7 July 2009 Design and testing of diffuse reflectance sensor for continuous monitoring of cutaneous blood perfusion
P. Zakharov, M. S. Talary, A. Caduff
Author Affiliations +
Abstract
A dual-wavelength reflectance optical sensor for monitoring cutaneous blood perfusion is presented as a part of multisensor glucose monitoring system. A Monte-Carlo simulation of partial differential pathlengths has been used for the optimization of the distance from light source to detector. The simulation indicated that the light pathlength within the upper vascularised skin layers increases before reaching saturation at separation distances larger than 3 mm. Thus the sensor sensitivity does not benefit from larger source-detector distances. At the same time with a higher separation of the detector from the source, the intensity exponentially decreases while undesirable sensitivity to the muscle perfusion increases. The hardware prototype has been developed based on the simulation findings and tested in a laboratory setting and in a home use study by patients with diabetes. For both testing procedures the optical sensor demonstrated high sensitivity to perfusion changes. The effect of initial cutaneous blood increase under the sensor has been observed which can be associated with pressure-induced vasodilation as a response to the sensor application.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
P. Zakharov, M. S. Talary, and A. Caduff "Design and testing of diffuse reflectance sensor for continuous monitoring of cutaneous blood perfusion", Proc. SPIE 7368, Clinical and Biomedical Spectroscopy, 736810 (7 July 2009); https://doi.org/10.1117/12.831578
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Skin

Blood

Sensors

Monte Carlo methods

Optical sensors

Temperature metrology

Plasma display panels

Back to Top