20 April 2005 Reflective and transmissive CR ScanHead technology on needle image plates
Jorg Frankenberger, Stephan Mair, Clemens Herrmann, Johan Lamotte, Robert Fasbender
Author Affiliations +
Abstract
The image quality of needle-image-plate (NIP) Computed Radiography (CR) scanners based on ScanHead technology was optimized. In order to get the best image quality for different applications, the influence of the phosphor layer thickness on the detective quantum efficiency (DQE) for different beam qualities was investigated. We compared a cassette-based, reflective CR-NIP-scanner to a new, transmissive flat-panel CR scanner with fixed, mounted NIP. The image quality was analyzed by DQE- and modulation transfer function (MTF) measurements supported by an observer study. The NIP systems reached DQE values up to three times higher than that of high-quality, state of the art CR scanners independent of the scanning principle. This allows a dose reduction by a factor of two to three without loss of image quality for both scanning systems. For high tube voltages, the variation of the phosphor layer thickness results in a DQE maximum at relatively large thicknesses. For lower tube voltages the DQE is less dependent on the layer thickness, reaching excellent values already at considerably lower thicknesses. Consequently, CR scanners can be adapted to different applications by using NIPs with different thicknesses. This could be easily realized for the cassette based system, but not for the flat-panel system with fixed IP. The latter demands a compromise with respect to the phosphor thickness, to yield superior image quality for all applications.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jorg Frankenberger, Stephan Mair, Clemens Herrmann, Johan Lamotte, and Robert Fasbender "Reflective and transmissive CR ScanHead technology on needle image plates", Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, (20 April 2005); https://doi.org/10.1117/12.597177
Lens.org Logo
CITATIONS
Cited by 10 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image quality

Chromium

Spatial frequencies

Modulation transfer functions

X-rays

Reflectivity

Chest

Back to Top