1 August 2003 Axicon as a retroreflector in open-path Fourier transform infrared spectrometry
Robert L. Richardson, Peter R. Griffiths
Author Affiliations +
Ray-tracing analysis of a particular reflecting axicon, a right circular cone, for use as a retroreflector in active open-path Fourier-transform IR (OP/FT-IR) spectrometry, and the results of testing a 0.305-m aperture right circular cone using a commercial active OP/FT-IR spectrometer are presented. The ray-tracing model is based on the optical characteristics of a commercial single-telescope monostatic OP/FT-IR spectrometer and models the off-axis behavior normally encountered under practical field conditions during field use. Of practical concern are the trends for the diameter of the beam reflected from the retroreflector as a function of the path length between spectrometer and retroreflector and misalignment of the retroreflector with respect to the transmitted beam. Construction and the results of field-testing two smaller epoxy-composite axicons replicated from the 0.305-m master are presented.
©(2003) Society of Photo-Optical Instrumentation Engineers (SPIE)
Robert L. Richardson and Peter R. Griffiths "Axicon as a retroreflector in open-path Fourier transform infrared spectrometry," Optical Engineering 42(8), (1 August 2003). https://doi.org/10.1117/1.1589756
Published: 1 August 2003
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Axicons

Retroreflectors

Spectroscopy

Sensors

Telescopes

Epoxies

Infrared spectroscopy

RELATED CONTENT

MIRS: an imaging spectrometer for the MMX mission
Proceedings of SPIE (August 06 2021)
8 - 25 eV High Regolution Soler Spectrometer
Proceedings of SPIE (November 27 1989)
ISAAC a 1 to 5 um imager spectrometer...
Proceedings of SPIE (March 21 1997)
Cerro Tololo infrared spectrometer
Proceedings of SPIE (July 01 1990)

Back to Top