Open Access
7 January 2013 Depth probing of diffuse tissues controlled with elliptically polarized light
Author Affiliations +
Abstract
Polarization gating is a popular technique in biomedical optics. It is widely used to inspect the surface of the tissues (under colinear or cocircular detection) or instead to probe the volume (cross-linear detection), without information on the probed depth. Elliptical polarization is introduced to explore the possibility of probing diffuse tissues at selective depths. A thorough Monte Carlo simulation study shows complete correlation between the probed depths and the ellipticity of the polarized light, for a medium with known optical properties. Within a wide range of optical parameters, a linear relation between the backscattered intensity and the depth extension of the probed volume was found whatever the polarization used, but with a controlled extension depending on the ellipticity.

1.

Introduction

Optical imaging in biomedical applications has been very popular since the early 1980s, for its noninvasiveness and for the simplicity of the techniques. The main problem of light interactions with biological tissues in the optical wavelength range is the strong scattering of light. These techniques, therefore, suffer from a lack of resolution in depth. As photons undergo multiple scattering events before being detected, it is difficult to associate a probed depth to the detected signal. To circumvent the problem, photon gating approaches have been introduced, among which time- and polarization-gating are very successful. Time-gating allows filtering the photons as a function of their pathlengths but has the disadvantage of needing very expensive and sophisticated equipment. Moreover, measurements at millimetric optical paths are challenging, as a 1 mm optical path corresponds approximately to a time delay of 4 to 5 picoseconds in tissues. The polarization gating method is a simple method using the information of the polarization state of the detected backscattered light to filter the unpolarized photons.

Intuitively, when traveling through the medium, polarized photons will maintain their state of polarization to a smaller or larger extent, according to the absorption or scattering events, that is, according to the transport mean free path (MFP). Initial polarization will be maintained only for a certain number of scattering events until total depolarization is reached.1 Commonly, linear polarization and more rarely circular polarization are used for polarization gating. It was shown that the method allows light extraction from superficial tissue,2 whereas circular polarization is sensitive to deeper layers.3 If the diffusing medium is composed of Mie scatterers as in most biological tissues, because of polarization-memory effects1,4 the depth defined by a circular-polarization gate is shown to be larger than that defined by a linear-polarization gate. Moreover, as circular polarization undergoes a helicity flip with mirror reflection, this technique thus allows screening of subsurface tissues deeper than linearly polarized light, without any blurring due to surface mirror reflection. Elliptical polarization has been seldom used in polarization gating. Preliminary studies3 have shown that elliptical polarization-gating is sensitive to tissue layers between those of the two latter mentioned polarizations.

In this paper, we show through the results of a Monte Carlo simulation study, which is exhaustive in terms of the range of optical properties tested, that using different polarization ellipticities allows probing at different depths. These depths are well-defined as the optical properties are known, and totally depend on the ellipticity of the polarization of the illumination. It is demonstrated that the maximum depth reached by polarized photons completely depends on the polarization used and the optical properties of the probed tissue. The technique thus offers the possibility of a continuous selective probing in depth by tuning the ellipticity of polarization. Moreover, the link between measured backscattered intensity and probed depth has been studied: a linear relation was found between the radial extension of the backscattered intensity and the probed depth for any polarization.

2.

Method and Definitions

2.1.

Simulating Polarized Light Propagation with Monte Carlo

The Monte Carlo method5,6 is used to simulate polarized light propagation in a quasi semi-infinite medium. The principle consists of evaluating statistically photons propagation through the tissues by taking into account local optical properties expressed in terms of probability distributions defining: (1) the step size of each photon movement between two scattering/absorption events, and (2) the angles of deflection in a photon's trajectory when a scattering event occurs. The evolution of the polarization state of each photon is tracked through the modifications of a Stokes vector, modified at each photon-tissue interaction by a Mueller matrix accounting for the optical properties of the medium. In the present study, a Quaternion Monte Carlo code, initially developed by J. Ramella-Roman,6 has been modified in order to take into account Fresnel reflections at the interface.7 The new code has been validated with comparison with published results (mainly Refs. 8 and 9). The final Stokes vector, describing the final polarization state with the contribution of all backscattered photons is saved.

2.2.

Polarization Gating

This allows calculating surface images representing different co- and cross-polarized illumination/detection schemes, also called imaging channels. A possible experimental setup is schematized in Fig. 1, allowing measurements under any polarization channel (co- and cross-linear, co-elliptical and co-circular).10 The definition of co-elliptical channels is, in fact, not trivial and must be detailed. The simulated imaging channels can be fully expressed with the Stokes vector depending on the rotation angles α of the horizontal axis of the linear polarizer and the rotation angle β of the fast axis of the quarter-wave plate (Fig. 1). The resulting ellipticity of the polarization is ε=βα.

Fig. 1

Sketch of an example of the optical imaging system with linear polarizers (P) and quarter-wave plates (QW). (a) Generation of the elliptically polarized light in the illumination path. (b) Analysis of elliptically polarized light in the detection path. The polarization ellipse of the illumination and the detection can be described by the rotation angles α and α of the horizontal axis of the linear polarizer and the rotation angle β and β of the fast axis of the quarter-wave plate. The ellipticities ε and ε are defined by ε=βα and ε=βα. Copolarization detection is obtained for α=α, β=β, and ε=ε.

JBO_18_1_016007_f001.png

Any elliptically co- and cross-polarized channel intensity I(α,β), including the special cases for linear and circular polarizations, can be described with the following equation depending on the rotation angles α of the polarizer and the rotation angle β of the quarter-wave plate as well as the Stokes vector Si=[Ii,Qi,Ui,Vi]t emerging from the probed medium:

Eq. (1)

I(α,β)=(Ii+cos(2β)cos[2(αβ)]Qi+sin(2β)cos[2(αβ)]Ui+sin[2(αβ)]Vi)/2.

As recommended in Ref. 9, by subtracting the cross-linear channel, measuring the multiple scattered (or totally depolarized) part of the signal, from the images obtained under copolarized channels, one obtains a good approximation of the polarization maintaining images, free from surface reflections if elliptical polarization is considered.

2.3.

Introduction of Signal Descriptors for Analyzing Depth and Radial Extension of the Polarized Backscattered Signals

In order to study the depth extent of each polarization gating measurement, the histogram of the mean visitation depth (MVD) of all emerging photons is specifically generated and weighted with the final Stokes vector, and normalized to signal maximum. The MVD is representative for the volume probed by the backscattered photons11 and therefore contains most of the information on the depth extension of the probed volume. Here, the optical properties of a semi-infinite medium are varied in order to scan all possible sets of optical properties in biological tissues: the scattering coefficient has a range of μs=10 to 200cm1, the absorption coefficients are μa=0.5 to 5cm1, and the mean cosines of the scattering angles are g=0.05 to 0.95, with size parameters (wave-number in the medium multiplied by the radius of the particle) in the range of 0.53 to 7.55, covering Mie scattering regime, and also approaching Rayleigh scattering regime. The refraction index of the medium is 1.4 and outside of the medium the refraction index of air is taken into account. The chosen range of optical properties for the simulations covers especially skin and vessel tissue, but includes also some properties of the Rayleigh regime.11,12 A pencil beam perpendicular to the surface is considered. The computed MVD of different polarization imaging channels as well as depolarized light is represented in Fig. 2(a) left, with a MFP scale, MFP=[μa+(1g)μs]1. This histogram demonstrates that the MVD depends on the illumination polarization. The peak of the normalized signal varies with the used ellipticity and allows, therefore, probing different depths.

Fig. 2

(a) Examples of (left) histograms of mean visitation depth (MVD) and (right) probabilities P for backscattered photons emerging at radial distances R, for different maintained polarizations [• linear, ▪ elliptical (ε=0.05π), ▴ elliptical (ε=0.1π), ▾ elliptical (ε=0.15π), ♦ circular (ε=0.25π), α=0.25π for all simulations, and + depolarized light] on a MFP’ scale (μa=0.5cm1, μs=100cm1, g=0.8: 1MFP=488μm). (b) Evolution of the depth-descriptor ZC for depolarized light (cross-linear measurement) as a function of μs, μa, and g. (c) Evolution of ZC with different polarizations [same markers as (a)] as a function of μs, μa and g.

JBO_18_1_016007_f002.png

A depth-descriptor ZC is introduced for each polarization histogram and defined as the MVD for which the intensity is 10% of the maximum, on the right hand side of the curve, I(ZC)=0.1. Amongst other possible descriptors (e.g., maximum or mean values), ZC was found as the most stable and producing the largest differentiation between the polarizations, for the simulated range of optical properties. The value of determined in this way fixes an upper bound of the extension of the mean depth probed by a majority of the detected polarized photons. The evolution of ZC is analyzed for different polarizations and with different optical properties and elliptically polarized light in Fig. 2(b) and 2(c). The MFP scale takes the scattering and absorption coefficients into account leading to a quasi independence of the descriptor on these parameters [Fig. 2(c)], while for depolarized light, a strong dependence on μa and μs persists [Fig. 2(b)]. Even if the MFP scale also accounts for the anisotropy factor g, there is still a relatively strong dependence on g for all polarizations as well as depolarized light, especially if the anisotropy factor is high (Mie scatterers), because of its influence on the MFP scale.

The backscattered intensity Io(r,ϕ) of the co- and cross-polarized channels is obtained by counting all backscattered photons collected in a given polarization channel. A probability P(R) can be defined for backscattered photons emerging at radial distance R per unit pixel length:

Eq. (2)

P(R)=r=0Rϕ=02πrIo(r,ϕ)drdϕ.

This integral probability [e.g., Fig. 2(a) right] takes into account the diversity of the different radial patterns of the polarized backscattered channel measurements. A radial-descriptor Rc has been defined in Ref. 13 for linear polarization channels:

Eq. (3)

0RCP(R)dR/0P(R)dR=0.9.

It has been adopted here for general elliptical polarization channels because it allows accounting for both radial extension of the backscattered signal and the diversity of the intensity profiles. The evolution the radial descriptor RC has been analyzed in the same way as ZC.

3.

Results and Discussion

A complete correlation between the two descriptors has been found: Fig. 3 gathers all the simulation results obtained under the large range of optical properties, for both descriptors.

Fig. 3

Evolution of ZC (probed depth) as a function of Rc (radial extension of the backscattered intensity). Color code represents the polarization state: linear (dark blue), elliptical (ε=0.05π) (clear blue), elliptical (ε=0.1π) (turquoise), elliptical (ε=0.15π) (green), circular (ε=0.25π) (orange) [α=0.25π for these simulations], depolarized (red). Three different markers are used for the different optical parameters: absorption coefficient μa (×), scattering coefficient μs (▵), and anisotropy factor g (○). Inset: Zoom in the region probed by polarized light. Red line: fitting curve obtained by linear regression.

JBO_18_1_016007_f003.png

The influence of optical properties and different polarizations on the relation between the MVD and the radial intensity distribution are summarized in this figure, with a description previously introduced by Liu et al.13 These results are consistent with those reported in Ref. 13, though for linear polarization and with a different depth indicator. To our knowledge, the present study reports for the first time that this behavior is maintained whatever the polarization. The correlation between the two descriptors is very linear for different maintaining polarizations and even for depolarized light. A linear curve can be fitted with the data having a slope 0.33, an ordinate at the origin of 0.34, and standard deviation of 0.10. Note that all results were obtained by considering an infinitely narrow pencil beam with normal incidence to the surface. The slope of this curve has to be considered as a maximum value as the depth accessible under a non-normal incidence illumination will be shallower. Interestingly, one can thus express in synthesized form the following approximation, when all quantities are expressed in metric units:

Eq. (4)

ZcRc/3+(3[μa+(1g)μs])1=Rc/3+D,
where D is the diffusion constant as defined in the diffusion approximation.12 We have thus introduced a general formula linking the radial extension of the measurable backscattered intensity to the depth extension of the probed volume, whatever the polarization channel used. Moreover, as the optical properties (anisotropy factor, absorption, and scattering coefficients) can be retrieved by analyzing the morphology of the backscattered polarized signals,14 one can estimate the value of MFP (or equivalently D). This leads unambiguously to an estimation of the probed depth by measuring Rc.

The limited depth extension of the volumes probed with different polarization channels is clearly highlighted with the chosen color code. It is shown that polarization maintaining channels will probe volumes larger as the polarization ellipticity increases. One can also verify that there is no intersection between volumes probed by depolarized and by polarized light. Interestingly, to illustrate this purpose, one can plot (Fig. 4) the evolution of the maximum and minimum values of ZC as a function of the polarization ellipticity ε: tuning the optical components in order to select a given polarization channel allows to probe scattering media up to specific depths, well defined if the optical properties are known.

Fig. 4

Evolution of ZC (probed depth) as a function of the polarization ellipticity ε. The extension of the bars is bounded by the maximum and minimum values found for the range of optical properties explored.

JBO_18_1_016007_f004.png

To fix the ideas, specific calculations have been done for different types of tissues such as skin or brain tissues. The choice of the tissues is motivated by the fact that, for these tissues, conventional two-dimensional (2-D) intrinsic optical imaging techniques is the main optical technique used. In these cases, the present imaging technique could offer new opportunities in probing selectively the depth of these tissues allowing a more precise identification of the sources of the measured signal. The results of the calculations are summarized in Table 1.

Table 1

Depth-indicator ZC calculated with optical properties corresponding to different types of tissues. Depths are reported in [MFP] unit and there corresponding values in µm. The values of the optical properties correspond approximately to the values reported in Ref. 15 and references therein, at the wavelength 633 nm.

TissueAbsorption coefficient μa(cm1)Scattering coefficient μs(cm1)Anisotropy factor g()MFP (μm)Depth-indicator ZC
LinearElliptical ε=0.1πCircularMultiple scattering
[MFP](μm)[MFP](μm)[MFP](μm)[MFP](μm)
Examplea01000.921007.7769.514.8147616.7166767.56750
Human skin3.2168.60.8159.33.4200.14.6271.85.8343.18.3491
Pork skin1492.60.9520.312.7256.923.5476.626.0528.640812
Aorta0.52316.40.8731.64.5142.37.9249.79.2291.222.8719
Gray matter2.7357.10.94289.1254.917.1477.918.8525.323.1647
White matter2.2534.80.8218.73.870.15.5102.86.6123.413.2246
Uterus0.35393.70.6925.42.665.93.177.33.588.310.5267

aWhich is an example representing the most favorable situation for light penetration in depth in biological tissues (nonabsorbing, forward scattering). These values are also those used by Stockford et al. in their simulations.9

This table shows the probed depths, accessible by polarization gating, can be of few hundreds of microns. This corresponds precisely to the depths extensions thought in many applications such as in skin cancers or exposed cortex screening, usually monitored with 2-D intrinsic optical imaging techniques, with no depth resolution. The present polarization gated technique can as well provide functional information and offers the advantage of being simple, with possibly fast data acquisitions.

4.

Conclusion

This paper has shown that the use of elliptical polarization allows selection of photons which probed a specific depth, in a range defined by: (1) a minimum depth obtained by measuring the linear polarization maintaining signal; and (2) a maximum depth with the circular polarization maintaining signal. This allows tuning elliptical polarization in order to select a given probed depth under the condition of known optical properties. Furthermore, it was demonstrated that the probed depth is fully related to the backscattered radial extension for all polarizations. The relation found for different polarized illuminations applies also to depolarized light. The formulation allows prediction of the expected probed depth if the optical properties of the investigated medium and the illumination polarization are known. The formulation is valid for a wide range of optical parameters and for all polarization channels.

References

1. 

F. C. MacKintoshet al., “Polarization memory of multiply scattered light,” Phys. Rev. B, 40 (13), 9342 –9345 (1989). http://dx.doi.org/10.1103/PhysRevB.40.9342 PRBMDO 1098-0121 Google Scholar

2. 

S. G. DemosR. R. Alfano, “Optical polarization imaging,” Appl. Opt., 36 (1), 150 –155 (1997). http://dx.doi.org/10.1364/AO.36.000150 APOPAI 0003-6935 Google Scholar

3. 

S. MorganM. Ridgway, “Polarization properties of light backscattered from a two layer scattering medium,” Opt. Express, 7 (12), 395 –402 (2000). http://dx.doi.org/10.1364/OE.7.000395 OPEXFF 1094-4087 Google Scholar

4. 

D. Bicoutet al., “Depolarization of multiply scattered waves by spherical diffusers: influence of the size parameter,” Phys. Rev. E, 49 (2), 1767 –1770 (1994). http://dx.doi.org/10.1103/PhysRevE.49.1767 PLEEE8 1063-651X Google Scholar

5. 

L. WangS. L. JacquesL. Zheng, “MCML—Monte Carlo modeling of light transport in multi-layered tissues,” Comput. Methods Program. Biomed., 47 (2), 131 –146 (1995). http://dx.doi.org/10.1016/0169-2607(95)01640-F CMPBEK 0169-2607 Google Scholar

6. 

J. Ramella-RomanS. PrahlS. Jacques, “Three Monte Carlo programs of polarized light transport into scattering media: part I,” Opt. Express, 13 (12), 4420 –4438 (2005). http://dx.doi.org/10.1364/OPEX.13.004420 OPEXFF 1094-4087 Google Scholar

7. 

F. JaillonH. Saint-Jalmes, “Description and time reduction of a Monte Carlo code to simulate propagation of polarized light through scattering media,” Appl. Opt., 42 (16), 3290 –3296 (2003). http://dx.doi.org/10.1364/AO.42.003290 APOPAI 0003-6935 Google Scholar

8. 

S. P. MorganI. Stockford, “Surface-reflection elimination in polarization imaging of superficial tissue,” Opt. Lett., 28 (2), 114 –116 (2003). http://dx.doi.org/10.1364/OL.28.000114 OPLEDP 0146-9592 Google Scholar

9. 

I. M. Stockfordet al., “Analysis of the spatial distribution of polarized light backscattered from layered scattering media,” J. Biomed. Opt., 7 (3), 313 –320 (2002). http://dx.doi.org/10.1117/1.1483316 JBOPFO 1083-3668 Google Scholar

10. 

W. S. Bickel, “Stokes vectors, Mueller matrices, and polarized scattered light,” Am. J. Phys., 53 (5), 468 –468 (1985). http://dx.doi.org/10.1119/1.14202 AJPIAS 0002-9505 Google Scholar

11. 

S.-H. Tsenget al., “Investigation of a probe design for facilitating the uses of the standard photon diffusion equation at short source-detector separations: Monte Carlo simulations,” J. Biomed. Opt., 14 (5), 054043 (2009). http://dx.doi.org/10.1117/1.3253386 JBOPFO 1083-3668 Google Scholar

12. 

V. V. Tuchin, Tissue Optics, SPIE, Bellingham, WA (2007). Google Scholar

13. 

Y. Liuet al., “Investigation of depth selectivity of polarization gating for tissue characterization,” Opt. Express, 13 (2), 601 –611 (2005). http://dx.doi.org/10.1364/OPEX.13.000601 OPEXFF 1094-4087 Google Scholar

14. 

J. Falconetet al., “Analysis of simulated and experimental backscattered images of turbid media in linearly polarized light: estimation of the anisotropy factor,” Appl. Opt., 47 (31), 5811 –5820 (2008). http://dx.doi.org/10.1364/AO.47.005811 APOPAI 0003-6935 Google Scholar

15. 

W. F. CheongS. A. PrahlA. J. Welch, “A review of the optical properties of biological tissues,” IEEE J. Quant. Electron., 26 (12), 2166 –2185 (1990). http://dx.doi.org/10.1109/3.64354 Google Scholar
© 2013 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2013/$25.00 © 2013 SPIE
Simon Rehn, Anne Planat-Chrétien, Michel Berger, Jean-Marc Dinten, Carole Deumié-Raviol, and Anabela da Silva "Depth probing of diffuse tissues controlled with elliptically polarized light," Journal of Biomedical Optics 18(1), 016007 (7 January 2013). https://doi.org/10.1117/1.JBO.18.1.016007
Published: 7 January 2013
Lens.org Logo
CITATIONS
Cited by 23 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Polarization

Tissues

Optical properties

Photons

Scattering

Monte Carlo methods

Light scattering

Back to Top