20 April 2005 Noise simulation in x-ray CT
Author Affiliations +
Abstract
A method is presented for accurately simulating the effects of dose reduction in x-ray computed tomography (CT) by adding synthetic noise to raw projection data. A model for realistic noise in projection data was generated, incorporating the mechanisms of stochastic noise in energy-integrating x-ray detectors, electronic system noise, and bowtie beam filtering (used for patient dose reduction). Parameters for the model were extracted from phantom measurements on a variety of clinical CT scanners (helical single row, four-row, and 16-row). Dose reduction simulations were performed by adding synthetic noise based on the noise model to raw data acquired from clinical scanners. Qualitative and quantitative validation of the process was accomplished by comparing phantom scans acquired under high and low dose conditions with simulated imagery. The importance of including alternative noise mechanisms (bowtie filter and system noise) was demonstrated. Henceforth, scans of clinical patients were acquired using conventional protocols; through simulations, image sets were presented at a variety of lower dose procedures. The methodology promises to be a useful tool for radiologists to explore dose reduction protocols in an effort to produce diagnostic images with radiation dose “as low as reasonably achievable”.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Parinaz Massoumzadeh, Orville A. Earl, and Bruce R. Whiting "Noise simulation in x-ray CT", Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, (20 April 2005); https://doi.org/10.1117/12.593235
Lens.org Logo
CITATIONS
Cited by 8 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Scanners

Image filtering

X-ray computed tomography

Data modeling

X-rays

Computed tomography

Diagnostics

Back to Top