1.IntroductionPolarization imaging can provide rich microstructural and optical information of tissues for diagnostic purposes.1–5 Since a Mueller matrix provides the most comprehensive characterization of the polarization features,6 it has been applied to differentiate various abnormal tissues, such as skin cancer,7 cervical cancer,8–10 colon cancer,11–13 liver fibrosis,14 and so on.15–18 For anisotropic tissues, previous studies have shown that the Mueller matrix elements may change significantly with the orientation of the sample, making quantitative characterization of the microstructural features very difficult.19–20 It was also pointed out that the structural information encoded in a Mueller matrix can be presented by other transformed parameters with more explicit physics meanings.10 In fact, the abundant information carried by the Mueller matrix may allow us to identify the characteristic features of abnormal tissues without using high-resolution images. The two-dimensional (2-D) images of Mueller matrix elements can be reduced into a group of quantitative or semiquantitative, orientation insensitive parameters that reveal clearly the key structural features of the samples. In this paper, we present a new way based on the statistical method to transform the 2-D images of the Mueller matrix elements into frequency distribution histogram (FDH), and central moment parameters. We study the quantitative influence of sample orientation on the frequency distributions of Mueller matrix elements, and then analyze the relations between the microstructures of tissues and the shapes of the FDHs. The experimental results show that the central moment analysis can provide us a group of orientation insensitive parameters representing the dominant features of tissues. The preliminary results of human cervical cancerous tissues show that the analysis method presented in this paper may serve as quantitative or semiquantitative criteria for cancerous tissue detection. 2.Methods and Materials2.1.Experimental SetupWe adopted the backscattering Mueller matrix measurement configuration based on the dual rotating retarder method.21–23 As shown in Figs. 1(a) and 1(b), the illuminating light from the light-emitting diode (Source, 633 nm, 1W) passes through the lens (L1, Thorlabs), and the polarization states generator consisting of a polarizer (P1, Thorlabs) and a quarter-wave plate (R1, Thorlabs). The photons backscattered from the sample pass through the polarization states analyzer (PSA) consisting of the analyzing quarter-wave plate (R2, Thorlabs) and polarizer (P2, Thorlabs), then are recorded by a charge-coupled device camera (QImaging 32-0122A, 12 bit, Canada) after passing through another lens (L2, Thorlabs). There is an oblique incident angle () between the illumination light and the detection direction to avoid the surface reflection of the sample. During the Mueller matrix measurements, the sample can be rotated in the imaging (X-Y) plane. As shown in Fig. 1(c), we can vary the angle from 0 deg to 180 deg while keeping the center of the sample unchanged. For the experimental setup shown in Fig. 1, the polarizers P1 and P2 are fixed in the horizontal direction, while the wave plates R1 and R2 are rotated with a fixed rate , where and are the rotation angles of the wave plates R1 and R2, respectively. The Fourier series intensities can be given by Eq. (1), where and are the Fourier coefficients. Using and , we can calculate the Mueller matrix elements according to Ref. 21. In order to record robust images, we measure the Mueller matrices of standard samples such as air and wave plate in the transmission mode and then calibrate the imaging system using the method proposed by Chenuault et al.24 The Mueller matrix elemental accuracy is tested by measuring the high extinction polarizers. After the calibration in transmission direction, the PSA arm of the system is rotated to the backscattered direction as shown in Fig. 1. We also measure the backscattering Mueller matrices of some samples with known polarization properties such as the microsphere solutions. In this work, the mean measurement errors of the diagonal and nondiagonal elements are less than 0.5% and 0.3%, respectively. It is shown in our analysis of experimental results that the elemental uncertainty of the Mueller matrix measurement does not change the main characteristic features of the FDH, as well as the central moment parameters.2.2.Biological Tissue SamplesIn previous studies, we have taken the 2-D backscattering Mueller matrix images of different tissues, and analyzed the relations between their characteristic microstructural features and the Mueller matrix elements.10 The tissue samples are shown in Fig. 2: (a) bovine skeletal muscle, (b) chicken heart muscle, (c) porcine liver, and (d) porcine fat. Both the bovine skeletal muscle and the chicken heart muscle are anisotropic, but the fibers are mostly aligned in the same direction for the bovine muscle sample while aligned concentrically around the ventricle for the chicken heart sample.25 Porcine liver and fat tissues are close to isotropic, but the liver sample contains many hexagonal boundaries of hepatic lobules, which are connective tissues of birefringence.10 The use of the animal tissues in this study was approved by the Administrative Committee on Animal Research of the Graduate School at Shenzhen, Tsinghua University. From the backscattering Mueller matrix images shown in Fig. 3, we can obtain abundant structural information of these samples. For instance, the anisotropy of tissues may originate from both the optical birefringence and cylindrical scatterers, which can be distinguished by the features in different Mueller matrix elements.10 The contributions due to the fibrous scatterers are encoded in the m12, m21, m13, and m31 elements (, ), while those due to the birefringence are encoded in the m24, m42, m34, and m43 elements (, ). The orientation of fibrous structures and the depolarization power can also be extracted from the Mueller matrix elements.10 Since the Mueller matrix contains abundant information on the tissue samples, we may not have to rely on the 2-D images to identify their characteristic features. It can be helpful to find a method to transform the 2-D images of Mueller matrix elements into a group of quantitative or semiquantitative, orientation insensitive parameters, which are crucial for the extraction of the dominant microstructural information of samples. 2.3.Central Moment AnalysisTo quantitatively evaluate the Mueller matrix elements, we adopt the central moment method for statistical analysis of frequency distributions.26 Suppose we have a random variable , whose central moments: expected value, variance, skewness, and kurtosis are exactly represented by Eq. (2).27 Here, the expected value P1 is the mean value of an FDH. The second central moment P2 is called the variance usually denoted by , where represents the standard deviation of the FDH. A small P2 indicates that the measured data tend to be distributed close to the expected value, while a large P2 indicates that the data points are spread out around the expected value and from each other. The third and fourth central moments P3 and P4 represent the skewness and kurtosis of the FDH, respectively. P3 (skewness) shows the asymmetry of the FDH. The skewness value can be positive or negative. A negative (or positive) skewness value indicates that the tail on the left side (or the right side) of the FDH is longer or fatter than the right side (or the left side). P4 (kurtosis) indicates the “peakedness” of the FDH. It is a descriptor of the shape of a probability distribution. The detailed physical interpretations of kurtosis, skewness, and the central moments can be found in Refs. 26 and 28. In this paper, we first record the 2-D backscattering Mueller matrix images of samples, and then transfer the pixel images to FDHs by statistical analysis. Finally, the central moment method is applied to the FDHs of Mueller matrix elements. 3.Results and Discussion3.1.Characteristic Features in Mueller Matrix Elements of Tissues at Different OrientationsFigure 4 shows the FDH of the backscattering Mueller matrix elements of the bovine skeletal muscle tissue. Previous studies have shown that the anisotropic muscle fibers can seriously affect the polarization measurement, resulting in difficulties of the structural information extraction.19 In this study, we rotate the skeletal muscle sample, measure the Mueller matrix images in different orientations, and then choose a square area of at the same location of the sample. In Fig. 4, the horizontal axis of each FDH represents the value of the pixel from the corresponding Mueller matrix element, while the vertical axis represents the distributing probability. There are four experimental curves of the bovine skeletal muscle sample along 30 deg (black lines), 60 deg (red lines), 120 deg (green lines), and 150 deg (blue lines) directions [the angle is indicated as in Fig. 1(c)]. To make the evaluation quantitative, we apply the central moment method to the FDHs, and calculate the parameters P1, P2, P3, and P4, which are listed in Table 1. Figure 4 and Table 1 show that the FDHs are transformed into the quantitative central moment parameters: the expected value P1, variance P2, skewness P3, and kurtosis P4 which all together characterize the position and shape of the FDH curves. Table 1Central moment parameters of the Mueller matrix elements for bovine skeletal muscle tissue.
It can be observed from Fig. 4 and Table 1 that, as the orientation direction of the fibrous sample changed, most FDHs of the elements move except the m11, m14, m41, and m44, showing the influence of sample orientation on Mueller matrix measurements. Compared with the 2-D images, the FDHs and central moment parameters still reveal clearly and quantitatively the same main structural features of tissues as summarized in our previous studies:10,18 (1) the Mueller matrix shown in Fig. 4 and Table 1 is nondiagonal, the m22 and m33 elements are not equal (for instance, for the muscle sample along 30 deg the P1 of m22 and m33 are 0.114 and 0.241, respectively), and testifying that the bovine skeletal muscle is anisotropic. (2) We also notice that the FDHs of the m24, m34, m42, and m43 elements represent slight variations (the absolute value of P1 varies from 0.001 to 0.022 in different angles), indicating the existence of birefringence in this sample. Further analysis of the positive or negative values of these elements can provide the orientation axis information.18 (3) The values of the diagonal elements m22, m33, and m44 are relatively small (P1 mostly distributed in 0.02 to 0.3), indicating a large depolarization property.10 The influences of sample orientation on the Mueller matrix elements are serious. Mueller matrix images of an anisotropic tissue sample along different directions may look like from different samples. However, Fig. 4 shows that the FDH curves can help us to distinguish that it is a different sample or just the same sample along a different orientation; as the sample rotated, the positions of the curves move but their shapes almost remain the same. Table 1 quantitatively confirms that when we rotate the muscle sample, the values of P1 change periodically, which indicates the variations of the positions for the FDHs. Meanwhile, the values of P2 (variance), P3 (skewness), and P4 (kurtosis) display very small changes, indicating the similar shapes (distributing width, asymmetry, and peakedness) of the FDHs. It should be pointed out that, compared to P2 and P4, the variation of P3 seems to be prominent in Table 1. It is because the length of the confidence interval of P3 is related to the absolute value of the skewness. When the skewness is very small, the confidence interval becomes wider.27 Therefore, according to the small values of P3 in this work, its confidence interval is about to 0.3. The values of P3 shown in Table 1 mean that the asymmetry of the FDH curves shown in Fig. 4 can be treated as almost the same. For comparisons, we also take the backscattering Mueller matrix images of porcine liver tissue along different sample orientations. The liver tissues are primarily isotropic, but contain many thin hexagonal structures around the isotropic liver tissues, which are identified as birefringent connective tissues.10 Figure 5 shows the FDHs of the liver sample at 30 deg (black lines), 60 deg (red lines), 120 deg (green lines), and 150 deg (blue lines) directions, respectively. The central moment parameters P1, P2, P3, and P4 are listed in Table 2. Table 2Central moment parameters of the Mueller matrix elements for porcine liver tissue.
It can be observed from Fig. 5 and Table 2 that, the FDHs of porcine liver tissues have isotropic structural features, which are different from those of the fibrous skeletal muscle shown in Fig. 4 and Table 1: (1) the Mueller matrix shown in Fig. 5 and Table 2 is nearly diagonal (the P1 values of the m12, m21, m13, and m31 are close to 0), and the m22 and m33 elements are equal, representing isotropic dominant properties. (2) The m24, m34, m42, and m43 elements show slight differences (almost , ), indicating the existence of birefringent connective tissues. (3) Compared to the skeletal muscle sample, the values of the diagonal elements of the liver tissue are larger (P1 mostly distributed in 0.25 to 0.5), reminding a smaller depolarization property. We also notice that when the sample rotates, the FDHs of the Mueller matrix elements all keep the same except the m24, m34, m42, and m43, confirming that for isotropic samples, the influence of orientation on polarization measurements is limited. Again, the FDHs and central moment parameters reveal more clearly and quantitatively the characteristic features of Mueller matrix elements for isotropic samples.10 From the data shown in Tables 1 and 2, we can conclude that: for the same sample placed along different orientations, the values of P2, P3, and P4 for the Mueller matrix elements almost do not change, while the value of P1 can be varied (for the anisotropic sample) or constant (for the isotropic sample). The central moment analysis of the FDHs provides us a tool to transform the complicated 2-D Mueller matrix images to a group of quantitative indicators of dominant structural properties of tissues. More importantly, using the P2, P3, and P4, we can obtain the main intrinsic properties of samples without the influence from orientation variations. 3.2.Comparisons of Tissue Samples with Distinctive MicrostructuresIn Sec. 3.1, we have found that the shapes of the FDHs of Mueller matrix elements are orientation insensitive, therefore may be used as indicators for intrinsic microstructural features of different tissues. To study the relationship between the FDHs and the structural properties, we take the Mueller matrices of the tissue samples as shown in Fig. 2. Figure 6 represents the experimental results of bovine skeletal muscle tissue (black lines), porcine liver tissue (red lines), chicken heart tissue (green lines), and porcine fat tissue (blue lines), the imaging area is a square of . During the measurements, the orientations of the samples are kept the same, for the bovine skeletal muscle sample its fibers are along the 30 deg direction (). We can see from Fig. 6 that the FDHs of different tissues have very different distributions. The corresponding central moment parameters are calculated and listed in Table 3. In the discussions above, we have summarized the characteristic features of anisotropic and isotropic tissues using Figs. 4, 5 and Tables 1, 2. Here, we can confirm the relations between the microstructures and the distributions of Mueller matrix elements from Fig. 6 and Table 3 more clearly. Table 3Central moment parameters of the Mueller matrix elements for different tissues.a
First, the anisotropic and isotropic tissues can be distinguished by using the diagonal elements. The porcine liver (red lines) and fat (blue lines) tissues are predominantly isotropic; therefore, their m22 and m33 curves are almost the same (for example, for the fat tissue, P1 of the m22 and m33 are 0.041 and 0.041, P2 are 0.013 and 0.012). The anisotropic bovine skeletal muscle (black lines) and chicken heart (green lines) tissues, however, display differences between the m22 and m33, which become more prominent as the anisotropy increases. Table 3 shows that the differences in P1 of m22 and m33 elements for anisotropic skeletal muscle and heart samples are 0.127 and 0.014, respectively. This is because that the fibers in skeletal muscle sample are well aligned in almost the same direction, while in heart sample the fibers are distributed in different orientations. For isotropic fat and liver tissues, the differences in P1 of m22 and m33 elements are 0 and 0.002. This is because the fat tissue is totally isotropic, while the liver sample has a small portion of birefringent connective tissues. Second, we also notice that the distribution widths of the FDHs (the values of P2) for bovine skeletal muscle, chicken heart, and porcine liver samples are larger than the fat sample, indicating more complicated microstructures for these metabolic exuberant tissues. The FDHs of the m24, m42, m34, and m43 elements for skeletal muscle, heart, and liver tissues show small positive or negative values, which are related to the birefringent structures in these tissues. The signs of the elements can be used to determine the aligned fibers directions.18 At last, the different depolarization power of tissues can also be observed from Fig. 6 and Table 3: the liver tissue sample has the largest P1 values of the diagonal elements, showing the smallest depolarization power, while the smallest P1 values of the diagonal elements indicate the most prominent depolarization property of the fat tissue.10 Although more studies are still needed to reveal the relationships between the derived parameters and tissue morphology, it has been shown that the parameter P2 should be sensitive to the complexity of a sample: a large value of P2 means that the measured polarization data are distributed in a wider range, indicating a complex structural feature of the tissue. The parameter P3 should be sensitive to the heterogeneity of a sample: a large value of P3 means that the measured polarization data are unequally distributed around the expected value. The parameter P4 can also be used to reflect the complexity of a sample: a large P4 shows that most measured polarization data are distributed very close to the mean value, meaning that the microstructural features are similar. In summary, from the results discussed above, we can conclude that: (1) the shapes of FDHs (values of P2, P3, and P4) are orientation insensitive, therefore can reflect some intrinsic structural properties of the samples. (2) The FDHs and corresponding central moment parameters of Mueller matrix elements are good quantitative indicators of the microstructures. Although the 2-D images contain more detailed structural information as described in our previous studies, transforming the images into FDHs and quantitative central moment parameters can reveal the dominant features of tissues. 3.3.Application to Human Cervical Cancerous TissuesTo testify the potential applications of the pixel FDHs and central moment parameters on diagnosis, we take the Mueller matrices of an unstained -thick slice of human cervical cancerous tissue prepared and provided by the Shenzhen Sixth People’s (Nanshan) Hospital (ID: 120900924), the detailed information of the tissue can be found in Ref. 10. This work was approved by the Ethics Committee of the Shenzhen Sixth People’s (Nanshan) Hospital. We also choose squares from both the normal and abnormal regions of the 2-D images, and then calculate the FDHs and central moments of Mueller matrix elements shown as shown in Fig. 7 and Table 4. Table 4Central moment parameters of the Mueller matrix elements for human cervical cancerous tissue.
It can be observed from Fig. 7 and Table 4 that the normal and abnormal cervical tissues represent different structural features. The normal region has larger anisotropy (more prominent difference between the m22 and m33 elements) and depolarization properties (smaller values of the m22, m33, and m44 elements) than the abnormal region. The positive and negative values of the m34 and m43 elements also indicate the existence of birefringence in normal tissue (P1 of the m34 and m43 are and 0.020), while for abnormal tissue, the birefringent effect becomes limited.9–10 Besides, we also notice that the values of P3 and P4 for normal and abnormal tissues represent large difference in some elements, such as the m22, m33, and m44. These preliminary studies show that the FDHs and central moment parameters may have the potential to give quantitative or semiquantitative criteria for cancerous tissues detections. 4.ConclusionIn this work, we take the backscattering Mueller matrices of tissues of distinctively different microstructures: bovine skeletal muscle tissue, porcine liver tissue, chicken heart tissue, and porcine fat tissue, then use the pixel FDH and central moment analysis to transform the 2-D Mueller matrix images to a group of quantitative indicators for characterizing the dominant structural properties of tissues. By rotating anisotropic skeletal muscle sample and isotropic liver tissue, we find that the central moment parameters P2, P3, and P4 are insensitive to sample orientation directions. Comparisons among different tissues testify that the distribution behavior and corresponding central moment parameters of Mueller matrix elements are good indicators of the microstructures of tissues. A preliminary application to human cervical cancerous tissues shows that the distribution curves and central moment parameters may have the potential to give quantitative or semiquantitative criteria for cancerous tissues detections. AcknowledgmentsThis work has been supported by the National Natural Science Foundation of China (NSFC) under Grant Nos. 11174178, 11374179, 61205199, 61405102, and the Science and Technology Project of Shenzhen Grant Nos. CXZZ20140509172959978 and JCY20130402145002404. ReferencesR. S. Gurjar et al.,
“Imaging human epithelial properties with polarized light-scattering spectroscopy,”
Nat. Med., 7
(11), 1245
–1248
(2001). http://dx.doi.org/10.1038/nm1101-1245 Google Scholar
B. Kunnen et al.,
“Application of circularly polarized light for non-invasive diagnosis of cancerous tissues and turbid tissue-like scattering media,”
J. Biophotonics, 8
(4), 317
–323
(2015). http://dx.doi.org/10.1002/jbio.201400104 Google Scholar
N. Ghosh and I. A. Vitkin,
“Tissue polarimetry: concepts, challenges, applications, and outlook,”
J. Biomed. Opt., 16
(11), 110801
(2011). http://dx.doi.org/10.1117/1.3652896 Google Scholar
S. L. Jacques, J. C. Ramella-Roman and K. Lee,
“Imaging skin pathology with polarized light,”
J. Biomed. Opt., 7
(3), 329
–340
(2002). http://dx.doi.org/10.1117/1.1484498 Google Scholar
R. R. Anderson,
“Polarized light examination and photography of the skin,”
Arch. Dermatol., 127
(7), 1000
–1005
(1991). http://dx.doi.org/10.1001/archderm.1991.01680060074007 Google Scholar
S. Alali and I. A. Vitkin,
“Polarized light imaging in biomedicine: emerging Mueller matrix methodologies for bulk tissue assessment,”
J. Biomed. Opt., 20
(6), 061104
(2015). http://dx.doi.org/10.1117/1.JBO.20.6.061104 Google Scholar
E. Du et al.,
“Mueller matrix polarimetry for differentiating characteristic features of cancerous tissues,”
J. Biomed. Opt., 19
(7), 076013
(2014). http://dx.doi.org/10.1117/1.JBO.19.7.076013 Google Scholar
P. Shukla and A. Pradhan,
“Mueller decomposition images for cervical tissue: Potential for discriminating normal and dysplastic states,”
Opt. Express, 17
(3), 1600
–1609
(2009). http://dx.doi.org/10.1364/OE.17.001600 Google Scholar
A. Pierangelo et al.,
“Polarimetric imaging of uterine cervix: a case study,”
Opt. Express, 21
(12), 14120
–14130
(2013). http://dx.doi.org/10.1364/OE.21.014120 Google Scholar
M. Sun et al.,
“Characterizing the microstructures of biological tissues using Mueller matrix and transformed polarization parameters,”
Biomed. Opt. Express, 5
(12), 4223
–4234
(2014). http://dx.doi.org/10.1364/BOE.5.004223 Google Scholar
A. Pierangelo et al.,
“Multispectral Mueller polarimetric imaging detecting residual cancer and cancer regression after neoadjuvant treatment for colorectal carcinomas,”
J. Biomed. Opt., 18
(4), 046014
(2013). http://dx.doi.org/10.1117/1.JBO.18.4.046014 Google Scholar
A. Pierangelo et al.,
“Ex vivo photometric and polarimetric multilayer characterization of human healthy colon by multispectral Mueller imaging,”
J. Biomed. Opt., 17
(6), 066009
(2012). http://dx.doi.org/10.1117/1.JBO.17.6.066009 Google Scholar
A. Pierangelo et al.,
“Ex-vivo characterization of human colon cancer by Mueller polarimetric imaging,”
Opt. Express, 19
(2), 1582
–1593
(2011). http://dx.doi.org/10.1364/OE.19.001582 Google Scholar
M. Dubreuil et al.,
“Mueller matrix polarimetry for improved liver fibrosis diagnosis,”
Opt. Lett., 37
(6), 1061
–1063
(2012). http://dx.doi.org/10.1364/OL.37.001061 Google Scholar
W. Wang et al.,
“Roles of linear and circular polarization properties and effect of wavelength choice on differentiation between ex vivo normal and cancerous gastric samples,”
J. Biomed. Opt., 19
(4), 046020
(2014). http://dx.doi.org/10.1117/1.JBO.19.4.046020 Google Scholar
J. Jagtap et al.,
“Quantitative Mueller matrix fluorescence spectroscopy for precancer detection,”
Opt. Lett., 39
(2), 243
–246
(2014). http://dx.doi.org/10.1364/OL.39.000243 Google Scholar
I. Ahmad et al.,
“Ex vivo characterization of normal and adenocarcinoma colon samples by Mueller matrix polarimetry,”
J. Biomed. Opt., 20
(5), 056012
(2015). http://dx.doi.org/10.1117/1.JBO.20.5.056012 Google Scholar
H. He et al.,
“Mapping local orientation of aligned fibrous scatterers for cancerous tissues using backscattering Mueller matrix imaging,”
J. Biomed. Opt., 19
(10), 106007
(2014). http://dx.doi.org/10.1117/1.JBO.19.10.106007 Google Scholar
M. Sun et al.,
“Probing microstructural information of anisotropic scattering media using rotation-independent polarization parameters,”
Appl. Opt., 53
(14), 2949
–2955
(2014). http://dx.doi.org/10.1364/AO.53.002949 Google Scholar
H. He et al.,
“Two-dimensional and surface backscattering Mueller matrices of anisotropic sphere-cylinder scattering media: a quantitative study of influence from fibrous scatterers,”
J. Biomed. Opt., 18
(4), 046002
(2013). http://dx.doi.org/10.1117/1.JBO.18.4.046002 Google Scholar
R. M. A. Azzam,
“Photopolarimetric measurement of the Mueller matrix by Fourier analysis of a single detected signal,”
Opt. Lett., 2
(6), 148
–150
(1978). http://dx.doi.org/10.1364/OL.2.000148 Google Scholar
D. H. Goldstein,
“Mueller matrix dual-rotating retarder polarimeter,”
Appl. Opt., 31
(31), 6676
–6683
(1992). http://dx.doi.org/10.1364/AO.31.006676 Google Scholar
D. H. Goldstein and R. A. Chipman,
“Error analysis of a Mueller matrix polarimeter,”
J. Opt. Soc. Am. A, 7
(4), 693
–700
(1990). http://dx.doi.org/10.1364/JOSAA.7.000693 Google Scholar
D. B. Chenuault, J. L. Pezzaniti and R. A. Chipman,
“Mueller matrix algorithms,”
Proc. SPIE, 1746 231
–246
(1992). http://dx.doi.org/10.1117/12.138793 Google Scholar
R. Liao et al.,
“Rotating linear polarization imaging for quantitative characterization of anisotropic tissues,”
J. Biomed. Opt., 15
(3), 036014
(2010). http://dx.doi.org/10.1117/1.3442730 Google Scholar
G. Grimmett and D. Stirzaker, Probability and Random Processes, Oxford University Press, New York
(2001). Google Scholar
X. D. Li,
“Value at risk based on the volatility, skewness and kurtosis,”
(1999). Google Scholar
V. A. Ushenko, O. V. Dubolazov and A. O. Karachevtsev,
“Two wavelength Mueller matrix reconstruction of blood plasma films polycrystalline structure in diagnostics of breast cancer,”
Appl. Opt., 53
(10), B128
–B139
(2014). http://dx.doi.org/10.1364/AO.53.00B128 Google Scholar
|