1.INTRODUCTIONAriel is the fourth medium-class mission under development in the framework of ESA “Cosmic Vision” Program. It was adopted in 2020 and launch is planned for 2029. During its 4-year of nominal mission duration, Ariel will conduct a survey of known exoplanets to characterize their atmospheres through transit spectroscopy and photometry in the waveband between 0.5 μm and 7.8 μm.1 The Ariel telescope is based on an off-axis, unobscured Cassegrain design with an elliptical primary mirror with an aperture of 1100mm (major axis) and 768mm (minor axis), corresponding to a light collecting area of approximately 0.6m2. The telescope was designed to be diffraction limited at the wavelength of 3 μm on a 30″ field of view.2 The required average telescope throughput is 96 %.3 Telescope and instruments will operate at a temperature below 50K. Following the heritage of the JWST MIRI instrument,4 aluminum alloy 6061 in the T651 forge has been chosen for mirrors substrates and supporting structures of the telescope, after a trade-off study5 on manufacturability and cost. To comply with throughput requirements, particularly in the visible portion of the operating waveband, and to protect the aluminum substrate from oxidation, the Consortium decided to apply a protected silver coating to the telescope mirrors. The coating, from CILAS* is qualified for space, but needed to be subjected to additional qualification tests to assess performance at the Ariel telescope operating temperature of <50K and because of the large size of the primary mirror, raising possible concerns on the uniformity of the deposition process and stability of the coating. An initial study was therefore devised to test optical performance and durability on Al6061-T651 substrates, consisting of a qualification campaign on coated aluminum samples,6,7 and a verification test on additional samples and on a full-scale demonstrator of the Ariel primary mirror denoted Pathfinder Telescope Mirror (PTM).8 After the successful completion of the qualification, the samples have been kept in storage in an cleanroom facility (ISO6), and are being periodically re-examined to detect signs of functional (reflectance) or visual deterioration due to oxidation or delamination. Silver coatings, although protected by a capping layer, Can be sensitive to damaging from contact with humidity, sulfur and chlorine pollutants, normally present even in the controlled atmosphere of a cleanroom.9 It is therefore necessary to ensure that the coating will not deteriorate in the time span from deposition on the Ariel telescope mirrors to launch. 2.MATERIALS AND PROCESSES2.1Samples DescriptionThe coating qualification campaign was performed on several samples of Al 6061-T651 in rolled plate form, the same aluminum alloy and forge currently foreseen for Ariel Telescope mirrors and supporting structure. The samples are shaped as discs of 25mm of diameter and 6mm thick (Figure 1 shows one of the samples being held for visual inspection before the coating run). The samples were procured, polished and cleaned by MediaLario† before delivery to CILAS for coating. Roughness of the optical surface was measured with a Taylor Hobson CCI White Light Interferometer on areas measuring 1.5mm × 1.5mm and 300 μm × 300 μm. All samples were within the 10 nm RMS specification. The qualification campaign (in fact, a delta-qualification), consisted in a series of environmental and mechanical resistance tests to ensure that the coating will reach the end of life of the instrument without significant performance degradation, following the European Cooperation for Space Standardization (ECSS) Q-ST-70-17C standard.10 A brief description of the set of tests is reported in Table 1. The qualification program was overall successful.6 Table 1.Summary of specifications of the coating qualification tests performed on the samples.
After the qualification campaing, a subset of the samples was retrieved by the authors and stored in airtight containers in a cleanroom environment (ISO6). The risk of exposure to excessive humidity is further minimized by placing silica gel dessicant bags inside the containers. The samples are being re-examined periodically in a normal laboratory environment. Table 2 identifies the samples being monitored, coming from two subsequent coating deposition runs: a first test run, performed on April 3rd, 2019, and the actual qualification run, on December 12th, 2019. Both runs were conducted with the nominal coating procedure and produced equivalent results. The set of treatments to which each sample was subjected during the qualification is also reported in the table. Table 2.List of samples under test, with coating deposition date and qualification tests performed.
All verification and measurements described in this paper have been performed on all samples listed Table 2, but for the reminder of the treatment we will focus on Samples SN01 and SN12, that have been in storage for a longer period, Sample SN06M since it was already examined in details in a previous work7 and Sample SN08M because it was subjected to the entire set of qualification tests. 2.2Coating ProcessThe coating process employed by CILAS for the samples is based on physical vapor deposition. The coating platform consists in a large magnetron sputtering chamber. Samples are inserted on a tray sliding beneath a set of cathodes.11 The process is suited to optical substrates up to 2m by 2m of footprint and 0.4m of thickness.12 The protected silver coating described in this paper is on average 350 nm thick, with a thickness uniformity measured at 10 %. The structure of the stack is illustrated in Figure 2 and consists of at least three layers: a NiCr adhesion layer of less than 10 nm of thickness, the reflecting silver layer, and a dielectric capping and protection layer. An additional intermediate adhesion layer may be present between the silver and the capping layer. The actual layers thicknesses and composition of the coating cannot be disclosed due to business confidentiality. 3.VERIFICATION METHODSThe ECSS Q-ST-70-17C standard10 defines a set of verification and acceptance criteria to be performed on coatings to assess their performance and compliance with requirements. More specifically, after each test, the coating shall present:
The following paragraphs describe the assessment methods employed to verify compliance of the samples under test according to the list above, except for item 3 (thickness) since it was already verified during qualification and it is not expected to be affected by storage. 3.1Relative Reflectance MeasurementsImmediately after coating deposition, the reflectance of all samples was measured by the manufacturer with a Perkin-Elmer Lambda 950 spectrophotometer with a reflectometry accessory, in the waveband 500 nm to 2500 nm. Accuracy, as reported in the instrument datasheet, is ±0.6% from 500 nm to 890 nm and ±1 % above 890 nm. Subsequent relative reflectance measurements were taken at the Institute for Photonics and Nanotechnologies of the National Research Council (CNR–IFN) in Padova with a custom built setup. Reproducibility of the measurements in the worst case has been determined to be better than ±1.2% over the 500 nm to 900 nm wavelength range, and better than ±3.8% over the 400 nm to 1000 nm wavelength range. In order to cross-calibrate the two setups, which appear to have a measurement bias between each other (see Figure 5 in the Results section), we used one of the earliest measurement of Sample SN01 as reference point, since it presented the shortest temporal gap between measurements. 3.2Visual InspectionThe ECSS standard suggests following Annex C of ISO 9211-4:2022 for Visual Inspection (VI), that mandates the use of two cool white 15W lamps positioned directly above the sample, and to look at the sample against a black matte background at a distance of ≤45cm and at a near grazing angle.13 The use of optical micrographs is suggested only in case of suspected degradation, to further qualify it. In our case it was employed to look for signs of oxidation, as explained in Section 4.1. Darkfield imaging using a compact digital camera (Canon IXUS 220 HS) and a custom built LED lighting setup was also employed to highlight the presence of light scattering defects. 3.3Atomic Force MicroscopyBesides optical imaging techniques, we employed an Atomic Force Microscope (AFM) for a qualitative analysis of surface topography and to measure surface roughness. AFM scans were taken with a Park System‡ XE-Series 70 microscope in non-contact mode and processed with the Gwyddion open source software14 (the processing pipeline consisted in removal of low spacial frequencies by fitting and subtraction of an x, y third order polynomial surface, rows alignment using “median” as statistic and “scars removal”). 4.RESULTS4.1Visual InspectionThe optical surface of all samples appeared visually unaltered after the storage period, as exemplified in Figure 3: no discernible signs of coating degradation, such as cracks, blistering or change in color/iridescence, nor other visible signs of delamination could be spotted. In particular, we could not detect signs of oxidation developing from surface grains, as described for example by Folgner et al. in their studies of protected silver coatings exposed to mixed flowing gas.9,15 Apart from scratches and occasional dust particles, the most prominent features on the optical surface continued to be the glue residues from adhesion tests, especially evident on sample SN01. It is worth noting that the uncoated back and lateral surfaces of the samples do exhibit a slight brownish coloration and faint whitish areas, compatible with the oxidation of bare aluminum, so the level of exposure to pollutants seems to be at least sufficient to cause slight degradation of the substrate. Darkfield photographs also show a mostly uniformly dark optical surface, indicative of low scattering (Figure 4). 4.2Reflectance MeasurementsFigure 5 shows the results of the reflectance measurements performed on the samples at different points in time. Samples reflectance was measured after coating and after qualification tests by the manufacturer (solid gray line), who found no degradation in performance.6 One of the samples (SN01) was also measured by the authors early on during the qualification campaign, so this measurement could be used as reference for the cross-calibration of the two setups (dashed gray line). Eventually two comprehensive measurement campaigns could be performed six months apart (in October 2021 and April 2022, orange and blue lines in the plots). Considering the difference between the initial measurement of the in-house setup and the manufacturer’s setup, and noting the estimated repeatability error, the results do not highlight a significant change in reflectance of the samples. More precise measurements may be performed in the future to further confirm this preliminary result. 4.3Atomic Force MicroscopyTwo sets of AFM scans of representative areas of samples SN01 and SN06M are presented in Figures 6 and 7. For each set, the first scan (on top) was performed at the beginning of the storage period, after the sample underwent the qualification tests, and the second one (at the bottom), in August 2022. As discussed in Section 3.3, the scans were performed to provide a qualitative assessment of surface morphology variations and to measure surface roughness. Since the time in storage did not produce any visible signs of surface degradation nor delamination, the sampling location for the AFM was chosen to be reasonably representative of the central area, where reflectance was also measured, without aiming at specific surface blemishes or scratches. Please also note that measurements of the same sample do not image the exact same portion of surface. A comparison of the AFM images does not indicate the appearance of new topological structures of relevance: most features are attributable to scratches and dents that were equally present before storage. The white areas in relief on SN01 are likely the residues of tape adhesive from the adhesion tests. These were in fact visible upon careful examination of the surface area affected by the test. Anecdotally, AFM scans of sample SN01 did result in frequent tip pollution that required replacement, possibly because of the residues. RMS roughness measurements also do not appear to change significantly before and after the cycles. 5.CONCLUSIONS AND NEXT STEPSIn the framework of the coating qualification program for the mirrors of the Ariel mission telescope, a series of samples of the mirrors substrate material, Al6061-T651, were tested and are currently being kept in storage and re-examined periodically for signs of degradation. After three years, results of visual inspections (both with direct and darkfield illumination) and reflectance measurements showed no alteration in appearance imputable to deterioration or delamination of the coating, nor a significant degradation in optical performance in the waveband 400 nm to 1000 nm, according to preliminary measurements. Additional AFM scans of the samples showed no qualitative morphology variations nor an increase in surface roughness. Further testing will be repeated in the future to confirm coating stability under cleanroom environmental conditions, equivalent or worse to those foreseen for Ariel telescope mirrors. Additional operational environment tests are also planned for the near future, in particular radiation testing with an ion bombardment program simulating Ariel L2 operating orbital environment. ACKNOWLEDGMENTSThis activity has been realized under the Implementation Agreement n. 2021-5-HH.0 of the Italian Space Agency (ASI) and the National Institute for Astrophysics (INAF) Framework Agreement “Italian Participation to Ariel mission phase B2/C” and was partly funded under the ESA contract with Centre Spatial de Liège, Belgium (CSL) and INAF n. 4000126124/18/NL/BW. REFERENCESTinetti, G. et al.,
“A chemical survey of exoplanets with ARIEL,”
Experimental Astronomy, 46 135
–209
(2018). https://doi.org/10.1007/s10686-018-9598-x Google Scholar
Da Deppo, V., Middleton, K., Focardi, M., Morgante, G., Claudi, R., Pace, E., and Micela, G.,
“The optical configuration of the telescope for the ARIEL ESA missionOptical, Infrared, and Millimeter Wave,”
[Space Telescopes and Instrumentation, 161 SPIE, AustinUnited States
(20182018). Google Scholar
Chioetto, P., Pascale, E., Zuppella, P., Pace, E., Tozzi, A., Malaguti, G., and Micela, G.,
“Initial estimation of the effects of coating dishomogeneities, surface roughness and contamination on the mirrors of Ariel mission telescopeOptical Design and Engineering VIII,”
46 SPIE, Spain
(2021). Google Scholar
Kroes, G., Oudenhuysen, A., Meijers, M., and Pel, J.-W.,
“MIRI-JWST spectrometer main optics optomechanical design and prototyping,”
in in [Proc. SPIE 5877, Optomechanics 2005,
58770
(2005). Google Scholar
Da Deppo, V., Morgante, G., Pace, E., Middleton, M., and Eccleston, P.,
“ARIEL Telescope Material Trade-Off,”
Technical Paper ARIEL-INAF-PL-TN-004,
(2017). Google Scholar
Chioetto, P., Zuppella, P., Da Deppo, V., Pace, E., Morgante, G., Terenzi, L., Brienza, D., Missaglia, N., Bianucci, G., Spinelli, S., Guerriero, E., Rossi, M., Grèzes-Besset, C., Bondet, C., Chauveau, G., Porta, C., Malaguti, G., Micela, G., the ARIEL Team,
“Qualification of the thermal stabilization, polishing and coating procedures for the aluminum telescope mirrors of the ARIEL mission,”
Experimental Astronomy,
(2022). https://doi.org/10.1007/s10686-022-09852-x Google Scholar
Chioetto, P., Zuppella, P., Da Deppo, V., Pace, E., Morgante, G., Terenzi, L., Brienza, D., Missaglia, N., Bianucci, G., Spinelli, S. M., Guerriero, E., Rossi, M., Bondet, C., Chauveau, G., Porta, C., Grezes-Besset, C., Malaguti, G., and Micela, G.,
“The primary mirror of the Ariel mission: Cryotesting of aluminum mirror samples with protected silver coating,”
in in [Proc. SPIE 11451, Advances in Optical and Mechanical Technologies for Telescopes and Instrumentation IV],SPIE,
114511A
(2020). Google Scholar
Chioetto, P., Zupella, P., Da Deppo, V., Nordera, S., Pace, E., Tozzi, A., Morgante, G., Terenzi, L., Brienza, D., Missaglia, N., Bianucci, G., Spinelli, S., Guerriero, E., Rossi, M., Grisoni, G., Bondet, C., Chauveau, G., Porta, C., Grezes-Besset, C., Malaguti, G., and Micela, G.,
“Test of protected silver coating on aluminum samples of ARIEL main telescope mirror substrate material,”
in in [International Conference on Space Optics — ICSO 2020]SPIE,
165
(2021). Google Scholar
Folgner, K. A., Chu, C.-T., Lingley, Z. R., Kim, H. I., Yang, J.-M., and Barrie, J. D.,
“Environmental durability of protected silver mirrors prepared by plasma beam sputtering,”
Applied Optics, 56 C75
(2017). https://doi.org/10.1364/AO.56.000C75 Google Scholar
European Cooperation for Space Standardization,
“ECSS-Q-ST-70-17C – Durability Testing of Coatings,”
European Space Agency,
(20182018). Google Scholar
Savin de Larclause, I., Valette, N., Chauveau, G., Grèzes-Besset, C., Costes, V., Gasc, K., and Lemarquis, F.,
“PACA2m magnetron sputtering silver coating: A solution for very big mirror dimensions,”
in in [Proc. SPIE 10563, International Conference on Space Optics — ICSO 2014]SPIE,
1056308
(2018). Google Scholar
Grèzes-Besset, C., Valette, N., Chauveau, G., Castelnau, M., Mathieu, K., Tatat, M., and Lemarquis, F.,
“High performance silver coating with PACA2M magnetron sputtering,”
in in [Proc. SPIE 11180, International Conference on Space Optics — ICSO 2018],SPIE,
1118083
(2019). Google Scholar
International Organization for Standardization,
“[ISO 9211-4:2022 Optics and Photonics — Optical Coatings — Part 4: Specific Test Methods: Abrasion, Adhesion and Resistance to Water], International Organization for Standardization,”
Vernier, Geneva, Switzerland
(2022). Google Scholar
Nečas, D. and Klapetek, P.,
“Gwyddion: An open-source software for SPM data analysis,”
Open Physics, 10
(2012). Google Scholar
Folgner, K. A., Chu, C.-T., Sitzman, S. D., Stuart, S. C., Lingley, Z. R., and Barrie, J. D.,
“Development and growth of corrosion features on protected silver mirrors during accelerated environmental exposure,”
Applied Optics, 59 A187
(2020). https://doi.org/10.1364/AO.375891 Google Scholar
|