22 October 1993 Testing of a scintillating fiber PET system
Gregory S. Nelson, Roy C. Chaney, Ervin J. Fenyves, H. Hammack, Cheri Hudson, Peter P. Antich, Jon A. Anderson
Author Affiliations +
Abstract
We are continuing the characterization of a PET module using plastic scintillating fibers coupled to position sensitive photomultipliers. The sensitivity in air and the resolution in air and water are tabulated for a source positioned at nine locations within the detector field of view. The full width half maximum resolution for a 0.5 mm diameter Na-22 source in air and water is approximately 2.5 m. The sensitivity is 19 c.p.s./(mu) Ci.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gregory S. Nelson, Roy C. Chaney, Ervin J. Fenyves, H. Hammack, Cheri Hudson, Peter P. Antich, and Jon A. Anderson "Testing of a scintillating fiber PET system", Proc. SPIE 2007, Scintillating Fiber Technology and Applications, (22 October 1993); https://doi.org/10.1117/12.161921