7 February 2012 Noise characteristics of heterodyne/homodyne frequency-domain measurements
DongYel Kang, Matthew A. Kupinski
Author Affiliations +
Abstract
We theoretically develop and experimentally validate the noise characteristics of heterodyne and/or homodyne measurements that are widely used in frequency-domain diffusive imaging. The mean and covariance of the modulated heterodyne output are derived by adapting the random amplification of a temporal point process. A multinomial selection rule is applied to the result of the temporal noise analysis to additionally model the spatial distribution of intensified photons measured by a charge-coupled device (CCD), which shows that the photon detection efficiency of CCD pixels plays an important role in the noise property of detected photons. The approach of using a multinomial probability law is validated from experimental results. Also, experimentally measured characteristics of means and variances of homodyne outputs are in agreement with the developed theory. The developed noise model can be applied to all photon amplification processes.

1.

Introduction

When an intensity-modulated beam is used to illuminate diffusive media, the modulation amplitude and phase of the beam are attenuated and delayed, respectively, due to absorption and scattering inside the medium. The amplitude attenuation is caused by both scattering and absorption, whereas the phase delay is mainly related to scattering. Because the typical modulation frequency range in frequency-domain diffusive imaging is from tens of megahertz up to a gigahertz,13 a very high bandwidth detector is required to correctly measure modulation amplitudes and phases. In additional to measuring the frequency-response information, it has been shown that obtaining large datasets using detector arrays also improves diffusive imaging performance.4,5 It is almost impossible to achieve the high bandwidths required in frequency-domain measurements with high-density detector arrays such as charge-coupled device (CCD) or complementary metal-oxide-semiconductor (CMOS) sensors. Heterodyne and/or homodyne (hetero/homodyne) methods extract the amplitude and phase information by converting the rapidly modulating output beam into a beat-frequency modulation mode using gain-modulated amplifiers, such as an image intensifier (IIN) and photo multiplier (PMT).6,7 When the gain modulation frequency is slightly different from that of a light source, the amplified output modulates at the frequency difference—a beat frequency, which is a heterodyne measurement. For a homodyne measurement, where a gain is modulated at the same frequency as a modulation source, the output is stable in time, but its irradiance is varied along the heterodyne output according to relative phases between source and gain modulations. Although time-domain measurements can also extract the frequency response of diffusive media, hetero/homodyne measurements are more widely used in diffusive imaging because of their relatively low cost and simplicity.1,6

In general, measuring intensity-modulated outputs improves the reconstruction algorithms and other algorithms designed to estimate tissue properties, when compared to measuring just diffused intensities (i.e., steady-state measurements).1 For example, diffuse optical tomography (DOT), which is considered a noninvasive and cost-effective medical imaging tool, can tomographically reconstruct images of scattering and absorption coefficients up to tens of millimeters in depth.2,3 DOT systems based on just a diffused intensity are simple and inexpensive but have significant disadvantages including the measurement variability caused by skin structure, hair, and surface roughness.1 Furthermore, the measured diffused intensity generally cannot reconstruct a unique solution, and the crosstalk between absorption and scattering coefficients cannot be effectively separated in reconstruction processes.8,9 A frequency-domain DOT can remedy these problems and thus improves detectability and reconstruction performance. Frequency-domain measurements are also required in fluorescent lifetime imaging (FLI), where the fluorophore lifetime is used as an important indicator of tissue characteristics.10 Since the typical fluorescent lifetimes are a few nanoseconds and cannot be measured directly, measuring attenuated amplitude and delayed phase of modulated fluorescent photons can be used effectively to estimate the lifetime.7,1113 Measuring the frequency-domain information in large tissue areas using a CCD and a gain-modulated IIN can be used to generate a spatial fluorescent lifetime map of the tissue.14

Noise is inherent in all measurements. Even though averaging multiple measurements or increasing detector exposure time can reduce the amount of noise, these approaches are prohibited or limited in some dynamic-imaging applications and clinical situations.1 In medical imaging, data are acquired for specific purposes or tasks that include classification, such as detection of abnormalities like a tumor, and estimation of biological quantities, such as a fluorescent lifetime. Noise-contaminated data is a limiting factor in an observer’s ability to perform these classification and/or estimation tasks. The noise is especially significant in diffusive medical imaging, such as DOT and FLI, because of low photon counts.1217 The performance of noise-contaminated systems can be evaluated quantitatively by considering figures of merit (FOMs) that are chosen to reflect an observer’s ability to perform these medial-relevant tasks.15,18 Task-based assessment of image quality evaluates and/or optimizes diffusive imaging systems by employing FOMs that assess observer performance.1820 Many task-based FOMs rely on the first and second moments of output data,18,21,22 so understanding the noise property of output data is important and necessary.

Because of the importance of noise analysis in diffusive imaging, the noise characteristics of a modulation amplitude and phase for frequency-domain diffusive imaging have been previously investigated. Toronov et al.23 mathematically derived variances of these quantities in frequency-domain DOT assuming a quantum shot noise model on the hetero/homodyne outputs. Conventional signal-to-noise ratios (SNRs) and a task-based FOM known as a Hotelling observer SNR were derived based on Toronov’s result, and an optimal modulation frequency was explored in DOT systems of a single point source and CCD-based phased array.16,2325 In FLI, a FOM called F-value is commonly used to evaluate system performance, where the F-value is defined by first- and second-order statistics of a fluorescent lifetime.12,13 Many investigators have optimized FLI systems by maximizing the F-value or minimally required fluorescent photon numbers for a meaningful F-number.12,13,26 It is reasonable that the F-value is strongly dependent on the noise characteristics of modulation amplitude and/or phase in frequency-domain FLI because the fluorescent lifetime is estimated from these quantities. For all previous studies including Toronov’s result, however, Gaussian or shot noise is simply considered for the statistics of modulation amplitude and phase or only shot noise is assumed for the hetero/homodyne outputs,9,23 which might not be valid in reality. For example, the irradiance of hetero/homodyne outputs is usually high due to the photon amplification process, which is enough to ignore the effect of shot noise in a detector. Furthermore, the beat-frequency modulated output might be temporally correlated, which is seriously harmful to the assumption of quantum shot noise.

The noise properties of amplified photons for the non-modulation case have already been investigated, where incident photons are assumed to be Poisson distributed. The photon amplification in IINs and intensified CCDs (ICCDs) should be considered a random process even for a constant gain, which generally induces non-Poisson outputs for Poisson inputs.18 Bell theoretically analyzed a noise figure for IINs with a constant gain,27 where the noise figure was defined as the ratio of SNRs of input and output photons. He considered photons incident on a micro channel plate (MCP) in IINs or ICCDs as governed by Poisson statistics, and each photoelectron collision on the wall of the MCP generates the photoelectron distribution called Furry distribution. Moran et al.28 theoretically derived a noise factor (NF) of ICCDs, which is defined as the square of the noise figure. He showed that variations of a NF and an output SNR are not linearly related to MCP gain, which is partially observed in experiments elsewhere.17,29 Moran adapted a Polya distribution as a single photoelectron pulse in a MCP. Instead of photoelectron distributions in a MCP, Apanasovich and Novikov30 analyzed the amplification process considering electron multiplication events in a MCP as random point processes and compared the theoretical model with experiments. Frenkel et al.31 investigated variance and noise-equivalent irradiance of outputs in an ICCD using the approach of spatial point processes. Although these previous noise analyses for randomly amplified photons are mathematically sound, to the author’s knowledge, noise analysis has not been performed for heterodyne/homodyne measurements where both incident beam and amplification gain are modulating. Spring and Clegg32 briefly touch on the SNR of homodyne outputs, but gain modulation is simply substituted for the constant gain in the conventional SNR expression for PMTs.

In this paper, we theoretically derive mean and covariance properties of hetero/homodyne outputs, which are expressed with systematic parameters of hetero/homodyne measurements, such as dc and ac parts of incident primary modulation beams and modulation gain. The theoretical development includes both temporal and spatial analyses. Temporally, a hetero/homodyne measurement process is considered random amplification of a modulated temporal point process. Spatially, a multinomial selection rule is applied to model the spatial distribution of the amplified photons from MCPs to detector array pixels. Resulting from this, it is found that hetero/homodyne outputs are temporally uncorrelated and the variance oscillates like the mean in some specific conditions. We suggest experimental results of homodyne measurements that show similar characteristics with what is expected from the developed theory. Discussions for the influence of the derived noise property on frequency-domain diffusive imaging are also presented.

2.

Theory for Temporal Noise Analysis

For hetero/homodyne detection processes in diffusive imaging, the diffused photons exiting an object (e.g., phantom) are captured by an imaging system, where the exit face is imaged onto a cathode of an IIN through the imaging system. The incident photons to the cathode generate photoelectrons that are injected into the MCP of the IIN, where the voltage difference between the cathode and the MCP determines the amount of injected photoelectrons. A MCP is typically composed of millions of small channels that are related to the overall resolution of an IIN. A single photoelectron collision that occurred on the wall of the MCP generates multiple photoelectrons, the number of which is proportional to the voltage applied to the MCP. The overall gain of the IIN is determined mainly by photoelectrons generated from the cathode and the MCP voltage that are independent of a cathode voltage. To achieve gain modulation in an IIN, a modulated voltage is applied to a cathode to generate modulated photoelectrons that are amplified in a MCP. The multiplied photoelectrons from the MCP are accelerated to an anode (phosphor) to generate intensified photons. The usual decay time of phosphors ranges from hundreds of nanoseconds to a few milliseconds.33 The photons discharged from the anode are transferred to a detector array, such as a CCD or CMOS, by an appropriate imaging system. Overall, a phantom exit face and a CCD detector surface are object and image planes, respectively, of the whole frequency-domain measurement system. Resulting from this view, each CCD detector pixel is conjugate to a detection area on the exit face, the size of which is determined by the CCD pixel size and the transverse magnification of the whole hetero/homodyne imaging system.

It is assumed that hetero/homodyne output photons are described by a temporal point process as

Eq. (1)

y(t)=n=1Nk=1knδ(ttnΔtnk),
where tn is the time that n’th primary photon (diffused photon from a phantom) is absorbed on a cathode of the IIN and Δtnk is the time of delay for secondary photons produced from an anode of the IIN by the n’th primary. In Eq. (1), N and kn are numbers of primaries and secondaries, respectively; these are all random variables. The mean and covariance of the point process of Eq. (1) have been derived for the case of non-modulation input and IIN gain.18,34 Referring to this known procedure, we will develop the mean and covariance of y(t) for hetero/homodyne measurements. A mean photon arrival rate from a phantom to a cathode of the IIN is expressed as

Eq. (2)

ap(t)=cp[1+mpcos(ωpt+ψ)],
where ωp=2πfp and cp, mp, and ψ indicate dc photon rate, modulation depth that is less than 1, and modulation phase, respectively. The term fp is a modulation frequency that is usually hundreds of megahertz in frequency-domain diffusive imaging, as mentioned in the introduction. Although cp, mp, and ψ are continuously varied across the phantom exit face or the cathode of the IIN, it is assumed that the amounts of variation are insignificant within the detection area conjugate to a CCD pixel, so they can be considered constant as in Eq. (2). If we assume that Poisson postulates are fulfilled for the diffused photons on the detection area, the probability density function (PDF) for the primary is18

Eq. (3)

prpri(tn|ap)=ap(tn)N¯(ap),
where

Eq. (4)

N¯(ap)=0Tap(t)dt.
Notice that the mean total number N¯(ap) also modulates because of the modulating photon arrival rate, ap(t). Without losing the generality, we assume the mean photon rate in Eq. (2) is not random, which indicates that noise external to the hetero/homodyne system, such as random biological motions and respiration, will not be considered in this theoretical development.

The shift-variant temporal point spread function (PSF) of the IIN with gain modulation can be described as

Eq. (5)

pd(t,tn)=cg[1+g=1mgcos(ωgtn+φg)]paI(t,tn)pmI(tn,tn)dtn=cg[1+g=1mgcos(ωgtn+φg)]pdI(t,tn),
where paI(t,tn) and pmI(tn,tn) are temporal PSFs of the anode and MCP of the IIN, respectively. The term c0[1+g=1mgcos(ωgtn+φg)] in Eq. (5) indicates a modulated photoelectron rate from the cathode for a single primary, which is equivalent to cathode gain modulation. The term ω1 is the same as or close to ωp in hetero/homodyne measurements. Because the modulation voltage should be applied to around the cutoff voltage of a cathode, the modulated cathode gain is generally similar to a bottom-truncated sinusoidal function rather than a perfect sinusoidal curve.32 It is noted that for this case, the largest coefficient m1 in the cathode modulation could be larger than 1 because c0 is defined by averaging the modulated cathode gain. The periodic cathode gain is expanded to Fourier cosine terms,32 as shown in Eq. (5), where the amplitude and phase of each term is indicated as c0mg and φg, respectively. From the assumption that the gain is an even and periodic function, it can be easily shown that there is no phase difference between cosine terms (we set all φg as zeros for simplicity) and ωg=gω1, where g is an integer. The usual photocathode response time of a cathode is picoseconds to femtoseconds,35 so the photoelectron conversion time on a cathode is ignored in Eq. (5). The time for multiplying photoelectrons in the MCP is also negligible compared with the photoelectron-photon conversion time on an anode, which is usually a few milliseconds. The random variable t for the secondary in Eq. (5) is not independent of each other because t is correlated to tn. If t in Eq. (5) is replaced by Δtnk as in Eq. (1), Δtnk can be considered statistically independent and Poisson postulates are satisfied for secondaries produced from each primary.18 With this replacement, the probability density function (PDF) for secondary photons can be written from Eq. (5) as

Eq. (6)

prΔt(Δtnk|tn)=[k¯n(tn)]1pd(Δtnk+tn,tn),
where the mean number of secondary photons produced by a primary is

Eq. (7)

k¯n(tn)=pd(t,tn)dt=c0[1+g=1mgcos(ωgtn)]pdI(t,tn)dt=c0[1+g=1mgcos(ωgtn)]k¯nI(tn).
In Eq. (7), k¯nI(tn) indicates the mean number of amplified photons for a single photoelectron injected into the MCP. The time dependence in k¯nI(tn) indicates a long-term MCP and/or anode variation caused by, for example, thermal instability. We are interested in the noise property of a hetero/homodyne output that is stationary in time, thus assuming k¯nI(tn)k¯nI. Also, photon and photoelectron losses, such as cathode quantum efficiency and an open area ratio of the MCP,27 could be included in c0 in Eq. (7).

Mean and covariance of Eq. (1) can be derived by statistically averaging the time for secondary Δtnk, the number of secondaries kn, the time for primary tn, and the number of primaries N(ap). As a final step, averaging ap(t) would be necessary if the primary is doubly stochastic, but this is not the case as we assumed before. As shown in the appendix, the mean and covariance are derived from Eqs. (2) to (7) as

Eq. (8)

y(t)=cpc0[1+mpcos(ωptn+ψ)][1+g=1mgcos(ωgtn+φg)]pdI(t,tn)dtn,
and

Eq. (9)

Ky(t,t)={k¯n2(tn)+Var[kn|tn]}ap(tn)prΔt(t,tn)prΔt(t,tn)dtn,
respectively, where Var[kn|tn] in Eq. (9) indicates variance of secondary photons produced by a single primary photon. Applying trigonometric identities to the integrand in Eq. (8) generates dc, ωp, ωg, Δωpg=ωpωg, and other high-frequency terms. Except the dc and Δωp1 (i.e., the integer g=1) terms, all other modulation terms are averaged out because the decay time of pdI(t,tn) in Eq. (5) is much larger than modulation periods of these terms. Also, the typical beat frequency for heterodyne measurements is much shorter than 2π/Δωp1, so pdI(t,tn)k¯nI(tn)δ(ttn) for the dc and Δωp1 terms. With these descriptions, Eq. (8) is simplified to

Eq. (10)

y(t)cpc0[1+mpm12cos(Δωp1tn+ψ)]k¯nI(tn)δ(ttn)dtn=cpc0k¯nI[1+mpm12cos(Δωp1t+ψ)].
The mean of Eq. (10) is physically reasonable and in agreement with conventionally experienced hetero/homodyne outputs. For the case of homodyne, where Δωp1=0, the phase difference ψ changes to measure the trajectory of the beat-frequency modulated output. The unit of Eq. (10) is the same as cp, a photon rate (i.e., units of c0 and k¯nI are photonelectron/primary and secondary/photoelectron, respectively), so multiplying a detector exposure time to Eq. (10) measures the mean output photons.

We define the Fano factor,18 F, for secondaries per primary as

Eq. (11)

F=Var[kn]k¯n,
which indicates the degree of random amplification of an IIN. It is commonly accepted that F is always greater than 1 in a random amplification process because F<1 means the secondary photons are sub-Poisson for Poisson primaries. Higher F means the amplified output photons are more randomly variable, so measured intensified outputs are less reliable. Considering F of Eq. (11) makes the integrand of Var[kn|tn] in Eq. (9) the same as Eq. (10) with the multiplication of F. The remaining mean square term in Eq. (9) is further developed by substituting Eq. (7) to k¯n. Removing ωg and higher modulation frequency terms, the integrand of k¯n2(tn) results in

Eq. (12)

cp(c0k¯nI)2[1+g=1g=1mgmgcos(ωgtn)cos(ωgtn)+2mpcos(ωptn+ψ)g=1mgcos(ωgtn)+mpcos(ωptn+ψ)g=1g=1mgmgcos(ωgtn)cos(ωgtn)].
Likewise, in the procedure for calculating the mean, only terms containing Δωp1=ωpω1, where ω1=ωg±ωg±1 in Eq. (12), are survived because of the relatively long decay time of prΔt(t) in Eq. (9). Therefore, the covariance of Eq. (9) becomes

Eq. (13)

Ky{cpc0k¯nIF[1+mpm12cos(Δωp1t+ψ)]+cp(c0k¯nI)2[1+12g=1mg2+mp2(2m1+oddgmgmg+1)cos(Δωp1t+ψ)]}δ(tt),
where prΔt(t,tn)δ(ttn) is applied to Eq. (9). Notice that this Dirac delta function approximation is valid only if the detector integration time ΔT for each measurement point is much longer than the decay time of prΔt(t). However, ΔT should be much shorter than 2π/Δωp1 of heterodyne outputs. Therefore, the range of ΔT validating the uncorrelation of the covariance depends on a phosphor of the IIN and a beat frequency of heterodyne outputs. If all mg are zero, which means non-modulated gain in the IIN, the variance of Eq. (13) becomes the well-known result of the Burgess Variance theorem18,36 for Poisson inputs. For gain modulation as in hetero/homodyne measurements, the covariance of Eq. (13) modulates at the beat frequency Δωp1 like the mean of Eq. (10), where dc (σDC2) and ac (σAC2) variances can be considered separately. Notice that σDC2 in Eq. (13) is increased by the factor of cp(c0k¯nI)2(1/2)g=1mg2, compared with the variance for a non-modulated input (mp=0), which is the effect of the gain modulation.. Both σDC2 and σAC2 are strongly dependent on F and c0k¯nI, which is an overall dc gain of the IIN, and mp affects only σAC2. Since we discuss mean and variance of output photons, Ky(t,t) is the photon number with the multiplication of ΔT, which is unitless, as the mean.

3.

Theory for Spatial Noise Analysis

The mean and covariance of Eqs. (10) and (13) are derived based on the temporal PSF of an IIN, as shown in Eq. (5), where the effect of spatial distribution of the PSF is omitted. The temporal consideration would be valid for photoelectrons from a MCP output pore because photoelectrons incident into each MCP input pore are Poisson and MCP channels are physically separated from each other. As shown in Fig. 1(a), however, secondaries from a single MCP output pore are spatially distributed to multiple CCD pixels, which should be considered for the noise analysis of photons detected on a CCD. It is reasonable to assume that the event of detecting one photon among secondaries from a specific MCP output pore by a certain CCD pixel does not affect other secondary photons from the MCP output pore. With this assumption, the spatial distribution of secondaries is theoretically described by a multinomial selection rule.18

Fig. 1

(a) Simple scheme for the spatial distribution of secondary photons. Experimentally measured profiles [(b) 2-D and (c) 1-D] of averaged normalized spatial covariance. In (c), circles, squares, and dots indicate exposure times of 60, 120, and 240 ms, respectively. In (d), the temporal autocovariance measured at a single CCD pixel is indicated.

JBO_17_1_015002_f001.png

To simplify the mathematics, simplified terminologies are used for the spatial noise analysis. If the number of total photons detected by the α’th CCD pixel during some measurement time T is Mα,

Eq. (14)

Mα=imαi,
where mαi indicates the number of photons detected on the α’th CCD pixel from i’th MCP output pore. The summation in Eq. (14) is for all MCP output pores contributing to Mα. Like the temporal noise analysis, T is assumed to be much longer than an anode decay time in the IIN. The mean Mα is derived as

Eq. (15)

Mα=imαi=0mαiPr(mαi)=imαi=0Ki=mαimαiPr(mαi|Ki)Pr(Ki)=iPαiK¯i,
where Pαi is the probability that photons from i’th MCP pore are detected on α’th CCD pixel and K¯I indicates mean total photons from the i’th MCP pore. In the procedure of Eq. (15), averaging is done for two random variables mαi and Ki with a multinomial PDF Pr(mαi|Ki) and Pr(Ki) that is unknown, respectively. For convenience, these two averages are denoted as m and K to calculate the covariance C[Mα,Mβ]=MαMβMαMβ. The second moment of MαMβ is

Eq. (16)

MαMβ=i,jmαimβjm,K=ijPαiPβjKiKjK+i=jPαiKiδαβPαiPβiKi+PαiPβiKi2K,
where the covariance of a multinomial PDF is applied to the second step and δαβ is 1 only if α=β, otherwise 0. Because a multinomial selection of Poisson results in Poisson,18 photoelectrons incident into each MCP input pore are also Poisson, which indicates that KiKjK in Eq. (16) can be approximated to K¯iK¯j. Then, from Eq. (16), the covariance is

Eq. (17)

C[Mα,Mβ]=i[PαiK¯iδαβ+PαiPβi(σKi2K¯i)],
where Ki2K=σKi2+(K¯i)2 is used in the second term of Eq. (16) to derive Eq. (17). The variance of photons detected on a single CCD pixel (i.e., α’th pixel) is given by α=β in Eq. (17). C[Mα,Mβ] of Eq. (17) shows that measured images on a CCD from an IIN are spatially correlated and the correlation length is mainly determined by Pαi. It is known that the efficiency of a photon transferring from an anode to a CCD in a typical IIN having a relay lens imaging system is 1015%. Therefore, the first term in Eq. (17) cannot be ignorable for the variance, unless σKi2 is much higher than K¯i. We insist that the spatial noise property shown in Eq. (17) is general in IINs and should be considered even for a PMT. In a PMT, only two probabilities of detected and lost photons are considered, so Pαi might be much higher than that in IINs. Moran et al.28 analyzed a spatial noise property in ICCDs by introducing a spatiotemporal PSF, but their procedure is very complicated and their result is different from Eq. (17). Experimental results showing the covariance of Eq. (17) in IIN outputs are presented in the next section.

Complete noise expressions of hetero/homodyne measurements are given by combining the temporal and spatial analyses. Resulting from this, dc and ac means and variances on α’th CCD pixel in the frequency-domain measurement are

M¯DC=cpiPαi(c0k¯nI)i,M¯AC=cpmp2iPαi(c0k¯nIm1)i,σDC2=cp{iPαi(c0k¯nI)i+iPαi2[c0k¯nI(F1)+(c0k¯nI)2(1+12g=1mg2)]i},
and

Eq. (18)

σAC2=cpmp2{iPαi(c0k¯nIm1)i+iPαi2[c0k¯nIm1(F1)+(c0k¯nI)2(2m1+oddgmgmg+1)]i},
respectively. For Eq. (18), we assumed that cp and mp of modulated output photons from a phantom slowly vary across the phantom exit surface or the cathode of the IIN, so these quantities are almost the same for each MCP channel contributing to the noise property of photons detected at α’th CCD pixel. This assumption is reasonable because the degree of diffusion (i.e., scattering) of tissues in medical imaging is usually large. It is important that both mean and variance modulate along cos(Δωp1t+ψ) with these dc and ac means and variances in Eq. (18). Equation (18) also shows that M¯AC and σAC2 are approximately different from M¯DC and σDC2, respectively, by the factor of mpm1/2. When the gain of the IIN increases, both F and c0k¯nI are increased, which indicates that the second terms of Pαi2 in both dc and ac variances are significant. It is expected that although σDC2 and σAC2 are increased as increasing the gain, the relative difference between these two variances becomes larger because the multiplication factor mpm1/2 in σAC2 is typically less than 1. Furthermore, Eq. (18) implies that as the modulation gain of the IIN deviates from a perfect sinusoidal function more and more, the noise of the hetero/homodyne is increased more due to higher values for mgs.

4.

Experiment

The CCD and the IIN used in the experiment are Point Grey F2-08S2M and Lambert Ins. II18MD, respectively. The source is the laser diode (LDH-M-C-650B) from PicoQuant with the modulated diode laser driver, MDL 300. This driver is run by external modulation signals from a Tektronix AFG3252 function generator that also generates modulation signals for the IIN cathode. The CCD is carefully adjusted onto the image plane of the relay lens of the IIN in Fig. 1(a) before the main experiment. Since the experiment of measuring variances usually takes a few hours, the experiment starts after about 2 h of warming-up to thermally stabilize all instruments. Especially, voltages directly measured from the cathode of the IIN show that the cathode is thermally unstable at the initial run, which might cause an unexpected long-term variation during the measurement. The laser diode (LD) beam incident into the IIN is almost spatially flat and speckle-free in intensity, which is achieved by lenses and ground glasses. Also, neutral density filters are used to decrease the power of the LD to avoid damage to the IIN cathode, because harmonic distortions of the modulated LD beam can be minimized by keeping the LD power high.

Figure 1(b) is the averaged variance-normalized spatial covariance calculated from images for non-modulated incident beams to the IIN with a non-modulated gain. The averaging is done with 49 variance-normalized spatial covariance matrices, where each of them is normalized by its own variance. The exposure time T is 60 ms, and 2000 images are captured to calculate the spatial covariance. The transverse magnification of the relay lens in the IIN is 0.5 and the CCD pixel size is 7.4×7.4μm2, so the pixel size on the anode surface in Fig. 1(b) is equivalent to 14.8×14.8μm2. As shown in Fig. 1(b), the data are spatially correlated, which is caused by the secondary photon distribution from MCP output pores to CCD pixels. Figure 1(c) shows 1-D profiles of calculated spatial covariance matrices for different MCP voltages. The exposure time T is adjusted for different MCP voltages so that measured means are almost the same, where Exp60 ms in the legend indicates 60 ms CCD exposure time and highest MCP voltage. As the MCP voltage gets higher (i.e., the IIN gain is higher), the variance becomes larger than the mean for a single MCP output pore, so the contribution of the second term of Eq. (17) to the variance for a CCD pixel is also increased. This is the reason that the width of variance-normalized correlation functions is decreased with decreasing MCP voltage, as shown in Fig. 1(c). It is noticed that the correlation length is almost the same for all three MCP voltages, which indicates that Pαi in Eq. (17) is almost independent of a MCP voltage. We observed that even modulating the cathode voltage does not have a large affect on these correlation functions. Figure 1(d) is a variance-normalized temporal auto covariance from data measured at a single CCD pixel for T=60ms. Since 60 ms of T is much longer than the decay time of the phosphor (1 ms to drop 1% of the peak output for P43 in II18MD), data are temporally uncorrelated as expected. The experimental results in Fig. 1 show that the approach of a multinomial PDF for the spatial distribution of secondaries is reasonable and that its effect can be independently considered from the temporal point process analysis.

Fig. 2

(a) Examples of measured homodyne profile. (b) Means and (c) variances of measured dc and ac. Labels on right y-axes in (b) and (c) indicate ratios of M¯AC/M¯DC and σAC12/σAC22, respectively. (d) Varied mp changes not dc variance, but ac variance of homodyne outputs. The x-axes for (b) and (c) indicate relative MCP voltages.

JBO_17_1_015002_f002.png

A homodyne method is used to experimentally measure σDC2 and σAC2 of Eq. (18), because it is more appropriate for measuring multiple images than a heterodyne method. Applying a high modulation frequency to the cathode in the IIN for a long time could permanently damage the cathode, so the modulation frequency for this experiment is set to 100 kHz. Although 100 kHz is several hundred times lower than a typical modulation frequency used in frequency-domain diffusive imaging, the noise characteristics shown in Eq. (18) can still be observed. Also, 4×4 CCD pixels are considered a single macro pixel, which increases the reliability of measuring variances without dramatically increasing the number of images, hence reducing the overall measurement time. As the first step, a homodyne profile is measured with 64 phase steps for 2π of the phase difference (ψ) between modulation signals applied to the IIN and LD. Homodyne profiles measured at three observer points (ObPs) of a macro pixel are shown in Fig. 2(a), indicating that these profiles are slightly different for different macro pixels. The maximum detector output (bin) is 212=4096. Each measurement point in these homodyne profiles is produced by averaging 50 homodyne images. Next, the averaged homodyne profile is calculated from 100 homodyne profiles measured at different macro pixels within an arbitrarily chosen local detector area. From this averaged homodyne profile, three ψ’s corresponding to maximum, dc, and minimum irradiances of the homodyne profile are sought. After these ψ’s are found, 1000 images are measured for each ψ to calculate means (M¯min, M¯DC, M¯max) and variances (σmin2, σDC2, σmax2) for all 100 macro pixels within the arbitrarily chosen local detector area. Finally, M¯AC and σAC2 are calculated from 1/2(M¯maxM¯min) and 1/2(σmax2σmin2), respectively. Macro pixels satisfying the ac mean ratio of M¯maxM¯DC and M¯DCM¯min is almost 1 are selected for investigating noise characteristics of Eq. (18), the number of which is 30 out of 100. Macro pixels showing ac mean ratios not close to 1 indicate that their homodyne profiles deviate much from the averaged homodyne profile.

Figure 2(b) shows measured M¯DC, M¯AC and their ratios to different MCP voltages that are indicated as numbers in the x-axis. T for these different MCP voltages are set to 60, 120, 240, and 480 ms. The MCP voltages for these T’s are chosen to make all measured M¯DC the same, where 1 MCP voltage in Fig. 2 corresponds to T=60ms. The primary photon number Tcp increases as T is increased, but it can be considered that primary photons are still Poisson for even T=480ms due to a very small output photon rate incident into the IIN. If mp and m1 are assumed almost constant within each measurement macro pixel, M¯AC/M¯DCmpm1/2 from Eq. (18). Since the separately measured mp in this experimental setup is 0.6, and M¯AC/M¯DC0.415 in Fig. 2(b), which indicates that m1 of the cathode modulation is 1.37. We observed that measured values of M¯AC/M¯DC are around 0.390.42 for all conducted experiments as long as applied modulation voltages to the cathode of IIN and LD remain the same. The different values of M¯AC/M¯DC indicate that the characteristics of gain modulation are slightly different, macro pixel by pixel. Measured σDC2, σAC2 and σAC12/σAC22 to different MCP voltages are shown in Fig. 2(c), where σAC12=σmax2σDC2 and σAC22=σDC2σmin2. The value σAC12/σAC221 indicates the validity of the experiment. Figure 2(c) shows that both σDC2 and σAC2 increase as the MCP voltage increases, although M¯DC and M¯AC are constant as in Fig. 2(b). This means that F and k¯nI in Eq. (18) dramatically increase as the MCP voltage (i.e., gain of the IIN) increases. From the other view, it is observed that the decreasing rates of both σDC2 and σAC2 are smaller for a lower MCP voltage, which indicates the mean terms contributing to variances in Eq. (18) are more dominant. It is also observed in Fig. 2(c) that the absolute difference between σDC2 and σAC2 is increased as the MCP voltage increases, which is caused by mpm1/20.415 in σAC2 of Eq. (18). Since the developed theory in Eq. (18) shows that mp affects not σDC2 but σAC2, we measure the variation of variances with different peak-to-peak voltages (Vpp) of the modulation signal applied to LD, which changes mp. As shown in Fig. 2(d), σDC2’s are almost invariant to Vpp, but σAC2’s are proportionally changed to Vpp. As shown in Fig. 2, experimental results show good agreement with the developed theory of Eq. (18).

5.

Discussion and Conclusion

A noise factor, NF, is commonly used for the study of noise properties in photon amplifiers, such as ICCDs or IINs.28 Although the noise factor NF is defined for the case of non-modulation gain in photon amplifiers, it is worth making a relationship between NF and F for applying NF to Eq. (18), which is

Eq. (19)

NFSNRi2SNRo2=miσo2mo2=Fc0k¯nI,
where the subscripts i and o indicate input and output, and mi and mo are means of primary and secondary photons, respectively. It is assumed that the primary photons are Poisson in the second step and mic0k¯nI=mo is applied to the third step of Eq. (19). Substituting Eq. (19) to Eq. (18) can express σDC2 and σAC2, including NF. For example, σDC2 becomes

Eq. (20)

σDC2cp{iPαi(c0k¯nI)i+iPαi2(c0k¯nI)i2[NF+1+12g=1mg2]i},
where 1/c0k¯nI in the second term of Eq. (20) is ignored assuming c0k¯nI1. It is reported that a NF for MCP outputs in an ICCD is inversely proportional to the MCP voltage.28 From the experimental results in Fig. 2 and Eq. (20), it can be conjectured that (c0k¯nI)[NF+1+12g=1mg2] is dramatically increased as the MCP voltage increases even though a NF is decreased. A similar phenomenon is experimentally observed in other study, where c0k¯nINF is increased as a MCP voltage increases although NF is decreased.28 As another issue, characterizing the noise property of photon amplifiers by measuring a NF as a function of gain should be done carefully. As shown in Eq. (20), the NF for MCP output photons might be significantly different from the NF for amplifiers’ outputs depending on the amount of Pαi when c0k¯nI is low.

The noise model of hetero/homodyne outputs is theoretically derived considering random amplification and multinomial selection. The noise characteristics determined from the developed theory are experimentally verified. Although this noise model is developed for hetero/homodyne measurements in this paper, it can be used for the case of non-modulation input beam and gain. Furthermore, it is reasonable to assume that the primary photons for most amplification processes are described by Poisson point process, as assumed in this paper. Therefore, the model in this paper can be applied to most photon-amplification processes including PMTs and Avalanche photodiodes (APDs) with or without gain modulation.

Some valuable properties are obtained from the developed noise model. The output is temporally uncorrelated, like shot noise, only if detector integration time for sampling is much longer than the decay time of phosphors in imaging intensifiers. For heterodyne measurements, the exposure time should be much less than the period of beat-frequency heterodyne output. Therefore, the heterodyne output is temporally uncorrelated for the specific range of the detector integration time. The amount of variance oscillates like the heterodyne and/or homodyne mean, so dc and ac variances can be considered separately. The dc variance indicates that the noise of the frequency-domain measurement output is amplified by the random amplification and gain modulation compared with the case of non-modulation gain. As the gain modulation deviates more from a perfect sinusoidal function, both dc and ac variances are increased. Another interesting point is that when both the gain of the image intensifier and the probability of detecting amplified photons from the image intensifier on a detector pixel are low, the variance approaches the mean—that is, the output photons are close to Poisson.

Hetero/homodyne measurements are widely used in frequency-domain diffusive imaging, so understanding the noise property of the frequency-domain outputs is beneficial in many ways. For example, it has recently been reported that estimated means of modulation amplitude and phase from noisy hetero/homodyne outputs are biased when the number of output photons is small.37 By adapting the expressions of mean and covariance developed in this paper, we have practically clarified the relationship between the bias and primary photon number. Furthermore, the developed noise model would be useful to precisely estimate modulation amplitude and phase using a maximum likelihood estimation method. As another example, the noise model in this paper could consolidate figures of merit for evaluating diffusive imaging performance, which is effective at optimizing frequency-domain diffusive imaging systems and helpful in developing new modalities of them. Finally, the performance of reconstruction algorithms can be enhanced with the appropriate noise model, so the theoretical results in this paper can be adapted to improve the quality of reconstructed images from frequency-domain tomographic diffusive imaging.

Appendices

Appendix:

For the mean of Eq. (1), averaging for Δtnk is

Eq. (21)

y(t){Δtnk}=n=1Nk=1knδ(ttnΔtnk)[k¯n(tn)]1pd(Δtnk+tn,tn)Δtnk=n=1Nk=1kn[k¯n(tn)]1pd(t,tn)=n=1Nkn[k¯n(tn)]1pd(t,tn),
where the marginal probability for all Δtnk except n=n and k=k for any particular n and k are conducted on the first line []. The last step in Eq. (21) is done by that [k¯n(tn)]1 is independent of k. The averaging for the secondary kn simply replaces kn in Eq. (21) to k¯n(tn), so

Eq. (22)

y(t){Δtnk},{kn}=n=1Npd(t,tn).
From Eq. (3), the averaging for secondary time tn produces

Eq. (23)

y(t){Δtnk},{kn},{tn}=n=1N[N(ap)]1pd(t,tn)ap(tn)dtn.
With the similar concept of independence conducted from Eqs. (21) to (22), further averaging Eq. (23) with N becomes

Eq. (24)

y(t){Δtnk},{kn},{tn},{N}=pd(t,tn)ap(tn)dtn.
Substituting Eqs. (2) and (5) to Eq. (24) finalizes the calculation for the mean, as shown in Eq. (8).

The covariance can be calculated with the similar averaging procedure from Eqs. (21)(22)(23)(24)(25)–(26), but it should be considered in three separate cases. The autocorrelation of the point process of Eq. (1) is expressed as

Eq. (25)

Ry(t,t)=n=1Nk=1knn=1Nk=1knδ(ttnΔtnk)δ(ttnΔtnk).

Case 1: n=n, k=k

The averaging over Δtnk for the autocorrelation with the condition of case 1 is

Eq. (26)

y(t)y(t){Δtnk}(1)=n=1Nk=1kn[k¯n(tn)]1pd(t,tn)δ(tt),
where a marginal probability law for Δtnk is similarly applied, as in Eq. (21). The superscript (1) indicates that averaging is performed for case 1. Further averaging Eq. (26) for kn, tn, and N produces

Eq. (27)

Ry(1)(t,t)=pd(t,tn)ap(tn)dtnδ(tt).
Substituting Eqs. (2) and (5) to Eq. (27) derives

Eq. (28)

Ry(1)(t,t)=cpc0[1+mpcos(ωptn+ψ)][1+g=1mgcos(ωgtn+ϕs)]pdI(t,tn)dtnδ(tt).
Similar with the procedure from Eqs. (8)(9)–(10), only the dc and Δωp1 terms survive in Eq. (28). Therefore,

Eq. (29)

Ry(1)(t,t)=cpc0k¯nI[1+mp2cos(Δωp1t+ψϕ1)g=1mg]δ(tt),
where it is assumed that pdI(t,tn)k¯nI(tn)δ(ttn).

Case 2: n=n, kk.

Considering a marginal probability law, averaging Eq. (25) over Δtnk produces

Eq. (30)

y(t)y(t){Δtnk}(2)=n=1Nk=1knk=1knpΔt(Δtnk,Δtnk)δ(ttnΔtnk)δ(ttnΔtnk)dΔtnkdΔtnk=n=1Nk=1knk=1kn[k¯n(tn)]2pd(t,tn)pd(t,tn)=n=1N(kn2kn)[k¯n(tn)]2pd(t,tn)pd(t,tn),
where the second line is given by that Δtnk and Δtnk are independent of each other. The kn2kn indicates the number of terms on the summations of k and k for kk, which can be switched to Var[kn|tn]+k¯n2(tn)k¯n(tn) through the procedure of averaging Eq. (30) over kn. Further averaging y(t)y(t)(2) for tn and N becomes

Eq. (31)

Rg(2)(t,t)=[Var[kn|tn]+k¯n2(tn)k¯n(tn)][k¯n(tn)]2ap(tn)pd(t,tn)pd(t,tn)dtn.
If the integration for the term k¯n(tn) in [Var[kn|tn]+k¯n2(tn)k¯n(tn)] is considered with the substitutions of Eqs. (2), (5), and (7),

Eq. (32)

[k¯n(tn)]1pd(t,tn)pd(t,tn)ap(tn)dtn=cpc0[k¯nI(tn)]1pdI(t,tn)pdI(t,tn)[1+mpcos(ωptn+ψ)][1+g=1mgcos(ωgtn+ϕg)]dtncpc0k¯nI(tn)δ(ttn)δ(ttn)[1+mp2cos(Δωp1t+ψϕ1)g=1mg]dtncpc0k¯nI[1+mp2cos(Δωp1t+ψϕ1)g=1mg]δ(tt).

The last line of Eq. (32) is equivalent to Eq. (29), so these are canceled out for the autocorrelation considering both case 1 and case 2. The autocorrelation is

Eq. (33)

Ry(1)+(2)(t,t)=[Var[kn|tn]+k¯n2(tn)]ap(tn)prΔt(t,tn)prΔt(t,tn)dtn,
where Eq. (6) is used for pd in Eq. (31).

Case 3: nn, kk.

Calculating the autocorrelation under this condition is so straightforward that the detailed mathematical procedure is not shown. Basically, after considering that there are N2N terms for the summations for n and n, two point processes g(t) and g(t) can be independently calculated. Resulting from this consideration is the multiplication of two means of Eq. (24). Because the covariance Ky(t,t)=Ry(t,t)y(t)y(t), the result Ry(t,t) of for case 3 is canceled out with y(t)y(t) in Ky(t,t). Therefore, Ky(t,t) is the same as Eq. (31), which is the same as Eq. (9).

References

1. 

A. P. GibsonJ. C. HebdenS. R. Arridge, “Recent advances in diffuse optical imaging,,” Phys. Med. Biol., 50 R1 –R43 (2005). http://dx.doi.org/10.1088/0031-9155/50/4/R01 PHMBA7 0031-9155 Google Scholar

2. 

D. A. Boaset al., “Imaging the body with diffuse optical tomography,” IEEE Signal Process. Mag., 1053 57 –75 (2001). http://dx.doi.org/10.1109/79.962278 ISPRE6 1053-5888 Google Scholar

3. 

U. J. NetzJ. BeuthanA. H. Hielscher, “Multipixel system for gigahertz frequency-domain optical imaging of finger joints,” Rev. Sci. Instrum., 79 (3), 034301 (2008). http://dx.doi.org/10.1063/1.2840344 RSINAK 0034-6748 Google Scholar

4. 

Z. M. Wanget al., “Experimental demonstration of an analytic method for image reconstruction in optical diffusion tomography with large data sets,” Opt. Lett., 30 (24), 3338 –3340 (2005). http://dx.doi.org/10.1364/OL.30.003338 OPLEDP 0146-9592 Google Scholar

5. 

G. Y. Panasyuket al., “Fluorescent optical tomography with large data sets,” Opt. Lett., 33 (15), 1744 –1746 (2008). http://dx.doi.org/10.1364/OL.33.001744 OPLEDP 0146-9592 Google Scholar

6. 

B. Chanceet al., “Phase measurement of light absorption and scatter in human tissue,” Rev. Sci. Instrum., 69 (10), 3457 –3481 (1998). http://dx.doi.org/10.1063/1.1149123 RSINAK 0034-6748 Google Scholar

7. 

A. B. ThompsonE. M. Sevick-Murace, “Near-infrared fluorescence contrast-enhanced imaging with intensified charge-coupled device homodyne detection: measurement precision and accuracy,” J. Biomed. Opt., 8 (1), 111 –120 (2003). http://dx.doi.org/10.1117/1.1528205 JBOPFO 1083-3668 Google Scholar

8. 

S. R. ArridgeW. R. Lionheart, “Nonuniqueness in diffusion-based optical tomography,” Opt. Lett., 23 (11), 882 –884 (1998). http://dx.doi.org/10.1364/OL.23.000882 OPLEDP 0146-9592 Google Scholar

9. 

G. Xuet al., “Direct-current-based image reconstruction versus direct-current included or excluded frequency-domain reconstruction in diffuse optical tomography,” Appl. Opt., 49 (16), 3059 –3070 (2010). http://dx.doi.org/10.1364/AO.49.003059 APOPAI 0003-6935 Google Scholar

10. 

J. R. Lakowiczet al., “Fluorescence lifetime imaging,” Anal. Biochem., 202 316 –330 (1992). http://dx.doi.org/10.1016/0003-2697(92)90112-K ANBCA2 0003-2697 Google Scholar

11. 

E. Grattonet al., “Fluorescence lifetime imaging for the two-photon microscope: time-domain and frequency-domain methods,” J. Biomed. Opt., 8 (3), 381 –390 (2003). http://dx.doi.org/10.1117/1.1586704 JBOPFO 1083-3668 Google Scholar

12. 

A. EspositoH. C. GerritsenF. S. Wouters, “Optimizing frequency-domain fluorescence lifetime sensing for high-throughput applications: photon economy and acquisition speed,” J. Opt. Soc. Am. A, 24 (10), 3261 –3273 (2007). http://dx.doi.org/10.1364/JOSAA.24.003261 JOAOD6 0740-3232 Google Scholar

13. 

A. ElderS. SchlachterC. F. Kaminski, “Theoretical investigation of the photon efficiency in frequency-domain fluorescence lifetime imaging microscopy,” J. Opt. Soc. Am. A, 25 (2), 452 –462 (2008). http://dx.doi.org/10.1364/JOSAA.25.000452 JOAOD6 0740-3232 Google Scholar

14. 

D. Y. Paithankaret al., “Imaging of fluorescent yield and lifetime from multiply scattered light reemitted from random media,” Appl. Opt., 36 (10), 2260 –2272 (1997). http://dx.doi.org/10.1364/AO.36.002260 APOPAI 0003-6935 Google Scholar

15. 

A. R. Pinedaet al., “Information content of data types in time-domain optical tomography,” J. Opt. Soc. Am. A, 23 (12), 2989 –2996 (2006). http://dx.doi.org/10.1364/JOSAA.23.002989 JOAOD6 0740-3232 Google Scholar

16. 

H. K. Kimet al., “Optimal source-modulation frequencies for transport-theory-based optical tomography of small-tissue volumes,” Opt. Express, 16 (22), 18082 –18101 (2008). http://dx.doi.org/10.1364/OE.16.018082 OPEXFF 1094-4087 Google Scholar

17. 

T. Tuet al., “Analysis on performance and optimization of frequency-domain near-infrared instruments,” J. Biomed. Opt., 7 (4), 643 –649 (2002). http://dx.doi.org/10.1117/1.1501562 JBOPFO 1083-3668 Google Scholar

18. 

H. H. BarrettK. J. Myers, Foundations of Image Science, Wiley, New York (2004). Google Scholar

19. 

H. H. Barrettet al., “Objective assessment of image quality. II. Fisher information, Fourier crosstalk, and figures of merit for task performance,” J. Opt. Soc. Am. A, 12 (5), 834 –852 (1995). http://dx.doi.org/10.1364/JOSAA.12.000834 JOAOD6 0740-3232 Google Scholar

20. 

J. Y. Hestermanet al., “Hardware assessment using the multi-module, multi-resolution system (M3R): a signal-detection study,” Med. Phys., 34 (7), 3034 –3044 (2007). Google Scholar

21. 

H. H. Barrett, “Objective assessment of image quality: effects of quantum noise and object variability,” J. Opt. Soc. Am. A, 7 (7), 1266 –1278 (1990). http://dx.doi.org/10.1364/JOSAA.7.001266 JOAOD6 0740-3232 Google Scholar

22. 

C. K. AbbeyH. H. Barrett, “Human- and model-observer performance in ramp-spectrum noise: effects of regularization and object variability,” J. Opt. Soc. Am. A, 18 (3), 473 –488 (2001). http://dx.doi.org/10.1364/JOSAA.18.000473 JOAOD6 0740-3232 Google Scholar

23. 

V. Toronovet al., “Optimization of the signal-to-noise ratio of frequency-domain instrumentation for near-infrared spectro-imaging of the human brain,” Opt. Express, 11 (21), 2717 –2729 (2003). http://dx.doi.org/10.1364/OE.11.002717 OPEXFF 1094-4087 Google Scholar

24. 

S. P. Morgan, “Detection performance of a diffusive wave phased array,” Appl. Opt., 43 (10), 2071 –2078 (2004). http://dx.doi.org/10.1364/AO.43.002071 APOPAI 0003-6935 Google Scholar

25. 

D. KangM. A. Kupinski, “Signal detectability in diffusive media using phased arrays in conjunction with detector arrays,” Opt. Express, 19 (13), 12261 –12274 (2011). http://dx.doi.org/10.1364/OE.19.012261 OPEXFF 1094-4087 Google Scholar

26. 

J. PhilipK. Carlsson, “Theoretical investigation of the signal-to-noise ratio in fluorescence lifetime imaging,” J. Opt. Soc. Am. A, 20 (2), 368 –379 (2003). http://dx.doi.org/10.1364/JOSAA.20.000368 JOAOD6 0740-3232 Google Scholar

27. 

R. L. Bell, “Noise figure of the MCP image intensifier tube,” IEEE Trans. Electron. Dev., 22 (10), 821 –829 (1975). IETDAI 0018-9383 Google Scholar

28. 

S. E. Moranet al., “Intensified CCD (ICCD) dynamic range and noise performance,” Proc. SPIE, 3173 430 –457 (1997). http://dx.doi.org/10.1117/12.294535 PSISDG 0277-786X Google Scholar

29. 

J. McGintyet al., “Signal-to-noise characterization of time-gated intensifiers used for wide-field time-domain FLIM,” J. Phys. D: Appl. Phys., 42 135103 (2009). http://dx.doi.org/10.1088/0022-3727/42/13/135103 JPAPBE 0022-3727 Google Scholar

30. 

V. V. ApanasovichE. G. Novikov, “Mathematical model of the microchannel plate single electron impulse formation,” IEEE Trans. Electron. Dev., 42 (21), 2231 –2235 (1995). http://dx.doi.org/10.1109/16.477784 IETDAI 0018-9383 Google Scholar

31. 

A. FrenkelM. A. SartorM. S. Wlodawski, “Photon-noise-limited operation of intensified CCD cameras,” Appl. Opt., 36 (22), 5288 –5297 (1997). http://dx.doi.org/10.1364/AO.36.005288 APOPAI 0003-6935 Google Scholar

32. 

B. Q. SpringR. M. Clegg, “Image analysis for denoising full-field frequency-domain fluorescence lifetime images,” J. Microsc., 235 (2), 221 –237 (2009). http://dx.doi.org/10.1111/jmi.2009.235.issue-2 JMICAR 0022-2720 Google Scholar

33. 

U. Schuhle, “Intensified solid state sensor cameras: ICCD and IAPS,” ISSI Scientific Reports Series, 419 –428 (2010). Google Scholar

34. 

H. H. BarrettR. F. WagnerK. J. Myers, “Correlated point processes in radiological imaging,” Proc. SPIE, 3032 110 –124 (1997). http://dx.doi.org/10.1117/12.273975 PSISDG 0277-786X Google Scholar

35. 

R. A. Loch, “Cesium-telluride and magnesium for high quality photocathodes: preparation and diagnostics,” Netherlands (2005). Google Scholar

36. 

R. E. Burgess, “Homophase and heterophase fluctuations in semiconducting crystals,” Discuss. Faraday Soc., 28 151 –158 (1959). http://dx.doi.org/10.1039/DF9592800151 DFSOAW 0014-7664 Google Scholar

37. 

D. KangM. A. Kupinski, “Effect of noise on modulation amplitude and phase in frequency-domain diffusive imaging,” J. Biomed. Opt., Google Scholar
© 2012 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2012/$25.00 © 2012 SPIE
DongYel Kang and Matthew A. Kupinski "Noise characteristics of heterodyne/homodyne frequency-domain measurements," Journal of Biomedical Optics 17(1), 015002 (7 February 2012). https://doi.org/10.1117/1.JBO.17.1.015002
Published: 7 February 2012
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Photons

Modulation

Microchannel plates

Charge-coupled devices

Homodyne detection

Signal to noise ratio

Heterodyning

RELATED CONTENT


Back to Top