Paper
15 June 1984 An Analysis Of The Fundamental Limitations Of Screen-Film Systems For X-Ray Detection. II. Model Calculations
Rodney Shaw, Richard L. VanMetter
Author Affiliations +
Proceedings Volume 0454, Application of Optical Instrumentation in Medicine XII; (1984) https://doi.org/10.1117/12.939324
Event: Application of Optical Instrumentation in Medicine XII, 1984, San Diego, United States
Abstract
A range of model DQE calculations illustrates the detailed influence of the most important screen-film parameters. The screen is characterized by typical values for the primary x-ray absorption, spread function associated with secondary photons, and statistics associated with the gain of the conversion process. An existing film DQE model is used to explore the effect of film parameters on screen-film performance. Examples are given of the conditions under which screen-film DQE becomes independent of both film DQE and screen-film MTF. The influence of all these factors on system performance is discussed.
© (1984) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Rodney Shaw and Richard L. VanMetter "An Analysis Of The Fundamental Limitations Of Screen-Film Systems For X-Ray Detection. II. Model Calculations", Proc. SPIE 0454, Application of Optical Instrumentation in Medicine XII, (15 June 1984); https://doi.org/10.1117/12.939324
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Fiber optic gyroscopes

X-rays

Spatial frequencies

Photons

Medicine

Optical instrument design

Systems modeling

Back to Top