14 March 2009 Dual energy with dual source CT and kVp switching with single source CT: a comparison of dual energy performance
M. Grasruck, S. Kappler, M. Reinwand, K. Stierstorfer
Author Affiliations +
Proceedings Volume 7258, Medical Imaging 2009: Physics of Medical Imaging; 72583R (2009) https://doi.org/10.1117/12.811534
Event: SPIE Medical Imaging, 2009, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
Stimulated by the introduction of clinical dual source CT, the interest in dual energy methods has been increasing in the past years. Whereas the potential of material decomposition by dual energy methods is known since the early 1980ies, the realization of dual energy methods is a wide field of today's research. Energy separation can be achieved with energy selective detectors or by varying X-ray source spectra. This paper focuses on dual energy techniques with varying X-ray spectra. These can be provided by dual source CT devices, operated with different kVp settings on each tube. Excellent spectral separation is the key property for use in clinical routine. The drawback of higher cost for two tubes and two detectors leads to an alternative realization, where a single source CT yields different spectra by fast kVp switching from reading to reading. This provides access to dual-energy methods in single source CT. However, this technique comes with some intrinsic limitations. The maximum X-ray flux is reduced in comparison to the dual source system. The kVp rise and fall time between each reading reduces the spectral separation. In comparison to dual source CT, for a constant number of projections per energy spectrum the temporal resolution is reduced; a reasonable trade of between reduced numbers of projection and limited temporal resolution has to be found. The overall dual energy performance is the guiding line for our investigations. We present simulations and measurements which benchmark both solutions in terms of spectral behavior, especially of spectral separation.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
M. Grasruck, S. Kappler, M. Reinwand, and K. Stierstorfer "Dual energy with dual source CT and kVp switching with single source CT: a comparison of dual energy performance", Proc. SPIE 7258, Medical Imaging 2009: Physics of Medical Imaging, 72583R (14 March 2009); https://doi.org/10.1117/12.811534
Lens.org Logo
CITATIONS
Cited by 15 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Dual energy imaging

Switching

Temporal resolution

Iodine

Bone

Sensors

Scanners

Back to Top