Paper
14 May 2010 Optical testing of bifocal diffractive-refractive intraocular lenses using Shack-Hartmann wavefront sensor
A. S. Gutman, I. V. Shchesyuk, V. P. Korolkov
Author Affiliations +
Abstract
Applicability of the Shack-Hartmann wavefront sensor for the bifocal diffractive-refractive intraocular lens testing is discussed. Measurement method based on quasi-continuous wavefront has been suggested. Light source requirements for testing of MIOL-Accord intraocular lens have been validated. The method has been realized in dioptrimeter including Shack-Hartman sensor and multi-wavelength coherent light source.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
A. S. Gutman, I. V. Shchesyuk, and V. P. Korolkov "Optical testing of bifocal diffractive-refractive intraocular lenses using Shack-Hartmann wavefront sensor", Proc. SPIE 7718, Optical Micro- and Nanometrology III, 77181P (14 May 2010); https://doi.org/10.1117/12.854484