23 February 2010 Accuracy validation for MRI-guided robotic prostate biopsy
Helen Xu, Andras Lasso, Siddharth Vikal, Peter Guion, Axel Krieger, Aradhana Kaushal, Louis L. Whitcomb, Gabor Fichtinger
Author Affiliations +
Abstract
We report a quantitative evaluation of the clinical accuracy of a MRI-guided robotic prostate biopsy system that has been in use for over five years at the U.S. National Cancer Institute. A two-step rigid volume registration using mutual information between the pre and post needle insertion images was performed. Contour overlays of the prostate before and after registration were used to validate the registration. A total of 20 biopsies from 5 patients were evaluated. The maximum registration error was 2 mm. The mean biopsy target displacement, needle placement error, and biopsy error was 5.4 mm, 2.2 mm, and 5.1 mm respectively. The results show that the pre-planned biopsy target did dislocate during the procedure and therefore causing biopsy errors.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Helen Xu, Andras Lasso, Siddharth Vikal, Peter Guion, Axel Krieger, Aradhana Kaushal, Louis L. Whitcomb, and Gabor Fichtinger "Accuracy validation for MRI-guided robotic prostate biopsy", Proc. SPIE 7625, Medical Imaging 2010: Visualization, Image-Guided Procedures, and Modeling, 762517 (23 February 2010); https://doi.org/10.1117/12.844251
Lens.org Logo
CITATIONS
Cited by 18 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Biopsy

Prostate

Image registration

Robotics

Robotic systems

Cancer

Magnetic resonance imaging

Back to Top