18 March 2008 Parameter optimization for a grating-based phase contrast x-ray system
Author Affiliations +
Abstract
Differential Phase Contrast Imaging (DPCI) has the potential to vastly increase soft tissue contrast. DPCI requires spatial and temporal coherence as generated by a synchrotron or a micro-focus X-ray source; however, recent research demonstrates DPCI can be implemented using a conventional X-ray source with three transmission gratings (Pfeiffer et al., Nature 2006). This paper describes the optimization of the essential system parameters (system size, delivered dose, spatial resolution) of this implementation from a theoretical perspective. The optimization of these parameters is an essential step in practical application of DPCI. We conclude that the minimum size of the system is approximately 700 mm, the minimum resolution is 100 um, and the dose is 1/1000 that of conventional absorption CT.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Bonsung Koo, Ming Jiang, Christopher L. Wyatt, and Ge Wang "Parameter optimization for a grating-based phase contrast x-ray system", Proc. SPIE 6913, Medical Imaging 2008: Physics of Medical Imaging, 691348 (18 March 2008); https://doi.org/10.1117/12.770548