28 June 2001 Method using a Monte Carlo simulation database for optimizing x-ray fluoroscopic conditions
Hironori Ueki, Kenichi Okajima
Author Affiliations +
Abstract
To improve image quality (IQ) and reduce dose in x-ray fluoroscopy, we have developed a new method for optimizing x-ray conditions such as x-ray tube voltage, tube current, and gain of the detector. This method uses a Monte Carlo (MC)-simulation database for analyzing the relations between IQ, x-ray dose, and x-ray conditions. The optimization consists of three steps. First, a permissible dose limit for each object thickness is preset. Then, the MC database is used to calculate the IQ of x-ray projections under all the available conditions that satisfy this presetting. Finally, the optimum conditions are determined as the ones that provide the highest IQ. The MC database contains projections of an estimation phantom simulated under emissions of single-energy photons with various energies. By composing these single-energy projections according to the bremsstrahlung energy distributions, the IQs under any x-ray conditions can be calculated in a very short time. These calculations show that the optimum conditions are determined by the relation between quantum noise and scattering. Moreover, the heat-capacity limit of the x-ray tube can also determine the optimum conditions. It is concluded that the developed optimization method can reduce the time and cost of designing x-ray fluoroscopic systems.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hironori Ueki and Kenichi Okajima "Method using a Monte Carlo simulation database for optimizing x-ray fluoroscopic conditions", Proc. SPIE 4320, Medical Imaging 2001: Physics of Medical Imaging, (28 June 2001); https://doi.org/10.1117/12.430959
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Photons

Databases

X-ray imaging

Monte Carlo methods

X-ray detectors

Sensors

RELATED CONTENT

Physics-based modeling of computed tomography systems
Proceedings of SPIE (March 18 2015)
Simulation of signal induction in the Caliste-SO detector
Proceedings of SPIE (September 11 2015)
Effects of scattered x-rays on cone-beam CT images
Proceedings of SPIE (May 01 1994)
MTF and DQE analysis of x-ray intensifying screens
Proceedings of SPIE (September 15 1999)

Back to Top