Paper
22 March 2007 3D ultrasound reconstruction based on rotational scanning coping with calibration uncertainty
Xiaoan Luo, Mingyue Ding, Chengping Zhou, Chao Cai, Wu Qiu, Hua Zhou
Author Affiliations +
Abstract
To obtain 3D ultrasound image, traditionally, a 1D ultrasound transducer and an orientation system are used to get a series of 2D images and their positions. From these 2D images and position information, a 3D image is reconstructed. In this paper, the accuracy of the 3D image is determined by the accuracy of the position information, furthermore it affects the result of the measure from the 3D image. When in rotational scanning mode, the reconstructed 3D image is sensitive to the calibration parameters, including the orientation difference as well as the offset between the central line of each 2D US image and the rotational axis. To address this type of effects, we developed a 3D reconstruction technique considering the calibration uncertainty by building a transformation model with a 360 degrees scanning, and an accurate 3D reconstruction with calibration uncertainty is achieved. The experiments with both synthetic and scanning phantom data demonstrated the feasibility of our approach.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Xiaoan Luo, Mingyue Ding, Chengping Zhou, Chao Cai, Wu Qiu, and Hua Zhou "3D ultrasound reconstruction based on rotational scanning coping with calibration uncertainty", Proc. SPIE 6509, Medical Imaging 2007: Visualization and Image-Guided Procedures, 65092G (22 March 2007); https://doi.org/10.1117/12.711751
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
3D image reconstruction

3D image processing

Calibration

3D modeling

Reconstruction algorithms

Ultrasonography

Lawrencium

Back to Top