Paper
1 July 1991 Space telescope imaging spectrograph 2048 CCD and its characteristics
Alan W. Delamere, Thomas H. Ebben, Debbie Murata-Seawalt, Morley M. Blouke, Richard Reed, Bruce E. Woodgate
Author Affiliations +
Abstract
Development of a 20482 CCD for a second-generation space telescope instrument has produced some very encouraging devices. The first experimental lot of 10 devices have very few defects, dark currents of less than 12 electrons/pixel/hour at -80 degree(s), readout noise levels of less than 4 electrons rms, and excellent charge transfer efficiency at signal levels of less than 10 electrons.
© (1991) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Alan W. Delamere, Thomas H. Ebben, Debbie Murata-Seawalt, Morley M. Blouke, Richard Reed, and Bruce E. Woodgate "Space telescope imaging spectrograph 2048 CCD and its characteristics", Proc. SPIE 1447, Charge-Coupled Devices and Solid State Optical Sensors II, (1 July 1991); https://doi.org/10.1117/12.45333