Open Access
14 January 2015 Swept source optical coherence tomography Gabor fusion splicing technique for microscopy of thick samples using a deformable mirror
Christopher Costa, Adrian Bradu, John Rogers, Pauline Phelan, Adrian Podoleanu
Author Affiliations +
Abstract
We present a swept source optical coherence tomography (OCT) system at 1060 nm equipped with a wavefront sensor at 830 nm and a deformable mirror in a closed-loop adaptive optics (AO) system. Due to the AO correction, the confocal profile of the interface optics becomes narrower than the OCT axial range, restricting the part of the B-scan (cross section) with good contrast. By actuating on the deformable mirror, the depth of the focus is changed and the system is used to demonstrate Gabor filtering in order to produce B-scan OCT images with enhanced sensitivity throughout the axial range from a Drosophila larvae. The focus adjustment is achieved by manipulating the curvature of the deformable mirror between two user-defined limits. Particularities of controlling the focus for Gabor filtering using the deformable mirror are presented.

1.

Introduction

Adaptive optics (AO)1 have been used to great effect in astronomy, eye imaging,2 and confocal microscopy (CM) systems by exploiting the capability to reduce system- and sample-induced optical aberrations to enhance resolution.3 Adding AO to an optical coherence tomography (OCT) setup enables further reduction of the voxel size of the imaging system, producing a much smaller three-dimensional volume per voxel due to enhanced transversal resolution, in addition to axial.4,5 Additionally, swept source OCT (SS-OCT) can provide exceptional imaging speeds due to the rate at which the wavelength can be tuned.6,7 The combination of these technologies has produced fast imaging systems capable of very high resolution, ideally suited for traditionally difficult clinical application, such as ophthalmology.8

1.1.

Problems Presented with AO/SS-OCT

There is a trend in developing long coherence length swept sources that can determine an axial range extending over 1 cm and more.9 On the other hand, an AO enhanced confocal microscope system can shrink the confocal profile width to values of tens of micrometers or even smaller. As the SS-OCT method presents the disadvantage that the reflectance of all points along the axial range are measured under a fixed focus, maximum sensitivity within the cross-sectional (B-scan) OCT image is achieved in the focus only. If the full width at half maximum (FWHM) of the confocal profile acting as a confocal gate (CG) is smaller than the OCT axial range (AR), then the OCT image presents high contrast and good transversal resolution within a spatially restricted stripe determined by the CG only.

1.2.

Possible Solution

The problem presented is that the region of highest quality (as determined by contrast and resolution) is limited by the width and position of the CG. Efforts have been made to extend the depth of focus (DOF) while maintaining high lateral resolution. One approach uses an axicon lens to produce a Bessel beam, extending the DOF when compared with a Gaussian beam. The DOF is enlarged at the expense of lowered sensitivity, making the method less suitable for imaging samples of low reflectance,10,11 although progress has been made to improve its efficiency.12

A solution was developed based on Gabor filtering13 that shifts the CG incrementally through the sample; an OCT cross-sectional image was acquired at five different focus positions (R) as manipulated by an electrical lens. From each of the five images, the in-focus region was extracted and then spliced together to form a single image wherein all regions are in focus.

It was determined that five repetitions of data acquisition with shifted focus were sufficient as the CG profile was approximately one-fifth of the AR. Due to the enhanced coherence length of swept sources, the ratio R=AR/CG may be much larger than 5, in which case the Gabor procedure may considerably slow the acquisition. This presents a problem as the process is largely manual, with the margin for error increasing as there may be variations in the imaging parameters over time, especially with living samples. The next logical step would be the design of a system that simplifies and expedites this procedure.

2.

Methodology

As detailed in Fig. 1, an AO configuration is presented where a deformable mirror (DM), normally employed by the AO closed loop to correct the wavefront, is also used to shift the focus for Gabor filtering. A DM was already reported as the element for focus control in a time-domain OCT (TD-OCT) system.14 The focus was moved in synchronism with the coherence gate, achieving what is known as dynamic focus.15 This procedure is not applicable in spectral domain OCT. In addition, in a closed-loop AO configuration, attempts at focus control have to be disconnected from the feedback loop operation which actively works to reduce aberrations (including defocus). In a related report,16 an AO system using a DM combined with spectrometer-based OCT was used to control the focus position and shape of the confocal gate by actuating on the spherical aberration introduced by the DM itself. Here, we actuate on a different aberration component, defocus, and use such a control to implement Gabor filtering as detailed.

Fig. 1

Schematic of the AO/SS-OCT system with imaging at 1060 nm and wavefront sensing plus confocal microscopy at 830 nm.

JBO_20_1_016012_f001.png

2.1.

Concept

Given the speed of SS-OCT imaging which allows frame rates as high as 100 Hz, there is room for sufficiently high values of R to still secure a video rate, which is important in live imaging systems.17 R B-scans are acquired where the region of maximum intensity as determined by the CG profile is shifted deeper within the object along the z coordinate. After acquisition, each image obtained in this way may then be cropped to isolate the in-focus regions in depth. The cropped images are subsequently assembled together across R into a single synthesized image of uniform intensity.

The focus control using a DM is achieved by generating a set of corrections by controlling the magnitude of the Zernike polynomial responsible for that of defocus only. Closed-loop correction will be conducted to optimize the wavefront, providing a base configuration for the DM. The focus will then be shifted and the closed loop executed again to achieve optimization at the new position. This will provide two states (base and modified states) from which intermediate positions may be interpolated. In essence, we would have a set of states where the corrections have been optimized for imaging at different axial positions in the sample.

2.2.

Optical Design for Minimizing System-Induced Aberrations

This implementation of DM control utilizes a system of passive and active aberration correction to improve image quality. Passive correction is achieved through an optimized system wherein the number of optical elements (specifically the curved mirrors required to resize the beam, known for inducing aberrations) has been minimized. Collimators (Schafter and Kirchhoff: FC-F-4-M15-37) launch beams of 3 mm diameter.

By using the two galvoscanners, X and Y (5 mm), close to each other, the number of curved mirrors is reduced. Two telescopes (pairs of curved mirrors with different focal lengths) are placed in the setup to increase the beam diameter to 15 mm to fully cover the reflective surface of a DM (Imagine Optics: Mirao-52e), and then to reduce the beam down to 7.5 mm on the final lens (L2) of the 25-mm focal length.

Furthermore, these telescopes are arranged in a nonplanar configuration with the beam propagating in a single plane, either horizontally or vertically. This is seen in Fig. 1 where the beam travels between curved mirrors CM1 and CM2 in the plane Y, with no deviation in X. The beam is then propagated horizontally across the surface of the DM and then vertically to CM4, at the same elevation with CM1. This technique has been shown to compensate for astigmatism aberrations induced by reflections from curved mirrors.18 We measured the sensitivity of the OCT setup to be 100 dB using previously reported techniques.19 The numerical aperture of the system was calculated to be 0.175. For imaging, we use a 100-kHz Axsun swept source with a central wavelength at 1060 nm.

We experimentally determine the field of view by blocking the reference arm (using the confocal signal only) to image a large bar on the United States Air Force (USAF) resolution target. For this measurement, the DC output of one photodetector from the balanced detector was used. The target is mounted on an XYZ translation stage, allowing movement of the bar to the limits of the displayed image and measurement of the displacement using the micrometer screws. From this, we determine that for an amplitude of 1 V of driving signals applied to both scanners we achieve a 700×700μm2 raster size, and with a 0.1 V amplitude the area is 70×70μm2.

2.3.

Active Aberration Compensation Using Deformable Mirror

Active wavefront correction is achieved using a single 52-actuator MEMS DM in conjunction with a Shack–Hartmann sensor (Imagine Optics: HASO-32). Wavefront sensing is performed using light from a 830-nm broadband super luminescent diode (SLD), whose beam path is almost completely shared with the 1060-nm imaging arm. Prior to imaging, using a scattering target such as paper, correction is achieved that brings the DM into an optimal starting configuration where any residual aberrations induced by the optical elements in the system itself are reduced.

With the inclusion of tilts and focus in the wavefront error calculations, we are able to achieve a wavefront roor-mean-square (RMS) value better than 0.2μm. After a stable correction is achieved, the voltage values applied to the DM are saved into an XML file. This file contains a one-dimensional (1-D) array of 52 numerical values, each corresponding to a single actuator on the DM. The values of this array represent the voltage settings of the 52 actuators on the DM that achieve the best wavefront correction. This set is considered a base state, i.e., the midpoint of the axial imaging range to be used next.

2.4.

Shifting the Confocal Gate

The next stage is to modify the amplitude of a single Zernike polynomial, in this case that of defocus is selected, to a different value. This causes the mirror to adjust its shape to add defocus aberration in addition to its corrected state. With defocus added, the closed loop is executed again to compensate for the possible increase in other aberrations caused by manipulation of the focus. No other types of aberrations are directly added in this way, so it can be assumed that the difference between the original base state and the modified state is the addition of defocus only. Again, the modified state file is a 1-D array of 52 values, this time representing the position of every actuator on the DM to form a state with a modified focus. This modified state file is also saved along with the original base state file.

Given that the linearity of the DM is greater than 95%,20 the difference between the base state and the modified state is a linear increase in defocus aberration. Custom software programmed in LabVIEW dynamically interpolates a two-dimensional (2-D) array of intermediate voltage values for every actuator using these two 1-D arrays. The size of the 2-D array is R by 52, where R is the number of steps required. Each successive column in this 2-D array contains a complete set of 52 actuator voltage values, representing an incremental shift in focus.

This 2-D array can be expanded to any size and any range of focus, limited only by the speed and stroke of the DM, respectively. A larger array may impact the rate at which the DM actuates the commands. Increasing the size of the array (number of columns) will adjust the number of incremental steps (R) between two user-defined limits. Shifting the confocal gate by 40μm in 50 steps covers a 2 mm range as experimentally determined using a mirror as an object and finding the new focus position for the deformed state of the DM.

Adjusting the limits of the array affects the magnitude of defocus aberration applied to the DM; in this way, the AR can be dynamically adjusted while scanning to suit the sample thickness. To produce a smooth focus sweep at a consistent rate, we typically make these limits symmetrical, i.e., between 1 and 1 mm with the midpoint representing zero defocus added. The visible effect on the image is a sweeping region of high intensity through the depth range of the sample.

2.5.

Analysis of the Confocal Profile and Signal Strength During Focus Sweep

It is important to determine how the confocal profile is affected as we shift the focus through the sample. To test the imaging performance, we set the DM to assume the base state corrected configuration. During the measurement, we set the AO to run in open loop, i.e., the corrections are maintained without feedback from the wavefront sensor (WFS). In this mode, we may still directly manipulate the mirror to add defocus.

A mirror is mounted on a XYZ translation stage and placed as the object in the imaging arm. We block the reference arm and use the confocal signal (with weak light emitted from the SLD to avoid saturating the detector) to determine the confocal profile of the system at this default state. The DM is then set to add different values of defocus in increments of 1μm RMS. For each new focus adjustment, the translation stage with the mirror was moved around the focussed position and the strength of the confocal signal was measured to construct the confocal gate profile.

Confocal profiles for three focus adjustments are shown in Fig. 2. From the graphs, we can ascertain two critical facts about the focus sweep. First, it is clear that the image intensity drops as the DM introduces more defocus.

Fig. 2

Chart showing how the confocal profile shifts in depth and changes shape as we add 1 and 2μm RMS focus.

JBO_20_1_016012_f002.png

Second, it is important to note that the width of the profile increases with defocus. We start with a FWHM of 60μm in the base state, which first spreads to a FWHM 70μm as defocus aberration is increased to 1μm RMS, and then spreads further to FWHM 80μm with 2μm RMS defocus. The enlargement of the confocal profile and reduction of its maximum with the defocus applied suggest deterioration of the correction. Naturally, this would decrease the depth resolution of the confocal microscope (the object arm), though this is not a concern with OCT imaging.

The sensitivity decay with defocus is quantified for larger defocus values in Fig. 3. In this case, for each defocus value applied by the DM, the object mirror was moved back into focus using the translation stage and the strength of the confocal signal was measured. There appears to be a sharp drop as the focus first moved from base state, and then recorded intensity decreases below 50% for ±3μm RMS defocus, reaching 10% for ±8μm RMS defocus. The power profile is shown as a symmetric around base state when adding positive or negative focus.

Fig. 3

Chart showing decay in signal intensity with added focus.

JBO_20_1_016012_f003.png

The chart in Fig. 3 can be used to select a useful range to operate the DM where the sensitivity loss can be tolerated for a given imaging object. This sensitivity loss is a limitation of this technique and is not observed with other methods of focus control such as using an electrical lens.

From the data shown in Fig. 2, it is clear that applying 1-μm RMS defocus equates to an axial shift of the focus by 250μm; therefore, a sweep between 2 and 2μm RMS would offer a 1-mm axial imaging range.

2.6.

Assessment of the Transversal Resolution

To test practical image resolution, an image is acquired of a USAF 1951 target, as shown in Fig. 4. We know from the data shown in Sec. 2.2 that for driving signals of 0.1 V amplitude applied to both scanners we achieve a scan area of 70×70μm2 on the sample. Given an image size of 250×250 pixels, it can be deduced that each pixel covers 0.28μm along the horizontal and vertical directions.

Fig. 4

High contrast confocal image of the USAF target showing the smallest 2μm bars at the bottom. Two sets of three bars are shown horizontally and vertically oriented. Image size is 250×250 pixels covering an actual area of 70×70μm on the target.

JBO_20_1_016012_f004.png

Figure 4 shows an image of the USAF target where two sets of three bars (each 2μm apart) are shown distributed both vertically and horizontally. Given the parameters described above, we may analyze the distance between the bars in pixels to determine the actual spacing in μm. Both the horizontal and vertical bars are shown to be 7 pixels apart: 7×0.32=1.96μm apart. This confirms our reported transversal resolution values.

To determine how lateral resolution is affected when axially shifting the CG, the USAF 1951 resolution target is used as the object and images are acquired for different defocus values. Initially, we assume the base state with no defocus, and position the object such that the set containing the smallest bars (of 2μm spacing) is within the center of the frame. Figure 5 shows four additional images taken of the same region of the target, with different levels of defocus applied to the DM. Each time defocus is applied to the DM, the USAF target is manually moved axially back into focus before acquiring the image. The distance between the bars in the images obtained with focuses 1, 1, and 2 (focus moved at 250, +250, and +500μm, respectively) are the same as in the image at focus 0. There appears to be a loss in signal intensity in the image acquired with focus 2 (500μm), which is not apparent in the others, but the transversal resolution seems intact.

Fig. 5

Effect of defocus on the image acquired from the USAF target, using the same area as in Fig. 4. The original image (c) is repeated in (d) and placed alongside images taken with 2 (a), 1 (b), 1 (e) and 2 (f) μm RMS focus added to the deformable mirror (DM).

JBO_20_1_016012_f005.png

Some lateral drift is also identified, seen by the bars moving across the image as the focus is changed. From 5(c) to 5(b), the bars moved up by 7 pixels and from 5(b) to 5(a) by a further 7 pixels. From 5(d) to 5(e), the bars moved down by 7 pixels, and from 5(e) to 5(f), by a further 7 pixels. Seven pixels’ vertical shift means 2μm, i.e., comparable and larger than the transversal resolution. In terms of the horizontal shift of the image, from 5(c) to 5(b) there is a slight shift of 1 pixel, while from 5(b) to 5(a) there is shift of 7 pixels to the left. From 5(d) to 5(e), there is a lateral shift to the right of 7 pixels and from 5(e) to 5(f) an additional 7 pixels to the right.

The experiments reported in Sec. 2.5 show that the variation of focus imprinted by the DM when obtaining the images in Fig. 5 covers a 1-mm axial range. Using the same process described above, the effect of applying larger defocus values to the DM is evaluated by the resolution in the images acquired, shown in Fig. 6.

Fig. 6

Images of the USAF target obtained using the DM deformed to create 3μm (b) and 4μm (c) RMS defocus, with the original 0μm focus image (a) for comparison.

JBO_20_1_016012_f006.png

In the image in Fig. 6(b), the bars moved to the right by 24 pixels (top of the bar barely visible), i.e., by 6.9μm. In the image in 6(c), the bars moved to the right by 28 pixels, i.e., by 8μm. In terms of vertical shift, the bars moved down in 6(b) by 21 pixels and in 6(c) by 28 pixels. On average, from images in Figs. 5 and 6, it can be inferred that each 1μm defocus shifts the image down and to the right by approximately 7 pixels and each 1μm defocus shifts the image up and to the left by approximately 7 pixels, i.e., 2μm or a resolution interval.

In conclusion, RMS focus control via the DM in the configuration assembled has a pronounced effect on the lateral drift. It is important to note here that the lateral resolution as determined by the horizontal bars appears to remain unaffected, though the vertical bars have become distorted and blurry. Practically, a good signal is obtained from the central part of the image only. We believe this is due to the fact that the correction file was taken with the scanners in a zeroed position, meaning the corrections are less valid for the periphery of the imaged area.

Some lateral shift is expected due to the off axis configuration, typical in AO systems using DMs. This can be further reduced by increasing the focal length of the spherical mirrors CM2 and CM3, with the obvious disadvantage of enlarging the size of the layout and a reduction in the axial range covered. Another factor is the imaging beam potentially being slightly off-axis to the USAF target. A combination of these two factors may be responsible for the slight horizontal shift between images in Figs. 5(b) and 5(c). We noticed that by tilting the USAF target, the lateral shifts change. This prevents postprocessing compensation by software.

2.7.

Effect upon Wavefront When Manipulating Focus

In-depth analysis of the effect of focus manipulation on other aberrations was conducted. WFS measurements are taken as the DM is set to varying levels of defocus. Figure 7 shows the overall wavefront error as detected by the WFS, with focus varied between 0.5 and 0.5 mm. It can be observed that tilt significantly contributes to the overall wavefront error in comparison to the other aberrations. It can also be inferred from the chart that the total wavefront error excluding defocus is five times less than the value with defocus. This demonstrates that the contribution of defocus to wavefront error is much greater than the aberrations that increase as a consequence of focus manipulation.

Fig. 7

Root mean square (RMS) wavefront error as focus is manipulated. Total error: green region. Error excluding focus: blue region. Error excluding tilts and focus: red region.

JBO_20_1_016012_f007.png

A breakdown of the first eight aberrations is shown in Fig. 8, where it can be seen that all these aberrations increase as defocus is applied in either direction. Defocus aberration is plotted for the reference (green), and appears linear as expected. The red and blue dashed lines represent fluctuating tilts 0 and 90, respectively, and seem to exhibit the highest variance.

Fig. 8

Effect of focus manipulation on defocus aberration (green) and on seven other high-order aberrations (according to color in the inset).

JBO_20_1_016012_f008.png

Figure 9 displays the RMS error for the data in Fig. 8, excluding the tilts and focus aberrations. From this graph, it is seen that the effect of focus manipulation on these aberrations is within a range between 0.1 and 0.07μm RMS. It is possible that these larger values at the extremeties of the range are due to the limited dynamic range of the DM used. If this is the case, these can be made smaller if a second wavefront corrector is employed.

Fig. 9

Magnified view of WFS RMS error showing effect of focus on wavefront excluding tilts and focus aberrations.

JBO_20_1_016012_f009.png

2.8.

Test Procedure

Measurements to test the procedure are conducted on a phantom target constructed using 10 layers of borosilicate glass microscope slide covers, each of thickness between 0.16 to 0.19 mm. Each slide was bound using a single layer of thin clear tape. As the clear tape is extremely thin, distances between the slides are not resolved and so, only 11 interfaces should be distinguished. As seen in Fig. 10, even after some tilt, extra layers appear at the bottom and some layers in between the main interfaces, due to multiple reflections between the slide facets. The total thickness of the sample is approximately 2 mm.

Fig. 10

Images acquired of a phantom constructed from ten 0.17-mm thick glass slides (bound top and bottom with sellotape of similar thickness) with focus set to 4 (a), 0 (b) and 4 (c) covering a total range of 2 mm. The final image (d) is synthesized through extraction of in-focus regions from a stack of images acquired with focus incrementally shifted.

JBO_20_1_016012_f010.png

During live imaging, a sequence of commands is sent to the DM in rapid succession, with an optional time delay between each command. The magnitude of defocus aberration is swept between the limits of ±4μm RMS using the DM (determining an axial range of 2 mm in depth). The procedure was tested with 50 steps of 40μm each at a rate of 200 commands sent to the DM per second. Oversampling by using a high number of small steps was found to produce a smoother focus transition. Transversal size was maintained at 1 mm and the refresh of the B-scan OCT imaging was 100 Hz.

3.

Optical Coherence Tomography Imaging Results

Figures 10(a) and 10(c) show images acquired with the focus position at respective negative (1mm) and positive (+1mm) extremes, while Fig. 10(b) shows an intermediate unmodified state. While images are constantly being acquired, we incrementally sweep the focus through 50 steps from one limit to the other, recording a video of the results. The final image in Fig. 10(d) is the Gabor image reconstructed using in-focus regions of all frames acquired during this sweep cycle. Figure 10(d) has mostly uniform contrast throughout the range.

We can clearly see in Figs. 10(a) to 10(c) a horizontal stripe of high intensity shifting vertically in the image as the commands are applied to the DM. Figures 10(a) to 10(c) allow an approximate visual evaluation of the CG profile under AO correction. The profile width in 10(a) looks similar to that in 10(b), where according to Fig. 2 the CG should exhibit a FWHM of 60μm. The profile in 10(c) displays a much wider width. This suggests that other aberrations are added by the intervening layers, affecting the CG profile at +4μm RMS more than at 4μm RMS.

Considering the FWHM of the CG given by the minimum width of 60μm and given the axial range of 2 mm, a minimum value of 33 is obtained for the ratio R. We used an experimental value of Rmax=50 for oversampling.

The procedure was repeated using an infra-red card as the object, with the DM configured to sweep ±375μm (total axial range of 0.75 mm). Figure 11(d) is assembled by stitching the in-focus regions of images acquired during the focus sweep [Fig. 11(a) through 11(c)]. It can be observed that the Gabor image in 11(d) exhibits details in both the superficial layers [(shown in Fig. 11(a)], intermediate layers [Fig. 11(b)] and deeper layers [Fig. 11(c)].

Fig. 11

Optical coherence tomography (OCT) B-scan images of an infra-red card, with focus set to 1.5 (a), 0 (b), and 1.5 (c) covering a range of 0.75 mm. The final image (d) is the composite Gabor image.

JBO_20_1_016012_f011.png

To verify that this procedure is suitable for in vivo imaging of live samples, we performed the same experiment on a wild-type Drosophila larvae. Several larvae were chosen at varying stages of development and secured using strong double-sided tape. All larvae are highly active and move constantly, potentially compromising the integrity of acquired images. This time we reduce the focus sweep to ±500μm, covering a total axial range of 1 mm. Total power to the sample was measured to be 2.3 mW, which is well within the safety range for a scanning beam at 1060 nm.

Raw images from a Drosophila larvae taken with the focus shifted at 0.5mm, 0 and 0.5 mm in air are presented in Figs. 12(a) to 12(c) respectively. Figure 12(d) shows the final Gabor filtered synthesized image, produced by compounding the regions of maximum intensity in each image from a stack of images recorded during a full sweep cycle.

Fig. 12

B-scan-OCT images of in vivo Drosophila with focus at 1 (a), 0 (b), and 1 (c) covering a total range of 1-mm axial range measured in air. The final image (d) is synthesized through extraction of in-focus regions from a stack of images acquired with focus incrementally shifted.

JBO_20_1_016012_f012.png

Finally, the procedure was tested upon the thumb of a volunteer using an axial sweep range of 1 mm. Figure 13(a) shows an image taken with static focus in the base state. The surface of the thumb is highly visible, as is the general position and structure of two sweat ducts. The distinctive spiral structure of the sweat duct is well documented in the literature.21 Figure 13(b) shows the thumb with focus swept deeper into tissue; not only have the sweat ducts become more clearly defined, but also the deeper structures of the epidermis.

Fig. 13

In vivo OCT B-scan images acquired of thumb with fixed focus (a) at the surface and after Gabor splicing (b). Visible are the distinctive spiral structures of the sweat ducts, and the deeper tissues are enhanced in (b) due to focus sweep. Vertical scale: 2 mm measured in air.

JBO_20_1_016012_f013.png

To produce the synthesized images in Figs. 10(d), 11(d), 12(d), and 13(b), ImageJ was used postacquisition. Given the speed of the image acquisition and DM command rate, it would be a simple evolutionary step to perform the splicing of cropped images live.

Artifacts may form in the synthesized image should the sample move during acquisition, which is a common problem in live imaging. One avenue of compensation is to increase the rate at which commands are sent to the DM, either by adjusting the delay between each command or simply reducing the number of incremental focus shifts to the minimum required for sufficient sampling as determined by R.

4.

Conclusions

The paper demonstrates that fine control over the curvature of a DM can be employed to perform Gabor splicing and yield better uniformity in terms of signal intensity throughout depth in a SS-OCT configuration.

This technique presents a clear advantage over possible implementations of focus control using a liquid crystal lens added to an AO configuration, as there exists potential to perform the focus shift while maintaining closed-loop corrections using the DM. Even without updating the correction, the loss of signal up to half of its maximum can be tolerated within a significant axial scanning range of more than 2 mm. In addition, we would emphasize the benefit and sophistication of the DM as we are able to make several hundreds of incremental high-precision focus adjustments every second.

However, a number of shortcomings should be considered in practice. In addition to variations in sensitivity with defocus as shown in Fig. 3, there is a marked lateral shift of the image in both vertical and horizontal directions. This was quantified here using confocal images of a USAF target, observing a lateral shift of approximately 2μm per each 1μm defocus applied to the DM. Large images were used here, where such lateral shifts could be ignored. Small size images, however, would require correction for the lateral shifts.

The lateral shifts induced by deforming the DM were stable in time; however, the dependence of such a shift to the object tilt prevents a universal compensation algorithm. An individual case solution can be implemented by first tracking the lateral shift for a given orientation and then correcting by applying corresponding bias values to the two transversal scanners. Knowing that 100 mV determines an image size of 70μm, a 2μm correction would require approximately 3 mV bias to each driver signal applied to the two transversal scanners. This would have some limited applicability, as individual layers with different inclinations inside the tissue will register different lateral shifts.

In the case presented here, a single DM is used to compensate for all aberrations and then used to manipulate defocus. Advanced designs may use multiple DMs to great effect; for example, one DM may exclusively control tilts to compensate lateral shift. Alternatively, one DM could be set to continuously run closed-loop correction (excluding focus control) while the second DM shifts the focus. This would eliminate the procedure of preprocessing the files for focus manipulation. Utilization of two deformable mirrors may improve the correction of the wavefront when limited by the dynamic range of a single corrector.

It can also be noticed by inspecting Figs. 10(d), 11(d), 12(d) and in 13(b) that the CG stripe can be seen axially moving with defocus aberration in Fig. 10(d), but less in Fig. 11(d) and much less in Figs. 12(d) and Fig. 13(b). This is a clear indication of enlargement of the CG profile due to aberrations created by the intermediate interfaces up to the depth of interest. Also, the variation of brightness among the sequence of images versus focus is not as sharp as illustrated by the profile in Fig. 3. This is another consequence of extra aberrations than those measured in Figs. 8 and 9; scattering caused by intermediate micro interfaces and scattering centers in the Drosophila and the thumb alter the shape and peak of the CG profile. This also shows that the correction determined by the WFS based on the dominant signal collected from the surface of a scattering sample is not so useful for imaging inside the volume of the sample.

Acknowledgments

C. Costa, A. Bradu, and A. Podoleanu acknowledge the support of the European Research Council ( http://erc.europa.eu), Grant No. 249889. A. Podoleanu also acknowledges support from the Biomedical Research Centre at Moorfields Eye Hospital NHS Foundation Trust and UCL Institute of Ophthalmology.

References

1. 

M. Booth, “Adaptive optics for high-resolution microscopy,” Imaging and Applied Optics, OSA, Washington, DC (2013). Google Scholar

2. 

A. Dubraet al., “Noninvasive imaging of the human rod photoreceptor mosaic using a confocal adaptive optics scanning ophthalmoscope,” Biomed. Opt. Express, 2 (7), 1864 –1876 (2011). http://dx.doi.org/10.1364/BOE.2.001864 BOEICL 2156-7085 Google Scholar

3. 

E. J. Fernandezet al., “Adaptive optics with a magnetic deformable mirror: applications in the human eye,” Opt. Express, 14 8900 (2006). http://dx.doi.org/10.1364/OE.14.008900 OPEXFF 1094-4087 Google Scholar

4. 

R. J. Zawadzkiet al., “Adaptive-optics optical coherence tomography for high-resolution and high-speed 3D retinal in vivo imaging,” Opt. Express, 13 8532 (2005). http://dx.doi.org/10.1364/OPEX.13.008532 OPEXFF 1094-4087 Google Scholar

5. 

B. Censeet al., “Volumetric retinal imaging with ultrahigh-resolution spectral-domain optical coherence tomography and adaptive optics using two broadband light sources,” Opt. Express, 17 4095 (2009). http://dx.doi.org/10.1364/OE.17.004095 OPEXFF 1094-4087 Google Scholar

6. 

W. Wieseret al., “Multi-megahertz OCT: high quality 3D imaging at 20 million A-scans and 4.5 GVoxels per second,” Opt. Express, 18 14685 –704 (2010). http://dx.doi.org/10.1364/OE.18.014685 OPEXFF 1094-4087 Google Scholar

7. 

W. Choiet al., “Phase-sensitive swept-source optical coherence tomography imaging of the human retina with a vertical cavity surface-emitting laser light source,” Opt. Lett., 38 338 –40 (2013). http://dx.doi.org/10.1364/OL.38.000338 OPLEDP 0146-9592 Google Scholar

8. 

M. Mujatet al., “High resolution multimodal clinical ophthalmic imaging system,” Opt. Express, 18 11607 –11621 (2010). http://dx.doi.org/10.1364/OE.18.011607 OPEXFF 1094-4087 Google Scholar

9. 

B. Potsaidet al., “Ultrahigh speed 1050 nm swept source/Fourier domain OCT retinal and anterior segment imaging at 100,000 to 400,000 axial scans per second,” Opt. Express, 18 20029 –20048 (2010). http://dx.doi.org/10.1364/OE.18.020029 OPEXFF 1094-4087 Google Scholar

10. 

Z. Dinget al., “High-resolution optical coherence tomography over a large depth range with an axicon lens,” Opt. Lett., 27 243 (2002). http://dx.doi.org/10.1364/OL.27.000243 OPLEDP 0146-9592 Google Scholar

11. 

R. A. Leitgebet al., “Extended focus depth for Fourier domain optical coherence microscopy,” Opt. Lett., 31 (16), 2450 (2006). http://dx.doi.org/10.1364/OL.31.002450 OPLEDP 0146-9592 Google Scholar

12. 

D. Lorenseret al., “Energy-efficient low-Fresnel-number Bessel beams and their application in optical coherence tomography,” Opt. Lett., 39 548 –551 (2014). http://dx.doi.org/10.1364/OL.39.000548 OPLEDP 0146-9592 Google Scholar

13. 

J. P. Rollandet al., “Gabor-based fusion technique for optical coherence microscopy,” Opt. Express, 18 (4), 3632 –3642 (2010). http://dx.doi.org/10.1364/OE.18.003632 OPEXFF 1094-4087 Google Scholar

14. 

B. Qiet al., “Dynamic focus control in high-speed optical coherence tomography based on a microelectromechanical mirror,” Opt. Commun., 232 123 –128 (2004). http://dx.doi.org/10.1016/j.optcom.2004.01.015 OPCOB8 0030-4018 Google Scholar

15. 

M. HughesA. G. Podoleanu, “A simplified dynamic focus method for time domain OCT,” Electron. Lett., 45 (12), 623 –624 (2009). http://dx.doi.org/10.1049/el.2009.0672 Google Scholar

16. 

K. Sasakiet al., “Extended depth of focus adaptive optics spectral domain optical coherence tomography,” Biomed. Opt. Express, 3 2353 –2370 (2012). http://dx.doi.org/10.1364/BOE.3.002353 BOEICL 2156-7085 Google Scholar

17. 

C. Tortiet al., “Adaptive optics optical coherence tomography at 120,000 depth scans/s for non-invasive cellular phenotyping of the living human retina,” Opt. Express, 17 19382 –19400 (2009). http://dx.doi.org/10.1364/OE.17.019382 OPEXFF 1094-4087 Google Scholar

18. 

A. DubraY. Sulai, “Reflective afocal broadband adaptive optics scanning ophthalmoscope,” Biomed. Opt. Express, 2 (6), 1757 –1768 (2011). http://dx.doi.org/10.1364/BOE.2.001757 BOEICL 2156-7085 Google Scholar

19. 

A. G. PodoleanuA. Bradu, “Master-slave interferometry for parallel spectral domain interferometry sensing and versatile 3D optical coherence tomography,” Opt. Express, 21 19324 –19338 (2013). http://dx.doi.org/10.1364/OE.21.019324 OPEXFF 1094-4087 Google Scholar

20. 

E. Ödlundet al., “Control of an electromagnetic deformable mirror using high speed dynamics characterization and identification,” Appl. Opt., 49 G120 (2010). http://dx.doi.org/10.1364/AO.49.00G120 APOPAI 0003-6935 Google Scholar

21. 

M. Ohmiet al., “Dynamic analysis for mental sweating of a group of eccrine sweat glands on a human fingertip by optical coherence tomography,” Skin Res. Technol., 18 378 –383 (2012). http://dx.doi.org/10.1111/srt.2012.18.issue-3 0909-752X Google Scholar

Biographies of the authors are not available.

© 2015 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2015/$25.00 © 2015 SPIE
Christopher Costa, Adrian Bradu, John Rogers, Pauline Phelan, and Adrian Podoleanu "Swept source optical coherence tomography Gabor fusion splicing technique for microscopy of thick samples using a deformable mirror," Journal of Biomedical Optics 20(1), 016012 (14 January 2015). https://doi.org/10.1117/1.JBO.20.1.016012
Published: 14 January 2015
Lens.org Logo
CITATIONS
Cited by 6 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

Confocal microscopy

Adaptive optics

Wavefronts

Deformable mirrors

Image fusion

Image resolution

Back to Top