3 March 2017 Validation of an image registration and segmentation method to measure stent graft motion on ECG-gated CT using a physical dynamic stent graft model
Author Affiliations +
Abstract
The application of endovascular aortic aneurysm repair has expanded over the last decade. However, the long-term performance of stent grafts, in particular durable fixation and sealing to the aortic wall, remains the main concern of this treatment. The sealing and fixation are challenged at every heartbeat due to downward and radial pulsatile forces. Yet knowledge on cardiac-induced dynamics of implanted stent grafts is sparse, as it is not measured in routine clinical follow-up. Such knowledge is particularly relevant to perform fatigue tests, to predict failure in the individual patient and to improve stent graft designs. Using a physical dynamic stent graft model in an anthropomorphic phantom, we have evaluated the performance of our previously proposed segmentation and registration algorithm to detect periodic motion of stent grafts on ECG-gated (3D+t) CT data. Abdominal aortic motion profiles were simulated in two series of Gaussian based patterns with different amplitudes and frequencies. Experiments were performed on a 64-slice CT scanner with a helical scan protocol and retrospective gating. Motion patterns as estimated by our algorithm were compared to motion patterns obtained from optical camera recordings of the physical stent graft model in motion. Absolute errors of the patterns' amplitude were smaller than 0.28 mm. Even the motion pattern with an amplitude of 0.23 mm was measured, although the amplitude of motion was overestimated by the algorithm with 43%. We conclude that the algorithm performs well for measurement of stent graft motion in the mm and sub-mm range. This ultimately is expected to aid in patient-specific risk assessment and improving stent graft designs.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Maaike A. Koenrades, Ella M. Struijs, Almar Klein, Henny Kuipers, Robert H. Geelkerken, and Cornelis H. Slump "Validation of an image registration and segmentation method to measure stent graft motion on ECG-gated CT using a physical dynamic stent graft model", Proc. SPIE 10134, Medical Imaging 2017: Computer-Aided Diagnosis, 1013418 (3 March 2017); https://doi.org/10.1117/12.2254262
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Motion measurement

Cameras

Computed tomography

Motion models

Device simulation

Image registration

Image segmentation

RELATED CONTENT


Back to Top