Paper
9 July 2009 Experimental determination of frequency dependent acoustic attenuation for photoacoustic imaging
Author Affiliations +
Abstract
In this work measurements of the acoustic attenuation coefficient for water in the frequency range of 20-40 Mhz have been performed. These measurements have been conducted with a photoacoustic setup, containing a nano second pulse laser and an integrating line detector. The nanosecond pulse laser is used to generate ultra sound waves in a target. Those waves are transmitted into water and are then detected by the line sensor. In this way the resulting ultra sound field was scanned and the absorption coefficient was evaluated. Different pulse energies, beam diameters and target materials were used. The resulting coefficients are in the range of the reported values found in literature.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Johannes Bauer-Marschallinger, Francisco Camacho-Gonzales, Thomas Berer, Hubert Grün, and Peter Burgholzer "Experimental determination of frequency dependent acoustic attenuation for photoacoustic imaging", Proc. SPIE 7371, Novel Optical Instrumentation for Biomedical Applications IV, 73711P (9 July 2009); https://doi.org/10.1117/12.831763
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.