1.INTRODUCTIONRadiation-induced effects on instrument detectors and other key instrument components are significant issues that ultimately impact the quality and quantity of the mission science return and the reliability of engineering sensor data critical to flight operations. A thorough assessment of radiation effects on sensors and detectors of science instruments is therefore needed to ensure data quality for any space mission. Space particle instruments (e.g., plasma and energetic particle sensors, neutral mass spectrometer, Energetic Neutral Atom imager) and remote-sensing spectrometers (e.g., UltraViolet spectrometer) include electron multiplier detectors that are subject to increased transient noise, long-term degradation, and even potential failure due to the substantial fluxes of high-energy particles that penetrate the instrument. The most commonly used electron multiplier detectors are Multi-Channel Plate (MCP, [1, 2]) and Channel Electron Multiplier (CEM). These detectors are sensitive not only to the incident energetic charged particles themselves but they are also sensitive to the final end-product energy deposited by energetic electrons, ions, and X-rays. The resulting radiation-induced background noise can potentially swamp the science signal [3]). This issue constitutes undoubtedly the main challenge for particle instruments onboard future missions to Jupiter like the European Space Agency Jupiter ICy moon Explorer (JUICE [4]), and requires dedicated and innovative radiation mitigation techniques (e.g., multiple coincidence, anti-coincidence) far beyond the simple passive shielding techniques commonly used to protect electronics and other subsystems against Total Ionizing Dose (TID). All plasma–oriented instrument on board the JUICE mission will be exposed to a very hard flow of high energy electrons that fill the Jovian magnetosphere. Since the plasma instruments are located outside the S/C body, it is very difficult to protect the instrument detectors against the penetrating electrons and gamma rays created by the incident electrons in instrument protecting shields. These electrons and gamma rays can create a background noise in the instrument detectors and, if the level of such a noise is rather high, contaminate and corrupt the scientific data. To mitigate such a problem, first of all, we need to know the sensitivity of the instrument detectors to the penetrating electrons and gamma particles. The accurate response (i.e., efficiency) of MCPs and CEMs detectors against high-energy particles is however not well known, with limited estimates available in the literature. This makes it complicated to reliably predict the Signal to Noise ratio of the instrument, and, hence, ensure that the instrument will return useful scientific data when operated in the Jovian magnetosphere. 2.EXPERIMENTAL SETUP2.1Experiment general plan and experiment limitationA strong flow of high-energy electrons in the energy range between 100keV and 100MeV exist in the Jovian environment. Any multilayer protection consisting of layers of Aluminium (Al) and Tantalum (Ta) can reduce the particle flux and modify the incident energy spectrum, but cannot really modify the energy range of the initial and secondary electrons bombarding the MCP or CEM detectors inside the instrument. Moreover we can expect many low-energy (<100 keV) secondary electrons in the vicinity of the MCP or CEM detectors and also photons of about 500 keV of energy. The last particles (gammas) are created in the Ta protection layer. The MCP efficiency for electrons less then 100 keV of energy is more or less known from the literature [2]. The MCP response to high-energy electrons are less known and can depend on the MCP glass, technology, etc. Thus an “ideal” experiment plan shall be as follows: 2.2The real experiment setup in ONERATo perform such measurements we have used the high-energy electron source GEODUR at ONERA, Toulouse, France. There are several limitations in the experiment realization:
2.3Detailed Experiment plan/procedure2.3.1Experiment outlineTo execute the experiment in the ONERA equipment we created a setup shown in the diagram in Figure 1. The original electron beam (on top) passes the attenuation foil and pseudo-uniformly irradiates a PCB with several detectors installed onto the PCB surface (see also Figure 5):
The Detectors PCB is located on the computer controlled turning stage (see also Figure 5). The preamplifiers of front-end electronics for MCPs and SSD are installed in the vicinity of the PCB inside the vacuum chamber (Figures 2 and 5). The other analog and digital electronics including:
The pictures of the vacuum chamber and the external equipment are shown in Figure 2. The MCP samples information is given in Annex 2. 2.3.2Experiment procedure (in general)
2.3.3Experiment detailsThe MCP2 protection shield design is shown in Figure 3. An example of the photon’s energy spectrum inside the box is shown in Figure 4. We can see that only 2% of the incident electrons generate the photons. The data are obtained by using the SPENVIS multilayer protection analysis tool. The real Detector PCB layout and the board installation in the vacuum chamber are shown in Figure 5. The beam intensity and the beam distribution are controlled also by the Al foil installed between the beam source and the detector PCB. The foil modifies the beam slightly. We uses the SSD to check the real incident beam, but it seems that the SSD spectrum sometimes is not absolutely correct due to a SSD saturation. Thus, to get the real incident beam we again rely on SPENVIS simulations. The example of the real beam spectrum is shown in Figure 6. We can see that the real peak intensity becomes smaller but the total beam flux is 12% greater then the original one because of the generation of low-energy secondary ions. We are not going to take into account this feature in the further data analysis. The additional details of the experiment setup are as follows:
As it was already mentioned, the MCP samples information is given in Annex 2. 3.EXPERIMENT PROGRESS AND EXPERIMENT DATAThe experiment has been performed on 17 – 21 November 2016 at ONERA. 3.1MCP Photonis France under the electron beamTable 1 shows the list of experiments with electrons (the red spectrum is the simulated one): Table 1
The experiment results are shown in Table 2. Table 2
We see that the PHD of the events generated by penetrating electron has an exponential shape. A high-energy electron can generate an initial secondary electron at any place inside the micro-channel plate, whereas the low energy particles trigger the secondary electrons at the entrance surface and cause a bell-shape PHD around about 107 of gain. We see that some high gain part of the distribution is contaminated by low-energy secondary ions events. We removed this part of the distribution from the efficiency calculations. 3.2MCP Photonis under the gamma photons beamThe Tables 3 shows the results of the experiment with the gamma particles (protected MCP2). Table 3
4.EXPERIMENT RESULTS AND CONCLUSIONSFigure 7 shows the test result summary. The red stars show the MCP efficiency to the high-energy electrons, the blue triangles show the same for the gamma radiation. We put here also the results from the publication [3] shown as green circles. Conclusions
AcknowledgmentsThese experimental tests were funded by the french space agency CNES in support of the JUICE mission and in particular for the contribution of the Institut de Recherche en Astrophysique et Planétologie to the Particle Environment Package (PEP) through grants Nos 4500046142 and 4500055493. AppendicesANNEX 1:CEM EFFICIENCY TO PENETRATING RADIATIONThere are two type of detectors used in space plasma instruments: channeltrons (CEM) and microchannel plates (MCP). Some JUICE plasma sensor teams have chosen the ceramic CEMs since they a priory should be less effective to detect penetrating radiation. To obtain ceramic channeltrons (CCEM) efficiency for electrons and X-rays for the energies, specified above, we have tested in 2012 a board with CCEMS from Dr. Sjuts (www.sjuts.com/, Figure 8) in the ONERA GEODUR electron accelerator for energies 0.4 – 1.6 MeV, as well as in the NIST/GSFC (USA) for an energy of 13 MeV. Our test board contained 3 CCEMs. One of them exposed a side to the incident beam and two others pointed their apertures toward the beam. Two CCEMs have been covered by a dedicated shielding of 3 mm Aluminum plus 2 mm Tantalum. During ONERA (low energy) test the side CCEMs was open and during GSFC test the aperture pointed CCEM was open for the incident beam. It each case the absolute value mono-energetic beam flux was known. The “effective aperture” of the CCEM should be a result of the test. The “effective aperture” can be different for the side and the aperture pointed CCEMs. The aperture is calculated as The results of the experiments are summarized in the right panel of Figure 8. The stars correspond to the direct electron beam striking to a “side” CCEM in the ONERA accelerator. The blue rectangle corresponds to the direct 13 MeV electron beam striking to the CCEM aperture (GSFC). The difference between GSFC and ONERA data are very small. The solid blue curve reproduces a possible profile of the “aperture pointed” CCEM effective area for high-energy electrons. The dashed line reproduces the same for the “side “ CCEM. The “side” curve start to deviate from the “aperture” one for low energies and quickly going down for electrons which cannot penetrate into the CCEM. GEANT-4 numerical simulations were used to estimate the electron and X-rays count rate of an idealized virtual omnidirectional detector of 1.5cm2 averaged surface and 100% efficiency, for an incident electron beam energy of 13 MeV. These simulations show that a CCEM under 3 mm Aluminum plus 2 mm Tantalum shield can detect only bremsstrahlung photons of 500 keV. Thus we can calculate the CCEM photon effective aperture for this important energy. The calculated value has been normalized to the ration between the physical and simulated detector effective area. The final “Photon effective aperture” is shown as a red rectangle in the right panel of Figure 8. One can see that the real CCEM aperture is very small. REFERENCESWiza, J.L.,
“Microchannel Plate Detectors,”
Nuclear Instruments and Methods, 162 587
–601
(1979). https://doi.org/10.1016/0029-554X(79)90734-1 Google Scholar
Blase, R.C., Benke, R.R., Cooke, C.M., Pickens, K.S.,
“Microchannel Plate Detector Detection Efficiency to Monoenergetic Electrons Between 0.4 and 2.6 MeV,”
IEEE Transactions on nuclear science, 62
(6),
(2015). https://doi.org/10.1109/TNS.2015.2488481 Google Scholar
Grasset, O.,
“JUpiter ICy moons Explorer (JUICE): An ESA mission to orbit Ganymede and to characterise the Jupiter system,”
Planetary and Space Science, 78 1
–21
(2013). https://doi.org/10.1016/j.pss.2012.12.002 Google Scholar
|