6 May 2004 Modeling alignment criteria of collimator to scintillation detector
Author Affiliations +
Abstract
In this paper, we present a simulation methodology that allows computed tomography (CT) detector designer to assess the criteria and tolerance requirements of positioning and alignment of x-ray collimators to scintillation detector arrays. Based on x-ray and optical transport simulation, we have developed an analytical model to predict the response of the measurement chain (collimator-scintillator-photodiode) in CT detector. We present the resulting effects of misalignments of the collimator array to detector array in the imaging chain through the simulation of beam hardening (water equivalent material and/or Bone) and light output using 120 kVp bremsstrahlung x-ray spectrum from a tungsten target anode source.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Aziz Ikhlef "Modeling alignment criteria of collimator to scintillation detector", Proc. SPIE 5368, Medical Imaging 2004: Physics of Medical Imaging, (6 May 2004); https://doi.org/10.1117/12.524044
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Sensors

Scintillators

Collimators

X-ray detectors

Monte Carlo methods

X-ray computed tomography

RELATED CONTENT

Volume CT (VCT) enabled by a novel diode technology
Proceedings of SPIE (April 20 2005)
Silicon drift chamber with extended energy range
Proceedings of SPIE (December 13 2000)
Performance of cone-beam CT using a flat-panel imager
Proceedings of SPIE (June 28 2001)
Crosstalk modeling of a CT detector
Proceedings of SPIE (May 06 2004)
Monte Carlo study of x ray cross talk in a...
Proceedings of SPIE (June 05 2003)
Small crystalline detector of high energy x ray and Beta...
Proceedings of SPIE (January 10 2003)

Back to Top