1 October 2012 Ex vivo determination of glucose permeability and optical attenuation coefficient in normal and adenomatous human colon tissues using spectral domain optical coherence tomography
Author Affiliations +
Abstract
Recent reports have suggested that spectral domain optical coherence tomography (SD-OCT) is a useful tool for quantifying the permeability of hyperosmotic agents in various tissues. We report our preliminary results on quantification of glucose diffusion and assessment of the optical attenuation change due to the diffusion of glucose in normal and adenomatous human colon tissues in vitro by using a SD-OCT and then calculated the permeability coefficients (PC) and optical attenuation coefficients (AC). The PC of a 30% aqueous solution of glucose was 3.37±0.23×10 −6   cm/s in normal tissue and 5.65±0.16×10 −6   cm/s in cancerous colon tissue. Optical AC in a normal colon ranged from 3.48±0.37 to 2.68±0.82  mm −1 and was significantly lower than those seen in the cancerous tissue (8.48±0.95 to 3.16±0.69  mm −1 , p<0.05 ). The results suggest that quantitative measurements of using PC and AC from OCT images could be a potentially powerful method for colon cancer detection.

1.

Introduction

It was estimated that more than one million new cases of colorectal cancer were clinically diagnosed in 2008. This disease kills nearly half a million people worldwide each year, with the majority of these patients suffering from adenocarcinomas.1 Early stages of colon adenocarcinoma can change the macro-architecture of the colon tissue, and mainly include an increase in microvascularisation, the blood content of the tissue lesion,2 and the distortion of collagen matrix.3 In addition, the significant increase in the concentration of blood and proteins in the cancerous tissue over that in normal tissue may lead to the differences in optical properties and permeability of an analyte between normal and malignant human colon tissues. In other words, the abnormalities in malignant tissues can alter physical and physiological properties of the specimen and can cause the change of diffusion rate of a certain analyte.4 Measurement and understanding of these changes could be utilized to detect and identify the cancerous tissue, which may be helpful for the early diagnosis of colon cancer.

Currently, colonoscopy is the most commonly technique used for early detection of colorectal cancer, but this method is only capable of surface visualization, which limits its ability to detect precancerous changes.5 Endoscopic ultrasound, which is capable of depth-resolved imaging, is currently used in the clinical and research settings, but its resolution in standard devices is limited.6 Hence, early detection of neoplastic changes before metastasis remains a critical objective in colorectal cancer diagnosis and treatment. Modern optics can provide real-time imaging of human tissues with resolutions that are comparable to that of histopathology, as well as having the potential ability to reveal biochemical and/or molecular information. These modern optical approaches could significantly improve the identification of malignancies at curable stages.7

Optical coherence tomography (OCT), a noninvasive technology for microstructure imaging of biological tissues, has been used to quantify the permeability of a hyperosmotic agent in atherosclerotic vascular tissue, cancerous gastric tissue, breast cancer tissue, and sclera and skin tissues ex vivo and in vivo.812 Additionally, OCT can be used for localized quantitative measurement of attenuation coefficients (AC), which can provide additional information for discrimination of the different tissues.1315 The scattering properties of the tissue depend on the difference of refractive index-matching between the cells and extracellular fluid.16 Partially applying an analyte can reduce the optical discontinuity in the extracellular fluid and the cell membrane, and therefore, reduce light scattering. OCT has been used to discriminate different structural features of the normal and atherosclerotic vascular tissues,17 apoptosis and necrosis in human fibroblasts. This shows that OCT is sensitive to the changes of the attenuation coefficient caused by analyte diffusion, which is induced by morphological changes of biological tissue.18 Thus, we can adopt the OCT technique for assessing the differences of permeability coefficients (PC) and AC in normal and cancerous tissues caused by hyperosmotic agents. This could assist in distinguishing cancerous from nonmalignant tissue, and possibly result in early colon diagnosis.

In this paper we report the results of our studies on the real-time monitoring, differentiation, and quantification of diffusion of the glucose solution in human normal and adenomatous colon tissues in vitro using spectral domain OCT (SD-OCT). Finally, we also compare the characteristics of the glucose permeability and optical AC properties of normal and malignant colon tissue. The results show PC and AC calculated from OCT images are helpful for the detection of cancerous colon tissue in the early stages.

2.

Materials and Methods

2.1.

SD-OCT System

Ex vivo experiments were performed with a custom-built SD-OCT system, the schematic of which is shown in Fig. 1. A broadband super luminescent diode light source (SLD, Inphenix Co. Ltd., Livermore, CA) was used as the optical source. The central wavelength, bandwidth, and output power were 830 nm, 40 nm and 5 mW, respectively, with the light source coupled into a 50/50 optical fiber coupler. Then the interference signal of the light beam reflected from the reference and sample arms was collected by using a spectrometer, which consisted of a transmitting grating (1200-line/mm), focus lens (f=200mm), and a linear CCD camera (2048 pixels, 14μm/pixel, e2V). The system operated at an A-scan frequency of 20 kHz, the signal-to-noise ratio was measured at 120 dB and the frame rate was 20 fps. The axial and transverse resolution was 12 and 15 μm, respectively. Two-dimensional (2-D) images were obtained by scanning the incident beam over the sample surface in the lateral direction. The SD-OCT system operation was entirely controlled by a personal computer. Two-dimensional OCT images were acquired from each experiment and stored for post-processing.

Fig. 1

Schematic drawing of the SD-OCT system: (CM-collimator, DG-diffraction grating, FL-focusing lens, RM-reference mirror and OL-objective lens) used in the experiment.

JBO_17_10_105004_f001.png

2.2.

Materials and Measurements

Patients volunteered for the research program conducted at the Second Affiliated Clinical Hospital of Guangzhou University of Traditional Chinese Medicine, China and signed consent forms, with the procedure being approved by the local Ethics Committee. Colon tissue samples were stored in 0.9% sodium chloride solution after the resection, and then placed on ice and transported to the laboratory for performing measurements as soon as possible. All the samples were divided into two groups according to pathological diagnosis: normal colon tissues (11 samples), adenomatous colon tissues (11 samples). In order to ensure the hydration and colon tissue samples integrity, each sample was rinsed briefly in saline to remove excess surface blood and fat tissue. Each colon sample tissue was stored in a refrigerator at 70°C, and then was cut into approximately 1.5×1.5cm size pieces for in vitro measurement. As in previous studies,1921 to guarantee the minimal changes in the physiological status and optical properties of low temperature samples, all tissue samples were prepared and measured within 12 h after removal. Prior to OCT imaging, tissue samples were brought to room temperature and defrosted in physiological saline for approximately 30 min. Baseline values were obtained from a selected region of the sample and monitored for approximately eight to 10 min prior to treatment with 30% glucose. The 30% glucose was then added to the colon tissues’ sample surface for 5 min, then solutions carefully removed prior to OCT imaging. Every 30 min during the measurement, glucose was applied again on the tissue for 5 min. The 2-D images were continuously obtained by scanning the sample in the lateral and axial direction by the OCT. Each sample was used only once. The entire ex vivo OCT imaging was monitored for approximately 2 h at room temperature (22°C) throughout the experiment. A 30% glucose (Tianjin Damao Chemical Reagent Factory, China) solution in dH2O, with the mean refractive index of 1.39,22 was used in the diffusion experiments. A reasonable assumption for the average refractive index of colon tissue is n>1.4.23

2.3.

Calculation Methods

The PC of the glucose in the tissues were determined by analyzing the changes in the slope of OCT signal intensity induced by the glucose diffusion in a specific depth region. This method was described in detail in previously reported works.2428A one-dimensional (1-D) curve displaying the distribution of OCT signal intensity in depth plotted in a logarithmic scale was created by averaging the 2-D images laterally. A region in the 1-D OCT signal intensity profile was chosen where the signal was relatively linear and with minimal oscillations. The slope was then calculated from the chosen intensity profile for further analysis,29,30 with the calculated OCT signal slopes (OCTSS) normalized and plotted as a function of time. The PC of 30% glucose in the normal and malignant colon tissue was calculated using the following equation: (P¯=zregion/tregion), where P is the permeability coefficient, zregion is the thickness of the chosen section, and tregion is the time for the glucose to diffuse through that section.8,11,31 The penetration time was measured from the point where the OCTSS started to decrease to the point at which the reverse process began.

The optical attenuation coefficient of the tissue can be quantified from the intensity of the detected light versus the depth. For media with absorption as described by the single-scattering approximation, the light travels in a ballistic way and Beer’s law can be applied to calculate the total OCT attenuation coefficient: μt=μa+μs, where μa is the absorption coefficient and μs is the scattering coefficient.32 These are physical properties unique to the biological tissue, which play a vital role in the assessment of the tissue feature.15,33,34 In this current OCT system case, the measured signal is defined as: 15,3335

Eq. (1)

[i2(z)]12(i20)12[exp(2μtz)]12,
where the i2(z) is the photo detector heterodyne signal current received by an OCT system from the probing depth z and the mean square heterodyne signal i20.The result of the OCT study is the measurement of optical backscattering or reflectance R(z)[i2(z)]1/2 from a tissue versus axial ranging distance, or depth, z. The reflectance depends on the optical properties of tissue, i.e., the total attenuation coefficient μt. Thus, combined with Eq. (1) and R(z) it follows that the reflected power can be approximately proportional to—μtz in exponential scale according to the single scattering model:

Eq. (2)

R(z)=I0a(z)exp(μtz).

Here I0 is the optical power launched into the tissue sample and a(z) is the reflectivity of the tissue sample at the depth of z. Therefore, measurement of OCT reflectance for depths z1 and z2 allows for approximately evaluating the attenuation coefficient and its temporal behavior. This evaluation is due to reduction of the tissue-scattering coefficient at the agent immersion if reflectivity a(z) is considered as weakly dependent on depth for a homogeneous tissue layer. The μt theoretically can be obtained from the reflectance intensity measurements at two different depths, z1 and z2:15,3335

Eq. (3)

μt=1Δzln[R(z1)R(z2)],
where Δz=|z1z2|. For more details about the entire formulas derivation process, check Ref. 36. Noise is inevitable in the measurement, thus a final result should be obtained using a least-square fitting method in order to improve the accuracy of determining μt value. An averaged intensity profile as a function of depth was obtained by averaging the 2-D images laterally over approximately 1 mm, which was enough for speckle noise suppression. A best-fit exponential curve was applied to the averaged intensity profiles of each group since the noise in the measurement is unavoidable.

2.4.

Statistical Analysis

The data from all samples were presented as means±SD and analyzed by an SPSS 16.0 software paired-test. The p<0.05 value indicated significant difference.

3.

Results and Discussion

Figure 2(a) to 2(c) showed OCT images of normal colon tissue 0, 15, and 30 min after the topical application of 30% glucose, and Fig. 2(d) to 2(f) showed images of the cancerous colon tissue. Normal colon tissue had a regular and compact appearance and layers was clearly visible, but the cancerous tissue structure appeared disorganized, nonuniform and had many dark crypts. Comparison of Fig. 2(a) and 2(d) demonstrated a remarkable difference between the normal and malignant colon tissue: the backscattering from the cancerous tissue appeared more heterogeneous than that of normal colon tissue. In addition, the cancerous tissue appeared to have a higher scattering than the normal colon tissue, which was most likely due to larger cell nuclei, higher nuclear-to-cytoplasmic ratio, and higher regional tumor cell density in tumor colon tissue.37 Figure 2 also showed that the visibility, contrast, and imaging depth in both tissues were significantly improved after the application of glucose. This change resulted from the diffusion of the topically applied analyte into the extracellular and intracellular space. A refractive index matching environment was created by matching the chemical agents with the main scattering components within the tissue, leading to the enhanced light penetration. And the light penetration was further enhanced with the reduced light scattering due to the dehydration effect after the glucose application.38,39

Fig. 2

OCT images of normal (a)–(c) and the adenomatous colon tissue (d)–(f) at 0, 15, and 30 min, (respectively) after the application of a 30% glucose solution.

JBO_17_10_105004_f002.png

The typical dynamic change of the 1-D OCT normalized signal intensity profiles and corresponding exponential best-fit curves in the experiments were shown in Fig. 3. The degree of change in the OCT signal intensity increased coincided with the glucose penetration of internal structures and a concomitant increase in glucose concentration. The cancerous tissues yielded a much broader but steeper OCT signal profile from Fig. 3(b) than that observed in the normal colon tissue between zero to 30min. These results may be due to the additive effect of the glucose diffusion, which enhances light transportation into the tissue. Thus more photons propagated to the deep reflective surfaces underneath the tissue give rise to a stronger back-reflected signal. OCTSS was calculated from an axial 95-µm region at a tissue depth of approximately 296 µm away from the tissue surface. As shown in Fig. 4, a similar trend was observed in the normal and malignant colon tissue. The dynamic OCTSS decreased due to a reduction of light scattering inside the tissue, which was caused by a local increase in glucose concentration with the increase in time. It is also apparent that the OCTSS decrease in cancerous tissue was faster than that of normal tissue, as shown in Fig. 4(a) and 4(b). In Fig. 4(a), glucose reached the monitored region approximately 20 min after application and took another 47 min for it to completely achieve the diffusion process. In contrast, for the cancer tissue, it only took about 13 min to reach the monitored region, and then another 41 min to completely diffuse the whole region in Fig. 4(b). At that point, a reverse process in the OCTSS was observed. This reverse process could be due to diffusion via concentration gradient differences on either side of the tissue. The net fluid (mainly water) movement from high concentration areas to lower concentration will occur, such that induced water re-enters the tissue after diffusing out until equilibrium is reached.10,11,40 The PC of 30% glucose for the normal colon tissue in Fig. 5 was significantly slower at 3.37±0.17×106cm/s compared with the PC in the adenomatous colon tissue 5.65±0.24×106cm/s (p<0.05).

Fig. 3

Normalized OCT intensity profiles with their corresponding exponential best fit curves zero, 15 and 30 min after the topical application of 30% glucose; (a) normal colon tissue and (b) adenomatous colon tissue.

JBO_17_10_105004_f003.png

Fig. 4

The typical OCTSS graphs for human normal (a) and adenocarcinoma (b) colon tissues during a 30% glucose diffusion experiment, respectively.

JBO_17_10_105004_f004.png

Fig. 5

Comparison of the mean permeability coefficient of 30% glucose diffusion in normal human and adenocarcinoma colon tissues.

JBO_17_10_105004_f005.png

Analysis of the decrease in the OCT signal allowed us to quantitatively determine the optical AC using the OCT technique.41,42 The mean OCT signal from a region of tissue was determined by the specific local optical properties of the region of interest. Quantification of these intrinsic optical properties can provide an extra objective, classification parameter.42 In both cases, the optical AC decreased with the length of time following topical application of the glucose. This may be due to the glucose diffusion into the tissue, wherein tissue contrast is caused mainly by the coefficients of tissue attenuation. And the coefficients of tissue attenuation depend on the tissue interstitial space volume fraction, cell diameters, and tissue architecture. The AC of normal tissue was found to differ significantly from other cancer components with the continuous diffusion of glucose into the tissues [Fig. 6(a) and 6(b)]. In our study the decrease of light attenuation was much more prominent in the cancer tissue than that of normal tissue in the same region where we determined the OCTSS. It also can be seen from Fig. 6(a) and 6(b) that the AC seemed to undertake periodic oscillations when the glucose was added to the tissues. This may be related to some exchange processes in the glucose and differently-sized and -hydrated (collagen, elastin) structures of tissue because of glucose osmotic impact inducing water flux from (dehydration) these structures and back (rehydration) to them inside tissue.43 The AC of OCT signal intensity reached its maximum (3.48±0.37mm1) in the top area and the minimum (2.68±0.82mm1) in the bottom area of the chose region in normal colon tissue. But for cancerous colon tissue, the corresponding values were 8.48±0.95mm1 and 3.16±0.69mm1, respectively. In both cases, the differences were significant (p<0.05). In our experiments the AC of cancerous tissue was quite higher than that of normal tissue in Fig. 6. Cancerous tissue had a higher concentration of hemoglobin, microvascular volume, blood content and protein that consisted mainly of red blood cells, hemachrome and fibrin, as compared with the surrounding unaffected tissue. These elements are known to have a high scattering property, which may explain that the higher AC values were seen in the cancerous tissue.41 The specificity and sensitivity were obtained based on the experimental data through comparing the average values of normal (n=11) and malignant (n=11) colon tissue groups. The specificity and sensitivity of PC was 81.8% and 72.7%, while specificity and sensitivity of AC was 81.8% and 63.6%, respectively. Although the specificity and sensitivity of PC and AC is helpful for the detection of normal and malignant colon tissues, combination of these values with excisional biopsy and histopathology is necessary to differentiate the tissues in practical applications.

Fig. 6

Mean attenuation coefficients of normal (a) and adenomatous colon tissues (b) after topical application of 30% glucose.

JBO_17_10_105004_f006.png

This experiment presented the preliminary results of glucose permeability and optical attenuation in normal and cancer tissues ex vivo using SD-OCT technology. The differentiation between different types of colon tissue is based on alterations in the morphology of the tissues, such as the structural appearance, tissue characterization, and concentration of blood. Alterations in the organization of cells and tissue that composed the adenomas could influence the permeability of agents and optical attenuation in the tissue. The results showed that there was considerable variation in glucose permeability and optical attenuation between normal and malignant colon tissues. Therefore, SD-OCT technology could assist in the early detection and discrimination of tumors by monitoring and quantifying the diffusion of a hyperosmotic agent, as well as the difference of optical AC in the normal and cancerous tissues by SD-OCT imaging.

However, it should be noted that our experiments were performed ex vivo rather than in vivo. The optical properties of excised tissue could change from live tissue. In addition, we measured the samples at room temperature, which may also have a little influence on the results since the optical properties of tissue are temperature-dependent.44 Therefore, an in vivo investigation is needed to indentify these differences on the measurement.

4.

Conclusions

In this paper, we have measured the optical AC and PC of glucose from normal and malignant colon tissue groups using the SD-OCT, and analyzed the changes of optical AC and PC during the region of interest. The results of this study have shown that the malignant colon tissue has higher optical AC and PC than that present in normal colon tissue. These findings also suggest that quantitative analysis of colon tissue optical properties by the SD-OCT technology could benefit the differentiation of the adenomatous human colon tissue from normal tissue and early diagnosis of gastrointestinal cancerous tissue.

Acknowledgments

The research is supported in part by the following grant: National Basic Research Program of China (Nos. 2011CB707504 and 2010CB933903), National Natural Science Foundation of China (81171377, 61273368). The authors thank Dr. T. FitzGibbon for comments and suggestions on earlier drafts of the manuscript.

References

2. 

S. A. SkinnerG. M. FrydmanP. E. O’Brien, “Microvascular structure of benign and malignant tumors of the colon in humans,” Digest, 40 (2), 373 –384 (1995). http://dx.doi.org/10.1007/BF02065424 DCILDN 0963-3308 Google Scholar

3. 

Y. FuruyaT. Ogata, “Scanning electron microscopic study of the collagen networks of the normal mucosa, hyperplastic polyp, tubular adenoma and adenocarcinoma of the human large intestine,” J. Exp. Med., 169 (1), 1 –19 (1993). JEMEAV 0022-1007 Google Scholar

4. 

J. F. Toussaintet al., “Water diffusion properties of human atherosclerosis and thrombosis measured by pulse field gradient nuclear magnetic resonance,” Arterioscler. Thromb. Vasc. Biol., 17 (3), 542 –546 (1997). http://dx.doi.org/10.1161/01.ATV.17.3.542 ATVBFA 1079-5642 Google Scholar

5. 

L. P. Haririet al., “Endoscopic optical coherence tomography and laser-in a murine colon cancer model,” Lasers Surg. Med., 38 (4), 305 –313 (2006). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar

6. 

A. Daset al., “High-resolution endoscopic imaging of the GI tract: a comparative study of optical coherence tomography versus high frequency catheter probe,” Gastrointest. Endosc., 54 (2), 219 –224 (2001). http://dx.doi.org/10.1067/mge.2001.116109 GAENBQ 0016-5107 Google Scholar

7. 

S. Yuanet al., “Simultaneous morphology and molecular imaging of colon cancer,” in IEEE/NIH Life Science Systems and Applications Workshop, 167 –169 (2009). Google Scholar

8. 

K. V. Larinet al., “Quantification of glucose diffusion in arterial tissues by using optical coherence tomography,” Laser Phys. Lett., 4 (4), 312 –317 (2007). http://dx.doi.org/10.1002/(ISSN)1612-202X 1612-2011 Google Scholar

9. 

H. Xionget al., “Application of hyperosmotic agent to determine gastric cancer with optical coherence tomography ex vivo in mice,” J. Biomed. Opt., 14 (2), 024029 (2009). http://dx.doi.org/10.1117/1.3103341 JBOPFO 1083-3668 Google Scholar

10. 

H. Q. Zhonget al., “Quantification of glycerol diffusion in human normal and cancer breast tissues in vitro with optical coherence tomography,” Laser Phys. Lett., 7 (4), 315 –320 (2010). http://dx.doi.org/10.1002/lapl.v7:4 1612-2011 Google Scholar

11. 

M. G. GhosnV. V. TuchinK. V. Larin, “Nondestructive quantification of analyte diffusion in cornea and sclera using optical coherence tomography,” Invest. Ophthalmol. Visual Sci., 48 (6), 2726 –2733 (2007). http://dx.doi.org/10.1167/iovs.06-1331 IOVSDA 0146-0404 Google Scholar

12. 

X. Guoet al., “In vivo quantification of propylene glycol, glucose and glycerol diffusion in human skin with optical coherence tomography,” Laser Phys., 20 (9), 1849 –1855 (2010). http://dx.doi.org/10.1134/S1054660X10170032 LAPHEJ 1054-660X Google Scholar

13. 

D. J. Faberet al., “Quantitative measurement of attenuation coefficients of weakly scattering media using optical coherence tomography,” Opt. Express, 12 (19), 4353 –4365 (2004). http://dx.doi.org/10.1364/OPEX.12.004353 OPEXFF 1094-4087 Google Scholar

14. 

C. Xuet al., “Characterization of atherosclerosis plaques by measuring both backscattering and attenuation coefficients in optical coherence tomography,” J. Biomed. Opt., 13 (3), 034003 (2008). http://dx.doi.org/10.1117/1.2927464 JBOPFO 1083-3668 Google Scholar

15. 

V. M. Kodachet al., “Determination of the scattering anisotropy with optical coherence tomography,” Opt. Express, 19 (7), 6131 –6140 (2011). http://dx.doi.org/10.1364/OE.19.006131 OPEXFF 1094-4087 Google Scholar

16. 

H. Q. Zhonget al., “In vitro study of ultrasound and different-concentration glycerol-induced changes in human skin optical attenuation assessed with optical coherence tomography,” J. Biomed. Opt., 15 (3), 036012 (2010). http://dx.doi.org/10.1117/1.3432750 JBOPFO 1083-3668 Google Scholar

17. 

F. J. van der Meeret al., “Apoptosis- and necrosis-induced changes in light attenuation measured by optical coherence tomography,” Lasers Med. Sci., 25 (2), 259 –267 (2010). http://dx.doi.org/10.1007/s10103-009-0723-y LMSCEZ 1435-604X Google Scholar

18. 

K. V. Larinet al., “Noninvasive blood glucose monitoring with opticalcoherence tomography: a pilot study in human subjects,” Diabetes Care, 25 (12), 2263 –2267 (2002). http://dx.doi.org/10.2337/diacare.25.12.2263 DICAD2 0149-5992 Google Scholar

19. 

P. KarandeS. Mitragotri, “Dependence of skin permeability on contact area,” Pharm. Res., 20 (2), 257 –263 (2003). http://dx.doi.org/10.1023/A:1022231406277 PHREEB 0724-8741 Google Scholar

20. 

Y. Liuet al., “Ultrasound-enhanced drug transport and distribution in the brain,” AAPS. Pharm. Sci. Tech., 11 (3), 1005 –1017 (2010). http://dx.doi.org/10.1208/s12249-010-9458-0 AAPHFZ 1530-9932 Google Scholar

21. 

Q. L. Zhaoet al., “Depth-resolved monitoring of diffusion of hyperosmotic agents in normal and malignant human esophagus tissues using optical coherence tomography in-vitro,” Quantum Electron., 41 (10), 950 –955 (2011). http://dx.doi.org/10.1070/QE2011v041n10ABEH014549 QUELEZ 1063-7818 Google Scholar

22. 

A. N. Bashkatovet al., “Monte Carlo study of skin optical clearing to enhance light penetration in the tissue: implications for photodynamic therapy of acne vulgaris,” Proc. SPIE, 7022 702209 (2008). http://dx.doi.org/10.1117/12.803909 PSISDG 0277-786X Google Scholar

23. 

F. P. Bolinet al., “Refractive index of some mammalian tissues using a fiber optic cladding method,” Appl. Opt., 28 (12), 2297 –2303 (1989). http://dx.doi.org/10.1364/AO.28.002297 APOPAI 0003-6935 Google Scholar

24. 

V. V. Tuchin, “Optical clearing of tissues and blood using the immersion method,” J. Phys. D., 38 (15), 2497 –2518 (2005). http://dx.doi.org/10.1088/0022-3727/38/15/001 JPAPBE 0022-3727 Google Scholar

25. 

K. V. Larinet al., “Specificity of noninvasive blood glucose sensing using optical coherence tomography technique: a pilot study,” Phys. Med. Biol., 48 (10), 1371 –1390 (2003). http://dx.doi.org/10.1088/0031-9155/48/10/310 PHMBA7 0031-9155 Google Scholar

26. 

R. O. Esenalievet al., “Noninvasive monitoring of glucose concentration with optical coherence tomography,” Opt. Lett., 26 (13), 992 –994 (2001). http://dx.doi.org/10.1364/OL.26.000992 OPLEDP 0146-9592 Google Scholar

27. 

M. G. Ghosnet al., “Differential permeability rate and percent clearing of glucose in different regions in rabbit sclera,” J. Biomed. Opt., 13 (2), 021110 (2008). http://dx.doi.org/10.1117/1.2907699 JBOPFO 1083-3668 Google Scholar

28. 

A. T. YehJ. Hirshburg, “Molecular interactions of exogenous chemical agents with collagen-implications for tissue optical clearing,” J. Biomed. Opt., 11 (1), 014003 (2006). http://dx.doi.org/10.1117/1.2166381 JBOPFO 1083-3668 Google Scholar

29. 

M. G. Ghosnet al., “Monitoring of glucose permeability in monkey skin in vivo using Optical Coherence Tomography,” J. Biophoton., 3 (1), 25 –33 (2010). http://dx.doi.org/10.1002/jbio.200910075 JBOIBX 1864-063X Google Scholar

30. 

Q. L. Zhaoet al., “Quantifying glucose permeability and enhanced light penetration in ex vivo human normal and cancerous esophagus tissues with optical coherence tomography,” Laser Phys. Lett., 8 (1), 71 –77 (2011). http://dx.doi.org/10.1002/lapl.v8.1 1612-2011 Google Scholar

31. 

M. G. GhosnV. V. TuchinK. V. Larin, “Depth-resolved monitoring of glucose diffusion in tissues by using optical coherence tomography,” Opt. Lett., 31 (15), 2314 –2316 (2006). http://dx.doi.org/10.1364/OL.31.002314 OPLEDP 0146-9592 Google Scholar

32. 

A. I. Kholodnykhet al., “Precision of measurement of tissue optical properties with optical coherence tomography,” Appl. Opt., 42 (16), 3027 –3037 (2003). http://dx.doi.org/10.1364/AO.42.003027 APOPAI 0003-6935 Google Scholar

33. 

D. Levitzet al., “Determination of optical scattering properties of highly-scattering media in optical coherence tomography images,” Opt. Express, 12 (2), 249 –259 (2004). http://dx.doi.org/10.1364/OPEX.12.000249 OPEXFF 1094-4087 Google Scholar

34. 

Y. Yanget al., “Optical scattering coefficient estimated by optical coherence tomography correlates with collagen content in ovarian tissue,” J. Biomed. Opt., 16 (9), 090504 (2011). http://dx.doi.org/10.1117/1.3625247 JBOPFO 1083-3668 Google Scholar

35. 

L. ThraneH. T. YuraP. E. Andersen, “Analysis of optical coherence tomography systems based on the extended Huygens-Fresenel principle,” J. Opt. Soc. Am. A, 17 (4), 484 –490 (2000). http://dx.doi.org/10.1364/JOSAA.17.000484 JOAOD6 0740-3232 Google Scholar

36. 

V. V. Tuchin, Handbook of Coherent-domain Optical Methods: Biomedical Diagnosis, Environmental Monitoring, 2nd ed.Springer-Verlag, New York (2012). Google Scholar

37. 

S. A. Boppartet al., “Optical coherence tomography: feasibility for basic research and image-guided surgery of breast cancer,” Breast Cancer Res. Treat., 84 (2), 85 –97 (2004). BCTRD6 Google Scholar

38. 

R. K. WangV. V. Tuchin, “Enhance light penetration in tissue for high resolution optical imaging techniques by use of biocompatible chemical agents,” J. X-Ray Science Tech., 10 (3), 167 –176 (2002). JXSTE5 0895-3996 Google Scholar

39. 

V. V. Tuchinet al., “Light propagation in tissues with controlled optical properties,” J. Biomed. Opt., 2 (4), 401 –417 (1997). http://dx.doi.org/10.1117/12.281502 JBOPFO 1083-3668 Google Scholar

40. 

M. Wojtkowskiet al., “Three-dimensional retinal imaging with high speed ultrahigh resolution optical coherence tomography,” Ophthalmolo., 112 (10), 1734 –1746 (2005). http://dx.doi.org/10.1016/j.ophtha.2005.05.023 Google Scholar

41. 

F. J. van der Meeret al., “Localized measurement of optical attenuation coefficients of atherosclerotic plaque constituents by quantitative optical coherence tomography,” IEEE Trans. Med. Imag., 24 (10), 1369 –1376 (2005). http://dx.doi.org/10.1109/TMI.2005.854297 ITMID4 0278-0062 Google Scholar

42. 

T. G. Van LeeuwenD. J. FaberM. C. Aalders, “Measurement of the axial point spread function in scattering media using single-mode fiber-based optical coherence tomography,” IEEE J. Sel. Topics Quantum Electron., 9 (2), 227 –233 (2003). http://dx.doi.org/10.1109/JSTQE.2003.813299 IJSQEN 1077-260X Google Scholar

43. 

V. V. Tuchin, “Coherent optical techniques for the analysis of tissue structure and dynamics,” J. Biomed. Opt., 4 (1), 100 –125 (1999). JBOPFO 1083-3668 Google Scholar

44. 

J. W. Pickeringet al., “Changes in the optical properties (at 632.8 nm) of slowly heated myocardium,” Appl. Opt., 32 (4), 367 –371 (1993). http://dx.doi.org/10.1364/AO.32.000367 APOPAI 0003-6935 Google Scholar
© 2012 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2012/$25.00 © 2012 SPIE
Qingliang Zhao, Chuanqing Zhou, Huajiang Wei, Yonghong He, Xinyu Chai, and Qiushi Ren "Ex vivo determination of glucose permeability and optical attenuation coefficient in normal and adenomatous human colon tissues using spectral domain optical coherence tomography," Journal of Biomedical Optics 17(10), 105004 (1 October 2012). https://doi.org/10.1117/1.JBO.17.10.105004
Published: 1 October 2012
Lens.org Logo
CITATIONS
Cited by 18 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Tissues

Colon

Glucose

Optical coherence tomography

Tissue optics

Signal attenuation

Diffusion

Back to Top