Paper
3 May 2002 Convolution estimation of spill-down and scatter in dual isotope (99mTc/111In) SPECT
Jian Feng, Bill C. Penney, Xiaochuan Pan
Author Affiliations +
Abstract
Accurate attenuation correction in dual isotope single photon emission computed tomography (SPECT) requires spill-down and scatter correction. Convolution methods that allow such correction were investigated. Three energy windows were used: 1) a Compton scatter window placed below the Tc-99m photopeak window, 2) the Tc-99m photopeak, and 3) both In-111 photopeaks. Spill-down of In-111 into the Tc-99m window was measured using a small source of In-111 at depths of 5, 7, 9, and 11cm. From these data, the convolution kernel relating the In-111 image and the Tc-99m window spill-down image was computed for each depth. Using a small Tc-99m source at the same depths, the convolution kernels describing its spill-down into the Compton window were computed. A factor to scale the Tc-99m spill-down estimate by for obtaining the self-scatter estimate was determined from the total counts in the Tc-99m photopeak vs. the Compton scatter window when primary photons were blocked with a small lead disk. Depths of 5 and 9cm were optimal for spill-down and Tc-99m self-scatter correction, respectively. It was demonstrated that the spill-down and scatter correction increased effective attenuation coefficient towards its 'no-scatter' value and dramatically decreased the intensity of an In-111 source in Tc-99m projection images.
© (2002) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jian Feng, Bill C. Penney, and Xiaochuan Pan "Convolution estimation of spill-down and scatter in dual isotope (99mTc/111In) SPECT", Proc. SPIE 4682, Medical Imaging 2002: Physics of Medical Imaging, (3 May 2002); https://doi.org/10.1117/12.465618
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Technetium

Point spread functions

Convolution

Single photon emission computed tomography

Signal attenuation

Cameras

Photons

RELATED CONTENT

A Modified Post Processing Correction Matrix For SPECT
Proceedings of SPIE (January 01 1986)
How many x ray photons can be scattered from a...
Proceedings of SPIE (March 13 2009)
Analysis Of Spect Using Monte Carlo Simulation
Proceedings of SPIE (November 01 1982)
SPECT reconstruction on the GPU
Proceedings of SPIE (March 18 2008)

Back to Top