PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.
To evaluate the clinical utility of Visible and near infrared optical coherence tomography (vnOCT) for glaucoma early detection, A total of 55 eyes from three groups of subjects (normal subjects, glaucoma suspects, glaucoma patients) were scanned by a custom-designed vnOCT device and Zeiss Spectralis OCT. The peripapillary retinal nerve fiber layer reflectivity in visible light OCT and the ratio between visible and NIR channel is more sensitive in separating suspect eyes from normal ones than clinical OCT thickness measurements. It could be a useful metric in early detection of glaucoma upon further longitudinally validation.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.
The alert did not successfully save. Please try again later.
Weiye Song, Sui Zhang, Natalie Sadlak, Marissa G. Fiorello, Manishi Desai, Ji Yi, "Visible and near-infrared optical coherence tomography (vnOCT) in normal, suspect, and glaucomatous eyes," Proc. SPIE 11623, Ophthalmic Technologies XXXI, 1162311 (5 March 2021); https://doi.org/10.1117/12.2578925