1 March 1992 Radiometric calibration and performance analysis of Fourier transform emission spectrometers
Daniel Lamarre, Jean Giroux, Henry L. Buijs
Author Affiliations +
Proceedings Volume 1575, 8th Intl Conf on Fourier Transform Spectroscopy; (1992) https://doi.org/10.1117/12.56316
Event: Eighth International Conference on Fourier Transform Spectroscopy, 1991, Lubeck-Travemunde, Germany
Abstract
FTIR instruments aimed at low temperature emission measurements require a two-point radiometric calibration because of significant emission from the instrument itself. Calibration measurements with two reference blackbodies are conducted for ground instruments. In space, the low temperature reference blackbody can be substituted with deep space observation. On the ground, pointing at a liquid nitrogen bath is used. Calibration operations involve complex arithmetics to obtain from reference and actual scene measurements a calibrated spectrum. Frequency and duration of calibration measurements are dependent upon instrument stability and noise budgets. When reference spectra are free of high resolution features, calibration measurements can be performed at a resolution lower than that of actual scene measurements. The Michelson interferometer for passive atmospheric sounding (MIPAS) will limb sound the atmosphere by measuring its mid-infrared emission with an apodized resolution of 0.05 cm-1. This instrument is used as an example to illustrate the trade-offs required in calibrating an emission FTIR. Given the current expected temperature variations of the instrument over the orbit, MIPAS will require a deep space reference measurement after every altitude scan, and a combined deep space and blackbody measurement at a lower frequency. For an adequate acquisition efficiency, low resolution calibration measurements are performed, at 0.2 cm-1 or 1 cm(superscript -1.
© (1992) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Daniel Lamarre, Jean Giroux, and Henry L. Buijs "Radiometric calibration and performance analysis of Fourier transform emission spectrometers", Proc. SPIE 1575, 8th Intl Conf on Fourier Transform Spectroscopy, (1 March 1992); https://doi.org/10.1117/12.56316
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Black bodies

Signal to noise ratio

Aerospace engineering

Spectral calibration

Spectrometers

Temperature metrology

Back to Top