3 February 1993 Calibration techniques for high flux x-ray detectors
George C. Idzorek, Robert G. Hockaday
Author Affiliations +
Abstract
A system is described for calibrating high flux (1011 W) x-ray detectors using moderate flux (10-4 W) synchrotron and low flux (10-7 W) conventional x-ray producing machines. The detectors have been designed for output currents of one milliampere to 100 amperes which results in output currents during calibration of approximately 10-16 to 10-12 amperes. The small calibration output currents led to the development of statistical methods to reduce the calibration measurement errors to acceptable levels. A personal computer is used to acquire, analyze, store, and display the calibration data in real time to allow the scientist to explore and correct the calibration technique or eliminate external noise sources.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
George C. Idzorek and Robert G. Hockaday "Calibration techniques for high flux x-ray detectors", Proc. SPIE 1736, X-Ray Detector Physics and Applications, (3 February 1993); https://doi.org/10.1117/12.140472
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Sensors

X-ray detectors

X-rays

Camera shutters

Control systems

Beam controllers

RELATED CONTENT

First results from the OGRESS sounding rocket payload
Proceedings of SPIE (September 18 2015)
Laboratory soft x-ray source with foil target
Proceedings of SPIE (February 03 1993)
Motion detection system for AXAF x-ray ground testing
Proceedings of SPIE (January 21 1993)
Techniques Of Absolute Low Energy X-Ray Calibration
Proceedings of SPIE (August 12 1986)
Vacuum Interface To A Soft X-Ray Synchrotron Beam Line
Proceedings of SPIE (August 12 1986)

Back to Top