Paper
25 July 1979 High Resolution Tv Fluoroscopy and Image Storing
J. Haendle, H. Sklebitz
Author Affiliations +
Proceedings Volume 0164, 4th European Electro-Optics Conf; (1979) https://doi.org/10.1117/12.965508
Event: Fourth European Electro-Optics Conference, 1978, Utrecht, Netherlands
Abstract
The investigation of the image-quality determining sections of the X-ray TV chain is an important matter for the optimization of the whole system. The modulation-transfer functions (MTF) of each component as the most important factor for the evaluation of the separate sections of the X-ray chain were measured. Special attention was given to gain exact measurements of MTF and other behavior of image pick-up tubes. In the second part of this paper image-quality determining factors and development tendencies of appropriate video stores for high resolution are discussed.
© (1979) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. Haendle and H. Sklebitz "High Resolution Tv Fluoroscopy and Image Storing", Proc. SPIE 0164, 4th European Electro-Optics Conf, (25 July 1979); https://doi.org/10.1117/12.965508
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Modulation

Video

X-rays

Modulation transfer functions

Image intensifiers

X-ray imaging

Image quality

Back to Top