1.IntroductionRapid advances in scanner technology have established computed tomography (CT) as an indispensable tool for the assessment of cardiac anatomy, coronary circulation, and the great vessels. Expanding applications coupled with increasing use have raised concerns over potential stochastic effects of radiation dose associated with CT scanning, in particular, the risk of radiation-induced carcinogenesis.1,2 Indeed, the use of radiation-based medical imaging has exploded over the past few decades and CT has emerged as the dominant contributor to medical radiation dose, accounting for of all the medical radiation dose to the United States population.3–7 Initial implementation of electrocardiographically (ECG) gated coronary CT angiography relied heavily on retrospectively gated acquisition mode, which was associated with relatively high radiation dose.8–11 Subsequent technologic evolution in both hardware and software CT technologies has helped lead a decreasing trend in mean radiation dose from cardiac CT.12 Several studies have demonstrated that the radiation dose from cardiac CT angiography (CCTA) can be reduced to lower than invasive catheter angiography with the use of techniques, such as prospective triggering, low tube potential technique, automatic exposure control (AEC), iterative reconstructive techniques, and so on.13–27 Conversely, despite the abundance of radiation dose-reduction strategies, several studies have reported a high variability in the radiation doses from cardiac CT.11,28–30 In an international, multicenter study, Hausleiter et al.11 reported that median doses from CCTA varied substantially among 21 participating institutions’ different CT systems, and that radiation reduction strategies were not adequately utilized. From the point of view of justification or appropriateness of CCTA, there is now a large body of evidence supporting its use in coronary artery disease including acute coronary syndromes.31–36 There is a need for raising awareness for its appropriate use and deployment of dose-reduction strategies available across different scan vendor platforms. We review various radiation dose-reduction strategies in the field of cardiac CT along with the scientific evidence in support of their value in the current clinical practice. 2.Scanner and Patient-Related Factors in Cardiac Computed Tomography AngiographyECG-gated cardiac CT images are acquired/reconstructed at time points in the cardiac cycle when the heart is relatively motion free (end systole or middiastole). To detect and predict these critical time points, CT data are acquired simultaneously with recording of the ECG data. ECG tagging of CT data enables image reconstruction from data acquired during these critical phases of the cardiac cycle. Phase of a reconstructed dataset refers to the position of the image reconstruction window within the cardiac cycle. Radiation dose may be switched on only during these phases (prospectively ECG triggered) or continuously with selective data filtering from acquired data (retrospective ECG gated). Prospective gating results in much lower radiation dose but cannot be applied with rapid heart rate with most of the scanners and can also result in stair-step artifacts and nonuniform contrast within the vessels. The stair-step artifact can also occur with ECG-gated cardiac CT due to gaps in data used for reconstruction. The prospectively triggered scanning does not provide full functional information; however, when cardiac functional information is obligatory, a prospectively ECG-triggered acquisition with a widened acquisition window (also referred to as “padding”) can provide limited functional information. The retrospective ECG gating provides multiple time points for image reconstruction, which allows functional assessment and may help mitigate the effects of motion artifacts, although at a cost of increased radiation dose. Prospective gating is applied in combination with sequential axial step and shoot scanning, whereas retrospective gating is applied in combination with a helical scanning mode. For helical scan mode, the pitch is defined as the ratio of table movement along the -axis during single tube rotation and the beam width in the -axis.8 Temporal resolution is a fundamental requirement of CCTA and requires fast-rotating gantries and multisegmented reconstructions (slice data acquired over more than one cardiac cycle). Faster gantry rotation and multisegmented reconstructions require a slower pitch in cardiac mode to avoid discontinuities in the anatomic coverage of the heart between images reconstructed from consecutive cardiac cycles. Multisegmented reconstruction is one of the most important factors responsible for high radiation dose. Due to radiation concerns and the limited clinical performance of multisegment reconstruction with variable heart rate, most scanners have these days switched to single-segment reconstruction. The helical pitch has to be still low in single-segment reconstruction, as the table should not move more than beam width during the time of one heart cycle to ensure gapless volume coverage. The pitch also has to be adapted to the patient’s heart rate and needs to be lower at lower heart rates. This leads to a relatively higher radiation dose with retrospectively gated scans in patients with lower heart rates. At our institution, a prospective gating technique is typically used in patients with heart rates lower than 75 per minute on a 128-slice dual-source scanner. 2.1.Other Scanner-Related FactorsThe radiation dose is primarily dependent on radiation output of the scanner, which is approximately proportional to the square of tube potential (kVp) and the product of tube current and gantry rotation time (mAs). Several scan factors have a major impact on radiation dose by affecting pitch during CCTA. These include number of tubes (single versus dual source), gantry rotation time and -axis coverage in single tube rotation (number of detector rows). Maintaining a low-pitch value is not necessary for dual-source CT and it could be increased to almost double at high heart rates allowing radiation dose savings.37 Recent scanners equipped with a large number of detector rows (256 to 320) may allow up to 16 cm -axis coverage, which is sufficient to cover the entire heart in single rotation eliminating the need of overlapping acquisitions. In combination with high gantry rotation speed/temporal resolution, these scanners allow single beat axial acquisition with a static table.38–41 The image reconstruction algorithms used for CT image reconstruction can further influence the choice of scan factors and thus the associated radiation dose. Most modern CT scanners are increasingly employing sophisticated iterative reconstruction (IR) techniques as compared with the previously used filtered back projection (FBP) algorithms, which are less computationally demanding. Advances in computation hardware and speed have enabled IR techniques into the mainstream to mitigate adverse effects of reduced radiation dose in FBP images while retaining diagnostic information.42 2.2.Patient-Related FactorsRadiation dose also depends on the body size as penetration of a larger body size demands a higher photon number and energy.43 Heart rate and rhythm often dictate the acquisition mode (prospective versus retrospective) and profoundly affect associated radiation dose with CCTA. Rapid heart rates (tachycardia) shorten the time window of relatively motion-free phases of the cardiac cycle (diastole), which can force the users to employ a retrospectively gated CCTA technique. Irregular rhythm with high heart rates may require application of the arrhythmia rejection technique (Fig. 1) with prospective triggering at end systole, which can result in increased radiation exposure (although still lower than a retrospectively gated scan).44–48 3.General Radiation Dose-Reduction Strategies in Cardiac Computed Tomography3.1.Clinical Indication-Driven ProtocolIndications such as assessment of coronary calcification, pulmonary veins, and myocardium, can be scanned with lower spatial and temporal resolutions and with much lower radiation doses. When CCTA is being performed primarily for anomalous origin and course of coronary arteries, aggressive dose reduction can be applied without sacrificing diagnostic information.49 3.2.Anatomic Scan Coverage (-Axis)This is an important factor determining the radiation dose, which is directly proportional to -axis coverage.43 Achieving optimal anatomic coverage for different indications requires careful planning. According to prior publications, each additional 1 cm of coverage at CCTA may increase the radiation dose by 40 to 100 mGy cm, although the additional dose per cm might not be this high with modern state-of-the-art scanners.11,49 Appropriate scan coverage for CCTA may be achieved either on the basis of the initial calcium score scan or with the use of anatomic landmarks on the planning radiograph (tracheal bifurcation on the frontal view to the bottom of the cardiac silhouette on the lateral view). Leschka et al.50 reported that scan length adjustments on the basis of a calcium scoring scan may allow a 16% radiation dose reduction. Scan range for CCTA for evaluation of postcoronary bypass grafts may vary often extending from the origin of the left internal mammary artery (at the root of the neck) to the base of the heart. The coverage for pulmonary veins extends from 2 cm above the aortic arch to the dome of the diaphragm.49 3.3.Automatic Exposure ControlThese techniques help adapt tube current according to the anatomic factors and specified image quality metric (angular, longitudinal, and combined types) or according to specified phases of ECG where reduced tube current is acceptable for image creation (ECG-based modulation). The angular AEC cannot be applied to ECG-gated scans but longitudinal AEC along the -axis based on the planning radiographs can be applied in cardiac imaging and helps adapt tube current to changing body habitus.51 ECG-triggered prospective axial scanning is being increasingly applied instead of retrospective ECG-gated helical scanning for radiation dose reduction in CCTA examinations.18,52 For retrospective ECG-gated CCTA examinations, the evaluation of ventricular function does not require high quality and the tube current can be modulated according to the ECG signal, with the full tube current applied during the end-systolic or mid-diastolic phases and the tube current kept at lower levels during the rest of the cardiac cycle (retrospective gating with ECG-based modulation).53,54 Based on the CT vendor, tube current can be reduced to a user-specified minimum level or 4% to 20% of the tube current applied in the phases with higher image quality requirement. A variable amount of dose reduction is noted with application of this type of modulation based on the length of the phase over which higher image quality is desired, and heart rate. A rapid heart rate leaves a smaller duration of cardiac cycle over which reduced tube current can be applied and vice versa. Compared to the use of fixed tube current-based retrospectively gated CCTA, up to 40% reduction in radiation dose from CCTA can be achieved with ECG-based tube current modulation.55 3.4.Tailoring Acquisition Modes to Heart RateHeart rate and its rhythm can affect both the ability to obtain motion-free coronary artery evaluation and reduce radiation dose with appropriate acquisition mode (prospectively ECG triggered versus retrospectively ECG gated). A target heart rate of up to 75 beats per minute is desirable with most vendors to minimize cardiac-motion artifacts and also reduce radiation dose associated with CCTA. Such a regular low heart rate enables use of prospectively ECG-triggered axial or helical high-pitched scan mode, which is associated with lower radiation dose (up to 90% lower) compared with retrospectively ECG-gated helical scan mode. Heart rate control medications, such as beta blockers, are frequently administered to reduce heart rate to apply prospective triggering and maximize the number of evaluable coronary segments. Although the need for rate control strongly depends on the CT system used and its temporal resolution, the modern CT systems may provide ECG-triggered prospective scans for patients with heart rates up to 70 bpm or even higher. 3.5.Parameter Selection Based on Body HabitusAEC and automatic tube potential selection techniques for CCTA examinations should be used when available to apply appropriate tube potential and current based on the patient’s body habitus. Most nonobese patients [Body mass index (BMI) of up to ] should be scanned 100 kV instead of the traditional 120 kV.25,56,57 In morbidly obese patients (), a combination of high potential (), high tube current, higher strength of IR, and retrospective ECG gating may be needed to obtain acceptable image quality.49 4.Recent Advances in Radiation Dose Reduction with Clinical EvidenceSeveral notable developments have transformed CCTA and enhanced its role in cardiac imaging while enabling substantial radiation dose reduction. These include wide-area detector arrays, dual-source CT, automatic tube potential selection, adaptive prospective ECG triggering, and IR. CT scanners with wide-area detector arrays (e.g., Toshiba 320 to 640-slice, GE 256-slice) enable 16 cm scan coverage of the entire heart in single gantry rotation and single heart beat, and eliminate banding or slab artifacts. The dual-source CT (64-, 128-, 192-slice, Siemens) has helped enhance the temporal resolution and enable high-pitch prospectively ECG-triggered helical mode of acquiring data in a single heart beat as well. In this section, we discuss some of these recent advances from the perspective of cardiac CT and radiation dose optimization. 4.1.Low Tube PotentialWhen other scan factors are held constant, low tube potential results in radiation dose reduction () and decreased requirement of iodinated contrast, but with an increase in image noise. This noise can be tolerated for two reasons: first, due to close proximity with the edge of iodinated contrast, low tube potential results in high contrast and secondly effective reconstruction algorithms, such as IR techniques.58,59 The application of lower tube potential requires careful adjustment of the tube output (mAs) to maintain the desired contrast to noise ratio, leading to high mAs demands in particular for bigger patients. Hence, the use of low kV scanning has to be limited to patients with smaller body size. To decrease the impact of higher image noise on image quality a large window width can be used on the monitor display for image interpretation. Combined low-voltage tube and IR techniques can also help in maintaining good signal-to-noise ratio comparable to using standard voltage tube and FBP techniques (Fig. 2).60 Low tube potential (80 or 100 kV) imaging for cardiac imaging was initially described for nonobese patients (, 80 kV; BMI 20 to , 100 kV) (Table 1).61 Experience from several clinical trials, such as PROTECTION I, PROTECTION II, and REALISE, demonstrated 53%, 31%, and 35% reductions in radiation dose, respectively, without substantial impairment of diagnostic image quality when the tube potential was lowered from 120 to 100 kV.25,56,57 These studies were done with the use of 64-slice dual-source scanners and IR techniques. Based on these observations, current guidelines recommend the use of low tube potential (80 to 100 kV) for CCTA in nonobese patients. Table 1Cardiac CT at low tube potential imaging: recent evidence (key: R, retrospective; P, prospective; DSCT, dual-source CT; MC, multicenter).
Two recent studies have reported further dose reductions and diagnostic image quality with further lowering of tube potential 70 kV in nonobese patients62,63 on the third-generation dual-source scanners (Siemens Somatom Force), which enables much higher tube current (up to 1300 mA for each x-ray tube) and IR technique (admire).63 On the same scanner, Mangold et al.64 reported diagnostic image quality for CCTA in overweight and obese patients at 120 kV, which were previously scanned with . Evaluating 102 overweight and obese patients with BMI ranging from 25 to , authors demonstrated diagnostic quality CCTA and triple rule out evaluation with 120 kV.64 Zhang et al.65 have reported the use of 70 kV with a 64-slice dual-source scanner, but these results must be interpreted cautiously as more than three quarters of the included patients had and there was a reduction in diagnostic accuracy on a per-patient, per-vessel, and per-segment basis. Lower tube potentials are also associated with increased attenuation of iodine on contrast CT. Therefore, low tube potential protocols might potentially reduce the contrast dose and are also more suitable for cardiovascular applications, such as myocardial CT, perfusion to better detect the regional differences in myocardial contrast attenuation. 4.2.High-Pitch Helical AcquisitionsSingle-source multidetector-row helical CT scanners are restricted to a pitch of up to (range 0.2 to ). The two simultaneous overlapping helices on dual-source CT with ECG-triggered prospectively gating enable image reconstruction at a high pitch of over at significantly lower than the radiation dose.66,67 High-pitch helical (HPH) acquisition mode with fast table speed and prospective ECG triggering enables coverage of the entire heart within one cardiac cycle. Several studies on HPH mode for CCTA in pediatric and adult populations16,63,68–70 have reported good diagnostic quality and consistently low mean radiation dose (0.8 to 1.3 mSv) (Table 2). The HPH is particularly advantageous in children or young adults for assessment of coronary artery anomalies and anatomic assessment in patients with congenital heart disease (when information regarding cardiac function is not needed). In particular, HPH can help avoid need for sedation or anesthesia in young children due to its rapidity. Table 2HPH acquisition mode in cardiac CT, recent evidence (key: R, retrospective; P, prospective; DSCT, dual-source CT; SSCT, single source CT; proECG, prospectively ECG-triggered axial mode; retroECG, retrospectively ECG-gated mode).
Important limitations of HPH mode include inconsistent diagnostic quality in patients with increased heart rates,16,68 motion artifacts and nonapplicability of retrospective reconstruction, and functional assessment of the heart.71 To mitigate the effects of motion on evaluation of coronary artery evaluation, a prior study has recommended use of subsequent acquisitions in two different cardiac cycles.71 This technique is typically avoided in morbidly obese patients with suspected or known coronary atherosclerotic disease due to higher likelihood of artifacts and excessive image noise. 4.3.Iterative ReconstructionsThe iterative reconstructions (IR) techniques have had a major impact on reducing radiation dose and have found applications in all domains of CT imaging. Today, major CT vendors offer multiple IR techniques on different generations of their scanners. Indeed, IR techniques are replacing conventional FBP techniques since they are less prone to artifacts and increased image noise at lower radiation doses as compared with FBP, which may have compromised diagnostic quality at low radiation doses.42,72 IR techniques remove image noise through a process of modeling of the imaging acquisition processes including photon statistics fluctuations, the optics system, and other aspects of x-ray interactions that are ignored by the conventional FBP techniques. IR techniques use and include a simple algebraic reconstruction technique, statistical IR, and more recent model-based IR (MBIR). Current IR algorithms typically combine statistical modeling with MBIR.73 Multiple studies have evaluated the impact of IR techniques on image quality and radiation dose for different clinical indications in both adults and pediatric patients including CCTA (Table 3).13–15,74,75 A recent meta-analysis reported that the IR techniques can lower radiation dose by up to 48% in comparison with the traditional FBP method for CCTA.14 Availability and application of IR in CCTA have helped in applying lower tube potential and tube current for reducing radiation dose while improving image contrast. Table 3Iterative reconstruction in cardiac CT: recent evidence (key: R, retrospective; P, prospective; M, meta-analysis; DSCT, dual-source CT).
There are several types of IR techniques available for commercial use but the most commonly used one involves blending of IR and FBP techniques (hybrid). Hybrid IR techniques involve prefixed blending of FBP and IR data and contrast to noise ratio has been shown to significantly improve with increasing strength or level of IR contribution (Fig. 3).76 A common concern or limitation at higher strength or level of IR algorithms pertains to the fact that radiologists might not feel comfortable with rather “glossy” images with increasing IR levels, which may be mitigated with increasing familiarity with the IR.77 The “glossy” appearance is more pronounced with pure model-based IR techniques, which also take substantially longer reconstruction time compared with hybrid IR techniques. 4.4.Single Heart Beat Scan with Wide Detector ArrayOther technical evolution in recent times has been the development of wide detector array scanners (up to 16 cm detector width along the patient length) with full cardiac coverage in single gantry rotation and single heart beat. These scanners eliminate banding artifacts (previously known as slab or misalignment artifacts) and provide uniform contrast enhancement in the coronary arteries. They also allow measurement of contrast attenuation gradients across atherosclerotic lesions (information somewhat similar to functional flow reserve) and are better suited for myocardial perfusion assessment.12 Common concerns pertaining to these scanners include the need for full 360-deg rotation, relatively higher radiation dose and cone beam geometry-related artifacts.9 Various studies have reported radiation doses associated with the use of wide detector array MDCT scanners for CCTA ranging from 4.4 to 15 mSv.9,38,39,41,78,79 In a single study with second-generation wide-volume 320-detector row MDCT, Chen et al.40 reported a substantially low radiation dose for CCTA examinations (0.9 mSv; interquartile range 0.6 to 1.7 mSv). 5.Pediatric Cardiac Computed TomographyDue to higher radiation risk and the need for serial follow-up imaging, pediatric cardiac CT warrants special attention. Echocardiography has a limited role in the characterization of complex congenital heart disease and coronary evaluation beyond proximal segments. MRI is an excellent modality for functional assessment; however, its use can also be limited by the presence of noncompatible devices, need for anesthesia, and long image acquisition time.80 Armed with technologic advances, cardiac CT has emerged as a useful alternative in pediatric heart diseases. As discussed above, most radiation dose-reduction techniques employed in adults are also applicable in children (Table 4). Most children can be and should be scanned at lower tube potential (70 to 80 kV) and reduced tube current.20,81 When information pertaining to cardiac function is not required or can be obtained with other imaging modalities, prospectively ECG-triggered axial or helical mode (such as high pitch on dual-source CT) should be used along with lower tube potential and IR techniques in order to reduce radiation dose. Most CCTA can be routinely obtained at sub-mSv radiation doses in children using prospective ECG-triggered high-pitch DSCT and iterative reconstruction to achieve consistent diagnostic scans.20,69,75,81,82 When cardiac functional information is obligatory, either prospectively ECG-triggered acquisition with a widened acquisition window (also referred to as “padding”) or retrospectively ECG-gated acquisition with an aggressive ECG-based tube current modulation should be used along with lower tube potential and optimal scan coverage.69 Table 4Pediatric cardiac CT: prospective studies on radiation dose reduction while maintaining diagnostic information. (Key: R, retrospective; P, prospective; M, meta-analysis; DSCT, dual-source CT; proECG, prospectively ECG-triggered axial mode; retroECG, retrospectively ECG-gated mode).
6.Miscellaneous Dose-Reduction StrategiesIn this section, we discuss several other measures and techniques that can aid in optimizing radiation dose for CT (Table 5).63,70,83 Scanners with faster rotation times (better temporal resolution) or an increased number of x-ray sources permit higher heart rate cutoffs for prospectively ECG-triggered acquisition mode, which is associated with substantially lower radiation dose as compared with retrospectively ECG-gated acquisition mode.63,70,83 Addition of “adaptive” or arrhythmia rejection feature in prospectively ECG-triggered acquisition mode enables the scanners to reject scanning in beats with irregular rhythm so that obtain diagnostic quality exams at reduced radiation dose even in patients with variable heart rates, such as atrial fibrillation. Without the arrhythmia rejection algorithms for prospectively ECG-triggered mode, patients with variable heart rates would require much higher radiation doses from retrospectively ECG-gated acquisition mode with conservative ECG-based tube current modulation. In fact, use of adaptive prospectively ECG-triggered acquisition mode in these patients can decrease radiation dose by up to 50% compared with retrospectively ECG-gated helical DSCT.48 Motion-correction algorithms can improve CCTA image quality and diagnostic capability in patients with high HR and HR variability at low radiation dose prospective scans.84 Table 5Additional literature on reducing radiation dose for CCTA (key: R, retrospective; P, prospective; DSCT, dual-source CT; NA, not applicable; proECG, prospectively ECG-triggered axial mode; retroECG, retrospectively ECG-gated mode; HPH, helical high pitch).
In female patients, radiation dose for cardiac CT can be substantially reduced by displacing the mobile portion of breasts using straps attached to the scanner table—a routinely performed exercise at our institution prior to acquisition of planning radiographs.85 In addition to dose reduction, this can help decrease image noise and/or artifacts in female patients with large breasts. Application of a bismuth-based breast displacement shield can effectively reduce breast exposure at the expense of increased noise and artifacts. Moreover, comparable breast radiation reductions can be achieved without shielding by globally reducing tube current with similar increase in image noise. At our institution, we neither use nor recommend the use of a bismuth shield for patients undergoing cardiac CT due to artifacts from edges of the bismuth shield. Bismuth shields can also affect the measured CT numbers, which is especially important in calcium scoring scans. To avoid these errors, a thick padding material needs to be placed between the patient’s skin and the shield, a time-consuming and cumbersome method. The more advanced methods of reducing exposure, including dose modulation and iterative reconstruction techniques, are superior to breast shielding and should be preferably used if available.86,87 Well-known limitations of CCTA pertain to evaluation of luminal stenosis in patients with dense coronary calcifications and stents. Recent studies have reported that combining CCTA and CT perfusion can provide better diagnostic accuracy for detecting hemodynamically significant coronary artery stenosis compared with CCTA alone in patients with dense coronary calcifications or stents.88,89 Dose reduction for such combined studies was achieved with a combination of low tube potential and prospectively ECG-triggered axial (for rest perfusion) and high-pitch (for stress perfusion) acquisition mode. Dual energy CT has also been evaluated for myocardial perfusion.90 Comparable radiation doses between single- and dual-energy CT have been reported although further research is necessary to compare doses with low tube potential single energy CT.90 Ability to provide additional information from generation of perfusion data might help in decreasing the need of additional radiation-based investigations and reduction in overall radiation dose. Finally, the dose tracking initiative of the American College of Radiology Dose Index Registry warrants special attention. The registry provides an opportunity for CT centers and institutions to compare their radiation doses with other practices and then tailor cardiac CT protocols according to median or average radiation doses across the United States.91 7.ConclusionLimiting radiation dose from CT is an important goal for the imaging community. Aforementioned techniques and strategies have helped in substantial reduction of radiation dose for cardiac CT examinations and make it an attractive and meaningful option for imaging suitable patients with suspected or known cardiac and coronary abnormalities. DisclosuresNo conflicts of interest, financial or otherwise, are declared by the authors SH, VB, and MK. Dr. BG serves as Educational consulting for Medtronic and Siemens. ReferencesD. J. Brenner and E. J. Hall,
“Computed tomography—an increasing source of radiation exposure,”
N. Engl. J. Med., 357
(22), 2277
–2284
(2007). http://dx.doi.org/10.1056/NEJMra072149 NEJMAGNEJMAG 0028-4793 Google Scholar
D. J. Brenner and H. Hricak,
“Radiation exposure from medical imaging: time to regulate?,”
JAMA, 304
(2), 208
–209
(2010). http://dx.doi.org/10.1001/jama.2010.973 JAMAAP 0098-7484 Google Scholar
F. A. Mettler et al.,
“Effective doses in radiology and diagnostic nuclear medicine: a catalog,”
Radiology, 248
(1), 254
–263
(2008). http://dx.doi.org/10.1148/radiol.2481071451 RADAPRADLAX 0033-8419 Google Scholar
H. Hricak et al.,
“Managing radiation use in medical imaging: a multifaceted challenge,”
Radiology, 258
(3), 889
–905
(2011). http://dx.doi.org/10.1148/radiol.10101157 RADAPRADLAX 0033-8419 Google Scholar
F. A. Mettler et al.,
“Radiologic and nuclear medicine studies in the United States and worldwide: frequency, radiation dose, and comparison with other radiation sources—1950–2007,”
Radiology, 253
(2), 520
–531
(2009). http://dx.doi.org/10.1148/radiol.2532082010 RADAPRADLAX 0033-8419 Google Scholar
E. C. Lin,
“Radiation risk from medical imaging,”
Mayo Clin. Proc., 85
(12), 1142
–1146
(2010). http://dx.doi.org/10.4065/mcp.2010.0260 MACPAJ 0025-6196 Google Scholar
J. A. Brink and R. L. Morin,
“Size-specific dose estimation for CT: how should it be used and what does it mean?,”
Radiology, 265
(3), 666
–668
(2012). http://dx.doi.org/10.1148/radiol.12121919 RADLAX 0033-8419 Google Scholar
B. Desjardins and E. A. Kazerooni,
“ECG-gated cardiac CT,”
Am. J. Roentgenol., 182
(4), 993
–1010
(2004). http://dx.doi.org/10.2214/ajr.182.4.1820993 AJROAM 0092-5381 Google Scholar
Z. Sun, G. H. Choo and K. H. Ng,
“Coronary CT angiography: current status and continuing challenges,”
Br. J. Radiol., 85
(1013), 495
–510
(2012). http://dx.doi.org/10.1259/bjr/15296170 BJRAAP 0007-1285 Google Scholar
A. J. Einstein, M. J. Henzlova and S. Rajagopalan,
“Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography,”
JAMA, 298
(3), 317
–323
(2007). http://dx.doi.org/10.1001/jama.298.3.317 JAMAAP 0098-7484 Google Scholar
J. Hausleiter et al.,
“Estimated radiation dose associated with cardiac CT angiography,”
JAMA, 301
(5), 500
–507
(2009). http://dx.doi.org/10.1001/jama.2009.54 JAMAAP 0098-7484 Google Scholar
A. Aghayev et al.,
“Recent developments in the use of computed tomography scanners in coronary artery imaging,”
Expert Rev. Med. Devices, 13
(6), 545
–553
(2016). http://dx.doi.org/10.1080/17434440.2016.1184968 1743-4440 Google Scholar
D. C. Benz et al.,
“Adaptive statistical iterative reconstruction-V: impact on image quality in ultralow-dose coronary computed tomography angiography,”
J. Comput. Assisted Tomogr., 40
(6), 958
–963
(2016). http://dx.doi.org/10.1097/RCT.0000000000000460 JCATD5 0363-8715 Google Scholar
A. M. Den Harder et al.,
“Dose reduction with iterative reconstruction for coronary CT angiography: a systematic review and meta-analysis,”
Br. J. Radiol., 89
(1058), 20150068
(2016). http://dx.doi.org/10.1259/bjr.20150068 BJRAAP 0007-1285 Google Scholar
Y. Iyama et al.,
“Submillisievert radiation dose coronary CT angiography,”
Acad. Radiol., 23
(11), 1393
–1401
(2016). http://dx.doi.org/10.1016/j.acra.2016.07.005 Google Scholar
M. Koplay et al.,
“Radiation dose and diagnostic accuracy of high-pitch dual-source coronary angiography in the evaluation of coronary artery stenoses,”
Diagn. Interventional Imaging, 97
(4), 461
–469
(2016). http://dx.doi.org/10.1016/j.diii.2015.10.008 Google Scholar
Y. Nishiyama et al.,
“Effect of the forward-projected model-based iterative reconstruction solution algorithm on image quality and radiation dose in pediatric cardiac computed tomography,”
Pediatr. Radiol., 46
(12), 1663
–1670
(2016). http://dx.doi.org/10.1007/s00247-016-3676-x PDRYA5 1432-1998 Google Scholar
B. Bischoff et al.,
“Comparison of sequential and helical scanning for radiation dose and image quality: results of the prospective multicenter study on radiation dose estimates of cardiac CT angiography (PROTECTION) I study,”
Am. J. Roentgenol., 194
(6), 1495
–1499
(2010). http://dx.doi.org/10.2214/AJR.09.3543 AJROAM 0092-5381 Google Scholar
S. Deseive et al.,
“Image quality and radiation dose of a prospectively electrocardiography-triggered high-pitch data acquisition strategy for coronary CT angiography: the multicenter, randomized PROTECTION IV study,”
J. Cardiovasc. Comput. Tomogr., 9
(4), 278
–285
(2015). http://dx.doi.org/10.1016/j.jcct.2015.03.001 Google Scholar
Q.-R. Hou et al.,
“A prospective evaluation of contrast and radiation dose and image quality in cardiac CT in children with complex congenital heart disease using low-concentration iodinated contrast agent and low tube voltage and current,”
Br. J. Radiol., 90
(1070), 20160669
(2017). http://dx.doi.org/10.1259/bjr.20160669 BJRAAP 0007-1285 Google Scholar
A. D. Choi et al.,
“Prospective evaluation of the influence of iterative reconstruction on the reproducibility of coronary calcium quantification in reduced radiation dose 320 detector row CT,”
J. Cardiovasc. Comput. Tomogr., 10
(5), 359
–363
(2016). http://dx.doi.org/10.1016/j.jcct.2016.07.016 Google Scholar
G. Sun et al.,
“Application of low tube voltage coronary CT angiography with low-dose iodine contrast agent in patients with a BMI of 26–30 kg/m2,”
Clin. Radiol., 70
(2), 138
–145
(2015). http://dx.doi.org/10.1016/j.crad.2014.10.002 Google Scholar
J. Stehli et al.,
“Accuracy of coronary CT angiography using a submillisievert fraction of radiation exposure: comparison with invasive coronary angiography,”
J. Am. Coll. Cardiol., 64
(8), 772
–780
(2014). http://dx.doi.org/10.1016/j.jacc.2014.04.079 JACCDI 0735-1097 Google Scholar
K. M. Chinnaiyan et al.,
“CT dose reduction using prospectively triggered or fast-pitch spiral technique employed in cardiothoracic imaging (the CT dose study),”
J. Cardiovasc. Comput. Tomogr., 8
(3), 205
–214
(2014). http://dx.doi.org/10.1016/j.jcct.2014.04.001 Google Scholar
J. Hausleiter et al.,
“Image quality and radiation exposure with a low tube voltage protocol for coronary CT angiography results of the PROTECTION II trial,”
JACC Cardiovasc. Imaging, 3
(11), 1113
–1123
(2010). http://dx.doi.org/10.1016/j.jcmg.2010.08.016 Google Scholar
T. M. LaBounty et al.,
“Coronary CT angiography of patients with a normal body mass index using 80 kVp versus 100 kVp: a prospective, multicenter, multivendor randomized trial,”
Am. J. Roentgenol., 197
(5), W860
–W867
(2011). http://dx.doi.org/10.2214/AJR.11.6787 AJROAM 0092-5381 Google Scholar
C. N. De Cecco et al.,
“Dual-source CT coronary angiography: prospective versus retrospective acquisition technique,”
Radiol. Med., 116
(2), 178
–188
(2011). http://dx.doi.org/10.1007/s11547-010-0584-2 Google Scholar
L. Husmann et al.,
“Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating,”
Eur. Heart J., 29
(2), 191
–197
(2008). http://dx.doi.org/10.1093/eurheartj/ehm613 EHJODF 0195-668X Google Scholar
N. R. Mollet et al.,
“High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography,”
Circulation, 112
(15), 2318
–2323
(2005). http://dx.doi.org/10.1161/CIRCULATIONAHA.105.533471 CIRCAZ 0009-7322 Google Scholar
“NCRP report no. 160, ionizing radiation exposure of the population of the United States,”
(2017). http://ncrponline.org/publications/reports/ncrp-report-160/ Google Scholar
P. S. Douglas et al.,
“Outcomes of anatomical versus functional testing for coronary artery disease,”
N. Engl. J. Med., 372
(14), 1291
–1300
(2015). http://dx.doi.org/10.1056/NEJMoa1415516 NEJMAG 0028-4793 Google Scholar
J. M. Miller et al.,
“Diagnostic performance of coronary angiography by 64-row CT,”
N. Engl. J. Med., 359
(22), 2324
–2336
(2008). http://dx.doi.org/10.1056/NEJMoa0806576 NEJMAG 0028-4793 Google Scholar
SCOT-HEART Investigators,
“CT coronary angiography in patients with suspected angina due to coronary heart disease (SCOT-HEART): an open-label, parallel-group, multicentre trial,”
The Lancet, 385
(9985), 2383
–2391
(2015). http://dx.doi.org/10.1016/S0140-6736(15)60291-4 Google Scholar
S. B. Puchner et al.,
“High-risk plaque detected on coronary CT angiography predicts acute coronary syndromes independent of significant stenosis in acute chest pain: results from the ROMICAT-II trial,”
J. Am. Coll. Cardiol., 64
(7), 684
–692
(2014). http://dx.doi.org/10.1016/j.jacc.2014.05.039 JACCDI 0735-1097 Google Scholar
A. J. Taylor et al.,
“ACCF/SCCT/ACR/AHA/ASE/ASNC/NASCI/SCAI/SCMR 2010 appropriate use criteria for cardiac computed tomography. A report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, the Society of Cardiovascular Computed Tomography, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the American Society of Nuclear Cardiology, the North American Society for Cardiovascular Imaging, the Society for Cardiovascular Angiography and Interventions, and the Society for Cardiovascular Magnetic Resonance,”
J. Am. Coll. Cardiol., 56
(22), 1864
–1894
(2010). http://dx.doi.org/10.1016/j.jacc.2010.07.005 Google Scholar
M. J. Budoff et al.,
“Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY (assessment by coronary computed tomographic angiography of individuals undergoing invasive coronary angiography) trial,”
J. Am. Coll. Cardiol., 52
(21), 1724
–1732
(2008). http://dx.doi.org/10.1016/j.jacc.2008.07.031 JACCDI 0735-1097 Google Scholar
A. N. Primak et al.,
“Relationship between noise, dose, and pitch in cardiac multi–detector row CT,”
RadioGraphics, 26
(6), 1785
–1794
(2006). http://dx.doi.org/10.1148/rg.266065063 Google Scholar
G. Sun et al.,
“320-detector row CT coronary angiography: effects of heart rate and heart rate variability on image quality, diagnostic accuracy and radiation exposure,”
Br. J. Radiol., 85
(1016), e388
–e394
(2012). http://dx.doi.org/10.1259/bjr/92160185 BJRAAP 0007-1285 Google Scholar
C.-M. Chen et al.,
“Radiation dose exposure of patients undergoing 320-row cardiac CT for assessing coronary angiography and global left ventricular function,”
Int. J. Cardiovasc. Imaging, 28
(Suppl. 1), 1
–5
(2012). http://dx.doi.org/10.1007/s10554-012-0065-x Google Scholar
M. Y. Chen, S. M. Shanbhag and A. E. Arai,
“Submillisievert median radiation dose for coronary angiography with a second-generation 320-detector row CT scanner in 107 consecutive patients,”
Radiology, 267
(1), 76
–85
(2013). http://dx.doi.org/10.1148/radiol.13122621 RADLAX 0033-8419 Google Scholar
J. Qin et al.,
“320-detector CT coronary angiography with prospective and retrospective electrocardiogram gating in a single heartbeat: comparison of image quality and radiation dose,”
Br. J. Radiol., 85
(1015), 945
–951
(2012). http://dx.doi.org/10.1259/bjr/29901700 BJRAAP 0007-1285 Google Scholar
A. Padole et al.,
“CT radiation dose and iterative reconstruction techniques,”
Am. J. Roentgenol., 204
(4), W384
–W392
(2015). http://dx.doi.org/10.2214/AJR.14.13241 AJROAM 0092-5381 Google Scholar
D. E. Litmanovich et al.,
“Dose reduction in cardiothoracic CT: review of currently available methods,”
RadioGraphics, 34
(6), 1469
–1489
(2014). http://dx.doi.org/10.1148/rg.346140084 Google Scholar
S. Oda et al.,
“A low tube voltage technique reduces the radiation dose at retrospective ECG-gated cardiac computed tomography for anatomical and functional analyses,”
Acad. Radiol., 18
(8), 991
–999
(2011). http://dx.doi.org/10.1016/j.acra.2011.03.007 Google Scholar
F. Cademartiri and P. Pavone,
“Advantages of retrospective ECG-gating in cardio-thoracic imaging with 16-row multislice computed tomography,”
Acta Bio-Med., 74
(3), 126
–130
(2003). Google Scholar
T. Techasith et al.,
“The effect of heart rhythm on patient radiation dose with dual-source cardiac computed tomography,”
J. Cardiovasc. Comput. Tomogr., 5
(4), 255
–263
(2011). http://dx.doi.org/10.1016/j.jcct.2011.05.003 Google Scholar
A. M. Lee et al.,
“Assessment of image quality and radiation dose of prospectively ECG-triggered adaptive dual-source coronary computed tomography angiography (cCTA) with arrhythmia rejection algorithm in systole versus diastole: a retrospective cohort study,”
Int. J. Cardiovasc. Imaging, 29
(6), 1361
–1370
(2013). http://dx.doi.org/10.1007/s10554-013-0208-8 Google Scholar
A. M. Lee et al.,
“Coronary computed tomography angiography during arrhythmia: radiation dose reduction with prospectively ECG-triggered axial and retrospectively ECG-gated helical 128-slice dual-source CT,”
J. Cardiovasc. Comput. Tomogr., 6
(3), 172
–183
(2012). http://dx.doi.org/10.1016/j.jcct.2012.04.003 Google Scholar
S. S. Halliburton et al.,
“SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT,”
J. Cardiovasc. Comput. Tomogr., 5
(4), 198
–224
(2011). http://dx.doi.org/10.1016/j.jcct.2011.06.001 Google Scholar
S. Leschka et al.,
“Scan length adjustment of CT coronary angiography using the calcium scoring scan: effect on radiation dose,”
Am. J. Roentgenol., 194
(3), W272
–W277
(2010). http://dx.doi.org/10.2214/AJR.09.2970 AJROAM 0092-5381 Google Scholar
M. K. Kalra et al.,
“Techniques and applications of automatic tube current modulation for CT,”
Radiology, 233
(3), 649
–657
(2004). http://dx.doi.org/10.1148/radiol.2333031150 RADLAX 0033-8419 Google Scholar
T. Maruyama et al.,
“Radiation dose reduction and coronary assessability of prospective electrocardiogram-gated computed tomography coronary angiography: comparison with retrospective electrocardiogram-gated helical scan,”
J. Am. Coll. Cardiol., 52
(18), 1450
–1455
(2008). http://dx.doi.org/10.1016/j.jacc.2008.07.048 JACCDI 0735-1097 Google Scholar
J. Menke et al.,
“Head-to-head comparison of prospectively triggered vs retrospectively gated coronary computed tomography angiography: meta-analysis of diagnostic accuracy, image quality, and radiation dose,”
Am. Heart J., 165
(2), 154
–163
(2013). http://dx.doi.org/10.1016/j.ahj.2012.10.026 Google Scholar
T. M. Labounty et al.,
“Effect of padding duration on radiation dose and image interpretation in prospectively ECG-triggered coronary CT angiography,”
Am. J. Roentgenol., 194
(4), 933
–937
(2010). http://dx.doi.org/10.2214/AJR.09.3371 AJROAM 0092-5381 Google Scholar
J. Hausleiter et al.,
“Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates,”
Circulation, 113
(10), 1305
–1310
(2006). http://dx.doi.org/10.1161/CIRCULATIONAHA.105.602490 CIRCAZ 0009-7322 Google Scholar
B. Bischoff et al.,
“Impact of a reduced tube voltage on CT angiography and radiation dose: results of the PROTECTION I study,”
JACC Cardiovasc. Imaging, 2
(8), 940
–946
(2009). http://dx.doi.org/10.1016/j.jcmg.2009.02.015 Google Scholar
W.-H. Yin et al.,
“Effect of reduced x-ray tube voltage, low iodine concentration contrast medium, and sinogram-affirmed iterative reconstruction on image quality and radiation dose at coronary CT angiography: results of the prospective multicenter REALISE trial,”
J. Cardiovasc. Comput. Tomogr., 9
(3), 215
–224
(2015). http://dx.doi.org/10.1016/j.jcct.2015.01.010 Google Scholar
T. A. Fuchs et al.,
“Coronary computed tomography angiography with model-based iterative reconstruction using a radiation exposure similar to chest X-ray examination,”
Eur. Heart J., 35
(17), 1131
–1136
(2014). http://dx.doi.org/10.1093/eurheartj/ehu053 EHJODF 0195-668X Google Scholar
A. Schuhbaeck et al.,
“Image quality of ultra-low radiation exposure coronary CT angiography with an effective dose <0.1 mSv using high-pitch spiral acquisition and raw data-based iterative reconstruction,”
Eur. Radiol., 23
(3), 597
–606
(2013). http://dx.doi.org/10.1007/s00330-012-2656-2 Google Scholar
Y. Iyama et al.,
“Cardiac helical CT involving a low-radiation-dose protocol with a 100-kVp setting: usefulness of hybrid iterative reconstruction and display preset optimization,”
Medicine, 95
(46), e5459
(2016). http://dx.doi.org/10.1097/MD.0000000000005459 MEDIAV 0025-7974 Google Scholar
D. Andreini et al.,
“Coronary CT angiography with 80 kV tube voltage and low iodine concentration contrast agent in patients with low body weight,”
J. Cardiovasc. Comput. Tomogr., 10
(4), 322
–326
(2016). http://dx.doi.org/10.1016/j.jcct.2016.06.003 Google Scholar
M. Meyer et al.,
“Closing in on the K edge: coronary CT angiography at 100, 80, and 70 kV-initial comparison of a second- versus a third-generation dual-source CT system,”
Radiology, 273
(2), 373
–382
(2014). http://dx.doi.org/10.1148/radiol.14140244 RADLAX 0033-8419 Google Scholar
P. M. Linsen et al.,
“Computed tomography angiography with a 192-slice dual-source computed tomography system: improvements in image quality and radiation dose,”
J. Clin. Imaging Sci., 6
(1), 44
(2016). http://dx.doi.org/10.4103/2156-7514.192840 Google Scholar
S. Mangold et al.,
“Coronary CT angiography in obese patients using 3rd generation dual-source CT: effect of body mass index on image quality,”
Eur. Radiol., 26
(9), 2937
–2946
(2016). http://dx.doi.org/10.1007/s00330-015-4161-x Google Scholar
L. J. Zhang et al.,
“Image quality, radiation dose, and diagnostic accuracy of prospectively ECG-triggered high-pitch coronary CT angiography at 70 kVp in a clinical setting: comparison with invasive coronary angiography,”
Eur. Radiol., 26
(3), 797
–806
(2016). http://dx.doi.org/10.1007/s00330-015-3868-z Google Scholar
M. A. Bolen et al.,
“Image quality, contrast enhancement, and radiation dose of ECG-triggered high-pitch CT versus non–ECG-triggered standard-pitch CT of the thoracoabdominal aorta,”
Am. J. Roentgenol., 198
(4), 931
–938
(2012). http://dx.doi.org/10.2214/AJR.11.6921 AJROAM 0092-5381 Google Scholar
T. G. Flohr et al.,
“First performance evaluation of a dual-source CT (DSCT) system,”
Eur. Radiol., 16
(2), 256
–268
(2006). http://dx.doi.org/10.1007/s00330-005-2919-2 Google Scholar
S. Achenbach et al.,
“Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition,”
Eur. Heart J., 31
(3), 340
–346
(2010). http://dx.doi.org/10.1093/eurheartj/ehp470 EHJODF 0195-668X Google Scholar
B. B. Ghoshhajra et al.,
“Radiation dose reduction in pediatric cardiac computed tomography: experience from a Tertiary Medical Center,”
Pediatr. Cardiol., 35
(1), 171
–179
(2014). http://dx.doi.org/10.1007/s00246-013-0758-5 PECAD4 1432-1971 Google Scholar
T. Selcuk et al.,
“Effectiveness of using dual-source CT and the upshot it creates on both heart rate and image quality,”
Balk. Med. J., 33
(3), 283
–293
(2016). http://dx.doi.org/10.5152/balkanmedj.2016.16220 Google Scholar
P. Kröpil et al.,
“Prospectively ECG-triggered high-pitch spiral acquisition for cardiac CT angiography in routine clinical practice: initial results,”
J. Thorac. Imaging, 27
(3), 194
–201
(2012). http://dx.doi.org/10.1097/RTI.0b013e318228223f JTIME8 0883-5993 Google Scholar
L. Fu et al.,
“Comparison between pre-log and post-log statistical models in ultra-low-dose CT reconstruction,”
IEEE Trans. Med. Imaging, 36 707
–720
(2016). http://dx.doi.org/10.1109/TMI.2016.2627004 ITMID4 0278-0062 Google Scholar
C. Naoum, P. Blanke and J. Leipsic,
“Iterative reconstruction in cardiac CT,”
J. Cardiovasc. Comput. Tomogr., 9
(4), 255
–263
(2015). http://dx.doi.org/10.1016/j.jcct.2015.04.004 Google Scholar
W.-H. Yin et al.,
“Iterative reconstruction to preserve image quality and diagnostic accuracy at reduced radiation dose in coronary CT angiography,”
JACC Cardiovasc. Imaging, 6
(12), 1239
–1249
(2013). http://dx.doi.org/10.1016/j.jcmg.2013.08.008 Google Scholar
L. Yang et al.,
“Optimization of hybrid iterative reconstruction level and evaluation of image quality and radiation dose for pediatric cardiac computed tomography angiography,”
Pediatr. Radiol., 47
(1), 31
–38
(2017). http://dx.doi.org/10.1007/s00247-016-3698-4 Google Scholar
P. Kröpil et al.,
“Impact of increasing levels of advanced iterative reconstruction on image quality in low-dose cardiac CT angiography,”
ROFO Fortschr. Geb. Rontgenstr. Nuklearmed., 186
(6), 567
–575
(2014). http://dx.doi.org/10.1055/s-0033-1356074 Google Scholar
L. L. Geyer et al.,
“State of the art: iterative CT reconstruction techniques,”
Radiology, 276
(2), 339
–357
(2015). http://dx.doi.org/10.1148/radiol.2015132766 RADLAX 0033-8419 Google Scholar
A. J. Einstein et al.,
“Radiation dose from single-heartbeat coronary CT angiography performed with a 320-detector row volume scanner,”
Radiology, 254
(3), 698
–706
(2010). http://dx.doi.org/10.1148/radiol.09090779 RADLAX 0033-8419 Google Scholar
F. Khosa et al.,
“Influence of image acquisition on radiation dose and image quality: full versus narrow phase window acquisition using 320 MDCT,”
Sci. World J., 2013 e731590
(2013). http://dx.doi.org/10.1155/2013/731590 Google Scholar
A. Prakash, A. J. Powell and T. Geva,
“Multimodality noninvasive imaging for assessment of congenital heart disease,”
Circ. Cardiovasc. Imaging, 3
(1), 112
–125
(2010). http://dx.doi.org/10.1161/CIRCIMAGING.109.875021 Google Scholar
M. Koplay et al.,
“Prospective ECG-gated high-pitch dual-source cardiac CT angiography in the diagnosis of congenital cardiovascular abnormalities: radiation dose and diagnostic efficacy in a pediatric population,”
Diagn. Interventional Imaging, 97
(11), 1141
–1150
(2016). http://dx.doi.org/10.1016/j.diii.2016.03.014 Google Scholar
W. Gao et al.,
“Diagnostic accuracy of sub-mSv prospective ECG-triggering cardiac CT in young infant with complex congenital heart disease,”
Int. J. Cardiovasc. Imaging, 32
(6), 991
–998
(2016). http://dx.doi.org/10.1007/s10554-016-0854-8 Google Scholar
dsct.editors@spiritlink.de, “Improvements in third generation dual-source CT and benefits for thoracic and cardiac CT in adult and pediatrics,”
(2017) http://www.dsct.com/index.php/improvements-in-third-generation-dual-source-ct-and-benefits-for-thoracic-and-cardiac-ct-in-adult-and-pediatrics/ February ). 2017). Google Scholar
D. Andreini et al.,
“Low-dose CT coronary angiography with a novel IntraCycle motion-correction algorithm in patients with high heart rate or heart rate variability,”
Eur. Heart J., 16
(10), 1093
–1100
(2015). http://dx.doi.org/10.1093/ehjci/jev033 CECED9 Google Scholar
H. Vadvala et al.,
“Coronary CTA using scout-based automated tube potential and current selection algorithm, with breast displacement results in lower radiation exposure in females compared to males,”
Cardiovasc. Diagn. Ther., 4
(6), 470
–479
(2014). http://dx.doi.org/10.3978/j.issn.2223-3652.2014.12.07 Google Scholar
A. J. Einstein et al.,
“Effect of bismuth breast shielding on radiation dose and image quality in coronary CT angiography,”
J. Nucl. Cardiol., 19
(1), 100
–108
(2012). http://dx.doi.org/10.1007/s12350-011-9473-x JNCAE2 1071-3581 Google Scholar
P. M. Colletti, O. A. Micheli and K. H. Lee,
“To shield or not to shield: application of bismuth breast shields,”
Am. J. Roentgenol., 200
(3), 503
–507
(2013). http://dx.doi.org/10.2214/AJR.12.9997 AJROAM 0092-5381 Google Scholar
B. Bischoff et al.,
“Myocardial ischemia detection with single-phase CT perfusion in symptomatic patients using high-pitch helical image acquisition technique,”
Int. J. Cardiovasc. Imaging, 33
(4), 569
–576
(2016). http://dx.doi.org/10.1007/s10554-016-1020-z Google Scholar
A. So et al.,
“Technical note: evaluation of a 160-mm/256-row CT scanner for whole-heart quantitative myocardial perfusion imaging,”
Med. Phys., 43
(8), 4821
–4832
(2016). http://dx.doi.org/10.1118/1.4957389 Google Scholar
I. Danad, B. Ó. Hartaigh and J. K. Min,
“Dual-energy computed tomography for detection of coronary artery disease,”
Expert Rev. Cardiovasc. Ther., 13
(12), 1345
–1356
(2015). http://dx.doi.org/10.1586/14779072.2015.1102055 Google Scholar
V. A. Murugan et al.,
“American College of Radiology Dose Index Registry: a user’s guide for cardiothoracic radiologists part 1,”
J. Thorac. Imaging, 30
(6), W66
–W68
(2015). http://dx.doi.org/10.1097/RTI.0000000000000184 Google Scholar
|