Open Access
30 September 2023 Influence of hematoxylin and eosin staining on linear birefringence measurement of fibrous tissue structures in polarization microscopy
Liangyu Deng, Chunyan Chen, Wenxi Yu, Conghui Shao, Zan Shen, Yonggang Wang, Chao He, Hongtao Li, Zhiyan Liu, Honghui He, Hui Ma
Author Affiliations +
Abstract

Significance

For microscopic polarization imaging of tissue slices, two types of samples are often prepared: one unstained tissue section for polarization imaging to avoid possible influence from staining dyes quantitatively and one hematoxylin-eosin (H&E) stained adjacent tissue section for histological diagnosis and structural feature identification. However, this sample preparation strategy requires high-quality adjacent tissue sections, and labeling the structural features on unstained tissue sections is impossible. With the fast development of data driven-based polarimetric analysis, which requires a large amount of pixel labeled images, a possible method is to directly use H&E stained slices, which are standard samples archived in clinical hospitals for polarization measurement.

Aim

We aim to study the influence of hematoxylin and eosin staining on the linear birefringence measurement of fibrous tissue structures.

Approach

We examine the linear birefringence properties of four pieces of adjacent bone tissue slices with abundant collagen fibers that are unstained, H&E stained, hematoxylin (H) stained, and eosin (E) stained. After obtaining the spatial maps of linear retardance values for the four tissue samples, we carry out a comparative study using a frequency distribution histogram and similarity analysis based on the Bhattacharyya coefficient to investigate how H&E staining affects the linear birefringence measurement of bone tissues.

Results

Linear retardance increased after H&E, H, or E staining (41.7%, 40.8%, and 72.5% increase, respectively). However, there is no significant change in the imaging contrast of linear retardance in bone tissues.

Conclusions

The linear retardance values induced by birefringent collagen fibers can be enhanced after H&E, H, or E staining. However, the structural imaging contrasts based on linear retardance did not change significantly or the staining did not generate linear birefringence on the sample area without collagen. Therefore, it can be acceptable to prepare H&E stained slices for clinical applications of polarimetry based on such a mapping relationship.

1.

Introduction

As a vectorial property, polarization encodes high-dimensional information of light.13 Polarimetry can provide abundant optical and structural information of tissue; thus it has shown great potential in biomedical and clinical applications.410 After interacting with tissue, the polarization state of light can be altered by different structures.11 By measuring the polarization state changes of output light from tissue, several properties, such as birefringence, dichroism, and depolarization, can be quantified.12 For microscopic polarization imaging of thin tissue slices, linear birefringence is a valuable and prevalently used metric for anisotropic structures characterization.1315 For instance, Wood et al.16 demonstrated that linear birefringence can be adopted for in vivo tissue fiber characterization. Furthermore, Pierangelo et al.17 showed that the reducing of linear birefringence resulting from the breaking down of well-ordered fibrous structures can be used for cervical cancer identification.

For polarization microscopy, two types of samples are often prepared: one unstained tissue section for polarization imaging and one adjacent stained tissue section for histological diagnosis and structural feature identification.1820 Currently, hematoxylin and eosin (H&E) staining is the most popular choice and the cornerstone of anatomical pathological diagnosis.21 This convenient and cost-effective dye combination is capable of revealing remarkable cellular details, to the extent that the ultrastructural features can be deduced.22 When measuring the linear birefringence property of tissue slices, unstained samples are often used to avoid possible influence from staining dyes quantitatively.23,24 However, this sample preparation strategy brings several other problems. First, the requirement for the preparation of high-quality adjacent tissue sections burdens the technicians. Second, labeling structural features on unstained tissue sections is impossible, and the distortions or rotations between adjacent slices make comparisons at the pixel level impractical. As the fast development of data driven-based polarimetric analysis, which requires a large amount of pixel labeled images, a possible method is to directly use H&E stained slices, which are standard samples archived in clinical hospitals for polarization measurement. There are very few studies that compare the optical properties of unstained and stained tissue samples.25,26 Although the assessment of polarized images of unstained and H&E stained tissues has been conducted previously, a quantitative study on the influence of hematoxylin staining, eosin staining, and H&E staining on linear birefringence measurement of fibrous tissue structures in polarization microscopy is still missing and necessary.

Here we examine the linear birefringence properties of four pieces of adjacent bone tissue slices with abundant collagen fibers, which are unstained, H&E stained, hematoxylin (H) stained, and eosin (E) stained. After obtaining the spatial maps of linear retardance values for the four tissue samples, we carry out a comparative study using a frequency distribution histogram (FDH) and similarity analysis based on the Bhattacharyya coefficient (BC) to investigate how H&E staining affects the linear birefringence measurement of bone tissues. The results reveal no significant change in the linear retardance imaging contrast induced by fibrous structures before and after H&E staining, which cannot generate linear birefringence without collagen fibers. It can be acceptable to directly prepare H&E stained slices to obtain normalized linear retardance images.

2.

Materials and Methods

2.1.

Polarimetry Imaging System and Analysis

A transmission Mueller matrix (MM) microscope27 based on dual-rotating retarders28 is employed in this study. As shown in Fig. 1, the illuminating beam from an LED (XLamp XP-E, 633 nm, 3.5 W, Δλ=20  nm, Cree Inc., United States) passes through the polarization state generator module consisting of a fixed linear polarizer (P2, extinction ratio 1000:1, Daheng Optics, China) and a rotatable quarter-wave plate (R2, Daheng Optics, China) to control the input state of polarization (SOP). Here 30 different elliptical SOPs are generated by rotating the R2. After interacting with the sample, the transmitted light passes through the objective lens (4×/0.1 NA, UPlanSApo, Olympus, Japan) and then is detected by the polarization state analyzer (PSA) module consisting of a rotatable quarter-wave plate (R1, Daheng Optics, China) and a fixed linear polarizer (P1, extinction ratio 1000:1, Daheng Optics, China). Finally, the light carrying the polarization related information of the sample is received by a CMOS camera (MV-CA016-10UM, 12-bit, Hikvision, China) and stored as a 1080×1440  pixels intensity image (3.45  μm×3.45  μm pixel size).

Fig. 1

Schematics of (a) transmission MM microscope; (b) the diagram of the device; and (c) flowchart of the MM calculation using the 30 intensity images.

JBO_28_10_102909_f001.png

During each measurement, the P1 and P2 are fixed in the horizontal direction, and both the R1 and R2 are synchronously rotated at different angular speeds (ωR1:ωR2=5:1) to achieve 30 different SOP modulations. Because the detected light intensity is a periodic signal, the MM is calculated based on the Fourier analysis using the coefficients αn and βn as shown in the following equation:

Eq. (1)

I=α0+n=112(αncos2nθ+βnsin2nθ),
where I is the light intensity collected by the camera, the Fourier coefficients αn and βn are the functions of the 16 MM elements, and θ is the rotation angle of the quarter-wave plate R1. Before measurement, the MM microscope was calibrated using standard samples to ensure that the maximum error of the individual element is within 1%. The detailed calibration procedure can be found in our previous work.29

For pathological tissue slices with several microns thickness, their depolarization and diattenuation properties are often limited.30 However, the birefringence induced by anisotropic structures in tissue sections can provide abundant useful information for label-free diagnosis.31 When propagating through fibrous structures with linear birefringence, polarized light undergoes a linear phase retardance δ, which is quantified as

Eq. (2)

δ=2πΔndλ,
where Δn is the refractive index difference, d is the optical path length, and λ is the wavelength of light. Thus for standard pathological tissue slices with relatively constant thickness d, the linear retardance δ is proportional to Δn and is used as a quantitative measure of abnormal tissue areas with different fibrous microstructures. Here to obtain the value of linear retardance δ, we adopt the Mueller matrix polar decomposition (MMPD) method prevalently used in tissue polarimetry.12 The MMPD method decomposes a MM into three main properties, namely diattenuation (D), retardation (R), and depolarization (Δ), as shown in the following equation:

Eq. (3)

M=MΔMRMD,
where MD, MR, and MΔ are the diattenuation, retardation, and depolarization submatrices, respectively. Then the retardation matrix MR is further decomposed to obtain the magnitude of linear retardance δ as shown in the following equation:12

Eq. (4)

δ=cos1((MR(2,2)+MR(3,3))2+(MR(3,2)MR(2,3))21),
where MR(i,j) are the elements of MR.

2.2.

Tissue Samples

For quantitative evaluation of the influence of H&E staining on the linear retardance measurement of tissue slices, in this study, the human bone tissue samples with strong birefringence were used.32 Specifically, 20 human bone tissue samples were included in this study, with 10 samples representing normal and 10 samples representing abnormal bone pathological conditions. For comparison, four adjacent 4-μm thick slices of each bone tissue dehydrated and embedded in paraffin were prepared to produce unstained, H&E stained, hematoxylin (H) stained, and eosin (E) stained versions by an experienced pathologist from the Shanghai Sixth People’s Hospital. The bright field images of the four adjacent bone tissue slices are shown in Fig. 2. It can be observed that the dewaxed unstained slice image shown in Fig. 2(a) only provides the contour information. However, in the stained slices images shown in Figs. 2(b)2(d), the nuclei and cytoplasm are highlighted as blue and pink colors by H and E staining, respectively. For the unpolarized images of the four adjacent slices shown in Fig. 2, the location and distribution of bone tissue structures are roughly the same, which permits direct comparisons between the corresponding polarimetric imaging results of the four slices to analyze how H&E staining affects the linear retardance measurement. This work was approved by the Ethics Committee of Shanghai Sixth People’s Hospital Affiliated to Shanghai Jiao Tong University School of Medicine.

Fig. 2

Bright field images of four adjacent bone tissue slices: (a) the dewaxed unstained slice; (b) the H&E stained slice; (c) the H stained slice; and (d) the E stained slice.

JBO_28_10_102909_f002.png

3.

Results and Discussion

The linear retardance δ images for the bone tissue slices are shown in Fig. 3. Obviously, bone tissue structures have prominent linear birefringence effect to produce significant contrast in parameter δ images. Previous studies have demonstrated that the strong birefringence effect mainly comes from collagen fibers, with density and orientation that can be quantitatively evaluated using linear retardance parameters.33 In this study, we choose the bone tissue specimens with abundant fibrous structures as the samples for comparative studies on the influence of H&E staining on linear birefringence measurements.

Fig. 3

Linear retardance parameter δ images of the bone tissue slices: (a1)–(d1) the linear retardance images for the unstained, H&E stained, H stained, and E stained slices, respectively, and (a2)–(d2) the corresponding normalized images.

JBO_28_10_102909_f003.png

It can be observed from the un-normalized linear retardance images shown in Figs. 3(a1)3(d1) that the distributions of the birefringent structures are basically the same, regardless of the pixel values differences. This indicates that the hematoxylin, eosin, and H&E staining cannot produce linear birefringence to the non-birefringent tissue areas. In other words, it suggests that the dye itself cannot generate significant changes in Δn of the tissue without linear birefringence effect, according to Eq. (2).

Meanwhile, Fig. 3 also shows that the linear birefringence of the fibrous structures may be enhanced by the attached dye. For a demonstration, we calculated the mean values of linear retardance δ for the bone tissue slices shown in Figs. 3(a1)3(d1). Here a pixel-by-pixel comparison of the values is impractical due to the displacement and slight rotation of different slices during preparation. The mean linear retardance δ values of the unstained [Fig. 3(a1)], H&E stained [Fig. 3(b1)], H stained [Fig. 3(c1)], and E stained [Fig. 3(d1)] slices are 0.0316, 0.0448, 0.0445, and 0.0545 rad, respectively. In other words, compared with the unstained bone tissue slice, the E and H staining enhanced bone tissue samples’ δ values by 72.5% and 40.8%, respectively. Hence, the mean linear retardance δ value of the H&E stained slice with both the hematoxylin and eosin dyes increased by 41.7%. This indicates that the combination of dye molecules and birefringence structures, which are mainly the collagen fibers in bone tissues, increases the linear birefringence property. Similarly, previous studies have shown that the dye of Sirius Red can also significantly enhance the birefringence of collagen fibers,3436 the mechanism of which is the elongated dye molecules attached to the collagen fibers with their long axes parallel to the collagen orientation. This parallel relationship between dye molecules and collagen fibers results in an enhanced linear birefringence. Here a possible reason for the results shown in Fig. 3 is that both the hematoxylin and eosin molecules can be attached to the collagen fiber in a similar parallel way as that of the Sirius Red molecules, leading to the enhancement of linear birefringence.

In addition to the global enhancement of linear birefringence values, it can also be noticed from Figs. 3(a1)3(d1) that the imaging contrasts between collagen fibers and other areas are kept in different stained bone tissue samples. For a demonstration, Figs. 3(a1)3(d1) images are normalized to generate images shown in Figs. 3(a2)3(d2). We can see that the normalized images of different stained tissue samples maintain very similar imaging contrasts, which reflect the density and distribution of collagen fibers. To evaluate such kind of imaging contrast consistency in detail, we performed FDH analysis on the four linear retardance images before and after normalization, as shown in Fig. 4. The FDHs of the four un-normalized images in Fig. 4(a) all represent the same fluctuating distributions of a sharp decline after reaching the peak value and then gently maintain a base value. Also it can be observed that the frequencies of linear retardance values are concentrated around different peaks for different stained bone tissues. The maximum deviation of the linear retardance value at the peak for Fig. 4(a) is 2.56%. After the normalization, the linear retardance images of different stained bone tissue slices tend to have more similar FDH distributions as shown in Fig. 4(b), with the maximum deviation of the linear retardance value at the peak being 2.2%. It confirms that the linear birefringence of the bone tissue slices was enhanced after H&E staining. However, the structural imaging contrasts based on linear retardance did not change significantly or the staining cannot generate linear birefringence on the sample area without collagen.

Fig. 4

Frequency distribution histogram curves of the linear retardance images of bone tissue slices: (a1), (a2) the frequency distribution histograms of the original linear retardance images, in which (a2) is the subgraph of (a1) counting in [0, 0.02]; (b1), (b2) the frequency distribution histograms of the normalized linear retardance images, in which (b2) is the subgraph of (b1) counting in [0, 0.2].

JBO_28_10_102909_f004.png

For a more detailed quantitative comparison, we then divided each normalized linear retardance image shown in Figs. 3(a2)3(d2) into nine subimages of the same size and performed similarity analysis on the corresponding subimages pair to judge whether they still maintain the imaging contrast consistency after staining. The original image of 1080×1440  pixels was linearly divided into nine equal-sized subimages of 360×480  pixels. Here we use the BC as shown in Eq. (5) for the similarity evaluation:

Eq. (5)

B(H1,H2)=1iH1(i)·H2(i)iH1(i)·iH2(i),
where H1(i) and H2(i) are the two frequency distribution functions of corresponding subimages pair and i represents the median value of each interval after the linear retardance value range is divided into 100 equal intervals.

The BC is a popular statistical method for measuring the similarity of two normalized distributions.37 It has been prevalently used in various fields ranging from classical statistics to artificial intelligence because of its simplicity and robustness to small outliers.38 Previous study indicated that a perfect overlap between two distributions yields a BC value of exactly 1. Also a BC value around 0.8 means that the two distributions still maintain the corresponding distribution peaks and valleys and the morphological deviation between the two is not significant.39 As we can see from Table 1, the average BC values between the unstained and different stained bone tissues are 0.8244, 0.8360, and 0.7849, respectively, indicating that the linear retardance distribution had a large overlap after the H&E, H, or E staining. Here we focus on the H&E stained subimages considering their broad pathological application. Except for subimages 3 and 6, the BC values calculated between the unstained and H&E stained slices are >0.8, which means that the pixels value distribution after H&E staining remains relatively consistent. The BC value lower than 0.8 in area 6 may be due to the artifacts induced by the occurrence of folds in the tissue section preparation process.

Table 1

Similarity analysis results using the BC on the corresponding normalized subimage pair.

H&EHE
UnstainedArea 10.85060.84970.8950
Area 20.87140.84260.7528
Area 30.67180.80010.7000
Area 40.80650.85220.8474
Area 50.89520.82000.8536
Area 60.73920.83000.7022
Area 70.81310.89940.8096
Area 80.86730.89530.7989
Area 90.90490.73430.7049
Average0.82440.83600.7849

Overall, a significant similarity exists in four different versions of the same tissue sample. As for the regions with higher similarity, they mainly reflect the birefringent pathological features, such as the fibrous bone structures, with linear retardance imaging contrast that is reserved in the measurement of stained slices. As for the regions with relatively low similarity, this may be due to the artifacts induced by the sample preparation process rather than specific pathological features.

In addition, we computed the BC among the four variants within each sample in a broader sample perspective, utilizing the identical calculation method as described earlier. As illustrated in Table 2, the coefficients represent the average BC values across 10 normal bone tissue samples, yielding an overall average similarity of 0.8081, 0.8101, and 0.8300 for H&E, H, and E stained samples compared with the unstained sample, respectively. Moreover, the global similarities for the abnormal bone tissue samples, as depicted in Table 3, reached 0.7835, 0.8004, and 0.7979 for H&E, H, and E staining, respectively. The statistical analysis results based on the expanded sample size indicate that H&E, H, or E staining has a minor impact on the overarching structural contrast in both normal and abnormal bone tissue samples.

Table 2

Similarity analysis results using the BC on the corresponding normalized image pair of normal human bone tissue samples.

Normal samplesH&EHE
UnstainedSample 10.78490.82440.8360
Sample 20.86870.87920.8700
Sample 30.75970.75750.7552
Sample 40.82120.82650.8367
Sample 50.76390.76820.8270
Sample 60.83750.83270.8580
Sample 70.81360.81290.8417
Sample 80.82170.81460.8424
Sample 90.79820.79590.8240
Sample 100.81180.78900.8088
Average0.80810.81010.8300

Table 3

Similarity analysis results using the BC on the corresponding normalized image pair of abnormal human bone tissue samples.

Abnormal samplesH&EHE
UnstainedSample 10.84030.82470.8232
Sample 20.74620.75990.7603
Sample 30.75510.76610.7518
Sample 40.79540.80330.7900
Sample 50.76580.76120.7596
Sample 60.80660.82850.8192
Sample 70.72130.76830.7998
Sample 80.83190.83870.8336
Sample 90.81860.77660.7922
Sample 100.75410.87670.8497
Average0.78350.80040.7979

To show the linear dichroism of the samples, we also calculated the MMPD diattenuation parameter D according to the following equation:

Eq. (6)

D=M122+M132+M142,
where Mij (i,j=1, 2, 3, 4) represents the corresponding MM element. The diattenuation parameter D images are shown in Fig. 5, where the average D values are 0.0060, 0.0084, 0.0082, and 0.0069 for the unstained, H&E stained, H stained, and E stained slices, respectively. The calculation also indicates that the maximal D values are 0.0181, 0.0288, 0.0274, and 0.0146 for the unstained, H&E stained, H stained, and E stained slices, respectively. According to the absorption spectra of the dyes in solution,4043 the spectra range of hemalum complexes, which are active ingredients in the hematoxylin staining process, is from 400 to 700 nm, producing a maximal peak at 566 nm, whereas the spectra range of eosin molecules is from 440 to 550 nm with the maximal peak at 517 nm. Thus when the 633 nm LED was used as the light source of the MM microscope in this study, the H&E and H stained tissue slices exhibited a slightly linear dichroism enhancement (0.002 for the average value and 0.01 for the maximal value) as shown in Figs. 5(b) and 5(c). It may be due to those dye molecules, especially hemalum complexes attaching to the collagen fibers. However, the results shown in Fig. 5 also confirm that both the intrinsic linear dichroism effects of the samples and the enhancement of diattenuation induced by staining are limited. It can produce very small additional contrast to the linear retardance images.

Fig. 5

Diattenuation D images of (a) unstained, (b) H&E stained, (c) H stained, and (d) E stained bone tissue slices.

JBO_28_10_102909_f005.png

In summary, both the FDH and BC analysis results reveal no significant change in the image contrast of linear retardance induced by the fibrous structure after H&E staining, which cannot generate linear birefringence without collagen fibers. Such a mapping relationship can be used for clinical application of polarimetry; for instance, for pathological tissue samples with significantly altered fibrous structure, the measurement of the linear retardance in the H&E stained slice can reflect the information of birefringent fibrous structure in the corresponding unstained slice. It may be acceptable to directly prepare H&E stained slices to obtain normalized linear retardance images for reducing sample preparation workload in pathological tissue diagnosis, which mainly focuses on the imaging contrast between different tissue structures rather than to obtain adjacent tissue slices for measuring and staining4446 or measure unstained slices first and then counterstaining them.47 For instance, the characteristic polarization imaging contrast produced by linearly birefringent collagen orientation is applicable in detecting common bone diseases such as fragility fractures and osteoporosis.48 However, it should be noted that the feasibility of directly utilizing H&E stained slices for polarization measurements may vary depending on the specific objectives, particularly if the focus is on absolute linear birefringence values, which can be changed by H&E staining. Meanwhile, polarization measurement data based on H&E stained slices can be more conveniently obtained, analyzed, and labeled because the pathological slices archived in the hospital are mainly H&E stained slices for retrospective research instead of unstained slices.49

4.

Conclusion

In this study, we quantitatively analyzed the effect of H&E staining on linear birefringence imaging results. We measured four adjacent pieces of bone tissue slices, which were unstained, H&E stained, hematoxylin (H) stained, and eosin (E) stained. The comparison results of mean retardance values showed that the linear retardance values induced by birefringent collagen fibers can be enhanced after H&E, H, or E staining. The increasing magnitudes were 41.7% for H&E staining, 40.8% for H staining, and 72.5% for E staining compared with that of the unstained tissue slice. Furthermore, the FDH and BC analysis results confirmed that the linear birefringence of the bone tissue slices was enhanced after H&E staining. However, the structural imaging contrasts based on linear retardance did not change significantly, or the staining did not generate linear birefringence on the sample area without collagen. Therefore, it can be acceptable to prepare H&E stained slices for some pathological polarized imaging situations based on such a mapping relationship, focusing on obtaining the imaging contrast between different tissue structures. With the fast development of data-driven based polarimetric analysis, which requires a large amount of pixel labeled images, an effective method is to directly use H&E stained slices, which are standard samples archived in clinical hospitals for polarization measurements.

Disclosures

The authors have no conflicts of interest to declare.

Code, Data, and Materials Availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

Acknowledgments

This study was partially supported by the Shenzhen Key Fundamental Research Project (Grant No. JCYJ20210324120012035), Science and Technology Innovation Plan of the Science and Technology Commission of Shanghai Municipality (STCSM) (Grant No. 20Z11900304), Nature Science Fund of STCSM (Grant No. 23ZR1448200), and clinical research funds from Shanghai Sixth People’s Hospital Affiliated to Shanghai Jiao Tong University School of Medicine (Grant No. ynhg201914).

References

1. 

C. He et al., “Revealing complex optical phenomena through vectorial metrics,” Adv. Photonics, 4 (2), 026001 https://doi.org/10.1117/1.AP.4.2.026001 AOPAC7 1943-8206 (2022). Google Scholar

2. 

J. L. Pezzaniti and A. C. Russell, “Mueller matrix imaging polarimetry,” Opt. Eng., 34 (6), 1558 –1568 https://doi.org/10.1117/12.206161 (1995). Google Scholar

3. 

J. C. Ramella-Roman, T. Novikova, “Polarized light,” Polarized Light in Biomedical Imaging and Sensing: Clinical and Preclinical Applications, 1 –24 Springer International Publishing, Cham (2022). Google Scholar

4. 

C. He et al., “Polarisation optics for biomedical and clinical applications: a review,” Light: Sci. Appl., 10 (1), 194 https://doi.org/10.1038/s41377-021-00639-x (2021). Google Scholar

5. 

J. J. Gil and R. Ossikovski, Polarized Light and the Mueller Matrix Approach, CRC Press( (2022). Google Scholar

6. 

A. Sanaz and I. A. Vitkin, “Polarized light imaging in biomedicine: emerging Mueller matrix methodologies for bulk tissue assessment,” J. Biomed. Opt., 20 (6), 061104 https://doi.org/10.1117/1.JBO.20.6.061104 JBOPFO 1083-3668 (2015). Google Scholar

7. 

M. D. Singh, N. Ghosh, I. A. Vitkin, “Mueller matrix polarimetry in biomedicine: enabling technology, biomedical applications, and future prospects,” Polarized Light in Biomedical Imaging and Sensing: Clinical and Preclinical Applications, 61 –103 Springer International Publishing, Cham (2023). Google Scholar

8. 

J. C. Ramella-Roman, I. Saytashev and M. Piccini, “A review of polarization-based imaging technologies for clinical and preclinical applications,” J. Opt., 22 (12), 123001 https://doi.org/10.1088/2040-8986/abbf8a (2020). Google Scholar

9. 

C. S. Joseph et al., “Optical phantoms for biomedical polarimetry: a review,” J. Biomed. Opt., 24 (3), 030901 https://doi.org/10.1117/1.JBO.24.3.030901 JBOPFO 1083-3668 (2019). Google Scholar

10. 

J. Qi and D. S. Elson, “Mueller polarimetric imaging for surgical and diagnostic applications: a review,” J. Biophotonics, 10 (8), 950 –982 https://doi.org/10.1002/jbio.201600152 (2017). Google Scholar

11. 

G. Nirmalya and I. V. Alex, “Tissue polarimetry: concepts, challenges, applications, and outlook,” J. Biomed. Opt., 16 (11), 110801 https://doi.org/10.1117/1.3652896 JBOPFO 1083-3668 (2011). Google Scholar

12. 

S. Y. Lu and R. A. Chipman, “Interpretation of Mueller matrices based on polar decomposition,” J. Opt. Soc. Am. A, 13 (5), 1106 –1113 https://doi.org/10.1364/JOSAA.13.001106 JOAOD6 0740-3232 (1996). Google Scholar

13. 

Y. Shen et al., “Comparative study of the influence of imaging resolution on linear retardance parameters derived from the Mueller matrix,” Biomed. Opt. Express, 12 (1), 211 –225 https://doi.org/10.1364/BOE.410989 BOEICL 2156-7085 (2021). Google Scholar

14. 

B. Chen et al., “Analysis and calibration of linear birefringence orientation parameters derived from Mueller matrix for multi-layered tissues,” Opt. Lasers Eng., 146 106690 https://doi.org/10.1016/j.optlaseng.2021.106690 (2021). Google Scholar

15. 

L. Deng et al., “A dual-modality imaging method based on polarimetry and second harmonic generation for characterization and evaluation of skin tissue structures,” J. Mol. Sci., 24 (4), 4206 https://doi.org/10.3390/ijms24044206 JMOSE7 0253-3677 (2023). Google Scholar

16. 

M. F. G. Wood et al., “Proof-of-principle demonstration of a Mueller matrix decomposition method for polarized light tissue characterization in vivo,” J. Biomed. Opt., 14 (1), 014029 https://doi.org/10.1117/1.3065545 JBOPFO 1083-3668 (2009). Google Scholar

17. 

A. Pierangelo et al., “Polarimetric imaging of uterine cervix: a case study,” Opt. Express, 21 (12), 14120 –14130 https://doi.org/10.1364/OE.21.014120 OPEXFF 1094-4087 (2013). Google Scholar

18. 

A. Pierangelo et al., “Ex-vivo characterization of human colon cancer by Mueller polarimetric imaging,” Opt. Express, 19 (2), 1582 –1593 https://doi.org/10.1364/OE.19.001582 OPEXFF 1094-4087 (2011). Google Scholar

19. 

M. Kim et al., “Optical diagnosis of gastric tissue biopsies with Mueller microscopy and statistical analysis,” J. Eur. Opt. Soc., 18 (2), 10 https://doi.org/10.1051/jeos/2022011 1990-2573 (2022). Google Scholar

20. 

Y. Dong et al., “A polarization-imaging-based machine learning framework for quantitative pathological diagnosis of cervical precancerous lesions,” IEEE Trans. Med. Imaging, 40 (12), 3728 –3738 https://doi.org/10.1109/TMI.2021.3097200 ITMID4 0278-0062 (2021). Google Scholar

21. 

J. K. C. Chan, “The wonderful colors of the hematoxylin–eosin stain in diagnostic surgical pathology,” Int. J. Surg. Pathol., 22 (1), 12 –32 https://doi.org/10.1177/1066896913517939 IJSPFL (2014). Google Scholar

22. 

A. T. Feldman, D. Wolfe, “Tissue processing and hematoxylin and wosin staining,” Histopathology: Methods and Protocols, 31 –43 Springer New York, New York (2014). Google Scholar

23. 

T. Liu et al., “Distinguishing structural features between Crohn’s disease and gastrointestinal luminal tuberculosis using Mueller matrix derived parameters,” J. Biophotonics, 12 (12), e201900151 https://doi.org/10.1002/jbio.201900151 (2019). Google Scholar

24. 

B. Chen et al., “Comparative study of modified mueller matrix transformation and polar decomposition parameters for transmission and backscattering tissue polarimetries,” Appl. Sci., 11 (21), 10416 https://doi.org/10.3390/app112110416 (2021). Google Scholar

25. 

W. Zhuo et al., “Tissue refractive index as marker of disease,” J. Biomed. Opt., 16 (11), 116017 https://doi.org/10.1117/1.3656732 JBOPFO 1083-3668 (2011). Google Scholar

26. 

V. P. Mantripragada et al., “A comprehensive dataset of histopathology images, grades and patient demographics for human osteoarthritis cartilage,” Data Brief, 37 107129 https://doi.org/10.1016/j.dib.2021.107129 (2021). Google Scholar

27. 

Y. Wang et al., “Differentiating characteristic microstructural features of cancerous tissues using Mueller matrix microscope,” Micron, 79 (8), 8 –15 https://doi.org/10.1016/j.micron.2015.07.014 MICNB2 0047-7206 (2015). Google Scholar

28. 

B. C. David, J. L. Pezzaniti and A. C. Russell, “Mueller matrix algorithms,” Proc. SPIE, 1746 231 –246 https://doi.org/10.1117/12.138793 PSISDG 0277-786X (1992). Google Scholar

29. 

J. Zhou et al., “Modulus design multiwavelength polarization microscope for transmission Mueller matrix imaging,” J. Biomed. Opt., 23 (1), 016007 https://doi.org/10.1117/1.JBO.23.1.016007 JBOPFO 1083-3668 (2018). Google Scholar

30. 

Y. Dong et al., “Quantitatively characterizing the microstructural features of breast ductal carcinoma tissues in different progression stages by Mueller matrix microscope,” Biomed. Opt. Express, 8 (8), 3643 –3655 https://doi.org/10.1364/BOE.8.003643 BOEICL 2156-7085 (2017). Google Scholar

31. 

T. Novikova and J. C. Ramella-Roman, “Is a complete Mueller matrix necessary in biomedical imaging?,” Opt. Lett., 47 (21), 5549 –5552 https://doi.org/10.1364/OL.471239 OPLEDP 0146-9592 (2022). Google Scholar

32. 

R. B. Martin and J. Ishida, “The relative effects of collagen fiber orientation, porosity, density, and mineralization on bone strength,” J. Biomech., 22 (5), 419 –426 https://doi.org/10.1016/0021-9290(89)90202-9 JBMCB5 0021-9290 (1989). Google Scholar

33. 

E. M. Spiesz, W. Kaminsky and P. K. Zysset, “A quantitative collagen fibers orientation assessment using birefringence measurements: calibration and application to human osteons,” J. Struct. Biol., 176 (3), 302 –306 https://doi.org/10.1016/j.jsb.2011.09.009 JSBIEM 1047-8477 (2011). Google Scholar

34. 

L. C. U. Junqueira, G. Bignolas and R. R. Brentani, “Picrosirius staining plus polarization microscopy, a specific method for collagen detection in tissue sections,” Histochem J., 11 (4), 447 –455 https://doi.org/10.1007/BF01002772 (1979). Google Scholar

35. 

P. B. Canham et al., “Layered structure of saccular aneurysms assessed by collagen birefringence,” Neurol. Res., 21 (7), 618 –626 https://doi.org/10.1080/01616412.1999.11740986 (1999). Google Scholar

36. 

C. M. López De Padilla et al., “Picrosirius red staining: revisiting its application to the qualitative and quantitative assessment of collagen type I and type III in tendon,” J. Histochem. Cytochem., 69 (10), 633 –643 https://doi.org/10.1369/00221554211046777 JHCYAS 0022-1554 (2021). Google Scholar

37. 

A. Bhattacharyya, “On a measure of divergence between two multinomial populations,” Sankhya, 7 (4), 401 –406 (1946). Google Scholar

38. 

T. Kailath, “The divergence and Bhattacharyya distance measures in signal selection,” IEEE Trans. Commun. Technol., 15 (1), 52 –60 https://doi.org/10.1109/TCOM.1967.1089532 IETCAF 0018-9332 (1967). Google Scholar

39. 

S. M. Kang and R. P. Wildes, “The n-distribution Bhattacharyya coefficient,” in EECS-2015-02, (2015). Google Scholar

40. 

E. T. Adam et al., “Nonlinear multicontrast microscopy of hematoxylin-and-eosin-stained histological sections,” J. Biomed. Opt., 15 (2), 026018 https://doi.org/10.1117/1.3382908 JBOPFO 1083-3668 (2010). Google Scholar

41. 

C. Bettinger and H. W. Zimmermann, “New investigations on hematoxylin, hematein, and hematein-aluminium complexes,” Histochemistry, 96 (3), 215 –228 https://doi.org/10.1007/BF00271540 HCMYAL 0301-5564 (1991). Google Scholar

42. 

C. Li et al., “Study on the interaction between verapamil hydrochloride and eosin y by absorption, fluorescence and resonance Rayleigh scattering spectra and their analytical applications,” J. Fluoresc., 21 (2), 723 –732 https://doi.org/10.1007/s10895-010-0762-6 JOFLEN 1053-0509 (2011). Google Scholar

43. 

R. W. Palmer and R. D. Lillie, “Spectroscopic and staining studies of the ripening and overripening of aluminum hematoxylins,” Histochemie, 5 (1), 44 –54 https://doi.org/10.1007/BF00307891 HICHAU 0018-2222 (1965). Google Scholar

44. 

J. Chang et al., “Division of focal plane polarimeter-based 3 × 4 Mueller matrix microscope: a potential tool for quick diagnosis of human carcinoma tissues,” J. Biomed. Opt., 21 (5), 056002 https://doi.org/10.1117/1.JBO.21.5.056002 JBOPFO 1083-3668 (2016). Google Scholar

45. 

A. E. Woessner et al., “Characterizing differences in the collagen fiber organization of skin wounds using quantitative polarized light imaging,” Wound Repair Regener., 27 (6), 711 –714 https://doi.org/10.1111/wrr.12758 (2019). Google Scholar

46. 

X. Zhou et al., “Automatic detection of head and neck squamous cell carcinoma on pathologic slides using polarized hyperspectral imaging and machine learning,” Proc. SPIE, 11603 116030Q https://doi.org/10.1117/12.2582330 PSISDG 0277-786X (2021). Google Scholar

47. 

Y. Wang et al., “Mueller matrix microscope: a quantitative tool to facilitate detections and fibrosis scorings of liver cirrhosis and cancer tissues,” J. Biomed. Opt., 21 (7), 071112 https://doi.org/10.1117/1.JBO.21.7.071112 JBOPFO 1083-3668 (2016). Google Scholar

48. 

M. G. Ascenzi, “Theoretical mathematics, polarized light microscopy and computational models in healthy and pathological bone,” Bone, 134 115295 https://doi.org/10.1016/j.bone.2020.115295 (2020). Google Scholar

49. 

L. Lei et al., “Clinicopathological characteristics of mucinous breast cancer: a retrospective analysis of a 10-year study,” PLoS One, 11 (5), e0155132 https://doi.org/10.1371/journal.pone.0155132 POLNCL 1932-6203 (2016). Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Liangyu Deng, Chunyan Chen, Wenxi Yu, Conghui Shao, Zan Shen, Yonggang Wang, Chao He, Hongtao Li, Zhiyan Liu, Honghui He, and Hui Ma "Influence of hematoxylin and eosin staining on linear birefringence measurement of fibrous tissue structures in polarization microscopy," Journal of Biomedical Optics 28(10), 102909 (30 September 2023). https://doi.org/10.1117/1.JBO.28.10.102909
Received: 26 April 2023; Accepted: 19 September 2023; Published: 30 September 2023
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
KEYWORDS
Tissues

Biological samples

Birefringence

Bone

Collagen

Polarization

Dyes

Back to Top