Poster + Paper
21 August 2024 The X-IFU focal plane assembly development model: evaluation of the global magnetic shielding factor
Author Affiliations +
Conference Poster
Abstract
This work describes the evaluation of the global shielding factor of the Development Model (DM) of the X-ray Integral Field Unit Focal Plane Assembly (X-IFU FPA) for the Athena X-ray observatory. The main task of the FPA is to provide an environment for the X-ray transition-edge sensors (TESs) microcalorimeter in which they can reach optimal performance, with stringent requirements on the thermal, electro-magnetic, mechanical, and electronic aspects of the environment. Within the FPA a two-stage thermal suspension system will isolate the detectors, operating at nominally 50 mK (T0 stage) from the FPA mechanical interfaces at 2 K (T2 stage). An intermediate temperature stage (T1 stage) operated at nominally 300 mK is implemented as a heat intercept.

The required magnetic shielding attenuation is obtained combining a cryogenic mu-metal (CMM) shield and a superconducting Niobium (Nb) shield. From a finite element model (FEM), the CMM shield alone should suppress the static field (for example Earth magnetic field) at the detector location with a factor 50; while the combination of the two has the goal of suppressing variable fields with a shielding factor of ∼106, defined as the ratio of the magnitude of the external field relative to the maximum field perpendicular to the sensor array. These numbers are referring to the component normal to the plane of the sensor array. In plane with sensor array the shielding factor can be a 104 times larger.

We use the TES detectors as very sensitive magnetometers. The TES critical current Ic exhibits oscillatory Fraunhofer-like behaviour, when magnetic field perpendicular is applied to it. Whether the variable external magnetic field perpendicular to the detectors penetrates the shields, the operating current ITES through the TES, which is stabilised in temperature and biased in a sensitive point in transition, may change and this variation can be used as witness for the shielding factor evaluation.

In this work we discuss the difficulties in performing such a measurement, where a Helmholtz coil outside the cryostat were used to apply an external magnetic field (< 200 μH) in order to avoid any permanent magnetisation of the CMM shield. Both AC and DC external magnetic field have been applied with different orientation respect to the detector plane. This improves the statistic of a very sensitive measurement, but also serves to maximise the signal measured by a number of TESs spread over the south-quadrant (closer to the Nb shield) of the array, where the field lines show different intensity. Preliminary results set a lower limit in the evaluation of the global FPA shielding factor of ∼105. Improvement in the measurement setup and in the analysis will be discussed in view of the future models of the FPA.
(2024) Published by SPIE. Downloading of the abstract is permitted for personal use only.
E. Taralli, R. den Hartog, J. Dercksen, H. van Weers, P. Khosropanah, J. Van der Kuur, W. J. Vreeling, M. D. Audley, G. Keizer, B. Jackson, L. Ferrari, P. Roelfsema, and J. W. A. den Herder "The X-IFU focal plane assembly development model: evaluation of the global magnetic shielding factor", Proc. SPIE 13093, Space Telescopes and Instrumentation 2024: Ultraviolet to Gamma Ray, 1309349 (21 August 2024); https://doi.org/10.1117/12.3018368
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Detector arrays

Superconductors

X-rays

Cryostats

Superconducting detectors

RELATED CONTENT

The transition-edge EBIT microcalorimeter spectrometer
Proceedings of SPIE (September 25 2014)
Multipixel readout of TES calorimeters
Proceedings of SPIE (March 11 2003)
Soft x ray response of a 36 pixel superconducting tunnel...
Proceedings of SPIE (November 10 1998)
Automatic setup of SCUBA-2 detector arrays
Proceedings of SPIE (July 18 2008)

Back to Top