27 December 1977 Data Processing Of Dual-Energy Scans
Rodney A. Brooks
Author Affiliations +
Proceedings Volume 0127, Application of Optical Instrumentation in Medicine VI; (1977) https://doi.org/10.1117/12.955956
Event: Application of Optical Instrumentation in Medicine VI, 1977, Boston, United States
Abstract
A method for analyzing dual-energy computed tomograms is presented. The method requires three calibration values at each energy: air, water, and a KI solution. The first two values are used to convert the CT numbers to the ideal Hounsfield scale. The KI value is used to calculate quality factor Q that describes the effective x-ray spectrum. By using the Q-values, the Compton and photoelectric coefficients of any tissue or material may be easily obtained. One can further calculate the electron density and effective atomic number, if desired, and perform chemical identification.
© (1977) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Rodney A. Brooks "Data Processing Of Dual-Energy Scans", Proc. SPIE 0127, Application of Optical Instrumentation in Medicine VI, (27 December 1977); https://doi.org/10.1117/12.955956
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Signal attenuation

Calibration

Chemical elements

Scanners

Medicine

Optical instrument design

Scattering

RELATED CONTENT

Reproducibility in dual energy CT the impact of a...
Proceedings of SPIE (October 17 2022)
Sensitometry In Computerized Tomography
Proceedings of SPIE (December 27 1977)
The Role Of CT Scanners In Radiotherapy
Proceedings of SPIE (December 23 1976)
A Comparison Of Some Models Of X-Ray Beam Hardening
Proceedings of SPIE (December 23 1976)

Back to Top