Poster + Presentation + Paper
5 March 2021 A new screening system for the estimation of ocular anterior chamber angle width
Author Affiliations +
Conference Poster
Abstract
Primary Angle Closure Glaucoma occurs more frequently in people with a narrower limbal anterior chamber depth (LACD) condition. Nowadays, clinical gold standard as an assessment technique, i.e. gonioscopy, is invasive and complex, whereas Van Herick (VH) technique is non-invasive, but subjective. The instrument, we propose, can automatically performs the VH procedure using a blue laser line, a piezo-actuator, and an image recognition algorithm embedded on a Raspberry Pi board. Preliminary measurements have been carried out on volunteers, and the results proved the feasibility of our approach. The final instrument unveils a high potential for early-stage diagnosis and screening applications.
Conference Presentation
© (2021) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Davide Cassanelli, Giovanni Gibertoni, Stefano Cattini, Luciano Quaranta, Ivano Riva, Carlo Bruttini, Giovanni De Angelis, and Luigi Rovati "A new screening system for the estimation of ocular anterior chamber angle width", Proc. SPIE 11623, Ophthalmic Technologies XXXI, 116231X (5 March 2021); https://doi.org/10.1117/12.2578215