Paper
6 August 2023 Research on calibration technology of relaxation testing machine force value system
Author Affiliations +
Proceedings Volume 12781, International Conference on Optoelectronic Information and Functional Materials (OIFM 2023); 127810V (2023) https://doi.org/10.1117/12.2686698
Event: 2023 International Conference on Optoelectronic Information and Functional Materials (OIFM 2023), 2023, Guangzhou, JS, China
Abstract
Stress relaxation performance is an important performance indicator for evaluating elastic materials or components. The accuracy of relaxation testing machines directly determines the accuracy of the detection results of relaxation tests, which affects the effectiveness and reliability of the detection results. For a long time, the calibration work of the relaxation testing machine has not been carried out, and the testing and testing institutions cannot grasp the accuracy of the relaxation testing machine equipment itself. To achieve the calibration of the relaxation test force value system, the requirements for calibration parameters were clarified, and a calibration test for the relative error of force value indication was designed, and its uncertainty was evaluated.
© (2023) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Lu Liu, Jiacheng Cai, Lei Cheng, and Bing Zhang "Research on calibration technology of relaxation testing machine force value system", Proc. SPIE 12781, International Conference on Optoelectronic Information and Functional Materials (OIFM 2023), 127810V (6 August 2023); https://doi.org/10.1117/12.2686698
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Elasticity

Metals

Deformation

Measurement devices

Measurement uncertainty

Error analysis