and calibrated against 100-μm path length ICG standards in DMSO. (D) 100μm Pam78-labeled HA crystals loaded with a 1μM ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) as viewed on a NIR fluorescence microscope.

014007_1_035601jbo5.jpg

A second NIR fluorescent inclusion used in this study is HA crystals loaded with Pam78, a NIR fluorescent derivative of the bisphosphonate pamidronate.13 Pam78 has been used previously for in vivo imaging of sites of osteoblastic activity and the simulated calcification of breast cancer.23 Pam78-loaded HA is an ideal fluorescent inclusion for tissue-like phantoms since it has biologic and medical relevance, and can be loaded with physiologically relevant concentrations of fluorophore (Fig. 5).

The stability of the NIR fluorescent inclusions was tested over 4 weeks with no significant changes in fluorescence emission found [Fig. 3C]. Given the stability of the base material and NIR fluorescent inclusions, phantoms could theoretically be shipped to any other laboratory for calibration or comparative assessment of different imaging systems.

3.8.

Imaging System Calibration and Performance Assessment

One of the major uses for these phantoms is in assessing the performance of tomographic and reflectance imaging systems. This is illustrated in Fig. 6 using a NIR fluorescence imaging system designed for large animal and human studies.17 The complex phantom shown is particularly relevant to in vivo conditions, where nearby tissues have distinctly different optical properties. The SBR, as well as visual discrimination of the inclusions, could be measured as a function of fluence rate, system optics, and camera integration time, and the precise 3-D position of inclusions measured using microcomputed tomography [Fig. 6C]. Of note, radiodense HA crystals require no additional contrast to be colocalized by microcomputed tomography (data not shown).

3.9.

Training in Imaging System Operation and Image-Guided Surgery Using NIR Fluorescent Phantoms

We previously described an intraoperative NIR fluorescence reflectance imaging system that enables the surgeon to view surgical anatomy (via color video) and tumor or lymph node location (via NIR fluorescence) simultaneously, in real-time, and without moving parts17, 21 [see also Fig. 6B]. As with any new imaging system, there is a learning curve associated with use of system software, and the surgeon must also be trained at NIR fluorescence-guided dissection, resection, and postresection inspection of the surgical field. The phantoms described in this paper may prove useful for such training and should permit quantitative assessment of surgical expertise.

Figure 7 shows a breast-shaped phantom with optical properties similar to the human breast. Embedded within the breast are nine 1-mm NIR fluorescent beads, corresponding to a tumor of approximately 8×105 cells (discussed later in the paper). Each bead has an ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) of 1μM , corresponding to approximately 6×105 fluorophore molecules per cell. Given the close index matching between inclusions and phantom, it is virtually impossible to find inclusions using the naked eye or color video enhancement. Using the NIR fluorescence channel, however, the inclusions are easily found (presurgery), and the surgeon can then resect the desired inclusion (resection) and inspect the field to ensure completeness (postresection).

Fig. 7

Training in image-guided surgery using NIR fluorescent phantoms. Nine 1-mm beads (each approximating a collection of 8×105 cells; arrows) with a 1μM ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) were placed concentrically, and 0.5cm below the surface, of a breast-shaped phantom (see oblique view). The phantom composition was 10% gelatin, 1% Intralipid, and 17μM hemoglobin, corresponding to μA=0.056mm1 , μS=0.310mm1 , and g=0.968 at 800nm . Shown is NIR fluorescence image-guided resection of one (red arrow) of the otherwise invisible beads using a scalpel and forceps. NIR fluorescence images have identical exposure times (100ms) and normalizations.

014007_1_035601jbo7.jpg

4.

Discussion

In this paper, we demonstrate how to construct simple, inexpensive phantoms for NIR fluorescence imaging applications. The significant advantages of these phantoms include complete control over geometry and optical properties, the ability to create both simple and complex conditions, the ability to independently confirm the 3-D position of fluorescent inclusions using computed tomography, the ability to incorporate any pigmented biomolecule (e.g., hemoglobin), high stability over time in the NIR, no artificial interface between volumes having different optical properties or between fluorescent inclusions and medium, and the representation of fluorescence emission as “ICG equivalence.” Moreover, these phantoms should be compatible with tomographic imaging systems having either circular24, 25, 26 or planar27 geometry (detailed in Fig. 1).

Interestingly, cross-linking of the phantoms for various periods of time (typically 1 to 2h ) with various concentrations of glutaraldehyde (0.1 to 1%) enables fine-tuning of their mechanical properties, with more highly cross-linked phantoms remarkably similar to living tissue when cut with a scalpel. Glutaraldehyde presumably cross-links both the buffer (Tris) and protein components of the phantom, and of course, requires remeasurement of all optical properties. Glutaraldehyde does not appear to affect significantly NIR fluorescence emission from either bead/ICG or HA/Pam78 inclusions (data not shown).

The fluorescent inclusions described in this paper were chosen for their relevance to important medical applications. For tumor cell receptor imaging, an abundant receptor will be present at 100,000 to 600,000 copies on the cell surface. Assuming a cell volume of 1pL , and no endocytosis, the maximum concentration of contrast agent achievable is 160nm to 1μM . As demonstrated in Fig. 5, the anion exchange beads used in this study can be loaded with ICG equivalence throughout this physiologic range. The beads are also available in volumes ranging from 45pL to 820nL , corresponding to roughly 45 to 820,000 cells. Pam78-loaded HA also has significant biologic relevance in light of recent reports that the malignant calcification of breast cancer is composed of HA, and not other calcium salts,14 and since the ICG equivalence(725775nm,5mWcm2,> 795nm,1mm) of Pam78-loaded HA is within the range obtained during in vivo imaging.13, 23 When combined with our results from tissue autofluorescence measurements, and published values for tissue optical properties,28 one should be able to use these phantoms to estimate, prior to in vivo experimentation, the performance of virtually any molecularly targeted NIR fluorophore having similar excitation/emission wavelengths as ICG in DMSO, embedded in virtually any simulated tissue of interest.

We previously described a simultaneous color video/NIR fluorescence reflectance imaging system for intraoperative use and demonstrated its utility in image-guided sentinel lymph node mapping, 29, 30, 31, 32, 33 optical angiography,17, 21 and tumor resection (discussed in Ref. 17). Others have described optical tomography systems for detecting embedded NIR fluorescent targets in the human breast (reviewed in Ref. 34). The translation of NIR fluorescence technology to the clinic will require the training of medical technologists (e.g., for tomographic breast imaging) and surgeons (e.g., for intraoperative reflectance imaging systems). Such training should be greatly facilitated by the tissue-like phantoms described in this paper. As optical imaging approaches clinical use, inexpensive training systems that do not require animal or human subjects during the steepest phase of the learning curve will be particularly helpful.

Acknowledgments

We thank Barbara L. Clough for editing and Grisel Rivera for administrative assistance. This work was supported by Department of Energy (Office of Biological and Environmental Research) Grant No. DE-FG02-01ER63188, a Clinical Scientist Development Award from the Doris Duke Charitable Foundation (nonanimal experiments), an Application Development Award from the Center for Integration of Medicine and Innovative Technology (CIMIT) and Grants No. R01-CA-115296 and No. R01-CA-110185 from the National Cancer Institute of the National Institutes of Health to JVF.

References

1. 

Y. T. Lim, S. Kim, A. Nakayama, N. E. Stott, M. G. Bawendi, and J. V. Frangioni, “Selection of quantum dot wavelengths for biomedical assays and imaging,” Molecular Imaging, 2 50 –64 (2003) Google Scholar

2. 

A. H. Hielscher, A. Y. Bluestone, G. S. Abdoulaev, A. D. Klose, J. Lasker, M. Stewart, U. Netz, and J. Beuthan, “Near-infrared diffuse optical tomography,” Dis. Markers, 18 313 –337 (2002). 0278-0240 Google Scholar

3. 

E. M. Sevick-Muraca, J. P. Houston, and M. Gurfinkel, “Fluorescence-enhanced, near infrared diagnostic imaging with contrast agents,” Curr. Opin. Chem. Biol., 6 642 –650 (2002). https://doi.org/10.1016/S1367-5931(02)00356-3 1367-5931 Google Scholar

4. 

V. Ntziachristos, C. Bremer, and R. Weissleder, “Fluorescence imaging with near-infrared light: new technological advances that enable in vivo molecular imaging,” Eur. Radiol., 13 195 –208 (2003). 0938-7994 Google Scholar

5. 

J. V. Frangioni, “In vivo near-infrared fluorescence imaging,” Curr. Opin. Chem. Biol., 7 626 –634 (2003). https://doi.org/10.1016/j.cbpa.2003.08.007 1367-5931 Google Scholar

6. 

R. Cubeddu, A. Pifferi, P. Taroni, A. Torricelli, and G. Valentini, “A solid tissue phantom for photon migration studies,” Phys. Med. Biol., 42 1971 –1979 (1997). https://doi.org/10.1088/0031-9155/42/10/011 0031-9155 Google Scholar

7. 

G. Wagnieres, S. Cheng, M. Zellweger, N. Utke, D. Braichotte, J. P. Ballini, and H. van den Bergh, “An optical phantom with tissue-like properties in the visible for use in PDT and fluorescence spectroscopy,” Phys. Med. Biol., 42 1415 –1426 (1997). https://doi.org/10.1088/0031-9155/42/7/014 0031-9155 Google Scholar

8. 

K. M. Quan, G. B. Christison, H. A. MacKenzie, and P. Hodgson, “Glucose determination by a pulsed photoacoustic technique: an experimental study using a gelatin-based tissue phantom,” Phys. Med. Biol., 38 1911 –1922 (1993). https://doi.org/10.1088/0031-9155/38/12/014 0031-9155 Google Scholar

9. 

R. G. Kelly, K. J. Jordan, and J. J. Battista, “Optical CT reconstruction of 3D dose distributions using the ferrous-benzoic-xylenol (FBX) gel dosimeter,” Med. Phys., 25 1741 –1750 (1998). https://doi.org/10.1118/1.598356 0094-2405 Google Scholar

10. 

C. A. Giller, H. Liu, P. Gurnani, S. Victor, U. Yazdani, and D. C. German, “Validation of a near-infrared probe for detection of thin intracranial white matter structures,” J. Neurosurg., 98 1299 –1306 (2003). 0022-3085 Google Scholar

11. 

T. Boehm, A. Hochmuth, A. Malich, J. R. Reichenbach, M. Fleck, and W. A. Kaiser, “Contrast-enhanced near-infrared laser mammography with a prototype breast scanner: feasibility study with tissue phantoms and preliminary results of imaging experimental tumors,” Invest. Radiol., 36 573 –581 (2001). https://doi.org/10.1097/00004424-200110000-00002 0020-9996 Google Scholar

12. 

S. Jiang, B. W. Pogue, T. O. McBride, and K. D. Paulsen, “Quantitative analysis of near-infrared tomography: sensitivity to the tissue-simulating precalibration phantom,” J. Biomed. Opt., 8 308 –315 (2003). https://doi.org/10.1117/1.1559692 1083-3668 Google Scholar

13. 

A. Zaheer, R. E. Lenkinski, A. Mahmood, A. G. Jones, L. C. Cantley, and J. V. Frangioni, “In vivo near-infrared fluorescence imaging of osteoblastic activity,” Nat. Biotechnol., 19 1148 –1154 (2001). https://doi.org/10.1038/nbt1201-1148 1087-0156 Google Scholar

14. 

A. S. Haka, K. E. Shafer-Peltier, M. Fitzmaurice, J. Crowe, R. R. Dasari, and M. S. Feld, “Identifying microcalcifications in benign and malignant breast lesions by probing differences in their chemical composition using Raman spectroscopy,” Cancer Res., 62 5375 –5380 (2002). 0008-5472 Google Scholar

15. 

S. A. Prahl, M. J. C. van Gemert, and A. J. Welch, “Determining the optical properties of turbid media by using the adding-doubling method,” Appl. Opt., 32 559 –568 (1993). 0003-6935 Google Scholar

16. 

X. Ma, J. Q. Lu, R. S. Brockv, K. M. Jacobs, P. Yang, and X. H. Hu, “Determination of complex refractive index of polystyrene microspheres from 370 to 1610nm,” Phys. Med. Biol., 48 4165 –4172 (2003). https://doi.org/10.1088/0031-9155/48/24/013 0031-9155 Google Scholar

17. 

A. M. De Grand and J. V. Frangioni, “An operational near-infrared fluorescence imaging system prototype for large animal surgery,” Technol. Cancer Res. Treat., 2 553 –562 (2003). 1533-0346 Google Scholar

18. 

S. Ohnishi, S. J. Lomnes, R. G. Laurence, A. Gogbashian, G. Mariani, and J. V. Frangioni, “Organic alternatives to quantum dots for intraoperative near-infrared fluorescent sentinel lymph node mapping,” Molecular Imaging, 4 (3), 172 –181 (2005). 1535-3508 Google Scholar

19. 

R. Rajagopalan, P. Uetrecht, J. E. Bugaj, S. A. Achilefu, and R. B. Dorshow, “Stabilization of the optical tracer agent indocyanine green using noncovalent interactions,” Photochem. Photobiol., 71 347 –350 (2000). https://doi.org/10.1562/0031-8655(2000)071<0347:SOTOTA>2.0.CO;2 0031-8655 Google Scholar

20. 

C. Benson and H. A. Kues, “Absorption and fluorescence properties of cyanine dyes,” J. Chem. Eng. Data, 22 379 –383 (1977). https://doi.org/10.1021/je60075a020 0021-9568 Google Scholar

21. 

A. Nakayama, F. del Monte, R. J. Hajjar, and J. V. Frangioni, “Functional near-infrared fluorescence imaging for cardiac surgery and targeted gene therapy,” Molecular Imaging, 1 365 –377 (2002) Google Scholar

22. 

J. R. English, R. Accorsi, J. D. Idoine, J. A. Parker, J. T. Renze, R. C. Lanza, and J. V. Frangioni, “Sub-millimeter technetium-99m calibration sources,” Mol. Imaging Biol., 4 380 –384 (2002) Google Scholar

23. 

R. E. Lenkinski, M. Ahmed, A. Zaheer, J. V. Frangioni, and S. N. Goldberg, “Near-infrared fluorescence imaging of microcalcification in an animal model of breast cancer,” Acad. Radiol., 10 1159 –1164 (2003). 1076-6332 Google Scholar

24. 

X. Intes, J. Ripoll, Y. Chen, S. Nioka, A. G. Yodh, and B. Chance, “In vivo continuous-wave optical breast imaging enhanced with indocyanine green,” Med. Phys., 30 1039 –1047 (2003). https://doi.org/10.1118/1.1573791 0094-2405 Google Scholar

25. 

H. Dehghani, B. W. Pogue, S. P. Poplack, and K. D. Paulsen, “Multiwavelength three-dimensional near-infrared tomography of the breast: initial simulation, phantom, and clinical results,” Appl. Opt., 42 135 –145 (2003). 0003-6935 Google Scholar

26. 

N. Iftimia, X. Gu, Y. Xu, and H. Jiang, “A compact, parallel-detection diffuse optical mammography system,” Rev. Sci. Instrum., 74 2836 –2842 (2003). https://doi.org/10.1063/1.1568558 0034-6748 Google Scholar

27. 

J. S. Reynolds, T. L. Troy, and E. M. Sevick-Muraca, “Multipixel techniques for frequency-domain photon migration imaging,” Biotechnol. Prog., 13 669 –680 (1997). https://doi.org/10.1021/bp970085g 8756-7938 Google Scholar

28. 

W. F. Cheong, S. A. Prahl, and A. J. Welch, “A review of the optical properties of biological tissues,” IEEE J. Quantum Electron., 26 2166 –2195 (1990). https://doi.org/10.1109/3.64354 0018-9197 Google Scholar

29. 

S. Kim, Y. T. Lim, E. G. Soltesz, A. M. De Grand, J. Lee, A. Nakayama, J. A. Parker, T. Mihaljevic, R. G. Laurence, D. M. Dor, L. H. Cohn, M. G. Bawendi, and J. V. Frangioni, “Near-infrared fluorescent type II quantum dots for sentinel lymph node mapping,” Nat. Biotechnol., 22 93 –97 (2004). https://doi.org/10.1038/nbt920 1087-0156 Google Scholar

30. 

C. P. Parungo, S. Ohnishi, A. M. De Grand, R. G. Laurence, E. G. Soltesz, Y. L. Colson, P. M. Kang, T. Mihaljevic, L. H. Cohn, and J. V. Frangioni, “In vivo optical imaging of pleural space drainage to lymph nodes of prognostic significance,” Ann. Surg. Oncol., 11 1085 –1092 (2004). 1068-9265 Google Scholar

31. 

C. P. Parungo, Y. L. Colson, S. W. Kim, S. Kim, L. H. Cohn, M. G. Bawendi, and J. V. Frangioni, “Sentinel lymph node mapping of the pleural space,” Chest, 127 1799 –1804 (2005). 0012-3692 Google Scholar

32. 

C. P. Parungo, S. Ohnishi, S. W. Kim, S. Kim, R. G. Laurence, E. G. Soltesz, F. Y. Chen, Y. L. Colson, L. H. Cohn, M. G. Bawendi, and J. V. Frangioni, “Intraoperative identification of esophageal sentinel lymph nodes with near-infrared fluorescence imaging,” J. Thorac. Cardiovasc. Surg., 129 844 –850 (2005). https://doi.org/10.1016/j.jtcvs.2004.08.001 0022-5223 Google Scholar

33. 

E. G. Soltesz, S. Kim, R. G. Laurence, A. M. DeGrand, C. P. Parungo, D. M. Dor, L. H. Cohn, M. G. Bawendi, J. V. Frangioni, and T. Mihaljevic, “Intraoperative sentinel lymph node mapping of the lung using near-infrared fluorescent quantum dots,” Ann. Thorac. Surg., 79 269 –277 (2005). 0003-4975 Google Scholar

34. 

V. Ntziachristos and B. Chance, “Probing physiology and molecular function using optical imaging: applications to breast cancer,” Breast Cancer Res., 3 41 –46 (2001) Google Scholar
©(2006) Society of Photo-Optical Instrumentation Engineers (SPIE)
Alec M. Degrand, Stephen J. Lomnes, Deborah S. Lee, Matthew Pietrzykowski, Shunsuke Ohnishi, Timothy Morgan, Andrew Gogbashian, Rita G. Laurence, and John V. Frangioni M.D. "Tissue-like phantoms for near-infrared fluorescence imaging system assessment and the training of surgeons," Journal of Biomedical Optics 11(1), 014007 (1 January 2006). https://doi.org/10.1117/1.2170579
Published: 1 January 2006
Lens.org Logo
CITATIONS
Cited by 127 scholarly publications and 9 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Near infrared

Luminescence

Imaging systems

Tissues

Optical properties

Tissue optics

Calibration