Paper
10 March 2006 Sequential intrinsic and extrinsic geometry calibration in fluoro CT imaging with a mobile C-arm
Arvi Cheryauka, Sebastien Breham, Wayne Christensen
Author Affiliations +
Abstract
Design of C-arm equipment with 3D imaging capabilitys involves retrieval of repeatable gantry positioning information along the acquisition trajectory. Inaccurate retrieval or improper use of positioning information may cause degradation of the reconstruction results, appearance of image artifacts, or indicate false structures. The geometry misrepresentation can also lead to the errors in relative pose assessment of anatomy-of-interest and interventional tools. Comprehensive C gantry calibration with an extended set of misalignment and motion parameters suffers from ambiguity caused by parameter cross-correlation and significant computational complexity. We deploy the concept of a waterfall calibration that comprises sequential intrinsic and extrinsic geometry calibration delineation steps. Following the image-based framework, the first step in our method is intrinsic calibration that deals with delineation of geometry of the X-ray tube-Detector assembly. Extrinsic parameters define motion of the C-arm assembly in 3D space and relate the Camera and World coordinate systems. We formulate both intrinsic and extrinsic calibration problems in vectorized form with total variation constraints. The proposed method has been verified by numerical design and validated by experimental studies. Sequential delineation of intrinsic and extrinsic geometries has demonstrated very efficient performance. The method eliminates the cross-correlation between cone-beam projection parameters, provides significantly better accuracy and computational speed, simplifies the structures of calibration targets used, and avoids the unnecessary workflow and image processing steps. It appears to be adequate for quality and cost derivations in an interventional surgery settings using a mobile C-arm.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Arvi Cheryauka, Sebastien Breham, and Wayne Christensen "Sequential intrinsic and extrinsic geometry calibration in fluoro CT imaging with a mobile C-arm", Proc. SPIE 6141, Medical Imaging 2006: Visualization, Image-Guided Procedures, and Display, 61412H (10 March 2006); https://doi.org/10.1117/12.653846
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

3D image processing

3D modeling

Data acquisition

Data modeling

Image processing

Image quality

RELATED CONTENT

Calibration of Geiger-mode lidar systems
Proceedings of SPIE (January 01 1900)
Stereo-based registration of ladar and color imagery
Proceedings of SPIE (October 06 1998)
Advances in image registration and fusion
Proceedings of SPIE (March 17 2008)
Computer Image Processing And Recognition
Proceedings of SPIE (May 28 1980)

Back to Top