Paper
18 November 1989 CW Doppler Systems For External Urodynamic Study, Comparison Between Electromagnetic Wave And Airborne Ultrasound
Y. Takeuchi
Author Affiliations +
Proceedings Volume 1039, 13th Intl Conf on Infrared and Millimeter Waves; (1989) https://doi.org/10.1117/12.978502
Event: 13th International Conference on Infrared and Millimeter Waves, 1987, Honolulu, HI, United States
Abstract
Urination behavior indicates a series of important information for urologist Doctors, to find, for example, prostatic hypertrophy. External, non-contact urodynamic monitors using CW-Doppler principle were built in similar wavelengths with 24GHz electromagnetic wave (λ=12.5mm) and 40KHz airborne ultrasound (λ8.5mm). Both. systems were found having almost equivalent excellence in detecting the Doppler return from. urinary flow simulated by water ejector. However, airborne ultrasound system. seems slightly less expensive although its electronics are more complicated.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Y. Takeuchi "CW Doppler Systems For External Urodynamic Study, Comparison Between Electromagnetic Wave And Airborne Ultrasound", Proc. SPIE 1039, 13th Intl Conf on Infrared and Millimeter Waves, (18 November 1989); https://doi.org/10.1117/12.978502
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.