3 July 1998 Factors affecting the accurate determination of cerebrovascular blood flow using high-speed droplet imaging
Stephen Rudin, Afshin Divani, Ajay K. Wakhloo, Baruch B. Lieber, William Granger, Daniel R. Bednarek, Chang-Ying J. Yang
Author Affiliations +
Abstract
Detailed cerebrovascular blood flow can be more accurately determined radiographically from the new droplet tracking method previously introduced by the authors than from standard soluble contrast techniques. For example, arteriovenous malformation (AVM) transit times which are crucial for proper glue embolization treatments, were shown to be about half when using droplets compared to those measured using soluble contrast techniques. In this work, factors such as x-ray pulse duration, frame rate, system spatial resolution (focal spot size), droplet size, droplet and system contrast parameters, and system noise are considered in relation to their affect on the accurate determination of droplet location and velocity.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Stephen Rudin, Afshin Divani, Ajay K. Wakhloo, Baruch B. Lieber, William Granger, Daniel R. Bednarek, and Chang-Ying J. Yang "Factors affecting the accurate determination of cerebrovascular blood flow using high-speed droplet imaging", Proc. SPIE 3337, Medical Imaging 1998: Physiology and Function from Multidimensional Images, (3 July 1998); https://doi.org/10.1117/12.312568
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Spatial resolution

Blood circulation

Signal attenuation

X-rays

Lead

Imaging systems

Iodine

Back to Top