Paper
18 December 2000 Integrated CT-SPECT system for small-animal imaging
Mark Bennett Williams, Guimin Zhang, Mitali J. More, Allen R. Goode, Stan Majewski, Randy F. Wojcik, Brian Kross, Vladimir Popov, Andrew G. Weisenberger, Martin J. Stanton, Walter Charles Phillips, Alexander X. Stewart, Thomas G. McCauley, Tao Wu, Edward V.R. DiBella
Author Affiliations +
Abstract
We are developing a scanner for simultaneous acquisition of x-ray computed tomography (CT) and single photon emission tomography (SPECT) images of small animals such as mice and rats. The scanner uses a cone beam geometry for both the x- ray transmission and gamma emission projections by using an area x-ray detector and pinhole collimator, respectively. The CT and SPECT data set are overlaid to form a coregistered structural-functional 3D image. The CT system includes a single CCD-based x-ray detector and a microfocus x-ray source. The SPECT scanner utilizes tungsten pinhole collimators and arrays of CsI(Tl) scintillation detectors. We describe considerations and the early performance of a prototype scanner.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Mark Bennett Williams, Guimin Zhang, Mitali J. More, Allen R. Goode, Stan Majewski, Randy F. Wojcik, Brian Kross, Vladimir Popov, Andrew G. Weisenberger, Martin J. Stanton, Walter Charles Phillips, Alexander X. Stewart, Thomas G. McCauley, Tao Wu, and Edward V.R. DiBella "Integrated CT-SPECT system for small-animal imaging", Proc. SPIE 4142, Penetrating Radiation Systems and Applications II, (18 December 2000); https://doi.org/10.1117/12.410571
Lens.org Logo
CITATIONS
Cited by 24 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Single photon emission computed tomography

Spatial resolution

X-ray computed tomography

Crystals

X-rays

Scanners

RELATED CONTENT

Measurement of x ray spectra using a recent YAP(Ce) MPPC...
Proceedings of SPIE (September 07 2017)
Evaluation of x ray detectors for dual modality CT SPECT...
Proceedings of SPIE (November 26 2002)
Medical imaging by fluorescent x ray CT its preliminary...
Proceedings of SPIE (January 07 2002)
Design and evaluation of a prototype volume CT scanner
Proceedings of SPIE (April 20 2005)
Signal statistics in x-ray computed tomography
Proceedings of SPIE (May 03 2002)

Back to Top