4 March 2011 Evaluation of 1D, 2D and 3D nodule size estimation by radiologists for spherical and non-spherical nodules through CT thoracic phantom imaging
Author Affiliations +
Abstract
The purpose of this work was to estimate bias in measuring the size of spherical and non-spherical lesions by radiologists using three sizing techniques under a variety of simulated lesion and reconstruction slice thickness conditions. We designed a reader study in which six radiologists estimated the size of 10 synthetic nodules of various sizes, shapes and densities embedded within a realistic anthropomorphic thorax phantom from CT scan data. In this manuscript we report preliminary results for the first four readers (Reader 1-4). Two repeat CT scans of the phantom containing each nodule were acquired using a Philips 16-slice scanner at a 0.8 and 5 mm slice thickness. The readers measured the sizes of all nodules for each of the 40 resulting scans (10 nodules x 2 slice thickness x 2 repeat scans) using three sizing techniques (1D longest in-slice dimension; 2D area from longest in-slice dimension and corresponding longest perpendicular dimension; 3D semi-automated volume) in each of 2 reading sessions. The normalized size was estimated for each sizing method and an inter-comparison of bias among methods was performed. The overall relative biases (standard deviation) of the 1D, 2D and 3D methods for the four readers subset (Readers 1-4) were -13.4 (20.3), -15.3 (28.4) and 4.8 (21.2) percentage points, respectively. The relative biases for the 3D volume sizing method was statistically lower than either the 1D or 2D method (p<0.001 for 1D vs. 3D and 2D vs. 3D).
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Nicholas Petrick, Hyun J. Grace Kim, David Clunie, Kristin Borradaile, Robert Ford, Rongping Zeng, Marios A. Gavrielides, Michael F. McNitt-Gray, Charles Fenimore, Z. Q. John Lu, Binsheng Zhao, and Andrew J. Buckler "Evaluation of 1D, 2D and 3D nodule size estimation by radiologists for spherical and non-spherical nodules through CT thoracic phantom imaging", Proc. SPIE 7963, Medical Imaging 2011: Computer-Aided Diagnosis, 79630D (4 March 2011); https://doi.org/10.1117/12.878265
Lens.org Logo
CITATIONS
Cited by 8 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Computed tomography

3D metrology

X-ray computed tomography

Spherical lenses

Lung

Scanners

Imaging systems

RELATED CONTENT


Back to Top