2 May 1997 DQE analysis for CCD imaging arrays
Author Affiliations +
Abstract
By consideration of the statistical interaction between exposure quanta and the mechanisms of image detection, the signal-to-noise limitations of a variety of image acquisition technologies are now well understood. However in spite of the growing fields of application for CCD imaging- arrays and the obvious advantages of their multi-level mode of quantum detection, only limited and largely empirical approaches have been made to quantify these advantages on an absolute basis. Here an extension is made of a previous model for noise-free sequential photon-counting to the more general case involving both count-noise and arbitrary separation functions between count levels. This allows a basic model to be developed for the DQE associated with devices which approximate to the CCD mode of operation, and conclusions to be made concerning the roles of the separation-function and count-noise in defining the departure from the ideal photon counter.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Rodney Shaw "DQE analysis for CCD imaging arrays", Proc. SPIE 3032, Medical Imaging 1997: Physics of Medical Imaging, (2 May 1997); https://doi.org/10.1117/12.273977