Paper
10 January 2003 Development of a model for gamma-ray spectra generation using pixelated mercuric iodide detectors
James E. Baciak, Zhong He
Author Affiliations +
Abstract
Recent studies have shown that pixelated HgI2 detectors with a thickness of 1 cm can yield energy resolutions of 1.4 - 2% FWHM at 662 keV γ-ray energy, which are similar to that obtained on 3-dimensional position sensitive CdZnTe detectors of the same thickness. However, there is a need to better understand the electron transportation and induced charge from the drift of charges within the detector in order to better understand the spectra obtained from HgI2 detectors. The results of a simple simulation model accounting for electron trapping and various weighting potentials is compared with recent physical spectra obtained from pixelated HgI2 detectors of 5 mm and 10 mm thickness.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
James E. Baciak and Zhong He "Development of a model for gamma-ray spectra generation using pixelated mercuric iodide detectors", Proc. SPIE 4784, X-Ray and Gamma-Ray Detectors and Applications IV, (10 January 2003); https://doi.org/10.1117/12.455979
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Electrodes

Gamma radiation

Electron transport

Image resolution

Mercury

Crystals

Back to Top