1 May 2010 Fibrin phantom for use in optical coherence tomography
Brendan F. Kennedy, Susanne Loitsch, Robert A. McLaughlin, Loretta Scolaro, Paul Rigby, David D. Sampson
Author Affiliations +
Abstract
This work presents a novel tissue-mimicking phantom for use in a range of optical coherence tomography (OCT) experiments. Such phantoms are critical in the development and assessment of new OCT techniques, but no previously published phantoms have become universally accepted. We present the first description of a phantom based on a fibrin matrix, which improves key attributes of previously published methods. It provides a biocompatible, optically transparent scaffold in which to incorporate organic and/or inorganic optical scattering materials. Its fabrication time is markedly shorter than many common phantoms, and its lifetime is longer than other biocompatible phantoms. The potential of fibrin phantoms incorporating IntralipidTM to introduce uniform optical scattering is demonstrated. The measured attenuation coefficient as a function of Intralipid concentration confirms the ability to control optical scattering. A bilayer phantom with distinct optical scattering in each layer is also presented.

1.

Introduction

An important aspect of research in optical coherence tomography (OCT) is the validation of new systems, novel techniques, and theoretical predictions using phantoms with optical properties that closely approximate those of tissue.1 Numerous such phantoms have been proposed;1, 2, 3 however, issues exist with each of these phantoms and no previously published phantom has become universally accepted within the OCT community. OCT phantoms generally comprise a bulk matrix, which provides a scaffold, and additives, which provide optical absorption and scattering. For many experiments, it is important to use a matrix that allows for the incorporation of the natural constituents of tissue, such as blood and fat, to closely approximate tissue optical properties.1

Room-temperature vulcanizing (RTV) silicone, epoxy resin, and polyvinyl alcohol (PVA) gels are commonly used matrix materials.1, 2, 3 However, these materials are not compatible with biological materials; therefore, inorganic scatterers must be used, typically white metal-oxide powders or microspheres. White metal-oxide powders, such as aluminum oxide (Al2O3) and titanium dioxide (TiO2) , have the disadvantage that their refractive index is much higher than that of tissue, resulting in phantoms with overall optical properties very different to those of tissue. An alternative is the more expensive dry silica microspheres, which have low size variations and a refractive index closer to tissue. However, it is difficult to achieve homogenously scattering phantoms with these particles due to aggregation.2 For PVA gels, scattering is deliberately introduced into the matrix material by performing a series of freeze/thaw cycles.3 However, PVA gels are not biologically compatible. Phantoms based on biologically compatible gelatin and agarose hydrogels have been used as phantom materials.1 Scattering from the gelatin matrix has been reported, which is undesirable as it limits the degree to which scattering is controlled by additive concentration. Such gelatin and agarose phantoms typically have short lifetimes of less than one week, and do not maintain their rigidity at room temperature.1

In this work, we present novel, biocompatible phantoms based on a low-scattering matrix material that improves key attributes associated with commonly used phantom materials, including time-efficient fabrication, longer lifetime, rigidity at room temperature, and low scattering. We define biocompatible as a material that is suitable for the addition of biological materials and substances such as the constituents of tissue. The matrix material used is fibrin, which is employed as an adhesive in surgical procedures and wound closures.4, 5 Fibrin is a naturally occurring protein in humans that provides structural support for blood clots. It is formed from the protein fibrinogen by proteolysis induced by the enzyme thrombin. To demonstrate fibrin’s potential, we used it to fabricate phantoms with optical scattering coefficients similar to those of tissue by incorporating it with Intralipid. The main advantages of Intralipid are its well-known optical properties6 and the similarity of its microparticles to lipid cell membranes and organelles that constitute the scattering sources in tissue.1 The result is a homogeneous, uniformly scattering phantom, with a scattering coefficient that can be controlled by the concentration of Intralipid. Single and bilayer phantoms were fabricated and imaged using OCT. This paper provides the first description of fibrin-based OCT phantoms, detailing both fabrication techniques and the resulting optical properties.

2.

Methods

Fibrin phantoms were fabricated using a mixture of fibrinogen and thrombin, both extracted from bovine plasma (Sigma-Aldrich, Germany). A quantity of 0.8mg of thrombin powder was dissolved in 1ml of saline, to which 40μmol of calcium chloride (CaCl2) was added. CaCl2 reduces the clotting time and increases the elasticity of fibrin gels.4 Separately, 70mg of fibrinogen powder (65% to 85% protein) was dissolved in 1ml of saline and mixed with a suitable volume of standard 20% Intralipid, where 20% signifies the percentage weight per volume (w/v) of soy bean oil. A 500-ml volume of 20% Intralipid contained 100g of soybean oil, 6g of egg lecithin, 11g of glycerol, and water. The fibrinogen/Intralipid and thrombin/ CaCl2 solutions were mixed in a 3:2 volume ratio. The solution was set in a cylindrical mold of diameter 10mm and thickness 3mm . After 10-min clotting time, the fibrin phantom was placed in saline and stored in the refrigerator. This procedure resulted in a homogeneous distribution of optical scatterers, without observable aggregation. Bilayer phantoms were fabricated by pouring fibrin into the base of the mold in two steps. By controlling the volume of solution added to the mold, a thin layer (with 1% w/v Intralipid) was fabricated, followed by a thicker layer (with 13% w/v Intralipid) added 10min later.

3.

Experiment

Scanning electron microscope (SEM) images of a single-layer fibrin phantom were obtained. To record the images, a 1-mm -thick section of fibrin phantom was prepared by freeze drying in liquid nitrogen, and a 6-nm platinum surface coating was applied. The magnification was 74K× and the accelerating voltage was 5kV . In addition, confocal fluorescence microscopy images of a similar phantom were obtained. The phantom was eosin stained for 10min to introduce fluorescent optical contrast, then rinsed in water for 15min and imaged at a magnification of 60× .

To quantify the attenuation coefficient μt in fibrin phantoms, we used the technique described in Ref. 7. The sample reflectance R(z) measured from the OCT signal is proportional to the product of the confocal point-spread function (PSF) with the single-scattering model for light attenuation in tissue7:

Eq. 1

R(z)1(zzcf/zR)2+1exp(2μtz),
where z represents physical path length, and z=0 corresponds to the sample interface, zcf is the position of the confocal gate, zR is the apparent Rayleigh length used to characterize the point-spread function (PSF), and μt represents the attenuation coefficient. The confocal PSF was measured experimentally by moving a reflector through the focus of the sample beam and recording the detector output. As R(z) and the confocal PSF were measured experimentally, Eq. 1 could be used to extract μt from measured OCT A-scans.

The attenuation coefficient of fibrin phantoms was measured using a time-domain OCT (TD-OCT) system. A broadband source ( λ=1310nm , Δλ=154nm , axial resolution 6.4μm ) with an optical power of 21mW was employed. The theoretical lateral resolution was 11.2μm .

3-D datasets (XYZ) were recorded with 512×600×2048pixels . The reflectance R(z) was extracted from the OCT signal by calibration from a series of reflectance standards. To reduce the effect of speckle on the calculation of μt , regions of 20pixels in the Y-Z plane were averaged. A swept-source OCT system (Thorlabs, Newton, New Jersey) was used to image the bilayer fibrin phantom. A broadband source ( λ=1300nm , Δλ=100nm , axial resolution 12μm ) with an optical power of 10mW was employed. The theoretical lateral resolution was 15μm .

4.

Results and Discussion

A scanning electron microscope image of a fibrin phantom, without Intralipid, is shown in Fig. 1 . During the clotting process, the fibrin fibers aggregate and branch to form a 3-D matrix. This fibrous structure is shown in Fig. 1. The nanoscale of the structure results in a sufficiently low optical scattering cross section for the phantom to be optically transparent at wavelengths used for OCT. Figure 1 shows a confocal fluorescence microscope image of a fibrin phantom prepared with 19% w/v Intralipid. Fig. 1 shows the homogeneous nature of the fluorescence. The air-phantom interface is visible in the top-left. The dark, largely circular features present throughout the image and ranging in size from 0.5to5μm are believed to be lipid droplets, which appear dark due to the absence of eosin.

Fig. 1

(a) SEM image of fibrin matrix; scale bar represents 100nm . (b) Confocal fluorescence microscope image of fibrin phantom containing Intralipid; scale bar represents 30μm .

030507_1_013003jbo1.jpg

The measured values of μt for phantoms with varying concentrations of Intralipid are presented in Fig. 2 . For each concentration, measurements were recorded at intervals of 600μm and averaged. The error bars represent one standard deviation (SD), calculated from five data points. A monotonic increase in scattering was measured for increasing Intralipid concentration. The values of μt presented in Fig. 2 are within the range of previously measured values for both tissue8 and Intralipid.6 In the inset of Fig. 2, the reflectance is plotted as a function of z . The slope (red line) is proportional to μt . The gradual increase in reflectance over the first 50μm is attributed to diffusion of index-matching glycerol into the phantom near the boundary. The measured values of μt were stable for a period of approximately two weeks. After this, leaching of the Intralipid resulted in a reduction in μt . However, the fibrin matrix remained intact for more than one month.

Fig. 2

Attenuation coefficient μt measured for phantoms with different % w/v concentrations of Intralipid. The inset shows the log-averaged OCT reflectance versus z , and the linear fit of attenuation (red line) for a phantom with 13% Intralipid. (Color online only.)

030507_1_013003jbo2.jpg

An OCT B-scan of a bilayer fibrin phantom is presented in Fig. 3 . Higher scattering in the thicker layer is visible. The values of μt measured using the TD-OCT system in the thin and thick layers were 1.7mm1 , SD 0.1mm1 , and 3.3mm1 , SD 0.2mm1 , respectively. Comparing these values with Fig. 2, μt in the thin layer (Intralipid concentration 1%) is higher than expected, while μt in the thick layer (Intralipid concentration 13%) is lower than expected. We speculate that these differences are due to leaching of Intralipid from the thick layer to the thin layer during fabrication. This leaching could potentially be reduced by fabricating the phantoms at lower temperature. We subsequently observed a constant μt from both layers over a two-week period. We verified the integrity of the phantom over time by visual inspection. The homogeneity of each layer is visible in Fig. 3. The small inhomogeneities visible in the first layer are caused by small variations in Intralipid concentration resulting from imperfect mixing. The interface between layers is located at a depth in optical path length of approximately 370μm .

Fig. 3

Bilayer fibrin phantom, with different Intralipid concentration in both layers. The scale bar represents dimensions 300×300μm .

030507_1_013003jbo3.jpg

The fibrin phantoms presented in this work are suitable for use in a wide range of experiments in OCT. The fabrication time for the phantom was approximately 30min . In comparison, bilayer silicone phantoms fabricated in our laboratory require up to 24-h fabrication time.9 The demonstrated homogeneity of fibrin phantoms will be important, for example, in studies of speckle in OCT, and measurement and use of the OCT scattering coefficient.7, 10 Investigations of the scattering properties of tissue constituents could be performed, and is the subject of ongoing work. Fibrin phantoms have the further advantage that they are pliable and exhibit a viscoelastic response that is in the range of biological tissues. It has been reported that the elastic modulus of fibrin can be changed by varying the thrombin concentration.5 Thus, fibrin phantoms could be useful in OCT elastography studies.9, 11 It has further been reported that the viscosity of fibrin can be varied by increasing the fibrinogen concentration,5 potentially mimicking the viscous properties of tissue. The ability to match a phantom’s mechanical properties to those of tissue is important for interstitial OCT measurements and/or procedures in solid tissues.12 Tumor-mimicking features can readily be incorporated into the fibrin matrix for such experiments.

5.

Conclusions

Novel fibrin-based tissue-mimicking phantoms for use in OCT are presented. Fibrin phantoms overcome problems that exist in other biocompatible phantoms, such as scattering from the matrix, loss of rigidity at room temperature, and a short lifetime, and have a shorter fabrication time than commonly used phantoms. The fibrin matrix is optically transparent and compatible with both organic and inorganic optical scatterers. We demonstrate that fibrin phantoms incorporating Intralipid are homogeneous, provide controllable optical attenuation, and can readily be fabricated into two layers with distinct optical properties. These phantoms are well suited to a range of OCT experiments, in particular, those requiring homogeneous scattering, the addition of biological scatterers, and potentially those requiring viscoelastic properties matched to those of tissue. Fibrin phantoms have the potential to be adopted in a wide range of OCT applications.

Acknowledgments

The authors acknowledge the facilities, and scientific and technical assistance (in particular Peter Duncan) of the Australian Microscopy and Microanalysis Research Facility at the Centre for Microscopy, Characterisation and Analysis, The University of Western Australia, a facility funded by The University, State, and Commonwealth Governments.

References

1. 

B. W. Pogue and M. S. Patterson, “Review of tissue simulating phantoms for optical spectroscopy, imaging and dosimetry,” J. Biomed. Opt., 11 (4), 041102 (2006). https://doi.org/10.1117/1.2335429 1083-3668 Google Scholar

2. 

C. E. Bisaillon, G. Lamouche, R. Maciejko, M. Dufour, and J. P. Monchalin, “Deformable and durable phantoms with controlled density of scatterers,” Phys. Med. Biol., 53 237 –247 (2008). https://doi.org/10.1088/0031-9155/53/13/N01 0031-9155 Google Scholar

3. 

C. U. Devi, R. M. Vasu, and A. K. Sood, “Design, fabrication, and characterization of a tissue-equivalent phantom for optical elastography,” J. Biomed. Opt., 10 (4), 1 –10 (2005). https://doi.org/10.1117/1.2003833 1083-3668 Google Scholar

4. 

D. S. Feldman, and D. H. Sierra, “Tissue adhesives in wound healing,” Encyclopedic Handbook of Biomaterials and Bioengineering, Part A: Materials, 1347 –1384 Marcel Dekker, New York (1995). Google Scholar

5. 

H. Kaetsu, T. Uchida, and N. Shinya, “Increased effectiveness of fibrin sealant with a higher fibrin concentration,” Int. J. Adhes. Adhes., 20 27 –31 (2000). https://doi.org/10.1016/S0143-7496(99)00011-1 0143-7496 Google Scholar

6. 

H. J. van Staveren, C. J. M. Moes, J. van Marle, S. A. Prahl, and M. J. C. van Gemert, “Light scattering in intralipid-10% in the wavelength range of 4001100nm,” Appl. Opt., 30 (31), 4507 –4514 (1991). https://doi.org/10.1364/AO.30.004507 0003-6935 Google Scholar

7. 

D. J. Faber, F. J. Van der Meer, and M. C. G. Aalders, “Quantitative measurements of attenuation coefficients of weakly scattering media using optical coherence tomography,” Opt. Express, 12 (19), 4353 –4365 (2004). https://doi.org/10.1364/OPEX.12.004353 1094-4087 Google Scholar

8. 

V. Tuchin, Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnosis, (2007) Google Scholar

9. 

B. F. Kennedy, T. R. Hillman, R. A. McLaughlin, B. C. Quirk, and D. D. Sampson, Opt. Express, 17 (24), 21762 –21772 (2009). https://doi.org/10.1364/OE.17.021762 1094-4087 Google Scholar

10. 

R. A. McLaughlin, L. Scolaro, P. Robbins, C. Saunders, S. L. Jacques, and D. D. Sampson, “Mapping tissue optical attenuation to identify cancer using optical coherence tomography,” Lect. Notes Comput. Sci., 5762 657 –664 (2009). https://doi.org/10.1007/978-3-642-04271-3_80 0302-9743 Google Scholar

11. 

J. M. Schmitt, “OCT elastography: imaging microscopic deformation and strain of tissue,” Opt. Express, 3 (6), 199 –211 (1998). https://doi.org/10.1364/OE.3.000199 1094-4087 Google Scholar

12. 

X. Li, C. Chudoba, T. Ko, C. Pitris, and J. G. Fujimoto, “Imaging needle for optical coherence tomography,” Opt. Lett., 25 (20), 1520 –1522 (2000). https://doi.org/10.1364/OL.25.001520 0146-9592 Google Scholar
©(2010) Society of Photo-Optical Instrumentation Engineers (SPIE)
Brendan F. Kennedy, Susanne Loitsch, Robert A. McLaughlin, Loretta Scolaro, Paul Rigby, and David D. Sampson "Fibrin phantom for use in optical coherence tomography," Journal of Biomedical Optics 15(3), 030507 (1 May 2010). https://doi.org/10.1117/1.3427249
Published: 1 May 2010
Lens.org Logo
CITATIONS
Cited by 36 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

Scattering

Tissue optics

Signal attenuation

Tissues

Confocal microscopy

Optical properties

Back to Top