Abstract

Purpose

High soft-tissue contrast imaging is essential for effective radiotherapy treatment. This could potentially be realized using both megavoltage and kilovoltage x-ray sources available on some therapy treatment systems to perform “MV-kV” dual-energy (DE) computed tomography (CT). However, noisy megavoltage images obtained with existing energy-integrating detectors (EIDs) are a limiting factor for clinical translation. We explore the potential for non-spectral photon-counting detectors (PCDs) to improve MV-kV image quality simply by equally weighting all MV photons rather than up-weighting the less informative, lower contrast high-energy photons as in an EID.

Approach

Three computational methods were applied to compare non-spectral PCDs with EIDs in MV-kV DE imaging. A single-line integral estimation theory approach was used to calculate the basis material signal-to-noise ratio (SNR) of tissue (1 to 50 cm) and bone (0.1 to 10 cm). CT images of a tissue cylinder with seven bone inserts (densities 1.0 to 2.2 g/cm3) were simulated to assess material decomposition accuracy. Multiple noisy simulations of an anthropomorphic phantom were performed to generate pixel-by-pixel noise profiles.

Results

PCDs yielded a 15% to 45% improvement in single-line integral SNR for both materials. In CT simulations, similar material decomposition accuracy was achieved, with both EIDs and PCDs slightly underestimating bone density. However, PCDs yield a higher contrast-to-noise ratio and more uniform noise texture than EIDs in virtual monoenergetic images.

Conclusions

We demonstrate the potential for improved MV-kV DE CT imaging using non-spectral PCDs and quantify the degree of improvement in a range of object compositions. This work motivates the experimental assessment of PCDs for megavoltage imaging and the potential clinical translation of PCDs to radiotherapy imaging.

1.

Introduction

“MV-kV” dual-energy (DE) computed tomography (CT) is a proposed imaging modality with potential application in radiotherapy settings.13 Modern radiation therapy systems are already equipped with the hardware necessary for single-shot DE-CT: two x-ray source-detector arrays that rotate in tandem. The first source operates in the megavoltage (MV) energy range. Its primary purpose is radiation therapy treatment, but it can also be detuned to operate with a softer energy spectrum (peak photon counts at 1  MeV) appropriate for imaging. The second source operates in the kilovoltage (kV) energy range (80 to 140 kVp) for on-board imaging during radiation therapy. By imaging with both sources simultaneously, one can conveniently realize the benefits of single-shot DE imaging available in diagnostic settings without the need to purchase or install specialized equipment.46

MV-kV DE-CT shows promise for improved soft-tissue contrast during routine radiotherapy imaging.1 Because images acquired with distinct x-ray energy spectra have distinct contrast levels, DE-CT enables the decomposition of raw image data into two or three basis material images, in which each pixel value corresponds to the density of a chosen material such as bone or tissue.7 These material images may in turn be used to generate virtual monoenergetic images (VMIs), synthetic estimates of the images that could be acquired with an ideal single-energy spectrum. VMIs can be generated at low energies with relatively high native soft-tissue contrast and additionally alleviate common single-kV imaging issues such as beam-hardening artifacts. These aspects are highly advantageous when aiming to visualize a tumor embedded in a similarly absorptive soft-tissue background.

In previous work, we demonstrated the ability of MV-kV DE-CT to generate images of the pelvis with higher native contrast than can be achieved with dose-matched single-kV images.1 The technique might also be used for metal artifact correction due to the high penetrability of megavoltage-energy photons. However, the limitations of MV imaging can curb the benefits of MV-kV DE-CT in common imaging scenarios. Though fewer MV x-rays are attenuated by the body relative to kV x-rays, they deposit a much greater dose per stopped photon. To deliver the same dose as a kV spectrum, an MV spectrum’s flux must be considerably reduced, resulting in much noisier images. Furthermore, MV detective efficiency is generally much lower than kV efficiency when using currently available x-ray detectors.810 When implementing MV-kV DE-CT imaging, the greater noise of the MV data can contaminate the kV data and result in VMIs with an overall lower contrast-to-noise ratio (CNR). Clinical translation of MV-kV DE-CT will likely require a technique that reduces the noise of MV images.

Though image noise might always be decreased by increasing incident flux, this results in a higher imaging dose, which is undesirable in radiotherapy imaging settings. Unlike in diagnostic imaging, a patient must be imaged repeatedly in a relatively short period of time to track their body morphology and tumor progression for precise and accurate treatment planning. The total dose of this required imaging can sum to the scale of a treatment fraction, motivating the search for accessible, high-contrast, low-dose imaging for radiotherapy.11 Ideally, the dose in the context of radiotherapy imaging would be equivalent to or lower than that in diagnostic settings to account for the patient’s repeat imaging. A recent literature review reported a diagnostic reference dose level range of 10 to 15 mGy for chest-abdominopelvic CT tumor imaging.12 Historically, MV CT for radiotherapy has been limited by the much higher dose it requires for sufficient image quality (IQ).13

An appealing avenue for improved MV IQ might be found on the detector side.3 The conventionally implemented MV imaging options are energy-integrating detectors (EIDs). Recent advances in MV imaging with EIDs have predominantly focused on improving detective efficiency. Most MV imaging uses flat-panel electronic portal imaging devices (EPIDs) with efficiencies as low as 1% to 2%.10 MV CT with these detectors requires a dose on the order of 300 mGy for sufficient IQ.14 Novel EPIDs are being developed with improved efficiency in the range of 5% to 20% through a variety of strategies, such as layering of multiple scintillators or development of new scintillating materials.10,11,15,16 With these emerging detectors, doses of 40 mGy can yield acceptable soft-tissue contrast.17 Modern tomotherapy systems are also capable of relatively high-efficiency MV CT imaging, with a fan-beam xenon gas detector that stops 20% of MV photons.8,9,18,19 These systems can achieve acceptable soft-tissue contrast with doses as low as 10 mGy.20

However, EIDs have intrinsic downsides aside from detective efficiency. EIDs can be designed with various electronic schemes, either direct or indirect conversion. A common EID used in medical imaging is the scintillating detector, which measures x-ray intensity through an indirect conversion to visible light.21 The uniting characteristic of EIDs is that they weigh detected photons by their incident energy and sum them into the final signal. This results in low-energy photons, which have greater soft-tissue contrast than high-energy photons, contributing less information to the final image. Although this reduces both kV and MV image contrast, it is especially unfortunate for MV imaging, in which the highest energy photons are weighted a full two orders of magnitude more than the most useful low-energy photons. A second downside of EIDs is that they suffer from electronic noise, which is independent of the unavoidable quantum noise associated with photon counting statistics. Because the flux of an MV treatment beam must be greatly decreased to deliver an acceptably low imaging dose to the patient, detected photon counts are generally much lower, and electronic noise can become a more significant component of the final image. These two effects result in both lower contrast and higher noise, doubly detrimental to CNR.

As an emerging alternative, photon-counting detectors (PCDs) avoid the energy weighting issue of EIDs. Medical imaging PCDs commonly utilize semiconductors that directly convert x-rays into electron-hole pairs.22 Incident photons are registered as electric pulses with amplitude proportional to their energy. One can define one or more pulse-height thresholds and register each signal surpassing the threshold as one count, bypassing the issue of energy weighting and also thresholding out the electronic noise background. PCDs face downsides that can degrade spatial and energy resolution, including K-escape, charge sharing, and pulse pile-up.23 The severity of these effects depends on the detector materials and geometry.24 Correction techniques are being developed to account for these factors and even apply them advantageously; for example, charge-sharing measurements have been used to achieve sub-micron resolution in a PCD for CT imaging.2527 Further, PCDs have recently debuted in diagnostic CT scanners and show great potential for improving IQ relative to conventional EIDs.28

A key promise of PCDs is that, if desired, one can extract spectral information by defining multiple pulse-height thresholds. These “spectral PCDs” enable multi-energy imaging without the historical need for repeat acquisitions, dual source-detector arrays, multi-layer detectors, fast-kVp switching sources, or other specialized equipment.23 With a spectral PCD, multi-energy imaging could be implemented on a radiation therapy treatment system using a single MV spectrum. However, the imposition of spectral thresholds significantly reduces the photon counts per energy bin, exacerbating the key challenge of MV CT—the high noise due to low flux at acceptable dose levels. We address the case of single-MV imaging with a spectral PCD in a Supplementary Material.

The benefits of PCDs might still be realized in the context of radiation therapy imaging by MV-kV DE-CT with “non-spectral PCDs.” These detectors essentially operate as a spectral PCD with one very low-energy threshold to remove electronic noise. Even without any energy resolution, non-spectral PCDs have great potential for improving MV imaging by alleviating the high-energy photon up-weighting issue and reducing low-dose image noise—the two key issues of EIDs. To gain multi-energy information with non-spectral PCDs, one requires a traditional dual-source setup, as in MV-kV DE-CT on a radiation therapy treatment system.

From a practical standpoint, the potential implementation of MV-kV DE-CT with non-spectral PCDs depends on the availability of a PCD with sufficient MV counting efficiency. The PCDs implemented in clinical and preclinical CT systems are typically made of cadmium telluride (CdTe), cadmium zinc telluride (CZT), or silicon (Si) semiconductors.24 CdTe and CZT have a higher atomic number and therefore can be made relatively thin (mm-scale) for diagnostic imaging. These detectors are likely too thin for MV imaging. Using an estimate based on the linear attenuation coefficient, a 3-mm CdTe slab will stop only 10% of 3-MeV photons. This is better than most commercial EPIDs but only half the reported 20% efficiency needed for sufficiently low-dose MV imaging.11 Si PCDs are a more affordable option due to the broader availability of silicon wafers, but for sufficient detective efficiency, they require centimeter-scale thickness due to their lower atomic number.29 For this reason, Si detectors have been developed using an “edge-on” approach. Perhaps surprisingly, this could be more advantageous for MV imaging: a 3-cm Si slab is expected to attenuate 20% of 3-MeV photons, which is comparable to the high-efficiency MV EID we modeled. Unfortunately, MV photons will produce larger charge clouds and thus are likely to suffer from more charge sharing. This would be especially prominent in Si detectors, where there is a greater proportion of Compton events contributing to overall photon attenuation.29 Although these PCDs may offer sufficient MV efficiency, they may suffer losses in spatial resolution. In an emerging application, some work has explored novel edge-on CZT detectors for positron emission tomography imaging.3032 These detectors are especially promising for MV imaging, as a 3-cm CdTe slab can stop 50% of 3-MeV photons. The drawback of edge-on CZT detectors relative to Si is likely manufacturing cost. These available and emerging diagnostic CT PCDs show promise for MV imaging and indicate the feasibility of MV-kV DE-CT.

The purpose of this work was to explore the potential utility of non-spectral PCDs in the context of MV-kV DE-CT imaging. The advancing clinical implementation of PCDs makes this topic of particular interest for MV imaging, which faces unique challenges relative to diagnostic kV imaging. Yet, there has been limited investigation of PCDs beyond the diagnostic CT imaging energy range, with tube potentials generally less than 150 kV. We aim to compare IQ achieved with non-spectral PCDs to that achieved with EIDs used for MV imaging on advanced radiotherapy treatment systems and to quantify the degree of potential improvement.

2.

Methods

We implemented three approaches for comparing MV-kV IQ achieved with non-spectral photon-counting and EID models. Spectral PCDs are discussed separately in the Supplementary Material.

A single-line integral toy model was used to assess a theoretical limit on achievable IQ and to optimize dose allocation between the MV and kV spectra (Sec. 2.2). Simulated MV-kV DE-CT images of an IQ phantom were used to assess material decomposition accuracy with the two detector models (Sec. 2.3). MV-kV DE-CT images of an anthropomorphic phantom were also simulated for qualitative noise analysis (Sec. 2.3). The two phantoms used for the CT simulations are shown in Fig. 1. We first describe the materials and parameters common to these approaches (Sec. 2.1).

Fig. 1

Two phantoms used for the CT simulations. (a) An IQ phantom of tissue with seven bone inserts of varying densities. (b) A section of the XCAT phantom. Pixel values are in HU evaluated at 80 keV and shown at a window level of 0 and width of 1000. Each image field-of-view is 50 cm.

JMI_11_S1_S12811_f001.png

2.1.

Materials

The modeled MV-kV DE system utilized a detuned MV x-ray spectrum and 80-kVp spectrum typical for diagnostic imaging. This MV source is based on a typical 6 MV radiotherapy treatment beam with its spectrum softened such that the average photon energy is below 3 MeV and the maximum flux is at 1  MeV.33,34 The detuned beam is achievable on existing treatment systems and more favorable for MV imaging by mitigating some drawbacks of high-energy x-rays, but it would not be used for radiotherapy treatment.

Both EID and PCD models used the same energy-dependent detective efficiency function η(E). This allowed us to isolate the effect of the different photon weighting schemes of the two detector models, which is of particular interest in the high-energy case of MV imaging. The detective efficiency function was modeled based on existing xenon gas EIDs with relatively high detective quantum efficiency in the MV energy range. The function is near unity in the keV energy range and decreases to 20% at energies up to 6 MeV.1 Such detectors are implemented in existing tomotherapy systems (Accuray Inc., Sunnyvale, CA, United States) for fan-beam MV imaging.18,19 For our theoretical comparison, we applied this same efficiency function to the PCD model, though such detectors do not exist in reality. Other high-efficiency MV detector options are also emerging, for example, utilizing stacks of EPIDs as EIDs or edge-on PCDs.11,2932

The raytracing CT simulation utilized a 55-cm source-to-isocenter distance and a 100-cm source-to-detector distance. The fan-beam detector had 800 channels and a total fan angle of 50 deg. For all simulations, the two input spectra magnitudes were rescaled such that they would deliver a total dose of 10 mGy to the center of a 40-cm water cylinder.1 A total of 1200 projection views were acquired over a 360-deg rotation. For each DE acquisition, sinogram-domain material decomposition into tissue and bone basis materials was performed using a Gauss-Newton algorithm.35 Images were reconstructed using fan-beam filtered back-projection with a matrix size of 512, field of view of 50 cm, and ramp filter with cutoff frequency at 80% of the Nyquist limit. VMIs were generated at a continuum of energies (E0=20120  keV) as a linear combination of the basis material images ρj:

Eq. (1)

VMI(E0)=ρ1(μ(E0)ρ)1+ρ2(μ(E0)ρ)2,
where the weights (μ(E0)/ρ)j are the known mass attenuation coefficients of basis material j at energy E0. The MV-kV DE-CT simulation workflow is summarized as a flowchart in Fig. 2, which includes example images demonstrating how VMIs can alleviate single-energy image artifacts.

Fig. 2

Flowchart summarizing the MV-kV simulation procedure. Single-energy MV and kV sinograms are simulated independently and then recombined for DE-CT (top row): sinogram-domain material decomposition is performed, then material images are reconstructed, and VMIs are generated using Eq. (1). Alternatively, the single-energy sinograms can be directly reconstructed (bottom row): the single-kV image (left) suffers from severe beam hardening, whereas the single-MV image (right) has poor contrast and high noise.

JMI_11_S1_S12811_f002.png

2.2.

Single-line Integral

We implement an estimation theory approach for computing the Cramèr-Rao lower bound (CRLB) on variance in the context of basis material decomposition along a single ray through two materials, summarized here.36 We consider two polychromatic x-ray spectra Ii(E) (i=1,2) incident on an object comprising two basis materials, each with mass thickness Aj=ρj×tj (j=1,2). The signal detected using a single spectrum λi can generally be described as

Eq. (2)

λi=EminEmaxIi(E)exp[jμj(E)tj]η(E)D(E)dE,
where μj is the linear attenuation coefficient of basis material j, η(E) is the detective efficiency, and D(E) is the detector response function. The signal model and a defined noise model are used to construct the log-likelihood function and, consequently, the Fisher information matrix F. Taking each basis material mass thickness as the parameter of interest, the CRLB σAj2 is then computed by inverting the Fisher matrix,

Eq. (3)

σAj2=Fjj1.
Just as the CRLB represents the lower bound on variance for an unbiased estimator, a theoretical upper bound on achievable IQ can be defined as the unitless value Aj/σAj. Interpreting the true mass thickness as the ideal signal, we refer to this parameter as the single-line integral model signal-to-noise ratio (SNR) for each basis material.

We previously applied this method for comparing MV-kV DE imaging with traditional diagnostic (kV-kV) DE imaging using EIDs only and neglecting electronic noise.1,2 However, the detector response function is distinct for integrating measurements [D(E)=E] and counting measurements [D(E)=1], resulting in different functional forms of the CRLB for either detector model. Furthermore, one major advantage of PCDs is their potential to eliminate electronic noise by applying some minimum pulse-height threshold greater than the noisy background. We assess the utility of this strategy by extending the EID CRLB model to include electronic noise by adding a factor of σe×E¯i to the EID noise model, where σe is the standard deviation in counts due to the electronic noise and E¯i is the intensity-weighted average energy of spectrum Ii. We chose σe=10 photons. By contrast, the PCD model used σe=0, simulating a low pulse-height threshold that eliminates background noise but does not yield other multi-energy information.

We computed SNRj for a range of tissue thicknesses (t1=1050  cm, ρ1=1.06  g/cm3) and bone thicknesses (t2=0.110  cm, ρ2=1.85  g/cm3). We also considered the effect of dose allocation between the MV and kV spectra. The magnitude of each spectrum was initially scaled such that the total dose delivered by a single polychromatic ray to the center of a 40-cm water cylinder would be 1  μGy.1 For a CT scan with 1000 views, this corresponds to a 1-mGy dose. To ensure the total dose remained fixed for each acquisition, the magnitude of the MV spectrum was rescaled by a factor r (ranging from 0.01 to 0.99), and the kV spectrum was rescaled by the remaining 1r. We compared the dose-optimized SNRs achieved with either an EID or non-spectral PCD model, the dose allocation factor r needed to achieve this optimal SNR, and the effect of basis material thickness.

The resulting SNRs are scalable to other dose levels. For the PCD case, which has a pure Poisson counting noise model, noise is proportional to the inverse square root of flux, so SNR increases with the square root of dose per ray [μGy] at any dose allocation factor. Although this square root rule-of-thumb is not exact for the EID case due to photon energy weighting and electronic noise, these nonidealities are mainly relevant at very low doses, so the rule is still widely applied.3741

2.3.

CT Simulations

As an extension of the single-line integral model, we simulated MV-kV DE-CT acquisitions of two phantoms. To assess basis material decomposition with the two detector schemes, we created a computational IQ phantom comprising a 40-cm tissue cylinder with seven 3-cm bone inserts with densities varying from 1.0 to 2.2  g/cm3. Twenty MV-kV DE-CT acquisitions were simulated using both EIDs and non-spectral PCDs as described above, providing several noisy realizations for statistical analysis. In the bone basis material image, the mean pixel value in each insert was measured and compared with the known ground truth. CNR measurements were taken in the region of each bone insert, defined as

Eq. (4)

CNRρ(E0)=|mρ(E0)mbg(E0)|σbg(E0)
where the subscript ρ indicates the masked bone insert at the density of interest, the subscript bg indicates the masked tissue background, m represents the mean, σ represents the standard deviation, and E0 indicates the energy of the VMI used for the measurements.

For qualitative analysis of image noise profiles using either detector scheme, we simulated MV-kV DE-CT acquisitions of an extended cardiac torso (XCAT) phantom.42 We generated twenty noisy CT images using both detector models and synthesized the basis material images into VMIs using Eq. (1). Noise profile images were generated by taking the pixel-by-pixel standard deviation over the multiple noisy realizations.

For both phantoms, the total DE-CT dose was fixed at 10 mGy. The dose allocation between the MV and kV spectra was approximately optimized using the results of the single-line integral. An initial 180-deg noiseless simulation with just one central detector channel was conducted to yield basis material thickness line profiles through the phantom, and a single-dose allocation factor r was chosen depending on the varying material thickness pairs over the several projection view angles. The thickness line profiles for the two phantoms are shown in Figs. 3 and 4. Though this method assumed a priori knowledge of the phantom composition, it was highly approximate. Similar if not better methods can likely be implemented in clinical scenarios using anterior-posterior and lateral scout scans, as is commonly used for tube current modulation. Based on these thickness profiles, we utilized 85% and 90% doses allocated to the MV spectrum for the IQ and XCAT simulations, respectively.

Fig. 3

Basis material thickness line profiles through the IQ phantom used for approximate dose allocation optimization. Tissue thickness ranged between 37 and 40 cm, and bone thickness ranged from 1.5- to 3.5-cm equivalent for the seven different density inserts.

JMI_11_S1_S12811_f003.png

Fig. 4

Basis material thickness line profiles through the XCAT phantom used for approximate dose allocation optimization. Tissue thickness ranged between 18 and 32 cm, and bone thickness varied up to 5 cm.

JMI_11_S1_S12811_f004.png

3.

Results

3.1.

Single-line Integral

Figure 5 shows the dose allocation factor ropt that optimizes basis material SNR using 40-cm tissue and a variety of bone thicknesses. Relative to the EID SNR, the PCD SNR is optimized with a slightly lower dose allocated to the MV spectrum, though the difference decreases as bone thickness increases. The optimal dose allocation is distinct for the bone and tissue, with bone generally having a slightly lower ropt than the tissue. The largest changes in ropt occur with respect to bone thickness: small bone thicknesses (<1  cm) are best with nearly 90% dose allocated to the MV spectrum, whereas higher bone thicknesses are optimized with a more equal 50% dose to the MV spectrum. Based on this, we infer the optimal dose allocation for a given image acquisition depends primarily on object composition, with additional but less pronounced dependence on the material of interest (tissue or bone) and detector (EID or PCD).

Fig. 5

Dose allocation to the MV spectrum maximizing basis material SNR for (a) the tissue and (b) bone as a function of bone thickness with fixed tissue thickness of 40 cm.

JMI_11_S1_S12811_f005.png

Figures 6 and 7 show heatmaps of dose-optimized basis material SNR at all bone and tissue thickness pairs assessed. Across all material compositions, PCDs produce between 15 and 45% improvement in SNR. The greatest percent improvements are seen for very thin objects, which also have the lowest overall SNRs due to their lower attenuation. The optimal dose factors that generated the maximum SNRs shown in these heatmaps were used to inform the approximate dose allocation optimization of the CT scans. For example, Fig. 5 shows ropt as a function of bone thickness for 40-cm tissue. For each phantom, characteristic average tissue thickness and non-zero bone thickness were approximated from the angle-dependent thickness profiles (Figs. 3 and 4). Average non-zero bone thickness was generally low (2  cm), where the ropt(tbone) curve is slowly varying (Fig. 5) and only 2% to 3% different for EIDs and PCDs. From this, we selected the constant values of ropt=0.85 and 0.90 for the IQ phantom and XCAT phantom, respectively.

Fig. 6

Dose-optimized tissue SNR using (a) PCD and (b) EID models for a variety of basis material thicknesses. Panel (c) shows the percent difference.

JMI_11_S1_S12811_f006.png

Fig. 7

Dose-optimized bone SNR using (a) PCD and (b) EID models for a variety of basis material thicknesses. Panel (c) shows the percent difference.

JMI_11_S1_S12811_f007.png

3.2.

CT Simulations

To assess basis material decomposition accuracy for the two detector schemes, Fig. 8 shows bone basis material measurements in the seven IQ phantom inserts. The measurements are similar using either PCD or EID at all densities, and both detector schemes result in a slight underestimation of true bone density. However, the basis material images of PCDs and EIDs produce distinct noise levels in VMIs, as illustrated in Fig. 9, which shows CNR in three bone density inserts at a continuum of monoenergies. PCDs consistently yield higher VMI CNR than EIDs at all monoenergies, with peak CNR generally occurring at a slightly lower monoenergy. The improvement in CNR with PCDs is especially pronounced at lower (<50  keV) VMI energies.

Fig. 8

Measured versus true bone density in the seven inserts of the IQ phantom. Error bars represent k=1 standard deviation over all noisy simulations.

JMI_11_S1_S12811_f008.png

Fig. 9

CNR measured in three bone inserts with density (a) 1.4  g/cm3, (b) 1.8  g/cm3, or (c) 2.2  g/cm3 as a function of VMI energy. The error bars represent k=2 standard deviations over all noisy simulations.

JMI_11_S1_S12811_f009.png

Figures 10Fig. 1112 show VMIs and their corresponding noise profiles using either the PCD or EID model. Broadly, EIDs result in noisier images. In the lower energy VMIs (50 keV), the noise profile is concentrated around bony regions and in streaks between bones. Although this is apparent for both detector models, it is more pronounced with EIDs. This results in a less isotropic noise texture in the EID images. As VMI energy increases to 60 and 70 keV (Figs. 11 and 12), the noise texture shifts to be more isotropic across the tissue regions of the phantom. Although the EID images still have greater noise magnitude than the PCD images, the two noise textures appear to be more similar.

Fig. 10

(a)–(b) 50-keV VMIs and (c)–(d) their noise profiles simulated with either PCD (a), (c) or EID (b), (d) schemes. Pixel units are HU. The noise profile is concentrated around the bone regions.

JMI_11_S1_S12811_f010.png

Fig. 11

(a)–(b) 60-keV VMIs and (c)–(d) their noise profiles simulated with either PCD (a), (c) or EID (b), (d) schemes. Pixel units are HU. The noise profile is distributed around the bones and lateral edges of the phantom.

JMI_11_S1_S12811_f011.png

Fig. 12

(a)–(b) 70-keV VMIs and (c)–(d) their noise profiles simulated with either PCD (a), (c) or EID (b), (d) schemes. Pixel units are HU. The noise profile has a more uniform texture across tissue regions.

JMI_11_S1_S12811_f012.png

4.

Discussion

MV-kV DE-CT imaging is a proposed modality that could be implemented in existing radiotherapy treatment systems, possibly bringing the advantages of multi-energy imaging to routine radiotherapy imaging over the course of a patient’s cancer treatment.

In this work, we explored the implications of exchanging the EIDs in a previously studied MV-kV DE-CT system with non-spectral PCDs of identical counting efficiency.1 This allowed us to assess the effect of the energy-weighting detection scheme and electronic noise. We expected the energy-weighting effect of EIDs is particularly detrimental for MV imaging due to the multiple order-of-magnitude increase in weighting for photons carrying the least contrast information. A 3-MeV photon would be weighted 100× more than a 30  keV photon. Furthermore, because the MV spectrum flux must be greatly reduced relative to the kV spectrum to reach acceptable dose levels, we expected the effect of a constant electronic noise background may be more pronounced. Our findings corroborate these expectations.

In the single-line integral model, for each basis material, dose-optimized SNR initially increases with thickness as the true mass thickness ρ×t (numerator) increases. Each SNR eventually peaks due to increases in the CRLB (denominator), likely due to a reduction in detected photon counts as they are attenuated by the object. This general relationship is true for the two detector models. Comparing PCD SNRs to EID SNRs, we observe a saddle surface shape in the percent improvement heatmap for all material thickness pairs (panel (c) of Figs. 6 and 7). The greatest improvement (approx. 45%) in SNR was seen at low thicknesses, where the signal is the weakest but the electronic noise background is less significant relative to the large number of detected photons. PCDs yield a smaller but still observable (approx. 15%) improvement at more moderate thickness pairs. At very high thicknesses, the percent improvement increases to near 30%, likely due to the high attenuation of the incident beams resulting in a more dominant electronic noise component in the EID signal.

The dose optimization factor ropt was slightly different depending on the detector model, and 0 to 5% less of the total dose should be allocated to the MV spectrum when using a PCD. Tables of ropt values for all material thickness pairs were used to choose the dose allocations of 0.85 and 0.90 for the IQ phantom and XCAT phantom CT simulations, respectively. More specifically, ropt varies depending on the specific thickness pair of each view angle as in Figs. 3 and 4. In a separate experiment, we simulated CT scans with ropt modulated as a function of view angle according to the thickness profiles for each phantom and detector type. We found that CNR did not significantly improve relative to the CNRs for the constant values of 0.85 and 0.90, indicating that these constant ropt choices are sufficient for this study. The scope of this paper focuses on fairly comparing EID and PCD performance, not necessarily optimizing performance as a function of dose allocation. This subject might be further investigated in future work.

In the MV-kV DE-CT simulations, the Gauss-Newton material decomposition method implemented had similar accuracy for either EID or PCD detector model.35 By combining the two basis material images into a VMI, greater CNR can be achieved at a range of monoenergies. As expected, CNR increases with bone density due to the greater attenuation of the signal relative to the background, and the monoenergy maximizing CNR increases with bone density due to beam hardening. For each individual bone density, PCD CNR was greater than EID CNR at all VMI energies examined (20 to 120 keV). The peak PCD CNR occurred at a slightly lower monoenergy than the peak EID CNR. This might be attributed to the improvement in PCD basis material image noise. In principle, the two material images’ noise is correlated such that when they are recombined using Eq. (1), the resulting VMI will have less noise. This depends on the energy. Because the mass attenuation coefficients in the VMI calculation decrease as energy increases, low-energy VMIs can amplify the noise in the material images. This effect is not as severe for the less noisy PCD images, so the peak CNR is at a lower monoenergy with higher native contrast. The greatest CNR improvements were observed at low monoenergies (<50  keV), whereas only a slight improvement was seen after peak PCD CNR. For a more clinically relevant assessment, we also simulated MV-kV VMIs of an anthropomorphic phantom. Qualitatively, for both detectors, lower monoenergies produced higher soft-tissue contrast but greater noise, especially around regions with bone and along the lateral patient axis (Fig. 10). The noise texture and magnitude are more prominent in the EID images. For both detector models, soft-tissue contrast decreased and noise texture became more uniform at higher monoenergies, but the noise magnitude was still higher using the EID (Figs. 11 and 12). The greatest improvements in IQ with the PCD appear to be due to improvements in noise magnitude and texture, whereas contrast is subjectively more similar.

The results of this computational assessment of non-spectral PCDs for MV-kV DE-CT motivate future experimental work. The CT simulation study provides an initial estimate of the noise reduction benefits achievable with PCDs. An experimental study would more realistically include other factors such as photon scatter, spatial resolution, and available PCDs. These results could be compared with simulated findings for a better understanding of the clinical translatability of our results. Simulations and estimation theory calculations allow us to conveniently explore a range of parameters relevant to radiotherapy imaging scenarios, complementing these experimental investigations. These computational methods could also be used to optimize image acquisition parameters to ensure fair comparison of different detector models.

5.

Conclusion

PCDs show promise for improving MV-kV DE CT imaging by improving MV image noise characteristics. In terms of single-line integral basis material SNR, we found that non-spectral PCDs yielded a 15% to 45% improvement over EIDs. In CT simulations, material decomposition accuracy was similar for both detectors, whereas non-spectral PCDs resulted in higher CNR and more uniform noise texture in VMIs. Relative to existing high-efficiency MV EIDs, similar or better MV detective efficiency might be achieved by emerging diagnostic PCDs if utilizing an edge-on setup. This computational study motivates the experimental assessment of PCDs for MV imaging and the potential clinical translation of PCDs for MV-kV DE-CT imaging on radiotherapy systems.

Disclosures

Patrick La Rivière has in the past received research funding from Accuray Inc., Canon Medical Research USA, Inc., and Toshiba Medical Research USA, Inc. He is on the Scientific Advisory Board and holds stock options in Metritrack, Inc., an ultrasound-guidance company.

Code and Data Availability

The code used to generate the data in this study can be freely accessed through GitHub at https://github.com/gjadick/dex-single-ray for the single-line integral toy model and https://github.com/gjadick/dex-ct-sim for the fan-beam CT simulation and basis material decomposition.

Acknowledgments

This work was supported by the National Science Foundation Graduate Research Fellowship (Grant No. 2140001) and the National Cancer Institute of the NIH (Grant No. R25CA221767).

References

1. 

G. Jadick, G. Schlafly and P. La Rivière, “Dual-energy computed tomography imaging with megavoltage and kilovoltage X-ray spectra,” J. Med. Imaging, 11 (2), 023501 https://doi.org/10.1117/1.JMI.11.2.023501 JMEIET 0920-5497 (2024). Google Scholar

2. 

G. Jadick and P. La Rivière, “Optimization of MV-kV dual-energy CT imaging for tomographic therapy,” Proc. SPIE, 12463 124632L https://doi.org/10.1117/12.2653674 PSISDG 0277-786X (2023). Google Scholar

3. 

M. Ventura, G. Jadick and P. La Rivière, “Comparison of energy-integrating detectors and photon-counting detectors for MV-kV dual-energy imaging on a tomographic therapy system,” Proc. SPIE, 12925 129252W https://doi.org/10.1117/12.3006854 PSISDG 0277-786X (2024). Google Scholar

4. 

M. F. Kruis, “Improving radiation physics, tumor visualisation, and treatment quantification in radiotherapy with spectral or dual-energy CT,” J. Appl. Clin. Med. Phys., 23 (1), e13468 https://doi.org/10.1002/acm2.13468 (2022). Google Scholar

5. 

W. van Elmpt et al., “Dual energy CT in radiotherapy: current applications and future outlook,” Radiother. Oncol., 119 137 –144 https://doi.org/10.1016/j.radonc.2016.02.026 RAONDT 0167-8140 (2016). Google Scholar

6. 

S. Sajja et al., “Technical principles of dual-energy cone beam computed tomography and clinical applications for radiation therapy,” Adv. Radiat. Oncol., 5 1 –16 https://doi.org/10.1016/j.adro.2019.07.013 (2020). Google Scholar

7. 

A. Macovski et al., “Energy dependent reconstruction in X-ray computerized tomography,” Comput. Biol. Med., 6 325 –336 https://doi.org/10.1016/0010-4825(76)90069-X CBMDAW 0010-4825 (1976). Google Scholar

8. 

H. Keller et al., “Monte Carlo study of a highly efficient gas ionization detector for megavoltage imaging and image-guided radiotherapy,” Med. Phys., 29 (2), 165 –175 https://doi.org/10.1118/1.1445414 MPHYA6 0094-2405 (2002). Google Scholar

9. 

K. J. Ruchala et al., “Megavoltage CT on a tomotherapy system,” Phys. Med. Biol., 44 2597 https://doi.org/10.1088/0031-9155/44/10/316 PHMBA7 0031-9155 (1999). Google Scholar

10. 

J. Rottmann et al., “A novel EPID design for enhanced contrast and detective quantum efficiency,” Phys. Med. Biol., 61 (17), 6297 https://doi.org/10.1088/0031-9155/61/17/6297 PHMBA7 0031-9155 (2016). Google Scholar

11. 

J. Star-Lack et al., “A piecewise-focused high DQE detector for MV imaging,” Med. Phys., 42 (9), 5084 –5099 https://doi.org/10.1118/1.4927786 MPHYA6 0094-2405 (2015). Google Scholar

12. 

G. Paulo et al., “Diagnostic reference levels based on clinical indications in computed tomography: a literature review,” Insights Imaging, 11 96 https://doi.org/10.1186/s13244-020-00899-y (2020). Google Scholar

13. 

J. Sillanpaa et al., “Low-dose megavoltage cone-beam computed tomography for lung tumors using a high-efficiency image receptor,” Med. Phys., 33 (9), 3489 –3497 https://doi.org/10.1118/1.2222075 MPHYA6 0094-2405 (2006). Google Scholar

14. 

B. Groh et al., “A performance comparison of flat-panel imager-based MV and kV cone-beam CT,” Med. Phys., 29 (6), 967 –975 https://doi.org/10.1118/1.1477234 MPHYA6 0094-2405 (2002). Google Scholar

15. 

Y.-H. Hu et al., “Characterizing a novel scintillating glass for application to megavoltage cone-beam computed tomography,” Med. Phys., 46 (3), 1323 –1330 https://doi.org/10.1002/mp.13355 MPHYA6 0094-2405 (2019). Google Scholar

16. 

T. C. Harris et al., “Impact of a novel multilayer imager on metal artifacts in MV-CBCT,” Phys. Med. Biol., 68 (14), 145009 https://doi.org/10.1088/1361-6560/ace09a PHMBA7 0031-9155 (2023). Google Scholar

17. 

Y. El-Mohri et al., “Low-dose megavoltage cone-beam CT imaging using thick, segmented scintillators,” Phys. Med. Biol., 56 (6), 1509 https://doi.org/10.1088/0031-9155/56/6/001 PHMBA7 0031-9155 (2011). Google Scholar

18. 

S. L. Meeks et al., “Performance characterization of megavoltage computed tomography imaging on a helical tomotherapy unit,” Med. Phys., 32 (8), 2673 –2681 https://doi.org/10.1118/1.1990289 MPHYA6 0094-2405 (2005). Google Scholar

19. 

K. Schombourg, F. Bochud and R. Moeckli, “Stability of the helical tomotherapy Hi-Art II detector for treatment beam irradiations,” J. Appl. Clin. Med. Phys., 15 (6), 119 –127 https://doi.org/10.1120/jacmp.v15i6.4897 (2014). Google Scholar

20. 

A. P. Shah et al., “Patient dose from megavoltage computed tomography imaging,” Int. J. Radiat. Oncol. Biol. Phys., 70 (5), 1579 –1587 https://doi.org/10.1016/j.ijrobp.2007.11.048 IOBPD3 0360-3016 (2008). Google Scholar

21. 

J. D. Kuttig et al., “Comparative investigation of the detective quantum efficiency of direct and indirect conversion detector technologies in dedicated breast CT,” Phys. Med., 31 (4), 406 –413 https://doi.org/10.1016/j.ejmp.2015.03.007 (2015). Google Scholar

22. 

T. Flohr et al., “Photon-counting CT review,” Phys. Med., 79 126 –136 https://doi.org/10.1016/j.ejmp.2020.10.030 (2020). Google Scholar

23. 

C. H. McCollough et al., “The technical development of photon-counting detector CT,” Eur. Radiol., 33 (8), 5321 –5330 https://doi.org/10.1007/s00330-023-09545-9 (2023). Google Scholar

24. 

M. Danielsson, M. Persson and M. Sjölin, “Photon-counting x-ray detectors for CT,” Phys. Med. Biol., 66 (3), 03TR01 https://doi.org/10.1088/1361-6560/abc5a5 PHMBA7 0031-9155 (2021). Google Scholar

25. 

A. S. Wang et al., “Pulse pileup statistics for energy discriminating photon counting X-ray detectors,” Med. Phys., 38 (7), 4265 –4275 https://doi.org/10.1118/1.3592932 MPHYA6 0094-2405 (2011). Google Scholar

26. 

M. Touch et al., “A neural network-based method for spectral distortion correction in photon counting X-ray CT,” Phys. Med. Biol., 61 (16), 6132 https://doi.org/10.1088/0031-9155/61/16/6132 PHMBA7 0031-9155 (2016). Google Scholar

27. 

C. Sundberg et al., “1-μm spatial resolution in silicon photon-counting CT detectors,” J. Med. Imaging, 8 (6), 063501 https://doi.org/10.1117/1.JMI.8.6.063501 JMEIET 0920-5497 (2021). Google Scholar

28. 

K. Rajendran et al., “First clinical photon-counting detector CT system: technical evaluation,” Radiology, 303 (1), 130 –138 https://doi.org/10.1148/radiol.212579 RADLAX 0033-8419 (2022). Google Scholar

29. 

C. Sundberg, M. Danielsson and M. Persson, “Compton coincidence in silicon photon-counting CT detectors,” J. Med. Imaging, 9 013501 https://doi.org/10.1117/1.JMI.9.1.013501 JMEIET 0920-5497 (2022). Google Scholar

30. 

Y. Gu et al., “Study of a high resolution, 3-D positioning cross-strip Cadmium Zinc Telluride detector for PET,” in IEEE Nucl. Sci. Symp. Conf. Record, 3596 –3603 (2008). https://doi.org/10.1109/NSSMIC.2008.4774098 Google Scholar

31. 

S. Abbaszadeh et al., “Characterization of a sub-assembly of 3D position sensitive Cadmium Zinc Telluride detectors and electronics from a sub-millimeter resolution PET system,” Phys. Med. Biol., 61 6733 https://doi.org/10.1088/0031-9155/61/18/6733 PHMBA7 0031-9155 (2016). Google Scholar

32. 

Y. Wang and S. Abbaszadeh, “Reduction of the low energy threshold through new modular data acquisition electronics for cross-strip Cadmium Zinc Telluride (CZT) based PET system,” in IEEE Nucl. Sci. Symp. and Med. Imaging Conf. (NSS/MIC), 1 –3 (2021). https://doi.org/10.1109/NSS/MIC44867.2021.9875680 Google Scholar

33. 

R. Jeraj et al., “Radiation characteristics of helical tomotherapy,” Med. Phys., 31 (2), 396 –404 https://doi.org/10.1118/1.1639148 MPHYA6 0094-2405 (2004). Google Scholar

34. 

T. R. Mackie, “History of tomotherapy,” Phys. Med. Biol., 51 (13), R427 https://doi.org/10.1088/0031-9155/51/13/R24 PHMBA7 0031-9155 (2006). Google Scholar

35. 

D. S. Rigie and P. La Rivière, “An efficient material decomposition method using the Gauss-Newton algorithm,” in IEEE Med. Imaging Conf., (2015). https://doi.org/10.13140/RG.2.1.1680.3289 Google Scholar

36. 

E. Roessl and C. Herrmann, “Cramér–Rao lower bound of basis image noise in multiple-energy X-ray imaging,” Phys. Med. Biol., 54 1307 https://doi.org/10.1088/0031-9155/54/5/014 PHMBA7 0031-9155 (2009). Google Scholar

37. 

M. Diwakar and M. Kumar, “A review on CT image noise and its denoising,” Biomed. Signal Process. Control, 42 73 –88 https://doi.org/10.1016/j.bspc.2018.01.010 (2018). Google Scholar

38. 

H. Lu et al., “Noise properties of low-dose CT projections and noise treatment by scale transformations,” in IEEE Nucl. Sci. Symp. Conf. Record (Cat. No. 01CH37310), 1662 –1666 (2001). https://doi.org/10.1109/NSSMIC.2001.1008660 Google Scholar

39. 

J. B. Solomon, X. Li and E. Samei, “Relating noise to image quality indicators in CT examinations with tube current modulation,” Amer. J. Roentgenol., 200 (3), 592 –600 https://doi.org/10.2214/AJR.12.8580 (2013). Google Scholar

40. 

S. Leng et al., “Noise reduction in spectral CT: reducing dose and breaking the trade-off between image noise and energy bin selection,” Med. Phys., 38 (9), 4946 –4957 https://doi.org/10.1118/1.3609097 MPHYA6 0094-2405 (2011). Google Scholar

41. 

J. Ma et al., “Variance analysis of x-ray CT sinograms in the presence of electronic noise background,” Med. Phys., 39 (7Part1), 4051 –4065 https://doi.org/10.1118/1.4722751 MPHYA6 0094-2405 (2012). Google Scholar

42. 

W. P. Segars et al., “4D XCAT phantom for multimodality imaging research,” Med. Phys., 37 (9), 4902 –4915 https://doi.org/10.1118/1.3480985 MPHYA6 0094-2405 (2010). Google Scholar

43. 

M. Myronakis et al., “Multi-layer imager design for mega-voltage spectral imaging,” Phys. Med. Biol., 63 (10), 105002 https://doi.org/10.1088/1361-6560/aabe21 PHMBA7 0031-9155 (2018). Google Scholar

44. 

A. S. Wang and N. J. Pelc, “Spectral photon counting CT: imaging algorithms and performance assessment,” IEEE Trans. Radiat. Plasma Med. Sci., 5 453 –464 https://doi.org/10.1109/TRPMS.2020.3007380 (2021). Google Scholar

45. 

T. R. Mackie et al., “Tomotherapy,” Semin. Radiat. Oncol., 9 108 –117 https://doi.org/10.1016/S1053-4296(99)80058-7 SRONEO (1999). Google Scholar

46. 

A. S. Wang and N. J. Pelc, “Impact of photon counting detector spectral response on dual energy techniques,” Proc. SPIE, 7622 76223L https://doi.org/10.1117/12.840052 PSISDG 0277-786X (2010). Google Scholar

47. 

J. Cammin et al., “A cascaded model of spectral distortions due to spectral response effects and pulse pileup effects in a photon-counting x-ray detector for CT,” Med. Phys., 41 (4), 041905 https://doi.org/10.1118/1.4866890 MPHYA6 0094-2405 (2014). Google Scholar

Biography

Giavanna Jadick is a PhD candidate in the graduate program in medical physics at the University of Chicago. She received her BS degree in physics from Duke University in 2020. Her research focuses on spectral or multi-energy x-ray imaging, computed tomography, and quantitative phase retrieval. She is a student member of SPIE.

Maya Ventura is an undergraduate student studying global health at Northwestern University. She is a Chicago EYES on Cancer summer student with the University of Chicago Medicine Comprehensive Cancer Center, through which she researched dual-energy computed tomography. She is a student member of SPIE.

Patrick J. La Rivière is a professor in the Department of Radiology and the Committee on Medical Physics at the University of Chicago. He received his AB degree in physics from Harvard University and his PhD from the graduate program in medical physics at the University of Chicago. His research interests include tomographic reconstruction in computed tomography, x-ray fluorescence computed tomography, and computational microscopy. He is a fellow of SPIE.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.

Funding Statement

Giavanna Jadick, Maya Ventura, and Patrick J. La Rivière "Utility of photon-counting detectors for MV-kV dual-energy computed tomography imaging," Journal of Medical Imaging 11(S1), S12811 (26 December 2024). https://doi.org/10.1117/1.JMI.11.S1.S12811
Received: 29 April 2024; Accepted: 10 December 2024; Published: 26 December 2024
Advertisement
Advertisement
KEYWORDS
Sensors

Bone

Signal to noise ratio

Photons

Imaging systems

Radiotherapy

Tissues

Back to Top