28 June 2001 Comparison of a-Si:H CsI flat-panel digital imaging systems with a CCD-based system, CR systems, and conventional screen-film systems: a contrast-detail phantom study
John X. Rong, Chris C. Shaw, Dennis A. Johnston, Michael R. Lemacks, Xinming Liu, Gary J. Whitman, Stephen K. Thompson, Kerry T. Krugh
Author Affiliations +
Abstract
Low-contrast detail detectability was evaluated and compared for a flat-panel digital chest system, a computed radiography (CR) system and a conventional screen/film (SF) system. Images of a contrast-detail phantom were acquired using these three systems under identical conditions. Additional images were acquired at varied exposures to study the potential for reduction of patient exposure using the flat-panel system. The results demonstrated that in chest imaging, the flat-panel system performed significantly better than the CR and the SF systems while the latter two performed about the same. Alternatively, an exposure reduction of at least 50% is possible if the same performance is maintained. For mammographic imaging, detectability for microcalcifications ((mu) Cs) was evaluated and compared for a flat-panel based full-field digital mammography (FFDM) system, a charge-coupled device (CCD) -based small-field system, a high resolution CR system and a conventional SF system. Images of simulated calcifications of three size ranges were acquired and evaluated by readers for detectability of the (mu) Cs. A Receiver Operating Characteristics (ROC) analysis was also performed to compare the overall detection accuracy for these four mammographic imaging systems. Our results show that in both the detectability analysis and the ROC analysis, the flat-panel systems performed the best followed by the screen/film system. The CCD based system showed better detection accuracy compared to the CR system in the ROC analysis. However, there was no significant difference between the CCD and the CR systems in the detectability analysis.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
John X. Rong, Chris C. Shaw, Dennis A. Johnston, Michael R. Lemacks, Xinming Liu, Gary J. Whitman, Stephen K. Thompson, and Kerry T. Krugh "Comparison of a-Si:H CsI flat-panel digital imaging systems with a CCD-based system, CR systems, and conventional screen-film systems: a contrast-detail phantom study", Proc. SPIE 4320, Medical Imaging 2001: Physics of Medical Imaging, (28 June 2001); https://doi.org/10.1117/12.430950
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Imaging systems

Chromium

Charge-coupled devices

Chest

Computing systems

Radiography

Chest imaging

RELATED CONTENT


Back to Top