Conventional OCT images, obtained using a focused Gaussian beam have a lateral resolution of approximately 30 μm and a depth of focus (DOF) of 2-3 mm, defined as the confocal parameter (twice of Gaussian beam Rayleigh range). Improvement of lateral resolution without sacrificing imaging range requires techniques that can extend the DOF. Previously, we described a self-imaging wavefront division optical system that provided an estimated one order of magnitude DOF extension. In this study, we further investigate the properties of the coaxially focused multi-mode (CAFM) beam created by this self-imaging wavefront division optical system and demonstrate its feasibility for real-time biological tissue imaging. Gaussian beam and CAFM beam fiber optic probes with similar numerical apertures (objective NA≈0.5) were fabricated, providing lateral resolutions of approximately 2 μm. Rigorous lateral resolution characterization over depth was performed for both probes. The CAFM beam probe was found to be able to provide a DOF that was approximately one order of magnitude greater than that of Gaussian beam probe. By incorporating the CAFM beam fiber optic probe into a μOCT system with ~1.5 μm axial resolution, we were able to acquire cross-sectional images of swine small intestine ex vivo, enabling the visualization of subcellular structures, providing high quality OCT images over more than a 300 μm depth range.
|