Paper
22 February 2013 Time-resolved wide-field optically sectioned fluorescence microscopy
Guillaume Dupuis, Nadia Benabdallah, Aurélien Chopinaud, Céline Mayet, Sandrine Lévêque-Fort
Author Affiliations +
Abstract
We present the implementation of a fast wide-field optical sectioning technique called HiLo microscopy on a fluorescence lifetime imaging microscope. HiLo microscopy is based on the fusion of two images, one with structured illumination and another with uniform illumination. Optically sectioned images are then digitally generated thanks to a fusion algorithm. HiLo images are comparable in quality with confocal images but they can be acquired faster over larger fields of view. We obtain 4D imaging by combining HiLo optical sectioning, time-gated detection, and z-displacement. We characterize the performances of this set-up in terms of 3D spatial resolution and time-resolved capabilities in both fixed- and live-cell imaging modes.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Guillaume Dupuis, Nadia Benabdallah, Aurélien Chopinaud, Céline Mayet, and Sandrine Lévêque-Fort "Time-resolved wide-field optically sectioned fluorescence microscopy", Proc. SPIE 8589, Three-Dimensional and Multidimensional Microscopy: Image Acquisition and Processing XX, 85890H (22 February 2013); https://doi.org/10.1117/12.2003966
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image fusion

Fluorescence lifetime imaging

Microscopy

Spatial frequencies

Confocal microscopy

Microscopes

Modulation

Back to Top