|
1.IntroductionPolarization imaging techniques are capable of probing the microstructural and optical properties of samples.1,2 They have been applied in different fields, including material characterization,3,4 medical diagnosis,5 remote sensing,6 target detection,7 and so on. The observation of microstructure makes it necessary to combine polarization imaging techniques with microscope systems. Oldenbourg8 established a polarization microscope by adding liquid crystal polarization modulator to a commercial transmission microscope. Other attempts for polarization microscopes have also been reported.9–11 These microscopes use monochromatic light sources and qualitatively analyze the birefringence properties of the samples. For complicated samples with strong depolarization and complex anisotropy like various biological tissues and cells, birefringence alone is not enough to describe different polarization characteristics. Mueller matrix measurements are more appropriate for polarization characterization of biological tissues because it describes all of the polarization properties, such as diattenuation, retardance, and depolarization.12–17 The hardware of Mueller matrix measurement is mostly similar to the corresponding nonpolarized measurements except for the polarization modulation optics. Therefore, a normal transmission or reflection mode microscope can be converted to a Mueller matrix microscope by adding a polarization-state generator (PSG) and a polarization-state analyzer (PSA) to the existing optical path. Different wavelengths of light behave differently in scattering, absorption, and propagation processes of the same biological tissue. Therefore, the feature parameters used to describe the sample, such as retardance ,18 that is extracted from different wavelength Mueller matrices of the same biological tissue are also different. Compared with a monochromatic Mueller matrix microscope, polarization measurements with different wavelengths will bring up extra information on the microstructure of samples. For different biological samples or different feature parameters, we can take advantage of multiple wavelengths to choose an optimal wavelength to measure the Mueller matrix. Previously, we reported a monochromatic Mueller matrix microscope by adding PSG and PSA to a commercial transmission-light microscope16 and demonstrated preliminary applications in detecting human cancerous tissues with fibrosis.2 In this paper, we report a multiwavelength version of the microscope and include detailed descriptions on how the microscope is designed, constructed, calibrated, and tested. The Mueller microscope is also based on an ordinary commercial optical microscope. We use a multiwavelength collimated LED light source module to substitute the ordinary halogen light source, add a set of polarizing and analyzing modules, and use a CCD detector as the image receiving module. The dual-rotating quarter-wave plates Mueller matrix measurement method is adopted to measure the Mueller matrix of samples by simultaneously rotating two quarter-wave plates at certain steps.19 For higher accuracy, optimizations and calibration processes are adopted to reduce the errors in the polarization measurements. With the advantages of simple structure, rapid measurement, and high precision, the modulus design polarization microscope for transmission Mueller matrix imaging shows promising prospects. 2.Theory of Operations and Selection Criteria of Components2.1.Theory of OperationsAfter comparing the division of focal plane polarimeter20,21 and rotating both polarizers and quarter-wave plates Mueller matrix measurement,22 we selected the dual-rotating quarter-wave plates Mueller matrix measurement method for the polarization microscope. In this measurement, there are only two quarter-wave plates synchronously rotating at a set of angular increments that introduce less error sources and are easy to be reconstructed and controlled by the computer. The main principle of the dual-rotating quarter-wave plates Mueller matrix measurement method is shown below. Figure 1 shows that the state-of-polarization (SoP) of detecting light can be expressed with the Mueller matrix of components and the light intensity value is the first component of ; therefore, the light intensity can be expressed as the following equation: where and represent the Stokes vector of the incident nature light and the detecting light, respectively, and represent the Mueller matrices of linear polarizer and quarter-wave plate in the PSG module, respectively, and represent the Mueller matrices of linear polarizer and quarter-wave plate R2 in the PSA module, respectively, represents the Mueller matrix of the sample, and represents the detector sensitivity.Let , which represents the Stokes vector of light leaving the PSG module and represents the elements in the Stokes vector. Let , which represents the first row of the Mueller matrix of PSA module, and represent the elements in the Stokes vector. Then, we can use to represent the elements in the vector product of and and obtain the following expression: where and represent the rotating angles of the two quarter-wave plates and , respectively. The detected light intensity can be expressed in a periodic form based on the Fourier analysis as shown below: where the fundamental frequency and () are Fourier amplitudes.By simultaneously solving Eqs. (2) and (4), we know that the Fourier amplitudes provide linear algebraic equations of the 16 unknown Mueller-matrix elements . Twenty-five Fourier amplitudes can be calculated if 30 measurements are taken. Five of the twenty five values are opposite to another five, and four values are always zero. Thus, we can derive the Mueller matrix based on the 16 independent amplitude values. 2.2.Selection Criteria of the Transmission MicroscopeAlthough Mueller matrix measurements can be realized by adding polarization optical elements to an ordinary transmission microscope, we need to choose microscopes suitable for the modification. According to our experience, we should choose the transmission microscope with the following three characteristics. First, the microscope should have an external light source that is easily replaced by a multiwavelength collimated light source (CLS) module. Second, the microscope should be an infinity optical system with infinity plan achromatic objectives, making it easier to add the PSA module in the detecting optical path. Finally, there should be a long moving-range sample stage that has spare space for adding the PSG module in the microscope. In this work, we chose a low cost commercial transmitted-light microscope (L2050, LISS Optical Co. Ltd., China). Based on the modulus design, the polarized light microscope can be easily disassembled and upgraded. 2.3.Selection Criteria of the Light SourceMonochrome LEDs are promising light sources for the microscopy measurements. They are small, lightweight, low cost, and convenient for working at low voltages.23,24 There are six different monochrome LEDs installed in the microscope; the center wavelengths of the LEDs are 455 nm (royal blue), 475 nm (blue), 490 nm (cyan), 535 nm (green), 590 nm (amber), and 630 nm (red), respectively, and each of their full-width at half-maximum is close to 15 nm. 2.4.Selection Criteria of the Polarization ModulesThere are many types of instrumental setups used for time sequential polarization modulation, including rotating polarized optical components,25,26 photoelastic modulators,27 and liquid crystal variable retarder.28 The photoelastic modulator is expensive, and the CCD detector that we used cannot take advantage of its high modulation speed. The liquid crystal variable retarder is sensitive to the ambient temperature. Thus, in our system, polarization-state modulations in both PSG and PSA are realized by rotating the wave plates around the optical axis. In the microscope, all the polarized components , and are placed horizontally. The transmission axes of the polarizers and the fast axes of the quarter-wave plates are parallel to each other; then, the PSG and PSA are modulated by synchronously rotating two quarter-wave retarders at fixed angles per step. 3.Experimental Setup3.1.General Design of HardwareFigure 2(a) shows a general schematic of the modulus design polarization microscope for transmission Mueller matrix imaging. The polarization microscope consists of several major modules: (i) CLS module, (ii) PSG module, (iii) PSA module, and (iv) image receiving module. In the CLS module, the diverging light from a monochrome LED (one of the six LEDs of different colors) is collimated by a set of fixed upright standing convex lenses (GCL-010158A, GCL-010111A, Daheng New Epoch Technology, Inc., China) and turned to the upright direction by a steering mirror M. The parallel beam passes through the PSG module, which consists of a fixed polarizer (LPVISE100-1, Thorlabs, Inc.) and an achromatic quarter-wave plate (WPA4225, 450-650 nm, Union Optic Co., Ltd., China) mounted on a computer controlled motorized rotational stage (PRM1Z8E, Thorlabs, Inc.). The parallel beam with well controlled SoP illuminates the sample S on the translation stage. The transmitted light from the sample is collected by the objective lens O and then passes though the PSA module, which also consists of an achromatic quarter-wave plate R2 (WPA4225, 450-650 nm, Union Optic, Inc., China) mounted on a computer controlled motorized stage (PRM1Z8E, Thorlabs, Inc.) and a fixed linear polarizer (LPVISE100-1, Thorlabs, Inc.). Finally, the analyzed light is focused and detected by a 12-bit CCD detector (74-0107A, Dynisco, Inc., China). The exposure time takes from 0.1 to 1 ms for imaging different samples, but the time for setting the PSG and PSA in position takes up to 3 s as it is limited by the speed of the rotational stage in PSA, as shown in Fig. 2(b). For each measurement, the system takes 30 polarization component images corresponding to different combinations of the PSG and PSA, and then evaluates the 16 Mueller matrix elements following Eqs. (2) and (4). It takes about one and a half minutes for a Mueller matrix measurement. 3.2.Software for System Control, Data Acquisition and AnalysisThe software system of the polarized microscope is capable of hardware control, data acquisition, data processing, and data analysis. A graphical user interface program for instrument control and data acquisition has been developed using the National Instruments’ Labview (Laboratory Virtual Instrument Engineering Workbench) graphical programming platform. We adopt Labview program to control the two high-precision motorized rotation stages and the CCD, as shown in Fig. 1. For each measurement, first, the system control program makes the rotation stages turn to a specified angle separately, making the transmission axes of the polarizers and the fast axes of the quarter-wave plates parallel. Second, the program repeats the image collecting process 30 times, during which the two rotation stages and rotate at preset angles, respectively. Then, the CCD takes one or several repeated images with suitable exposure time ranging from 0.1 to 1 ms. Finally, the Labview program takes the two rotation stages back to their initial positions ready for the next measurement. In this Labview program, we can set the number for repetitive data collections, rotation angles for each step, exposure time, and data storage location. With a computer controlling the program, the measurement time is mainly limited by the rotating speeds of the motors. The current microscope takes a minimum of 91 s to complete a Mueller matrix measurement, 90 s for the rotations of the wave plates, and about 1 s for calculating the Mueller matrix from the 30 polarization component images. We use MATLAB for both data processing and for calibrations that will be discussed later. 4.System Optimization and Calibration4.1.Optimization of Imaging Receiving ModuleThe calculation of the Mueller matrix is based on the light intensity value from the CCD. To ensure accurate light intensity detection and avoid overexposure, the illumination intensity is controlled to make the maximum pixel values of the polarization component images in CCD below 4000. The error of light intensity value is mainly related to the stability of light source and the noise of optoelectronic detectors. In this section, we take the light intensity value of a single measurement and the average value of 2, 4, 6, 8, 10, 12, 14, and 16 repeated measurement results as experiment results. Then, we calculate the mean value and the variance of 10 parallel experiments to estimate the stability of the measured light intensity values. Using 630-nm (red) LED as illumination light, we display the values and the variance of pixels in the middle of the image with the form of error-bar in Fig. 3. There is a significant decrease in the variance of the direct intensity values and the average values of 2, 4, 6, 8, and 10 repeated images, as shown in Figs. 3(a)–3(f). In addition, Figs. 3(f)–3(i) indicate that the average value of over 10 repeated measurements results in a very small improvement to the stability of light intensity value. We get the same results with other wavelengths of illumination light. Thus, we use the average value of 10 repeated results to substitute the single measurement result in the Mueller matrix measurements. 4.2.Modeling Method of Polarization CalibrationCalibrations of the polarization optics are very important for accurate Mueller matrix measurements. The polarization elements , and have errors associated with their initial azimuthal alignment with respect to the transmission axis of . In addition, the retarders may slightly deviate from the retardance of 90 deg. We build a simple model based on the study of Goldstein,29 which include five potential error sources, respectively, represented by , , , , and , as shown in Fig. 1, and can be expressed as shown below: The Mueller matrix and containing the errors can be expressed as shown below: and can be expressed by analogizing . By contrasting Eqs. (2) and (5), we can derive the 16 unknown Mueller-matrix elements with Fourier amplitudes and errors. When air is used as the reference sample, the Mueller matrix should be strictly diagonal. Through inverse operations, the five errors can be calculated from the Fourier coefficient in the following equations:By taking many measurements, we can reduce , , and to 1 deg, and the maximum errors of the Mueller matrix elements with angular calibration can be less than 0.01. 4.3.Eigenvalue Calibration MethodAfter the modeling method calibration, the errors associated with the polarized optics in the Mueller matrix measurements can be mostly eliminated. It is noteworthy that the objective lens also affects Mueller matrix measurements. Although objective lenses are usually regarded as nonpolarization optics, sometimes their polarization properties show up as birefringence and depolarization. The birefringence of glass lens is usually small but could be big enough to affect the accuracy of Mueller matrix measurements if miniaturized or organic material lenses are used. Wolfe and Chipman30 found the annealing of lens generated by depolarization phenomenon and Wood and Elson31 found that the optical quartz window had an impact on polarization imaging. We use the Mueller matrix to characterize the polarization property of an objective. Five standard samples, i.e., air, polarizer in 45 deg, polarizer in 90 deg, quarter-wave plate in 45 deg, and quarter-wave plate in 90 deg, are used to evaluate the Mueller matrix of an objective using the eigenvalue calibration method.32,33 The Mueller matrix of air is a unit matrix, and the measured results of air and other samples can be expressed as shown below: where represents the experimental Mueller matrix of the air and represents the experimental Mueller matrix of the th type of standard samples. represents the Mueller matrix of the air and represents the Mueller matrix of the th standard sample. represents the unknown Mueller matrix of the objective. We used experimental result and to define a new matrix and obtain the following equation: by multiplying the inverse of matrix on both sides of the equation. Equation (11) formed a linear system whose optimal solution can be derived by the least square method. We define the linear operator to describe the different value.For convenience of calculation, and are arranged in rows and projected into vectors, which are shown as and separately in the following equation: and the coefficient matrix , whose size is , is calculated. For convenience, we use to represent , as shown in Eq. (12). The square sum of the difference values can be expressed as shown below: and vector is the eigenvector of the real symmetric positive semidefinite matrix .We define to represent the sum of and find the smallest eigenvalue and the corresponding eigenvector, which can be rearranged into the Mueller matrix of the objective, as the calculated result of the 4× objective shown in Table 1. Table 1The calculated optimal Mueller matrix of the 4× objective.
4.4.Optimization of Rotation AnglesAfter the calibration as above, the accuracy of the Mueller matrix measurement is enhanced, but errors still exist in the system. Fortunately, a good measurement system can reduce the effects of errors to the results, and the condition number (CN) of the measurement matrix is a metric of the stability of the system. We calculate the CN of the Mueller matrix microscope with the Mueller matrix of the objective when the rotation angles of the two quarter-wave plates change, as shown in Fig. 4. For the smallest possible CN and the convenience of operation, we chose the smallest and , so , , and . 4.5.Solution to Beam DriftingWe find that the images received by the detector cannot line up perfectly. The cause of this is that there are two rotation polarization devices in the system, and the surface of the polarization components may be unparallel. The beam of light drifts, and the measurement regions form dislocations because of rotation, as shown in Fig. 5. We calculate in Eq. (14) the relationship between the offset of the image and the angle γ between the optical path and the central axis of the rotating optical component in PSA. We enhance the accuracy of image matching before calculation and show the original image and the improved image of MMD parameter of porous alumina34 in Fig. 6. Comparing the results shown in Figs. 6(a) and 6(b), the MMD parameter of porous alumina with image registration has less noise and a clearer boundary, which demonstrates that our strategy can reduce part of the effects due to light beam drift: 5.System Test Result and Experiment Result5.1.System Test ResultIn addition to air, we also measured a polarizer as a standard sample. Table 2 shows the measurement results of the polarizer for six different wavelengths, which demonstrate that the error for each wavelength is below 0.01 after calibration. Therefore, the measurement results of our polarized light microscope are reliable. Table 2(a)–(f) The final Mueller matrix of polarizer after calibration with light in 455 nm (royal blue), 475 nm (blue), 490 nm (cyan), 535 nm (green), 590 nm (amber), and 630 nm (red) respectively.
5.2.Experimental ResultIn this study, we use histopathologic slides as samples. The histopathologic samples used in this study are nonstained, dewaxing sections of human liver cancer tissues, which were prepared and provided by Shenzhen Sixth People’s (Nanshan) Hospital. We display the images of the Mueller matrix polarization decomposition (MMPD)35 retardance of human liver tissue with different wavelengths in Fig. 7. During the development process of liver cancerous tissues, the accompanied inflammatory reactions can result in fibrosis and cirrhosis, which can be reflected by the values of the MMPD parameter retardance . Therefore, compared with the normal regions, the abnormal regions of the liver tissue slides contain much more fibrous structures, resulting in larger MMPD parameter retardance δ values. The normal regions in liver tissue are mainly the blue regions with low retardance, while the abnormal regions are the red regions with high retardance and marked with black lines in Fig. 7(a). For quantitative analysis, we calculate the mean retardance value of the normal regions and abnormal regions. When the wavelength of light decreases from 630 to 455 nm, retardance of normal regions increases from 0.18 to 0.28, retardance of abnormal regions increases from 0.54 to 0.81, and the retardance difference between the abnormal and normal regions increases from 0.36 to 0.53 correspondingly, as shown in Fig. 8. We can use retardance as a feature parameter to distinguish abnormal regions from normal regions. In this experiment, the retardance difference of 455 nm light is 47% higher than that of 630-nm light, which means that 455-nm light is a better choice because the shorter wavelength is more sensitive to the change of retardance . Because the scattering and absorption characteristics of different samples are not the same, longer wavelength light may have advantages in some other cases. For different samples and different feature parameters, we need to choose a suitable wavelength to measure the Mueller matrix to achieve the best effect. Fortunately, a multiwavelength polarization microscope can give us multiple choices and provide much more useful information compared with a monochromatic Mueller matrix microscope. This work was approved by the Ethics Committee of the Shenzhen Sixth People’s (Nanshan) Hospital. 6.ConclusionsIn conclusion, we have demonstrated that an ordinary optical microscope can be transformed to a multiwavelength polarization microscope for Mueller matrix measurements just by replacing the light source and adding a set of PSG and PSA. The maximum errors for the absolute values of Mueller matrix elements can be reduced to about 0.01 after calibration. By measuring air, polarizer, and histopathologic slides, we verify the reliability of the polarized light microscope and find that we can obtain rich information on the microstructure and optical properties of samples. The Mueller matrix microscope is a very versatile instrument and will have a wide range of applications. DisclosuresThe authors have no relevant financial interests in this article and no potential conflicts of interest to disclose. AcknowledgmentsThis research was supported by the National Natural Science Foundation of China (NSFC) under Grant Nos. 11374179, 61527826, and 61405102 and Science and Technology Project of Shenzhen under Grant Nos. JCYJ20170412170814624 and GJHZ20150316160614844. ReferencesR. A. Chipman,
“Handbook of optics,”
Polarimetry, 2 2nd ed.McGraw Hill, New York
(1995). Google Scholar
Y. Wang et al.,
“Mueller matrix microscope: a quantitative tool to facilitate detections and fibrosis scorings of liver cirrhosis and cancer tissues,”
J. Biomed. Opt., 21
(7), 071112
(2016). http://dx.doi.org/10.1117/1.JBO.21.7.071112 JBOPFO 1083-3668 Google Scholar
G. C. Giakos et al.,
“Electro-optical polarimetric imaging system for characterization of resident space objects,”
J. Spacecr. Rockets, 52
(4), 1112
–1128
(2015). http://dx.doi.org/10.2514/1.A33167 JSCRAG 0022-4650 Google Scholar
B. Peng, T. Ding and P. Wang,
“Propagation of polarized light through textile material,”
Appl. Opt., 51
(26), 6325
–6334
(2012). http://dx.doi.org/10.1364/AO.51.006325 APOPAI 0003-6935 Google Scholar
H. Wang et al.,
“Polarization sensitive optical coherence microscopy for brain imaging,”
Opt. Lett., 41
(10), 2213
–2216
(2016). http://dx.doi.org/10.1364/OL.41.002213 OPLEDP 0146-9592 Google Scholar
S. F. Pellicori and E. Burke,
“Passive multispectral imaging polarimeter for remote atmospheric and surface studies: design based on optical coatings,”
Appl. Opt., 55
(6), 1291
–1301
(2016). http://dx.doi.org/10.1364/AO.55.001291 APOPAI 0003-6935 Google Scholar
F. Goudail and M. Boffety,
“Optimal configuration of static polarization imagers for target detection,”
J. Opt. Soc. Am. A, 33
(1), 9
–16
(2016). http://dx.doi.org/10.1364/JOSAA.33.000009 JOAOD6 0740-3232 Google Scholar
R. Oldenbourg,
“A new view on polarization microscopy,”
Nature, 381
(6585), 811
–812
(1996). http://dx.doi.org/10.1038/381811a0 Google Scholar
S. Ross et al.,
“Quantitative image analysis of birefringent biological material,”
J. Microsc., 187
(1), 62
–67
(1997). http://dx.doi.org/10.1046/j.1365-2818.1997.2160776.x JMICAR 0022-2720 Google Scholar
R. Weaver,
“Rediscovering polarized light microscopy,”
Am. Lab., 35
(20), 55
–61
(2003). Google Scholar
R. Oldenbourg,
“Polarized light field microscopy: an analytical method using a microlens array to simultaneously capture both conoscopic and orthoscopic views of birefringent objects,”
J. Microsc., 231
(3), 419
–432
(2008). http://dx.doi.org/10.1111/jmi.2008.231.issue-3 JMICAR 0022-2720 Google Scholar
M. Antonelli et al.,
“Mueller matrix imaging of human colon tissue for cancer diagnostics: how Monte Carlo modeling can help in the interpretation of experimental data,”
Opt. Express, 18
(10), 10200
–10208
(2010). http://dx.doi.org/10.1364/OE.18.010200 OPEXFF 1094-4087 Google Scholar
A. Pierangelo et al.,
“Polarimetric imaging of uterine cervix: a case study,”
Opt. Express, 21
(12), 14120
–14130
(2013). http://dx.doi.org/10.1364/OE.21.014120 OPEXFF 1094-4087 Google Scholar
N. Ghosh et al.,
“Mueller matrix decomposition for polarized light assessment of biological tissues,”
J. Biophotonics, 2
(3), 145
–156
(2009). http://dx.doi.org/10.1002/jbio.v2:3 Google Scholar
M. Dubreuil et al.,
“Mueller matrix polarimetry for improved liver fibrosis diagnosis,”
Opt. Lett., 37
(6), 1061
–1063
(2012). http://dx.doi.org/10.1364/OL.37.001061 OPLEDP 0146-9592 Google Scholar
Y. Wang et al.,
“Differentiating characteristic microstructural features of cancerous tissues using Mueller matrix microscope,”
Micron, 79 8
–15
(2015). http://dx.doi.org/10.1016/j.micron.2015.07.014 MICNB2 0047-7206 Google Scholar
D. Dixit et al.,
“Optical critical dimension metrology for directed self-assembly assisted contact hole shrink,”
J. Micro/Nanolith. MEMS MOEMS, 15
(1), 014004
(2016). http://dx.doi.org/10.1117/1.JMM.15.1.014004 Google Scholar
Y. Dong et al.,
“Quantitatively characterizing the microstructural features of breast ductal carcinoma tissues in different progression stages by Mueller matrix microscope,”
Biomed. Opt. Express, 8
(8), 3643
–3655
(2017). http://dx.doi.org/10.1364/BOE.8.003643 BOEICL 2156-7085 Google Scholar
R. Azzam,
“Photopolarimetric measurement of the Mueller matrix by Fourier analysis of a single detected signal,”
Opt. Lett., 2
(6), 148
–150
(1978). http://dx.doi.org/10.1364/OL.2.000148 OPLEDP 0146-9592 Google Scholar
J. Chang et al.,
“Division of focal plane polarimeter-based 3×4 Mueller matrix microscope: a potential tool for quick diagnosis of human carcinoma tissues,”
J. Biomed. Opt., 21
(5), 056002
(2016). http://dx.doi.org/10.1117/1.JBO.21.5.056002 JBOPFO 1083-3668 Google Scholar
Y. Liu et al.,
“Complementary fluorescence-polarization microscopy using division-of-focal-plane polarization imaging sensor,”
J. Biomed. Opt., 17
(11), 116001
(2012). http://dx.doi.org/10.1117/1.JBO.17.11.116001 JBOPFO 1083-3668 Google Scholar
H. He et al.,
“Two-dimensional and surface backscattering Mueller matrices of anisotropic sphere-cylinder scattering media: a quantitative study of influence from fibrous scatterers,”
J. Biomed. Opt., 18
(4), 046002
(2013). http://dx.doi.org/10.1117/1.JBO.18.4.046002 JBOPFO 1083-3668 Google Scholar
S. Ho et al.,
“Review of recent progress in multilayer solution-processed organic light-emitting diodes,”
J. Photonics Energy, 5
(1), 057611
(2015). http://dx.doi.org/10.1117/1.JPE.5.057611 Google Scholar
G. Li et al.,
“GaN-based light-emitting diodes on various substrates: a critical review,”
Rep. Prog. Phys., 79
(5), 056501
(2016). http://dx.doi.org/10.1088/0034-4885/79/5/056501 RPPHAG 0034-4885 Google Scholar
E. Bernabeu and J. J. Gil,
“An experimental device for the dynamic determination of Mueller matrices,”
J. Opt., 16
(3), 139
–141
(1985). http://dx.doi.org/10.1088/0150-536X/16/3/006 JOOPDB 0150-536X Google Scholar
K. Ichimoto et al.,
“Photopolarimetric measurement system of Mueller matrix with dual rotating waveplates,”
Publ. Natl. Astron. Obs. Jpn., 9 11
–19
(2016). Google Scholar
O. Arteaga et al.,
“Mueller matrix polarimetry with four photoelastic modulators: theory and calibration,”
Appl. Opt., 51
(28), 6805
–6817
(2012). http://dx.doi.org/10.1364/AO.51.006805 APOPAI 0003-6935 Google Scholar
A. De Martino et al.,
“Optimized Mueller polarimeter with liquid crystals,”
Opt. Lett., 28
(8), 616
–618
(2003). http://dx.doi.org/10.1364/OL.28.000616 OPLEDP 0146-9592 Google Scholar
D. H. Goldstein,
“Mueller matrix dual-rotating retarder polarimeter,”
Appl. Opt., 31
(31), 6676
–6683
(1992). http://dx.doi.org/10.1364/AO.31.006676 APOPAI 0003-6935 Google Scholar
J. Wolfe and R. A. Chipman,
“Reducing symmetric polarization aberrations in a lens by annealing,”
Opt. Express, 12
(15), 3443
–3451
(2004). http://dx.doi.org/10.1364/OPEX.12.003443 OPEXFF 1094-4087 Google Scholar
T. C. Wood and D. S. Elson,
“Polarization response measurement and simulation of rigid endoscopes,”
Biomed. Opt. Express, 1
(2), 463
–470
(2010). http://dx.doi.org/10.1364/BOE.1.000463 BOEICL 2156-7085 Google Scholar
E. Compain, S. Poirier and B. Drevillon,
“General and self-consistent method for the calibration of polarization modulators, polarimeters, and Mueller-matrix ellipsometers,”
Appl. Opt., 38
(16), 3490
–3502
(1999). http://dx.doi.org/10.1364/AO.38.003490 APOPAI 0003-6935 Google Scholar
A. D. Martino et al.,
“General methods for optimized design and calibration of Mueller polarimeters,”
Thin Solid Films, 455 112
–119
(2004). http://dx.doi.org/10.1016/j.tsf.2003.12.052 THSFAP 0040-6090 Google Scholar
Y. Fan et al.,
“Optical waveguide sensor based on silica nanotube arrays for label-free biosensing,”
Biosens. Bioelectron., 67 230
–236
(2015). http://dx.doi.org/10.1016/j.bios.2014.08.021 BBIOE4 0956-5663 Google Scholar
S. Y. Lu and R. A. Chipman,
“Interpretation of Mueller matrices based on polar decomposition,”
J. Opt. Soc. Am. A, 13
(5), 1106
–1113
(1996). http://dx.doi.org/10.1364/JOSAA.13.001106 JOAOD6 0740-3232 Google Scholar
|